AggreGuide A-100 Instrument

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 00856893004065; Software Versions 5.1 or 5.2; Serial Numbers Between xx0001 and xx0160
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Aggredyne, Inc.
Reason for Recall:
The firm identified a software design issue where the improper shut down of the instrument resulted in speeds.csv file contents of the software version 5.1 being erased. Due to this issue, the target RPM for the assay is 0 and the motor output required to achieve this RPM is also 0. Therefore, when the assay is run, the software applies 0 Volts to the motor and performs a 0 RPM assay. Because of this, there is no mixing of the blood with agonist and no moving of aggregates happens in front of the optical detectors. The resulting PAI is 0 and the result is displayed as Low.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
completed

Product Information

Full Description:

AggreGuide A-100 Instrument

Product Codes/Lot Numbers:

UDI-DI: 00856893004065; Software Versions 5.1 or 5.2; Serial Numbers Between xx0001 and xx0160

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1639-2022