CareFusion AVEA ventilator, all models. Product Usage: A continuous ventilator (respirator), intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial No. AKV01874 AGV01443 AEV02197 AKV02580 AEV02375 BBV01304 BBV01303 AJV02510 AJV02502 AGV02650 AFV01802 AFV01792 AFV01400 AFV01361 AEV01622 AEV01608 AKV01915 BBV03730 BBV03557 BBV03556 BBV03554 BBV03553 BBV 3558 BBV01406 BBV01396 AHV01880 AGV02205 BBV01432 AFV01657 AFV01655 AFV01647 BAV05234 AHV03069 AGV02249 BBV03797 BAV06514 BDV01011 BDV01000 BCV01938 BAV04011 BAV04010 BAV04009 BAV04005 BAV04004 BAV04003 BAV04002 BAV01313 AKV04053 AJV02289 BAV04021 AEV02264 AEV02225 AEV02205 ADV06091 AKV02569 AHV01206 AKV04351 AKV04349 AKV02550 AGV01385 AHV03128 ADV04116 BBV06059 BBV06057 BAV05153 AFV01492 BBV01620 BDV01371 BAV06447 BAV06446 BAV06445 AEV01724 AEV01712 BAV06463 AEV01694 AJV02446 AGV02139 BAV04188 BAV04187 AKV01153 AKV01147 AEV02720 AEV02717 AEV02713 BAV04191 BAV04993 AHV02129 ADV06297 ADV01637 ADV01592 AFV01922 AFV02277 AHV01274 BAV05235 BCV02186 BCV01822 BBV06061 BDV01249 BDV01242 ADV03032 AJV01685 AHV01967 AHV01722 AHV01701 AKV02230 AKV02226 AKV02224 AKV02219 AKV02217 AKV02215 AKV02214 BEV01326 BEV01325 BEV01324 BEV01306 BEV01205 BDV02001 BDV01994 BDV01810 AKV01279 AKV04320 BAV06124 BBV03028 BBV03027 BBV01483 BBV01019 BBV01018 BBV01016 BAV01015 BAV01020 BBV01010 AKV01029 AKV01012 AKV01007 AKV01000 BCV02865 BAV01088 BAV01026 BAV01018 AKV01018 AKV01017 AHV03287 AHV03061 AFV02243 AFV02230 ADV01638 AGV01808 BFY01003 AGV02074 AFV01176 AKV04033 AEV01588 BAV01089 AKV01009 AKV01005 AFV02205 BBV01005 ACV01718 AKV01028 BFV01053 BEV02183 BAV01099 ADV06397 ADV03003 BAV04113 BAV04111 BEV02185 BEV02184 BDV01816 ajv01601 AGV01333 AGV01278 AFV01962 AHV01326 AJV02876 AEV02362 AEV02359 AEV02340 AFV01616 AFV01744 AFV01661 AFV01633 AFV01632 AFV01631 AFV01629 AFV01625 AFV01545 AFV01481 AFV01449 BEV01356 BEV01352 BDV02093 BCV01220 BCV01206 AGV02366 BBV06021 BBV06016 BBV06007 BBV06006 AEV02120 AKV01274 AKV01240 BCV01810 BCV01619 BCV01176 BCV01072 BCV01059 BCV01057 BBV03951 BBV03598 BBV03570 BAV04368 AKV01068 AKV01047 BAV04362 BDV01515 BDV01480 BEV01663 BEV01648 BCV02271 BCV02270 AJV02640 AHV01549 AEV02727 BBV01207 BBV01206 AKV01466 AKV01442 AKV01367 AKV01046 AKV01003 AJV02681 BAV04017 BAV04016 BAV04015 BAV04014 BAV01306 AKV04488 AGV02552 AGV02441 AGV02432 AJV01532 AJV01476 AJV01424 BBV03552 AJV02441 AJV02950 AJV02504 AJV02467 AJV02432 AJV02427 AJV04226 AJV04195 AJV04219 AJV04225 AJV04192 AJV04228 AJV04232 AJV04223 AJV04229 AJV04201 AJV04205 AJV04203 AJV04202 AJV04221 AJV04189 AJV04176 AJV04046 AJV04233 AJV04224 AKV04118 AHV01285 AJV04196 BEV01544 BEV01543 BEV01542 BEV01541 BEV01540 BEV01539 BEV01534 BEV01531 BEV01522 BEV01519 BEV01515 AJV01591 AKV02755 ADV06008 BDV02092 BDV02087 BDV02022 BDV02021 BDV02002 BDV01980 BDV01668 AFV02626 AKV04119 AFV01567 BCV02763 BCV02757 AFV02102 BBV01459 BDV01474 BDV01461 BCV01732 BCV01729 BCV01722 BAV04354 ADV01559 BBV01300 BBV01292 BBV01278 BDV01818 BDV01646 BBV01017 BEV02141 BEV02137 BEV02136 BEV02135 BEV02134 BEV02133 BEV02131 BEV02127 BEV02123 BEV02113 BDV02011 BDV02006 BDV01995 BDV01982 BDV01971 AKV04324 AJV01076 AKV04325 BCV02209 BCV02208 BCV02205 AJV02216 AJV02173 AFV02439 AFV02368 AFV02299 AFV02281 AFV02183 AFV02182 AFV02170 AFV02129 AFV02000 AFV01688 AFV01668 AFV01296 AGV01584 AGV01527 BAV01218 BAV01217 BAV06470 BEV01359 BCV02187 BCV01831 BCV01830 AJV01531 AJV01524 AHV01705 AGV01413 AFV02402 AFV01572 AFV01546 AFV01250 AFV01247 AFV01165 AFV01106 AEV02698 AGV01762 BDV01385 BDV01375 AHV01434 BEV01140 BEV01135 BCV01708 BCV01705 BCV01704 BCV01703 BCV01702 BCV01701 BBV01480 BBV01472 BBV01471 BBV01470 BBV01468 BBV01465 BBV01462 BAV05231 BAV05230 BAV05229 BAV05228 BEV01131 AJV02530 AJV02529 AJV02505 AGV02710 AKV04061 AKV04060 AKV01478 AHV01745 ADV01758 AGV01287 AKV01475 AKV01022 AJV03054 AJV03053 AJV01139 AHV03369 AHV01916 AGV02297 AFV02424 ADV04002 ADV02151 AGV01352 AKV04114 AKV04116 BDV01830 BDV01745 BAV05103 BAV05102 BAV05101 AKV04163 AKV04129 AKV01926 AKV01906 AKV04104 AKV01402 BAV05001 BAV05002 BBV01176 AJV02144 AJV02130 AJV02121 AHV01778 AHV01748 BAV01041 BCV01785 BCV01734 BBV03213 BBV03206 BBV03112 AKV01964 AKV01959 AKV01957 AGV01497 BDV01264 BFV01058 BEV01865 BDV01285 BEV01235 BEV01233 BEV01226 BEV01224 BEV01217 BEV01208 BEV01202 BDV01974 BDV01970 BDV01357 BDV01349 BDV01347 BDV01338 BDV01336 BDV01333 BDV01318 BDV01317 BDV01181 BDV01175 BDV01104 BCV02335 BAV05008 BAV05007 BAV05006 BAV05005 BAV05004 BAV05003 BAV05000 BAV04999 AGV02364 BBV01409 BBV01380 BBV01379 BBV01357 BBV01197 BBV01181 BBV01173 BBV01167 BBV01164 BAV04256 BAV04254 BCV02306 BCV02304 BCV02303 AEV02657 AJV01699 AGV01964 BBV03145 AHV01184 AGV01160 BEV01264 BEV01227 BEV01028 BEV01017 AGV01986 AGV01259 AGV01255 AGV01242 AGV01241 AGV01156 AGV01138 AGV01037 AFV02619 AFV02616 BBV01213 BBV01203 BBV01201 BBV01000 AHV01575 AHV01591 AGV02223 AGV01237 AGV01222 AGV01198 BCV02262 BCV02261 BCV02260 BCV02259 BCV02258 BCV02257 BCV02256 BCV02254 BCV02253 BCV02252 BCV02249 BCV02248 BCV02247 BCV02246 BCV02245 BCV02244 BCV02243 BCV02241 BCV02239 BCV02238 BCV02237 BCV02236 BCV02235 BCV02234 BCV02231 BCV02222 BCV02220 BBV03803 BBV03796 AFV01043 BAV01150 AJV02703 AJV01645 AHV01370 AGV01811 AFV01503 AEV02548 AEV02551 BDV01601 BCV02633 BBV03599 AGV02040 AGV01257 BCV02240 AKV02765 AKV02763 AKV02762 AKV02761 