Artis Q floor- Model No. 10848280 Artis Q ceiling- Model No. 10848281 Artis Q biplane- Model No. 10848282 Artis Q zeego- Model No. 10848283 Artis Q.zen floor- Model No. 10848353 Artis Q.zen ceiling- Model No. 10848354 Artis Q.zen biplane- Model No. 10848355

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Artis Q floor- Model No. 10848280 UDI-DI: 4056869009971 Serial Numbers 103221 103217 104049 103409 103232 104005 104027 132508 132510 103359 104051 103220 104029 103365 104025 103286 103222 132507 104047 132502 132509 132500 103241 103212 103254 103255 ***Updated 9/23/22*** 103035 ***Updated 1/9/23*** 103444 103050 Artis Q ceiling- Model No. 10848281 UDI-DI: 4056869009988 Serial Numbers 109552 110088 110199 110059 110103 133503 110166 109123 110005 109344 109628 110058 110156 110176 110180 110147 110181 110056 110158 109074 109479 109197 109421 109792 109738 109405 110195 109437 109645 109166 110067 110161 110184 110169 110119 109747 110144 109760 110095 110167 110192 110131 109169 110093 109769 133202 109202 110109 110128 109378 109384 110096 109621 133504 110198 109015 110069 110106 110139 109211 109726 110185 110075 109674 109627 109780 110179 110063 110047 110102 109656 109309 110023 110110 110159 110065 110052 109057 109915 110191 109604 110136 110201 109457 110072 109034 109129 110145 109181 109884 109643 109678 110081 109330 109058 110182 109380 109222 109435 110054 109188 109518 109458 109155 109379 109538 109752 109659 110053 110107 109591 109630 109393 110196 110175 109920 110193 110187 110077 109164 110079 109362 109762 110099 110113 109717 110134 109564 109813 110049 110117 110066 109088 110186 109397 109685 109096 109159 110154 110183 109715 109364 110100 110129 110188 110190 109463 110083 110091 109048 109742 109301 109020 109036 109818 110115 109664 109163 109105 110055 109854 109069 109010 109756 110152 133501 109156 109157 109307 110177 110057 109391 110132 110104 110174 110149 ***Updated 9/23/22*** 110223 110227 110224 110228 109225 109358 133508 133509 ***Updated 1/9/23*** 109417 109066 110048 110014 110233 Artis Q biplane- Model No. 10848282 UDI-DI: 4056869009995 Serial Numbers 122049 121665 121564 122025 121627 121185 121336 122012 121471 121467 121612 122022 121094 122040 121006 122021 122028 121013 121501 121208 121459 121011 122037 121054 121422 121209 121325 121357 121369 121133 121200 121163 121364 122039 121154 121418 121442 122046 121205 122034 121137 121012 121049 122042 121332 121183 121455 121019 121381 121352 122050 121434 121165 121625 121415 121355 121010 121210 121573 121215 121323 121370 121556 121384 121487 122017 121395 121405 121513 121342 121301 121081 121649 121190 121424 134505 121629 121529 121138 121039 121521 121600 121193 121554 121179 121216 121566 121417 121151 121188 121168 121106 134502 134500 122033 121411 121120 121121 121541 122010 122048 121436 121359 121622 121141 121648 121535 122016 121175 121439 121419 121008 121152 121155 121160 121449 122013 121576 121379 121085 121361 121206 121431 121544 121211 121187 121197 121375 121377 121505 121601 121506 121616 121594 121510 121533 122024 122027 121041 121502 122014 ***Updated 1/9/23*** 121015 121309 121633 122032 121653 121393 134506 Artis Q zeego- Model No. 10848283 UDI-DI: 4056869010007 Serial Numbers 117316 117330 117006 117034 117026 117123 117111 117024 117309 117113 117322 117040 117041 ***Updated 9/23/22*** 117136 Artis Q.zen floor- Model No. 10848353 UDI-DI: 4056869010014 Serial Numbers 105007 105503 105018 105119 105125 Artis Q.zen ceiling- Model No. 10848354 UDI-DI: 4056869010021 Serial Numbers 111162 111163 111137 111502 111523 111174 111510 111527 111155 111140 111507 111124 111512 111521 111133 111110 111101 111524 111526 111149 111109 111151 111130 111517 111513 111017 111120 111127 111157 111142 111525 111107 111514 111128 ***Updated 1/9/23*** 111529 111530 111528 ***Updated 1/13/23*** 111535 Artis Q.zen biplane- Model No. 10848355 UDI-DI: 4056869010038 Serial Numbers 123245 123026 123031 123050 123008 123255 123247 123254 123221 123508 123505 123510 123205 123509 123230 123239 123010 ***Updated 1/9/23*** 123225 123513
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Artis Q floor- Model No. 10848280 Artis Q ceiling- Model No. 10848281 Artis Q biplane- Model No. 10848282 Artis Q zeego- Model No. 10848283 Artis Q.zen floor- Model No. 10848353 Artis Q.zen ceiling- Model No. 10848354 Artis Q.zen biplane- Model No. 10848355

