Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-10012-200, SIZE 2-LEFT MEDIAL; d) REF 1-10012-250, SIZE 2-RIGHT MEDIAL; e) REF 1-10012-300, SIZE 3-LEFT MEDIAL; f) REF 1-10012-350, SIZE 3-RIGHT MEDIAL; g) REF 1-10012-400, SIZE 4-LEFT MEDIAL; h) REF 1-10012-450, SIZE 4-RIGHT MEDIAL; i) REF 1-10012-500, SIZE 5-LEFT MEDIAL; j) REF 1-10012-550, SIZE 5-RIGHT MEDIAL; k) REF 1-10012-600, SIZE 6-LEFT MEDIAL; l) REF 1-10012-650, SIZE 6-RIGHT MEDIAL; Unicondylar knee prothesis

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a) REF 1-10012-100, UDI/DI 00885556873113; b) REF 1-10012-150, UDI/DI 00885556873120; c) REF 1-10012-200, UDI/DI 00885556873137; d) REF 1-10012-250, UDI/DI 00885556873144; e) REF 1-10012-300, UDI/DI 00885556873151; f) REF 1-10012-350, UDI/DI 00885556873168; g) REF 1-10012-400, UDI/DI 00885556873175; h) REF 1-10012-450, UDI/DI 00885556873182; i) REF 1-10012-500, UDI/DI 00885556873199; j) REF 1-10012-550, UDI/DI 00885556873205; k) REF 1-10012-600, UDI/DI 00885556873212; l) REF 1-10012-650, UDI/DI 00885556873229; ALL BATCHES
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Smith & Nephew, Inc.
Reason for Recall:
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-10012-200, SIZE 2-LEFT MEDIAL; d) REF 1-10012-250, SIZE 2-RIGHT MEDIAL; e) REF 1-10012-300, SIZE 3-LEFT MEDIAL; f) REF 1-10012-350, SIZE 3-RIGHT MEDIAL; g) REF 1-10012-400, SIZE 4-LEFT MEDIAL; h) REF 1-10012-450, SIZE 4-RIGHT MEDIAL; i) REF 1-10012-500, SIZE 5-LEFT MEDIAL; j) REF 1-10012-550, SIZE 5-RIGHT MEDIAL; k) REF 1-10012-600, SIZE 6-LEFT MEDIAL; l) REF 1-10012-650, SIZE 6-RIGHT MEDIAL; Unicondylar knee prothesis

Product Codes/Lot Numbers:

a) REF 1-10012-100, UDI/DI 00885556873113; b) REF 1-10012-150, UDI/DI 00885556873120; c) REF 1-10012-200, UDI/DI 00885556873137; d) REF 1-10012-250, UDI/DI 00885556873144; e) REF 1-10012-300, UDI/DI 00885556873151; f) REF 1-10012-350, UDI/DI 00885556873168; g) REF 1-10012-400, UDI/DI 00885556873175; h) REF 1-10012-450, UDI/DI 00885556873182; i) REF 1-10012-500, UDI/DI 00885556873199; j) REF 1-10012-550, UDI/DI 00885556873205; k) REF 1-10012-600, UDI/DI 00885556873212; l) REF 1-10012-650, UDI/DI 00885556873229; ALL BATCHES

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1672-2023

Related Recalls

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

Mar 18, 2025 Other Medical Devices Nationwide View Details →