AKV02760 AKV02759 AKV02758 AKV02757 AKV02756 AKV02225 AKV01893 AJV02967 AHV01636 AHV01621 AHV01585 AHV01526 AHV01648 AGV02704 BEV01182 BEV01181 BEV01180 BEV01179 BEV01178 BEV01176 BEV01175 BEV01173 BEV01172 BEV01169 BEV01168 BEV01150 BDV01308 BDV01307 BDV01306 BDV01303 BDV01302 BDV01301 BDV01147 BCV02338 BCV02266 BCV02265 BCV02264 BCV02263 BCV01748 BAV04434 BAV04432 BAV04431 BAV04430 BAV04429 BAV04428 BAV04427 AKV04029 AKV04028 AKV04026 AKV04025 AKV04024 AKV04022 AKV04020 AKV04019 AKV04018 AKV04017 AKV04016 AKV04015 AKV04014 AKV04013 AKV04010 AKV04009 AKV04008 AKV04007 AKV04006 AKV04005 AKV04003 AHV01620 AHV01615 AGV02580 AGV02556 AGV02545 AGV02481 AGV01940 BAV04433 AGV02590 AKV04023 AKV04027 AKV04021 AKV04001 AKV04011 AKV04004 BDV01645 BDV01559 BAV04772 BAV04768 BFV01055 BFV01054 BFV01051 BBV01223 BAV06784 BAV06780 BAV06777 BAV06796 BAV06782 BAV06783 BAV06781 BAV06778 BAV06775 BBV03005 BBV03000 BAV04489 BAV04488 BAV04487 BAV04486 BAV04485 BAV04481 AJV02381 BCV01312 BCV01262 BDV01684 BCV01738 BCV02176 BCV02174 AKV04316 AKV04315 AKV04314 AKV02719 AKV02718 AKV02201 AKV02175 AKV01616 AEV01908 BCY01036 BBV06068 AJV01609 AHV01343 AGV01327 AEV02473 AEV01808 AGV01318 AEV02648 AJV02165 AJV01665 AFV01117 AEV02712 AEV02700 AEV02581 AEV02380 BDV01263 BDV01259 AKV04120 AGV02357 AEV01520 AGV01855 AGV01852 AGV01851 AGV01850 AGV01849 ADV06048 ADV05035 ADV05033 ADV05031 AGV01857 BCV02063 AKV04318 AKV04317 BCV01930 BCV01924 BCV01920 BCV01917 BCV01902 BCV01895 BCV01894 BCV01888 BCV01886 BCV01885 BCV01784 BCV01782 BCV01775 BCV01770 BCV01766 BCV01762 BCV01760 BCV01758 BCV01757 BCV01756 BCV01754 BCV01751 BCV01683 BCV01663 BCV01657 AKV01619 AKV01606 AKV01604 AKV01602 AKV01500 AJV02688 AJV02473 AJV02425 BAV04112 BAV04094 BCV02558 AJV01073 AJV01072 AJV01067 AJV01065 AHV03215 AJV01625 BDV01981 BDV01848 BDV01842 BDV01841 BDV01836 BDV01824 BDV01822 BDV01820 BDV01815 BDV01748 BCV01771 BCV01763 BCV01710 BCV01699 BCV01696 BCV01689 BAV04770 AKV04460 AKV04458 BAV04777 BAV04771 BAV04769 BAV04776 AKV04459 AKV04462 AKV04461 AKV04457 BAV04792 BAV04790 BAV04789 BAV04786 AKV02753 AKV02752 AKV02581 AKV02751 BDV01029 BCV02805 AKV01614 AKV01591 BEV01304 BEV01195 BEV01193 BEV01192 BEV01191 BEV01190 BEV01188 BEV01186 BEV01185 BEV01184 BEV01183 BEV01171 BEV02119 BEV02116 BEV02114 BEV02110 BEV02099 BEV02090 BEV01879 BEV01857 BEV01290 BEV01288 BEV01287 BEV01285 BEV01282 BEV01281 BEV01280 BEV01278 BEV01277 BEV01275 BEV01274 BEV01273 BEV01272 BEV01271 BEV01270 BEV01269 BEV01268 BEV01260 BEV01259 BEV01258 BEV01257 BEV01256 BEV01255 BEV01245 BEV01241 BEV01240 BEV01222 BBV06026 BBV06025 AFV02157 BAV04061 AJV02390 AJV01101 AGV01322 AGV01056 AFV02070 AFV01653 AEV01780 AEV01726 ADV06195 ADV02030 ADV01587 AEV01701 AJV01633 BAV04063 AHV01492 AHV01477 AFV01372 ADV01052 AJV02992 BDV01297 BDV01295 BDV01294 BDV01292 BDV01287 BDV01280 BDV01279 BDV01276 BDV01274 AKV04494 AKV04483 AKV04481 AKV04480 AKV04479 AHV01409 AKV04472 AJV02713 AKV02717 ADV06065 AHV01976 BEV01358 BEV01357 AKV04058 AKV04055 AKV01767 AKV01760 AKV01757 AKV01756 AKV01753 AKV01752 AKV01751 AKV01750 AKV01716 AKV01715 AKV01708 AKV01706 AKV01758 AGV02163 AGV01212 AGV01209 AGV01201 AGV01203 AFV02211 AFV01582 AFV01580 AFV01352 AKV04456 AKV04455 AHV01488 agv01509 AKV04454 AKV01697 AGV01493 AGV01104 BCV02331 BCV02327 BCV02320 AKV04117 AKV04115 BEV01321 BEV01318 BEV01315 BEV01314 BEV01311 BEV01310 BEV01309 BEV01305 BEV01303 BEV01301 BEV01300 BEV01299 BEV01298 BEV01296 BEV01295 BEV01294 BEV01293 BEV01292 BEV01291 BEV01284 BCV02747 BCV02681 BCV02678 BCV02677 BCV02669 BCV02529 BCV02519 BCV02328 BCV02321 BCV01678 AGV02025 AFV01454 AFV01419 AFV01336 BAV04700 AKV01165 AJV02961 AHV03116 AFV02187 ADV06096 ADV01808 ADV01689 BCV02861 BEV01267 BEV01263 BEV01238 BEV01232 BEV01230 BEV01228 BEV01225 BEV01221 BEV01220 BEV01219 BEV01218 BEV01215 BEV01214 BEV01213 BEV01210 BEV01209 BEV01207 BEV01206 BEV01201 BEV01200 BBV03128 BBV03125 BBV03124 AHV01899 BEV01470 BDV01991 BAV01281 AKV01587 AKV01585 AKV01582 ADV02050 BDV01804 BCV01903 BCV01899 BCV01897 BCV01753 BBV06044 BAV04797 BAV04796 AHV01917 ADV06365 BBV06035 ADV01872 ADV01696 ADV01025 AKV01438 AEV02580 ADV01575 BCV01315 BBV01601 BAV06476 BAV04732 AJV02146 AJV02135 AJV02131 AJV02116 AJV02065 AJV02037 AJV02117 BFV01007 BFV01006 BFV01004 BFV01001 BFV01000 BEV01978 AHV02249 AHV02202 BDV01863 BDV01847 BDV01845 BDV01844 BDV01838 BDV01837 BDV01806 BDV01801 BDV01799 BDV01585 BCV01390 BCV01388 AHV01466 AFV01557 AFV01554 AFV01482 AFV01383 AFV01280 AEV02172 AFV01462 AKV01302 AKV01273 AKV01269 AKV01268 AKV01267 AKV01266 AKV01264 AKV01261 AKV01260 AKV01258 AKV01255 AKV01252 AKV01250 AGV02679 AGV02677 AGV02559 BBV03127 AKV01783 AKV01778 ADV01031 BAV06457 AKV04276 BAV06459 AKV01855 AEV01621 BBV06099 BBV06069 BBV06043 BEV01466 BEV01329 BEV01328 BAV05138 AEV01528 BCV01139 BCV01137 BCV01135 BCV01134 BCV01133 BCV01132 BCV01131 BCV01130 BCV01129 BCV01128 BCV01127 BCV01117 BBV06117 BBV06113 BBV06095 AHV01637 AKV04348 AHV03316 AHV01926 AEV01577 AEV01638 AEV02667 BBV01421 BBV01408 BBV01400 BBV01393 BBV01381 BBV01163 AKV04453 AKV04450 AKV04449 AKV04446 AKV04445 AKV04443 AKV04442 AKV04441 AKV04440 AKV04439 AKV04438 AKV04437 AKV04434 AKV04432 AKV04431 AKV04430 AKV04429 AJV01545 AJV01519 AJV01492 AJV01483 AGV02078 AGV02000 AFV01003 AEV02693 AEV02670 AEV02644 AEV02513 AEV02505 AEV02477 AEV02472 AEV02432 