Recall Details
- Company:
- Maquet Cardiovascular, LLC
- Reason for Recall:
- Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Maquet Getinge-BEQ-T 5915 TCH 3/8 Inch Change Kit Material: 709000280
Product Codes/Lot Numbers:
Batch Numbers: 3000071199 3000081153
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1609-2019
Related Recalls
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Heartstring III Proximal Seal System. Intravascular anastomosis occluder.
Maquet Cardiovascular
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
Maquet Cardiovascular
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.