ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip; 51 count MedBen strip 51 count Mail Order strip; Sample Kit 16 per case. The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Compact Plus Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Roche Diagnostics Operations, Inc.
- Reason for Recall:
- Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip; 51 count MedBen strip 51 count Mail Order strip; Sample Kit 16 per case. The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Compact Plus Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm.
Product Codes/Lot Numbers:
5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1616-2014
Related Recalls
cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine
Roche Diagnostics Operations
Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination of creatine kinase (CK) in human serum and plasma. Catalog Number: 07190794190
Roche Diagnostics Operations
Creatine Kinase (CK) Assay Abnormal calibrations on the cobas¿ c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.
Tina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Material Numbers 03001245322 - (cobas c 311, 501 and 502; INTEGRA 400 plus), 08105618190 - (cobas c 303 and 503)
Roche Diagnostics Operations
Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.