AEV02423 AEV02422 AEV02419 AEV02656 ADV06102 ADV06081 AJV01536 ajv01521 AJV01533 ADV01570 AEV02613 AKV04436 AKV04452 AKV04451 AKV04444 AHV02082 AGV02700 AJV02631 AJV02629 AJV02633 BCV01399 BCV01398 BCV01396 BCV01393 BCV01392 BCV01389 BCV01967 BCV01884 AJVO2812 AHV02047 AGV01206 AFV01017 BEV01212 BEV01203 AEV01542 ADV06284 AKV04127 AJV02379 AEV01549 AEV01545 AEV01509 AEV01503 AEV01002 ADV06398 ADV01706 ADV01531 AHV01927 BDV01051 BCV02792 BCV02790 BCV02760 BBV03260 BBV03070 AGV02488 BBV01159 BBV01168 BBV06104 BAV05149 ADV05026 BDV01563 BDV01290 BDV01639 BDV01628 BDV01582 BDV01577 BDV01574 BDV01568 BBV03690 BDV01558 AGV02330 BCV02881 BCV02423 AFV01463 AFV01425 AKV01807 AKV01795 AKV01720 AGV01462 BCV01530 BCV01529 BCV01947 BCV01946 BCV01945 BCV01944 BCV01943 BCV01942 BCV01941 BCV01940 BCV01939 BCV01937 BCV01936 BCV01925 BCV01922 BCV01921 BCV01919 BCV01915 BCV01914 BCV01913 BCV01912 BCV01911 BCV01910 BCV01909 BCV01908 BCV01907 BCV01906 BCV01905 BCV01904 BCV01898 BCV01893 BCV01891 BCV01890 BCV01887 BCV01879 BCV01783 BCV01755 BCV01666 BCV01535 BCV01531 BCV01892 BAV06757 BAV06691 BAV06686 BAV06685 BAV06683 BAV06680 BAV06679 BAV06677 BAV06675 BAV06751 BAV06692 BAV06688 BAV06830 BAV06672 BAV06684 BAV06676 BAV06673 BAV06690 BAV06674 BAV06678 BAV06762 BAV06760 BAV06756 BAV06755 BAV06752 BAV06750 BAV06748 BAV06695 BAV06694 BAV06689 BAV06577 BAV06758 BAV06746 BAV06747 BAV06753 BAV06681 AKV02754 BCV01995 AHV01903 ADV03053 AGV01662 AFV01496 AFV01485 AEV02110 AEV02106 BAV04309 BAV04299 BAV04277 BAV04259 BAV04236 BAV04186 ADV03045 AKV04081 AKV04071 AKV04069 AKV04068 AKV04066 AHV01253 AHV01196 AHV01149 ADV06302 ADV06205 ADV06034 ADV02071 AKV04067 AKV04070 AJV02623 BAV04706 AHV01486 AFV02609 AFV02610 BBV03010 AFV01191 AEV01974 AEV01777 ADV01692 AFV02006 AKV02507 BCV02706 BCV02687 BCV02682 BCV02671 AKV04344 AKV04343 AKV04342 AKV04341 AKV04340 AKV04338 AKV04327 AKV04326 AKV04339 AJV02892 AJV01695 AJV02201 BEV01244 AJV01610 AEV02390 BEV01137 BEV01127 BEV01126 BEV01124 BEV01123 BEV01118 BEV01117 BEV01116 BEV01111 BCV02354 BCV02346 BCV02340 AGV02721 AGV01009 AGV01004 AEV02594 ADV02004 ADV02021 AFV02576 BAV04255 BAV04252 BBV06033 BBV06003 BAV06741 AJV01275 AHV02084 AJV04035 AGV01074 AFV02577 AFV02319 BAV05164 BAV01024 BDV01421 BDV01403 BDV01401 BBV01586 BBV01572 BAV04049 BCV02020 BCV02019 BCV02018 BCV02017 BCV02016 BCV02014 BCV02013 BCV02012 BCV02011 BCV02010 BCV02009 BCV02008 BCV02007 BCV02006 BCV02005 BCV02004 BAV05033 BAV05031 BAV04997 BAV04984 BBV01347 BBV06019 BBV06009 BBV06008 AGV01165 AFV02624 AFV02615 AFV02595 AFV02155 AGV01146 AFV02574 AFV02572 AFV02540 AFV02529 AFV02490 AFV02605 BDV01589 BDV01583 BDV01572 BDV01571 BDV01564 BDV01553 BDV01551 BDV01289 BDV01273 BDV01260 BCV02719 AJV02150 AJV02132 AJV02026 AJV02021 AJV02013 AJV02012 AJV02011 AJV01711 AJV02154 AKV01476 BDV01986 BDV01811 BDV01809 BEV01071 BBV03561 AGV02553 AJV02483 AJV02480 BCV01308 BCY01026 BCY01025 BEV01813 AJV01806 AJV01687 BCV02495 BCV02491 BCV02329 BCV02315 BCV02312 BCV01318 BCV01317 BCV01043 BBV3596 BBV03595 AFV02283 AEV01046 ADV06282 AGV01553 AEV02745 AEV01727 ADV01538 ADV01535 AGV01338 AGV01312 BAV06268 AGV01721 ADV05015 AFV01327 BCV01542 BCV01540 BCV01537 BCV01536 BCV01523 AKV02737 AKV02736 BCV01255 BEV01469 BAV04245 AKV01648 BAV04069 BCV01313 AHV01633 AHV01617 AKV01451 AKV01280 AKV01167 AKV01452 AKV04131 AKV04130 AKV04128 AKV04063 AKV04062 AKV04059 AKV04057 AKV04052 AKV04051 AKV04050 AKV04049 AKV04048 AKV04047 AKV04046 AKV04044 AKV04043 AKV04042 AKV04039 AKV04038 AKV04037 AKV04036 AKV04035 AKV04034 AKV04032 AKV04031 AKV04030 AKV01775 AHV01091 AJV02743 BAV01154 BAV01158 BAV01156 BAV01164 AJV04113 AFV01717 AEV02043 AEV01997 AEV01986 AEV01979 AEV01968 BCV01210 BCV01204 BCV01200 BCV01194 BCV01193 BCV01192 BCV01184 BCV01180 BCV01174 AGV02698 AGV01882 AGV01798 AFV01386 AEV02080 AEV02079 AFV01375 AGV02522 BBV03061 AJV02315 BBV03062 BEV01343 BEV01342 BEV01341 BEV01337 BEV01334 BEV01818 BEV01812 BAV06650 BAV06649 BAV06646 BAV06652 AKV01736 AKV01723 BAV04747 BAV04745 BAV04751 BAV04750 BAV04755 BAV04748 AHV01867 BAV04805 BAV04806 BAV04816 BAV04813 BAV04810 BAV04804 AKV01737 AKV01840 BAV04817 BEV01327 BEV01262 AKV01462 BDV01860 BDV01851 BDV01840 BCV02283 BCV02282 BAV04366 BAV04360 BAV04359 BAV04369 BAV04365 AEV01711 ADV03111 ADV01585 ADV03042 ADV01899 BAV06024 BAV06018 BAV06012 BAV05295 BAV05262 BAV05257 BAV05246 BAV05244 BAV06014 BAV06022 BAV06007 BAV06025 AFV02585 ADV02092 AJV01640 BBV06100 BAV06017 BAV05249 BAV05245 AKV04499 AKV04123 ADV06308 AJV02468 AFV02359 AEV02222 AEV02207 AEV02182 ADV06238 BEV02138 BEV02130 BEV02129 BEV02128 BEV02126 BEV02122 BEV02121 BEV02118 BEV02115 BEV02094 BEV01506 BEV01505 BEV01504 BEV01503 BEV01501 BEV01500 BEV01499 BEV01498 BEV01497 BEV01495 BEV01493 BEV01492 BEV01491 BEV01490 BEV01489 BEV01488 BEV01487 BEV01486 BEV01485 BEV01484 BEV01483 BEV01482 BEV01481 BEV01480 BEV01479 BEV01478 BEV01476 BEV01474 BEV01473 BEV01354 AJV02428 AGV02405 AKV01694 BAV06019 BAV06011 BAV06010 BAV06008 BAV05258 BAV05255 BAV05254 BAV05248 BAV05260 BAV06015 BAV05241 BAV06184 AKV01693 AKV01692 AKV01690 AKV01688 AKV01686 AKV01683 AKV01669 AKV01661 AKV01660 AGV02015 AKV04500 AJV02782 AJV02780 AJV02775 AJV02771 AJV02744 BDV01316 BCV01397 BBV03241 BBV03239 BBV03238 BBV03237 AEV01829 BAV06518 BAV05177 BCV01391 ADV04115 BAV01139 BEV01964 BDV01862 AEV02170 AEV02169 