Delta Vaporizer An anesthetic vaporizer is a device used to vaporize liquid anesthetic and deliver a controlled amount of the vapor to the patient.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All; recalled products are traced by serial numbers D0312-XXXX to D1112-XXXX.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Penlon, Ltd.
- Reason for Recall:
- It has come to Penlon Ltd's attention of an instance where the installation and pre-use checks listed in the User Manual may not have been followed prior to the clinical use of the vaporizer. Penlon Ltd has updated the Installation and daily Pre-Use Check sections of the User Manual to ensure that the instructions are explicit and clear.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Delta Vaporizer An anesthetic vaporizer is a device used to vaporize liquid anesthetic and deliver a controlled amount of the vapor to the patient.
Product Codes/Lot Numbers:
All; recalled products are traced by serial numbers D0312-XXXX to D1112-XXXX.
Distribution:
Distributed in: FL, KY, PA, ND, MN, IL, NJ, TX, TN, VA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1653-2013