Aesculap Inc.(AIC) Miethke Ventricular Catheter with Deflector/Miethke Shunt System The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Catalog No: FV431T, FV434T, FV433T, FV459T, FV078P, FV494T, FV676T, FV324T Lot No: 4505150452,4505156642,4505164365,4505134636,4505150384,4505156634,4505246780,4505394315,4505150371,4505174213,4505187583,4505366616,4505134635,4505187582,4505134629,4505212089,4505199667,4505150340,4505171900,4505168804
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Aesculap, Inc.
Reason for Recall:
AIC (USA) received information regarding 3 complaints in which the deflector did not move freely on the ventricular catheter.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Aesculap Inc.(AIC) Miethke Ventricular Catheter with Deflector/Miethke Shunt System The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Product Codes/Lot Numbers:

Catalog No: FV431T, FV434T, FV433T, FV459T, FV078P, FV494T, FV676T, FV324T Lot No: 4505150452,4505156642,4505164365,4505134636,4505150384,4505156634,4505246780,4505394315,4505150371,4505174213,4505187583,4505366616,4505134635,4505187582,4505134629,4505212089,4505199667,4505150340,4505171900,4505168804

Distribution:

Distributed in: US, PA, KY, NY, RI, SC, KS, TX, CT, TN, WA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1676-2014

Related Recalls