AEV02148 AEV02051 AEV01866 AJV02948 BDV01374 BBV01612 BAV06800 BAV06797 BAV06794 BAV06793 BAV06788 BAV06787 BAV06786 BAV06785 BAV06776 BAV06095 BAV06094 AJV04187 AHV03024 AEV02229 AEV02158 AEV02147 AEV02126 AEV02095 AEV01675 ADV06223 ADV06140 ADV02130 ADV02081 AEV02160 BDV01454 BBV03122 BCV02855 BCV02606 BCV02345 BCV02268 BCV02267 AJV02829 AKV01224 AKV01180 ADV02134 BDV01342 BDV01331 BDV01319 BDV01304 BDV01173 BDV01471 BDV01465 BDV01358 BDV01355 BDV01354 BDV01350 BDV01348 BDV01321 BAV06795 BAV06792 BAV06790 BAV06789 BAV06779 BAV06006 BAV05298 AEV02121 BAV06806 BAV06799 BAV05300 BEV01521 BEV01516 BEV01194 BEV01189 BEV01187 BEV01115 BEV01110 BDV01757 BDV01755 BDV01537 BDV01536 BDV01532 BDV01527 BDV01458 AJV02827 BBV01614 BAV06456 BAV04059 BAV04056 AKV01925 AKV01923 AKV01912 AKV01910 AKV01909 AKV01908 AKV01904 BEV02203 BEV02202 BEV02201 BEV02200 BEV02199 BEV02198 BEV02196 BEV02195 BEV02194 BEV02192 BEV02190 AJV01182 AJV02598 AHV03106 AHV03104 AFV02030 BDV01363 AKV01443 AKV01434 AKV01417 AJV02872 ADV01827 BAV01170 BAV01167 BAV01127 BDV01507 BDV01499 BDV01493 BDV01492 BDV01486 BDV01484 BDV01483 BDV01475 BDV01473 BDV01468 BDV01282 BCV02703 BCV02702 BCV02679 BAV01168 AJV02600 AEV02704 BBV06103 AJV01106 BCV01138 BCV01250 BAV05171 BAV04333 AKV04125 BAV04339 AJV02585 AJV02583 AJV02581 AJV02548 AJV01107 AHV01151 AHV01145 AJV04209 BBV03549 BBV03548 BBV03547 BBV03546 BBV03545 BBV03543 BBV03542 BBV03535 BBV03514 BBV03513 BBV03512 BBV03511 BBV03409 BBV03408 BBV03407 BBV03406 BBV03405 BBV03404 BBV03401 BBV03400 BBV03399 BBV03397 BBV03396 BBV03395 BBV03392 BBV03391 BBV03390 BBV03387 BBV03386 BBV03383 BBV03382 BBV03381 BBV03380 BBV03249 BBV03248 BBV03247 BDV02016 BDV02015 BDV02014 BDV02003 BDV01997 BDV01985 BDV01979 BDV01972 BDV01875 BDV01813 BBV03240 BBV03130 ADV01645 BCV02879 BCV02876 BCV02859 BAV06693 BAV06464 AKV04484 AKV04482 BAV06772 BAV06771 BAV06040 BAV05159 BAV05158 BAV05157 BAV05154 BAV05152 BAV05151 BAV05148 BAV05146 BAV05145 BAV05144 BAV05143 BAV05142 BAV05141 BAV05140 BAV05137 BAV05136 BAV05135 BAV05134 BAV05133 BAV05132 BAV05130 AKV02195 AKV02194 AKV02178 AKV02169 AKV02135 AHV03331 AHV03327 AHV03302 BAV05147 BBV01007 AJV01502 BAV04152 BAV04151 BAV04150 BAV04144 BAV04140 BAV04131 BAV04128 BBV03387 BAV04838 BAV04811 BAV04145 AHV02016 AEV01765 AEV01753 ADV04100 ADV04008 BAV06763 BAV06761 BAV06759 ADV06327 ADV06309 ADV06190 BAV06773 AGV02291 AEV01703 BBV01026 BBV01024 AKV01780 AKV01373 AKV01385 BDV01610 BDV01597 BDV01505 BDV01497 BDV01298 BDV01271 BDV01270 BDV01266 BDV01265 BDV01250 BCV02707 BCV02680 BCV02662 BCV02650 BCV01261 BCV01260 BCV01259 BCV01258 BCV01254 BCV01252 BCV01251 BCV01149 BCV01148 BCV01147 BCV01146 BCV01145 BCV01144 BCV01143 BCV01142 BCV01141 BCV01140 BCV01136 BCV01111 BBV06111 BBV06074 BBV01059 BBV01058 BBV01057 BBV01055 BBV01054 BBV01003 BBV01056 BAV05168 BAV05166 BAV05165 BAV04340 BAV04338 BAV04337 BAV04335 BAV04275 AJV02490 AJV02486 AHV01166 AGV02304 AGV02226 AGV02193 AGV02081 AGV02217 AEV02266 AEV02214 AEV01831 BBV01053 BAV04721 AEV01822 AHV02140 BDV01356 BDV01345 BDV01312 BEV01464 BAV06197 AJV02391 AJV01693 AJV01690 AJV01569 AJV01207 AKV04346 BCV01709 BCV01621 BCV01618 BCV01615 BCV01610 BCV01604 AKV02547 BAV05100 BAV04825 BAV04798 AKV01474 AKV01473 AKV01467 BAV04814 AKV01455 AKV01454 AKV01458 AKV01427 AKV01318 AKV01314 AKV01967 AKV01460 AKV01461 AKV01464 AKV01414 AKV01457 AKV01344 AKV01292 BCV02870 BCV02868 BCV02866 BCV02864 BCV02857 BDV01777 BDV01764 BDV01533 BDV01530 AKV02721 AKV02720 AKV02706 AKV02705 ADV01522 AKV01300 BDV01977 BDV01814 BEV01109 BEV01106 BEV01105 BEV01102 BEV01100 BDV01861 BDV01858 BDV01857 BDV01854 BDV01513 BDV01511 BDV01496 BDV01464 BDV01455 BDV01389 BDV01386 BDV01352 BDV01346 BDV01332 BDV01498 BAV01108 BAV01109 BAV01106 BAV01107 BBV06116 BBV06119 AHV01395 AHV01116 AHV01018 AHV01002 BAV06475 ADV03113 BDV01414 BDV01404 BBV03138 AKV01930 AGV01809 BBV06142 BBV06141 BBV06140 BBV06139 BBV06137 BBV06136 BBV06135 BBV06134 BBV06133 BBV06122 BBV06121 BBV06118 BBV06115 BBV06114 BBV06110 BBV06087 BBV01049 BBV01022 BBV01021 BAV06728 BAV06715 BAV06714 BAV06711 BAV06707 BAV06705 BAV06704 BAV06699 BAV06698 BAV06697 BAV06696 AHV01482 AHV01404 AHV01394 AHV01357 AFV01523 AFV01002 BAV06710 ADV06406 ADV04031 ADV03055 ADV05003 BBV06138 BAV06701 AJV02501 AJV02179 AJV02164 AJV02159 BCV01690 BCV01626 BBV03591 BAV05119 BDV01654 BDV01652 BDV01651 BDV01447 BDV01441 BDV01439 AKV01489 AEV02030 AEV01948 AEV01714 AEV01682 AEV01610 AEV02187 BDV01988 BDV01596 AKV02597 BEV02167 BEV02164 BEV02158 BEV02156 BEV02152 BEV02151 BEV02002 BEV01496 BEV01351 BEV01350 BDV01216 BDV01213 BDV01204 BCV1009 BCV02251 BBV03246 BBV03245 BBV03146 BBV03123 BBV03001 BBV01635 BBV01634 BBV01630 BAV05233 BAV04812 AKV04074 AKV04073 AKV01670 AKV01668 AKV01662 AKV01295 AJV01742 AJV01135 AJV01104 AJV01064 AHV01308 AGV01618 AGV01535 AGV01532 AGV01464 AGV01410 BBV03002 BAV06429 AGV01569 AGV01463 BDV01555 BDV01615 BDV01595 BDV01584 BDV01323 BDV01322 BCV02630 BBV03544 BBV03534 BBV03533 BBV03532 BBV03531 BAV04593 AHV01910 AFV02473 AFV02332 AHV03304 BAV04592 BCV02305 BAV01115 AGV01871 AKV01931 AJV02769 AJV02735 AJV02734 AHV03158 AGV01968 AGV01873 AGV01870 AFV01499 AFV01471 AFV01468 AFV01467 AFV01441 AFV01430 AFV01395 ADV02154 AFV01435 AJV01700 AEV02747 AEV02617 AJV01248 BDV02009 BDV02008 BDV02000 BDV01999 AJV01196 AJV01222 AHV01060 AGV02662 