BD WAVELINQ 4F EndoAVF System, Arterial Catheter 50 cm, Venous Catheter 43 cm, Sterile EO, Rx only, Mfr. Bard Peripheral Vascular, Inc., Ref W04200, Lot S0053, UDI (01)00801741182754(17)210215(10)S0053 - Product Usage: The WavelinQ 4F EndoAVF System in indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code: W04200 Lot/Serial Number: S0053
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Bard Peripheral Vascular Inc
- Reason for Recall:
- The firm has become aware of a potential problem with their venous and arterial magnetic catheters may be at risk of experiencing a magnetic deficiency, which may result in the magnets failing to attract to one another. This may result in procedural delay in obtaining or creating a functional fistula and need for additional contrast to perform the necessary fluoroscopy. This can pose an incremental risk of harm to a diabetic patient's kidney function.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BD WAVELINQ 4F EndoAVF System, Arterial Catheter 50 cm, Venous Catheter 43 cm, Sterile EO, Rx only, Mfr. Bard Peripheral Vascular, Inc., Ref W04200, Lot S0053, UDI (01)00801741182754(17)210215(10)S0053 - Product Usage: The WavelinQ 4F EndoAVF System in indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.
Product Codes/Lot Numbers:
Product Code: W04200 Lot/Serial Number: S0053
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1637-2020
Related Recalls
Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 60060
Bard Peripheral Vascular
Due to damage to outer tray that can potentially compromise the sterile barrier
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Bard Peripheral Vascular
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.