Product Codes/Lot Numbers:

Artis Q floor- Model No. 10848280 UDI-DI: 4056869009971 Serial Numbers 103221 103217 104049 103409 103232 104005 104027 132508 132510 103359 104051 103220 104029 103365 104025 103286 103222 132507 104047 132502 132509 132500 103241 103212 103254 103255 ***Updated 9/23/22*** 103035 ***Updated 1/9/23*** 103444 103050 Artis Q ceiling- Model No. 10848281 UDI-DI: 4056869009988 Serial Numbers 109552 110088 110199 110059 110103 133503 110166 109123 110005 109344 109628 110058 110156 110176 110180 110147 110181 110056 110158 109074 109479 109197 109421 109792 109738 109405 110195 109437 109645 109166 110067 110161 110184 110169 110119 109747 110144 109760 110095 110167 110192 110131 109169 110093 109769 133202 109202 110109 110128 109378 109384 110096 109621 133504 110198 109015 110069 110106 110139 109211 109726 110185 110075 109674 109627 109780 110179 110063 110047 110102 109656 109309 110023 110110 110159 110065 110052 109057 109915 110191 109604 110136 110201 109457 110072 109034 109129 110145 109181 109884 109643 109678 110081 109330 109058 110182 109380 109222 109435 110054 109188 109518 109458 109155 109379 109538 109752 109659 110053 110107 109591 109630 109393 110196 110175 109920 110193 110187 110077 109164 110079 109362 109762 110099 110113 109717 110134 109564 109813 110049 110117 110066 109088 110186 109397 109685 109096 109159 110154 110183 109715 109364 110100 110129 110188 110190 109463 110083 110091 109048 109742 109301 109020 109036 109818 110115 109664 109163 109105 110055 109854 109069 109010 109756 110152 133501 109156 109157 109307 110177 110057 109391 110132 110104 110174 110149 ***Updated 9/23/22*** 110223 110227 110224 110228 109225 109358 133508 133509 ***Updated 1/9/23*** 109417 109066 110048 110014 110233 Artis Q biplane- Model No. 10848282 UDI-DI: 4056869009995 Serial Numbers 122049 121665 121564 122025 121627 121185 121336 122012 121471 121467 121612 122022 121094 122040 121006 122021 122028 121013 121501 121208 121459 121011 122037 121054 121422 121209 121325 121357 121369 121133 121200 121163 121364 122039 121154 121418 121442 122046 121205 122034 121137 121012 121049 122042 121332 121183 121455 121019 121381 121352 122050 121434 121165 121625 121415 121355 121010 121210 121573 121215 121323 121370 121556 121384 121487 122017 121395 121405 121513 121342 121301 121081 121649 121190 121424 134505 121629 121529 121138 121039 121521 121600 121193 121554 121179 121216 121566 121417 121151 121188 121168 121106 134502 134500 122033 121411 121120 121121 121541 122010 122048 121436 121359 121622 121141 121648 121535 122016 121175 121439 121419 121008 121152 121155 121160 121449 122013 121576 121379 121085 121361 121206 121431 121544 121211 121187 121197 121375 121377 121505 121601 121506 121616 121594 121510 121533 122024 122027 121041 121502 122014 ***Updated 1/9/23*** 121015 121309 121633 122032 121653 121393 134506 Artis Q zeego- Model No. 10848283 UDI-DI: 4056869010007 Serial Numbers 117316 117330 117006 117034 117026 117123 117111 117024 117309 117113 117322 117040 117041 ***Updated 9/23/22*** 117136 Artis Q.zen floor- Model No. 10848353 UDI-DI: 4056869010014 Serial Numbers 105007 105503 105018 105119 105125 Artis Q.zen ceiling- Model No. 10848354 UDI-DI: 4056869010021 Serial Numbers 111162 111163 111137 111502 111523 111174 111510 111527 111155 111140 111507 111124 111512 111521 111133 111110 111101 111524 111526 111149 111109 111151 111130 111517 111513 111017 111120 111127 111157 111142 111525 111107 111514 111128 ***Updated 1/9/23*** 111529 111530 111528 ***Updated 1/13/23*** 111535 Artis Q.zen biplane- Model No. 10848355 UDI-DI: 4056869010038 Serial Numbers 123245 123026 123031 123050 123008 123255 123247 123254 123221 123508 123505 123510 123205 123509 123230 123239 123010 ***Updated 1/9/23*** 123225 123513

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1645-2022

Related Recalls

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →
Class I - Dangerous

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA

Class I - Dangerous

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Mar 12, 2026 Other Medical Devices View Details →