AJV01197 aev01792 BEV01250 BCV01003 BCV01000 BAV06660 AKV01525 AJV02915 AJV01151 AEV01536 AJV02815 BCV01001 BBV03835 BBV01205 AGV01526 AGV01524 AKV01698 AKV01710 AKV01709 AKV01699 BDV01642 AEV01586 BEV02140 BBV06106 AJV02376 AGV02282 AFV01217 aev01921 BAV05098 BAV05094 BAV05099 AKV01530 AGV02089 ADV06381 AFV02500 ADV01553 BEV01764 BEV01760 AJV03041 AJV03026 AJV03007 AJV02989 AJV02907 AJV02895 AJV02666 AGV01846 AEV01764 BCV02163 BCV02161 BCV02843 BCV02842 BBV03129 BBV03108 AKV01223 AKV01174 AKV01533 AKV01528 AKV01527 AKV01523 AKV01522 AKV01520 BBV01280 AGV02564 AFV02061 AEV02092 AEV01619 AEV01617 BBV01204 BEV01266 BEV01265 BEV01261 BEV01253 BEV01252 BEV01249 BEV01248 BEV01247 BEV01246 BEV01243 BEV01239 BEV01237 BEV01236 BEV01234 BEV01229 BEV01223 BEV01216 BEV01204 BDV01768 BDV01706 BDV01641 BDV01638 BDV01637 BDV01636 BDV01634 BDV01630 BDV01617 BDV01580 BCV02025 BCV02024 BCV02023 BCV02001 BCV02000 BCV01998 BCV01270 BCV01268 BCV01267 BBV01011 BAV06048 BAV06433 BAV06426 BAV06052 BAV06051 BAV06049 AKV01771 AKV01743 AKV01218 AKV01217 AKV01216 AKV01214 AKV01194 AKV01163 AKV01158 AKV01149 AKV01144 AHV03310 AHV03213 AKV01215 BAV06428 BEV01165 BEV01162 BEV01160 BEV01145 BEV01119 BCV02608 BCV02607 BCV02604 BCV02598 BCV02590 BCV02567 BBV03712 BBV03711 BBV03710 BBV03659 BBV03658 BBV03657 BBV03656 BBV03655 BBV03653 BBV03652 BBV03649 BBV03648 BBV03647 BBV03646 BBV03645 BBV03644 BBV03643 BBV03642 BBV03641 BBV03640 BBV03639 BBV03638 BBV03637 BBV03636 BBV03635 BBV03634 BBV03633 BBV03632 BBV03631 BBV03630 BBV03629 BBV03628 BBV03627 BBV03555 BBV03366 BBV03361 BBV03357 BBV03356 BBV03354 BBV03353 BBV03347 BBV01271 BBV01270 BBV01269 BBV01268 BBV01267 BBV01266 BBV01265 BBV01264 BBV01263 BBV01262 BAV04803 BAV04802 BAV04801 BAV04799 BAV04626 BAV04625 BAV04624 BAV04623 BAV04621 BAV04620 BAV04619 AKV01944 AKV01920 AKV01362 AKV01358 AKV01355 AKV01351 AKV01348 AKV01345 AKV01336 AKV01334 AKV01333 AKV01329 AKV01328 AKV01326 AKV01325 AKV01324 AKV01323 AKV01313 AKV01312 AKV01311 AKV01310 AKV01309 AKV01308 AKV01307 AKV01306 AKV01305 AKV01304 AKV01303 AKV01272 AKV01257 AJV02342 AGV01024 AJV02325 BAV04629 AKV01327 BAV04800 AJV02326 BBV03654 BAV04618 BAV04617 AKV01347 AKV01350 AKV01339 AKV01338 AKV01337 BBV06102 AJV01605 BDV01644 BAV01268 AJV01692 BDV02012 BDV02007 BDV01983 BDV01978 BDV01976 BDV01649 AHV01161 AFV02497 AFV01981 AFV01907 AFV01893 AFV01870 AFV01835 AFV01819 BDV01812 BBV01276 BCV02193 BCV02185 BCV01834 BCV01833 BCV01265 BCV01264 AKV01789 AKV01787 AKV01786 AKV01784 AKV01782 AKV01779 AHV02014 AGV01812 AGV01782 AGV01764 AGV01757 AGV01780 BAV06512 AKV04312 BEV01807 BEV01802 BCV01513 BAV04224 BAV04051 BAV04050 BAV04052 BAV04053 ADV05172 AHV01082 AHV01013 AHV01009 AGV02717 AJV01674 AJV01656 AJV01682 BCV01849 BAV06513 BAV06511 AKV04313 AFV02053 AJV02434 BCV01517 AFV01267 AFV01284 AFV01275 AFV01271 AFV01262 AFV01241 AEV02351 AHV01296 AHV03336 AHV01123 AHV01061 AHV01006 AGV02708 BAV04268 AHV03271 BAV05170 BAV05163 AKV01079 AKV01062 AJV01325 AFV02530 AFV02083 AEV02332 AEV02288 AFV02329 BDV01296 BDV01272 BDV01269 BDV01268 BDV01267 AFV02260 BAV04702 BAV04467 AJV02957 AJV02880 AJV01038 AFV02416 AFV02264 AFV02201 AEV02387 BEV01477 BEV01465 BEV01254 BEV01251 BEV01242 BEV01231 BEV01655 BEV01652 BEV01650 BEV01649 BEV01645 BEV01177 BDV01850 BDV01788 BDV01360 BDV01343 BCV01583 BCV01582 BCV01581 BCV01580 BCV01579 BCV01577 BCV01574 BCV01566 BCV01506 BCV01395 BCV01386 BAV04783 AKV01792 AKV01790 AKV01733 AKV01729 AJV04075 AJV04071 AJV04065 AHV01036 AGV02400 AGV02324 BAV06534 BAV06626 BAV06535 BAV06769 BAV06765 BAV06768 BAV06767 AKV01419 AKV01343 AKV01331 AKV01330 BBV03216 BBV03215 BEV01752 BEV01747 BEV01679 BEV01340 BEV01339 BEV01338 BEV01336 BEV01335 BEV01332 BEV01331 BEV01330 BEV01302 BEV01297 AFV02357 BAV06042 AKV04322 AGV01020 AKV04323 AKV04321 BCV02309 BCV02308 BCV02295 BCV02292 BCV02286 BCV01314 BCV01311 BCV01310 BCV01302 BCV01049 BCV01032 BCV01026 BCV01018 BCV01017 BBV03811 BBV03707 BBV03705 BBV03702 BBV03700 BBV03502 BBV03501 BBV03500 BBV03491 BBV03490 BBV03489 BBV03488 BBV03487 BBV03482 BBV03473 AKV04497 AKV04496 AKV04495 AKV04493 AKV01973 AJV01621 AJV01618 AHV03262 AHV03248 AHV03245 BCV01030 BAV06659 AEV01597 BAV06658 BAV06656 BAV06653 BAV06655 BAV06754 BCV01781 BCV01780 BCV01779 BCV01764 BCV01778 BCV01949 BCV01948 BBV01513 BBV01512 BBV01488 BBV01485 BBV01327 BBV01509 AKV01058 AHV02064 AHV02040 AHV01849 AHV01837 BEV01502 BDV01834 BDV01829 BDV01828 BDV01794 BCV02609 BCV02602 BCV02592 BBV03236 BBV03234 BBV03233 BBV03230 BCV02391 BCV02376 BCV02367 BCV02362 BCV01744 BCV01695 BCV01687 BCV01686 BCV01643 BCV01629 BCV01082 BDV01643 AJV02374 AJV01115 BDV02004 BBV06042 BBV06058 BAV04336 BAV04315 BAV04314 BAV04311 BAV04307 BAV04303 BAV04298 BAV04297 BAV04296 BAV04284 BDV01989 BDV01973 BDV01817 BCV02716 BCV02688 BCV02686 BCV01698 BCV01697 BCV01688 BBV03594 BBV03593 BBV03592 BBV03141 BBV01610 BAV06766 AKV02227 AKV01086 AJV04222 AHV02124 AHV01137 AGV
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Carefusion 211 Inc dba Carefusion
- Reason for Recall:
- A pressure transducer failure can develop, activating a false Extended High Ppeak or Circuit Occlusion Alarm. The Safety Valve will open to ambient air and the unit will stop ventilating, allowing spontaneous breathing patients to breathe. The malfunction may delay initiation or cease ventilation. Nonbreathing patients will need manual ventilation or to be connected to another ventilator.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CareFusion AVEA ventilator, all models. Product Usage: A continuous ventilator (respirator), intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
