SIEMENS Luminos Agile Max; a universal imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile is applicable to emergency treatment on an outpatient basis, as well as for bedside examinations.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number of device 10762472 with serial numbers: 61034 61052 61045 61013 61054 61056 61021 61050 61053 61032 61015 61046 61028 61040 61011 61029 61048 61060 61042 61035 61014 61025 61043 61036 61038 61022 61037 61049 61044 61062 61030 61018 61027 61061 61033 61039 61047 61031 61041 61055 61019 61057 61059 61017 61051
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SIEMENS Luminos Agile Max; a universal imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile is applicable to emergency treatment on an outpatient basis, as well as for bedside examinations.
Product Codes/Lot Numbers:
Model Number of device 10762472 with serial numbers: 61034 61052 61045 61013 61054 61056 61021 61050 61053 61032 61015 61046 61028 61040 61011 61029 61048 61060 61042 61035 61014 61025 61043 61036 61038 61022 61037 61049 61044 61062 61030 61018 61027 61061 61033 61039 61047 61031 61041 61055 61019 61057 61059 61017 61051
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1669-2015
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.