Philips V60 Ventilators with Power Management PCBA part number 1055906
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 100002908 to 100027093; 100027096 to 100085389; 100085399 to 100109746; 100109868 to 201003390; 201003392 to 201010952
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Respironics California, LLC
- Reason for Recall:
- Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Philips V60 Ventilators with Power Management PCBA part number 1055906
Product Codes/Lot Numbers:
Serial Numbers: 100002908 to 100027093; 100027096 to 100085389; 100085399 to 100109746; 100109868 to 201003390; 201003392 to 201010952
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1622-2020
Related Recalls
Philips Respironics V680 Ventilator (All Models, including: P/N 1103044 (850011)), *Distributed outside the United States
Respironics California
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
Philips Respironics V60 Plus Ventilator (All Models, including: 1138747, 1137276)
Respironics California
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, 1053615, 1053616, 1053617, 1053618, 1076709, 1076716, 1076717, 1137292, DU1053617, U1053617, R1053618, R1076709)
Respironics California
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.