Product Codes/Lot Numbers:
Serial No. AKV01874 AGV01443 AEV02197 AKV02580 AEV02375 BBV01304 BBV01303 AJV02510 AJV02502 AGV02650 AFV01802 AFV01792 AFV01400 AFV01361 AEV01622 AEV01608 AKV01915 BBV03730 BBV03557 BBV03556 BBV03554 BBV03553 BBV 3558 BBV01406 BBV01396 AHV01880 AGV02205 BBV01432 AFV01657 AFV01655 AFV01647 BAV05234 AHV03069 AGV02249 BBV03797 BAV06514 BDV01011 BDV01000 BCV01938 BAV04011 BAV04010 BAV04009 BAV04005 BAV04004 BAV04003 BAV04002 BAV01313 AKV04053 AJV02289 BAV04021 AEV02264 AEV02225 AEV02205 ADV06091 AKV02569 AHV01206 AKV04351 AKV04349 AKV02550 AGV01385 AHV03128 ADV04116 BBV06059 BBV06057 BAV05153 AFV01492 BBV01620 BDV01371 BAV06447 BAV06446 BAV06445 AEV01724 AEV01712 BAV06463 AEV01694 AJV02446 AGV02139 BAV04188 BAV04187 AKV01153 AKV01147 AEV02720 AEV02717 AEV02713 BAV04191 BAV04993 AHV02129 ADV06297 ADV01637 ADV01592 AFV01922 AFV02277 AHV01274 BAV05235 BCV02186 BCV01822 BBV06061 BDV01249 BDV01242 ADV03032 AJV01685 AHV01967 AHV01722 AHV01701 AKV02230 AKV02226 AKV02224 AKV02219 AKV02217 AKV02215 AKV02214 BEV01326 BEV01325 BEV01324 BEV01306 BEV01205 BDV02001 BDV01994 BDV01810 AKV01279 AKV04320 BAV06124 BBV03028 BBV03027 BBV01483 BBV01019 BBV01018 BBV01016 BAV01015 BAV01020 BBV01010 AKV01029 AKV01012 AKV01007 AKV01000 BCV02865 BAV01088 BAV01026 BAV01018 AKV01018 AKV01017 AHV03287 AHV03061 AFV02243 AFV02230 ADV01638 AGV01808 BFY01003 AGV02074 AFV01176 AKV04033 AEV01588 BAV01089 AKV01009 AKV01005 AFV02205 BBV01005 ACV01718 AKV01028 BFV01053 BEV02183 BAV01099 ADV06397 ADV03003 BAV04113 BAV04111 BEV02185 BEV02184 BDV01816 ajv01601 AGV01333 AGV01278 AFV01962 AHV01326 AJV02876 AEV02362 AEV02359 AEV02340 AFV01616 AFV01744 AFV01661 AFV01633 AFV01632 AFV01631 AFV01629 AFV01625 AFV01545 AFV01481 AFV01449 BEV01356 BEV01352 BDV02093 BCV01220 BCV01206 AGV02366 BBV06021 BBV06016 BBV06007 BBV06006 AEV02120 AKV01274 AKV01240 BCV01810 BCV01619 BCV01176 BCV01072 BCV01059 BCV01057 BBV03951 BBV03598 BBV03570 BAV04368 AKV01068 AKV01047 BAV04362 BDV01515 BDV01480 BEV01663 BEV01648 BCV02271 BCV02270 AJV02640 AHV01549 AEV02727 BBV01207 BBV01206 AKV01466 AKV01442 AKV01367 AKV01046 AKV01003 AJV02681 BAV04017 BAV04016 BAV04015 BAV04014 BAV01306 AKV04488 AGV02552 AGV02441 AGV02432 AJV01532 AJV01476 AJV01424 BBV03552 AJV02441 AJV02950 AJV02504 AJV02467 AJV02432 AJV02427 AJV04226 AJV04195 AJV04219 AJV04225 AJV04192 AJV04228 AJV04232 AJV04223 AJV04229 AJV04201 AJV04205 AJV04203 AJV04202 AJV04221 AJV04189 AJV04176 AJV04046 AJV04233 AJV04224 AKV04118 AHV01285 AJV04196 BEV01544 BEV01543 BEV01542 BEV01541 BEV01540 BEV01539 BEV01534 BEV01531 BEV01522 BEV01519 BEV01515 AJV01591 AKV02755 ADV06008 BDV02092 BDV02087 BDV02022 BDV02021 BDV02002 BDV01980 BDV01668 AFV02626 AKV04119 AFV01567 BCV02763 BCV02757 AFV02102 BBV01459 BDV01474 BDV01461 BCV01732 BCV01729 BCV01722 BAV04354 ADV01559 BBV01300 BBV01292 BBV01278 BDV01818 BDV01646 BBV01017 BEV02141 BEV02137 BEV02136 BEV02135 BEV02134 BEV02133 BEV02131 BEV02127 BEV02123 BEV02113 BDV02011 BDV02006 BDV01995 BDV01982 BDV01971 AKV04324 AJV01076 AKV04325 BCV02209 BCV02208 BCV02205 AJV02216 AJV02173 AFV02439 AFV02368 AFV02299 AFV02281 AFV02183 AFV02182 AFV02170 AFV02129 AFV02000 AFV01688 AFV01668 AFV01296 AGV01584 AGV01527 BAV01218 BAV01217 BAV06470 BEV01359 BCV02187 BCV01831 BCV01830 AJV01531 AJV01524 AHV01705 AGV01413 AFV02402 AFV01572 AFV01546 AFV01250 AFV01247 AFV01165 AFV01106 AEV02698 AGV01762 BDV01385 BDV01375 AHV01434 BEV01140 BEV01135 BCV01708 BCV01705 BCV01704 BCV01703 BCV01702 BCV01701 BBV01480 BBV01472 BBV01471 BBV01470 BBV01468 BBV01465 BBV01462 BAV05231 BAV05230 BAV05229 BAV05228 BEV01131 AJV02530 AJV02529 AJV02505 AGV02710 AKV04061 AKV04060 AKV01478 AHV01745 ADV01758 AGV01287 AKV01475 AKV01022 AJV03054 AJV03053 AJV01139 AHV03369 AHV01916 AGV02297 AFV02424 ADV04002 ADV02151 AGV01352 AKV04114 AKV04116 BDV01830 BDV01745 BAV05103 BAV05102 BAV05101 AKV04163 AKV04129 AKV01926 AKV01906 AKV04104 AKV01402 BAV05001 BAV05002 BBV01176 AJV02144 AJV02130 AJV02121 AHV01778 AHV01748 BAV01041 BCV01785 BCV01734 BBV03213 BBV03206 BBV03112 AKV01964 AKV01959 AKV01957 AGV01497 BDV01264 BFV01058 BEV01865 BDV01285 BEV01235 BEV01233 BEV01226 BEV01224 BEV01217 BEV01208 BEV01202 BDV01974 BDV01970 BDV01357 BDV01349 BDV01347 BDV01338 BDV01336 BDV01333 BDV01318 BDV01317 BDV01181 BDV01175 BDV01104 BCV02335 BAV05008 BAV05007 BAV05006 BAV05005 BAV05004 BAV05003 BAV05000 BAV04999 AGV02364 BBV01409 BBV01380 BBV01379 BBV01357 BBV01197 BBV01181 BBV01173 BBV01167 BBV01164 BAV04256 BAV04254 BCV02306 BCV02304 BCV02303 AEV02657 AJV01699 AGV01964 BBV03145 AHV01184 AGV01160 BEV01264 BEV01227 BEV01028 BEV01017 AGV01986 AGV01259 AGV01255 AGV01242 AGV01241 AGV01156 AGV01138 AGV01037 AFV02619 AFV02616 BBV01213 BBV01203 BBV01201 BBV01000 AHV01575 AHV01591 AGV02223 AGV01237 AGV01222 AGV01198 BCV02262 BCV02261 BCV02260 BCV02259 BCV02258 BCV02257 BCV02256 BCV02254 BCV02253 BCV02252 BCV02249 BCV02248 BCV02247 BCV02246 BCV02245 BCV02244 BCV02243 BCV02241 BCV02239 BCV02238 BCV02237 BCV02236 BCV02235 BCV02234 BCV02231 BCV02222 BCV02220 BBV03803 BBV03796 AFV01043 BAV01150 AJV02703 AJV01645 AHV01370 AGV01811 AFV01503 AEV02548 AEV02551 BDV01601 BCV02633 BBV03599 AGV02040 AGV01257 BCV02240 AKV02765 AKV02763 AKV02762 AKV02761 AKV02760 AKV02759 AKV02758 AKV02757 AKV02756 AKV02225 AKV01893 AJV02967 AHV01636 AHV01621 AHV01585 AHV01526 AHV01648 AGV02704 BEV01182 BEV01181 BEV01180 BEV01179 BEV01178 BEV01176 BEV01175 BEV01173 BEV01172 BEV01169 BEV01168 BEV01150 BDV01308 BDV01307 BDV01306 BDV01303 BDV01302 BDV01301 BDV01147 BCV02338 BCV02266 BCV02265 BCV02264 BCV02263 BCV01748 BAV04434 BAV04432 BAV04431 BAV04430 BAV04429 BAV04428 BAV04427 AKV04029 AKV04028 AKV04026 AKV04025 AKV04024 AKV04022 AKV04020 AKV04019 AKV04018 AKV04017 AKV04016 AKV04015 AKV04014 AKV04013 AKV04010 AKV04009 AKV04008 AKV04007 AKV04006 AKV04005 AKV04003 AHV01620 AHV01615 AGV02580 AGV02556 AGV02545 AGV02481 AGV01940 BAV04433 AGV02590 AKV04023 AKV04027 AKV04021 AKV04001 AKV04011 AKV04004 BDV01645 BDV01559 BAV04772 BAV04768 BFV01055 BFV01054 BFV01051 BBV01223 BAV06784 BAV06780 BAV06777 BAV06796 BAV06782 BAV06783 BAV06781 BAV06778 BAV06775 BBV03005 BBV03000 BAV04489 BAV04488 BAV04487 BAV04486 BAV04485 BAV04481 AJV02381 BCV01312 BCV01262 BDV01684 BCV01738 BCV02176 BCV02174 AKV04316 AKV04315 AKV04314 AKV02719 AKV02718 AKV02201 AKV02175 AKV01616 AEV01908 BCY01036 BBV06068 AJV01609 AHV01343 AGV01327 AEV02473 AEV01808 AGV01318 AEV02648 AJV02165 AJV01665 AFV01117 AEV02712 AEV02700 AEV02581 AEV02380 BDV01263 BDV01259 AKV04120 AGV02357 AEV01520 AGV01855 AGV01852 AGV01851 AGV01850 AGV01849 ADV06048 ADV05035 ADV05033 ADV05031 AGV01857 BCV02063 AKV04318 AKV04317 BCV01930 BCV01924 BCV01920 BCV01917 BCV01902 BCV01895 BCV01894 BCV01888 BCV01886 BCV01885 BCV01784 BCV01782 BCV01775 BCV01770 BCV01766 BCV01762 BCV01760 BCV01758 BCV01757 BCV01756 BCV01754 BCV01751 BCV01683 BCV01663 BCV01657 AKV01619 AKV01606 AKV01604 AKV01602 AKV01500 AJV02688 AJV02473 AJV02425 BAV04112 BAV04094 BCV02558 AJV01073 AJV01072 AJV01067 AJV01065 AHV03215 AJV01625 BDV01981 BDV01848 BDV01842 BDV01841 BDV01836 BDV01824 BDV01822 BDV01820 BDV01815 BDV01748 BCV01771 BCV01763 BCV01710 BCV01699 BCV01696 BCV01689 BAV04770 AKV04460 AKV04458 BAV04777 BAV04771 BAV04769 BAV04776 AKV04459 AKV04462 AKV04461 AKV04457 BAV04792 BAV04790 BAV04789 BAV04786 AKV02753 AKV02752 AKV02581 AKV02751 BDV01029 BCV02805 AKV01614 AKV01591 BEV01304 BEV01195 BEV01193 BEV01192 BEV01191 BEV01190 BEV01188 BEV01186 BEV01185 BEV01184 BEV01183 BEV01171 BEV02119 BEV02116 BEV02114 BEV02110 BEV02099 BEV02090 BEV01879 BEV01857 BEV01290 BEV01288 BEV01287 BEV01285 BEV01282 BEV01281 BEV01280 BEV01278 BEV01277 BEV01275 BEV01274 BEV01273 BEV01272 BEV01271 BEV01270 BEV01269 BEV01268 BEV01260 BEV01259 BEV01258 BEV01257 BEV01256 BEV01255 BEV01245 BEV01241 BEV01240 BEV01222 BBV06026 BBV06025 AFV02157 BAV04061 AJV02390 AJV01101 AGV01322 AGV01056 AFV02070 AFV01653 AEV01780 AEV01726 ADV06195 ADV02030 ADV01587 AEV01701 AJV01633 BAV04063 AHV01492 AHV01477 AFV01372 ADV01052 AJV02992 BDV01297 BDV01295 BDV01294 BDV01292 BDV01287 BDV01280 BDV01279 BDV01276 BDV01274 AKV04494 AKV04483 AKV04481 AKV04480 AKV04479 AHV01409 AKV04472 AJV02713 AKV02717 ADV06065 AHV01976 BEV01358 BEV01357 AKV04058 AKV04055 AKV01767 AKV01760 AKV01757 AKV01756 AKV01753 AKV01752 AKV01751 AKV01750 AKV01716 AKV01715 AKV01708 AKV01706 AKV01758 AGV02163 AGV01212 AGV01209 AGV01201 AGV01203 AFV02211 AFV01582 AFV01580 AFV01352 AKV04456 AKV04455 AHV01488 agv01509 AKV04454 AKV01697 AGV01493 AGV01104 BCV02331 BCV02327 BCV02320 AKV04117 AKV04115 BEV01321 BEV01318 BEV01315 BEV01314 BEV01311 BEV01310 BEV01309 BEV01305 BEV01303 BEV01301 BEV01300 BEV01299 BEV01298 BEV01296 BEV01295 BEV01294 BEV01293 BEV01292 BEV01291 BEV01284 BCV02747 BCV02681 BCV02678 BCV02677 BCV02669 BCV02529 BCV02519 BCV02328 BCV02321 BCV01678 AGV02025 AFV01454 AFV01419 AFV01336 BAV04700 AKV01165 AJV02961 AHV03116 AFV02187 ADV06096 ADV01808 ADV01689 BCV02861 BEV01267 BEV01263 BEV01238 BEV01232 BEV01230 BEV01228 BEV01225 BEV01221 BEV01220 BEV01219 BEV01218 BEV01215 BEV01214 BEV01213 BEV01210 BEV01209 BEV01207 BEV01206 BEV01201 BEV01200 BBV03128 BBV03125 BBV03124 AHV01899 BEV01470 BDV01991 BAV01281 AKV01587 AKV01585 AKV01582 ADV02050 BDV01804 BCV01903 BCV01899 BCV01897 BCV01753 BBV06044 BAV04797 BAV04796 AHV01917 ADV06365 BBV06035 ADV01872 ADV01696 ADV01025 AKV01438 AEV02580 ADV01575 BCV01315 BBV01601 BAV06476 BAV04732 AJV02146 AJV02135 AJV02131 AJV02116 AJV02065 AJV02037 AJV02117 BFV01007 BFV01006 BFV01004 BFV01001 BFV01000 BEV01978 AHV02249 AHV02202 BDV01863 BDV01847 BDV01845 BDV01844 BDV01838 BDV01837 BDV01806 BDV01801 BDV01799 BDV01585 BCV01390 BCV01388 AHV01466 AFV01557 AFV01554 AFV01482 AFV01383 AFV01280 AEV02172 AFV01462 AKV01302 AKV01273 AKV01269 AKV01268 AKV01267 AKV01266 AKV01264 AKV01261 AKV01260 AKV01258 AKV01255 AKV01252 AKV01250 AGV02679 AGV02677 AGV02559 BBV03127 AKV01783 AKV01778 ADV01031 BAV06457 AKV04276 BAV06459 AKV01855 AEV01621 BBV06099 BBV06069 BBV06043 BEV01466 BEV01329 BEV01328 BAV05138 AEV01528 BCV01139 BCV01137 BCV01135 BCV01134 BCV01133 BCV01132 BCV01131 BCV01130 BCV01129 BCV01128 BCV01127 BCV01117 BBV06117 BBV06113 BBV06095 AHV01637 AKV04348 AHV03316 AHV01926 AEV01577 AEV01638 AEV02667 BBV01421 BBV01408 BBV01400 BBV01393 BBV01381 BBV01163 AKV04453 AKV04450 AKV04449 AKV04446 AKV04445 AKV04443 AKV04442 AKV04441 AKV04440 AKV04439 AKV04438 AKV04437 AKV04434 AKV04432 AKV04431 AKV04430 AKV04429 AJV01545 AJV01519 AJV01492 AJV01483 AGV02078 AGV02000 AFV01003 AEV02693 AEV02670 AEV02644 AEV02513 AEV02505 AEV02477 AEV02472 AEV02432 AEV02423 AEV02422 AEV02419 AEV02656 ADV06102 ADV06081 AJV01536 ajv01521 AJV01533 ADV01570 AEV02613 AKV04436 AKV04452 AKV04451 AKV04444 AHV02082 AGV02700 AJV02631 AJV02629 AJV02633 BCV01399 BCV01398 BCV01396 BCV01393 BCV01392 BCV01389 BCV01967 BCV01884 AJVO2812 AHV02047 AGV01206 AFV01017 BEV01212 BEV01203 AEV01542 ADV06284 AKV04127 AJV02379 AEV01549 AEV01545 AEV01509 AEV01503 AEV01002 ADV06398 ADV01706 ADV01531 AHV01927 BDV01051 BCV02792 BCV02790 BCV02760 BBV03260 BBV03070 AGV02488 BBV01159 BBV01168 BBV06104 BAV05149 ADV05026 BDV01563 BDV01290 BDV01639 BDV01628 BDV01582 BDV01577 BDV01574 BDV01568 BBV03690 BDV01558 AGV02330 BCV02881 BCV02423 AFV01463 AFV01425 AKV01807 AKV01795 AKV01720 AGV01462 BCV01530 BCV01529 BCV01947 BCV01946 BCV01945 BCV01944 BCV01943 BCV01942 BCV01941 BCV01940 BCV01939 BCV01937 BCV01936 BCV01925 BCV01922 BCV01921 BCV01919 BCV01915 BCV01914 BCV01913 BCV01912 BCV01911 BCV01910 BCV01909 BCV01908 BCV01907 BCV01906 BCV01905 BCV01904 BCV01898 BCV01893 BCV01891 BCV01890 BCV01887 BCV01879 BCV01783 BCV01755 BCV01666 BCV01535 BCV01531 BCV01892 BAV06757 BAV06691 BAV06686 BAV06685 BAV06683 BAV06680 BAV06679 BAV06677 BAV06675 BAV06751 BAV06692 BAV06688 BAV06830 BAV06672 BAV06684 BAV06676 BAV06673 BAV06690 BAV06674 BAV06678 BAV06762 BAV06760 BAV06756 BAV06755 BAV06752 BAV06750 BAV06748 BAV06695 BAV06694 BAV06689 BAV06577 BAV06758 BAV06746 BAV06747 BAV06753 BAV06681 AKV02754 BCV01995 AHV01903 ADV03053 AGV01662 AFV01496 AFV01485 AEV02110 AEV02106 BAV04309 BAV04299 BAV04277 BAV04259 BAV04236 BAV04186 ADV03045 AKV04081 AKV04071 AKV04069 AKV04068 AKV04066 AHV01253 AHV01196 AHV01149 ADV06302 ADV06205 ADV06034 ADV02071 AKV04067 AKV04070 AJV02623 BAV04706 AHV01486 AFV02609 AFV02610 BBV03010 AFV01191 AEV01974 AEV01777 ADV01692 AFV02006 AKV02507 BCV02706 BCV02687 BCV02682 BCV02671 AKV04344 AKV04343 AKV04342 AKV04341 AKV04340 AKV04338 AKV04327 AKV04326 AKV04339 AJV02892 AJV01695 AJV02201 BEV01244 AJV01610 AEV02390 BEV01137 BEV01127 BEV01126 BEV01124 BEV01123 BEV01118 BEV01117 BEV01116 BEV01111 BCV02354 BCV02346 BCV02340 AGV02721 AGV01009 AGV01004 AEV02594 ADV02004 ADV02021 AFV02576 BAV04255 BAV04252 BBV06033 BBV06003 BAV06741 AJV01275 AHV02084 AJV04035 AGV01074 AFV02577 AFV02319 BAV05164 BAV01024 BDV01421 BDV01403 BDV01401 BBV01586 BBV01572 BAV04049 BCV02020 BCV02019 BCV02018 BCV02017 BCV02016 BCV02014 BCV02013 BCV02012 BCV02011 BCV02010 BCV02009 BCV02008 BCV02007 BCV02006 BCV02005 BCV02004 BAV05033 BAV05031 BAV04997 BAV04984 BBV01347 BBV06019 BBV06009 BBV06008 AGV01165 AFV02624 AFV02615 AFV02595 AFV02155 AGV01146 AFV02574 AFV02572 AFV02540 AFV02529 AFV02490 AFV02605 BDV01589 BDV01583 BDV01572 BDV01571 BDV01564 BDV01553 BDV01551 BDV01289 BDV01273 BDV01260 BCV02719 AJV02150 AJV02132 AJV02026 AJV02021 AJV02013 AJV02012 AJV02011 AJV01711 AJV02154 AKV01476 BDV01986 BDV01811 BDV01809 BEV01071 BBV03561 AGV02553 AJV02483 AJV02480 BCV01308 BCY01026 BCY01025 BEV01813 AJV01806 AJV01687 BCV02495 BCV02491 BCV02329 BCV02315 BCV02312 BCV01318 BCV01317 BCV01043 BBV3596 BBV03595 AFV02283 AEV01046 ADV06282 AGV01553 AEV02745 AEV01727 ADV01538 ADV01535 AGV01338 AGV01312 BAV06268 AGV01721 ADV05015 AFV01327 BCV01542 BCV01540 BCV01537 BCV01536 BCV01523 AKV02737 AKV02736 BCV01255 BEV01469 BAV04245 AKV01648 BAV04069 BCV01313 AHV01633 AHV01617 AKV01451 AKV01280 AKV01167 AKV01452 AKV04131 AKV04130 AKV04128 AKV04063 AKV04062 AKV04059 AKV04057 AKV04052 AKV04051 AKV04050 AKV04049 AKV04048 AKV04047 AKV04046 AKV04044 AKV04043 AKV04042 AKV04039 AKV04038 AKV04037 AKV04036 AKV04035 AKV04034 AKV04032 AKV04031 AKV04030 AKV01775 AHV01091 AJV02743 BAV01154 BAV01158 BAV01156 BAV01164 AJV04113 AFV01717 AEV02043 AEV01997 AEV01986 AEV01979 AEV01968 BCV01210 BCV01204 BCV01200 BCV01194 BCV01193 BCV01192 BCV01184 BCV01180 BCV01174 AGV02698 AGV01882 AGV01798 AFV01386 AEV02080 AEV02079 AFV01375 AGV02522 BBV03061 AJV02315 BBV03062 BEV01343 BEV01342 BEV01341 BEV01337 BEV01334 BEV01818 BEV01812 BAV06650 BAV06649 BAV06646 BAV06652 AKV01736 AKV01723 BAV04747 BAV04745 BAV04751 BAV04750 BAV04755 BAV04748 AHV01867 BAV04805 BAV04806 BAV04816 BAV04813 BAV04810 BAV04804 AKV01737 AKV01840 BAV04817 BEV01327 BEV01262 AKV01462 BDV01860 BDV01851 BDV01840 BCV02283 BCV02282 BAV04366 BAV04360 BAV04359 BAV04369 BAV04365 AEV01711 ADV03111 ADV01585 ADV03042 ADV01899 BAV06024 BAV06018 BAV06012 BAV05295 BAV05262 BAV05257 BAV05246 BAV05244 BAV06014 BAV06022 BAV06007 BAV06025 AFV02585 ADV02092 AJV01640 BBV06100 BAV06017 BAV05249 BAV05245 AKV04499 AKV04123 ADV06308 AJV02468 AFV02359 AEV02222 AEV02207 AEV02182 ADV06238 BEV02138 BEV02130 BEV02129 BEV02128 BEV02126 BEV02122 BEV02121 BEV02118 BEV02115 BEV02094 BEV01506 BEV01505 BEV01504 BEV01503 BEV01501 BEV01500 BEV01499 BEV01498 BEV01497 BEV01495 BEV01493 BEV01492 BEV01491 BEV01490 BEV01489 BEV01488 BEV01487 BEV01486 BEV01485 BEV01484 BEV01483 BEV01482 BEV01481 BEV01480 BEV01479 BEV01478 BEV01476 BEV01474 BEV01473 BEV01354 AJV02428 AGV02405 AKV01694 BAV06019 BAV06011 BAV06010 BAV06008 BAV05258 BAV05255 BAV05254 BAV05248 BAV05260 BAV06015 BAV05241 BAV06184 AKV01693 AKV01692 AKV01690 AKV01688 AKV01686 AKV01683 AKV01669 AKV01661 AKV01660 AGV02015 AKV04500 AJV02782 AJV02780 AJV02775 AJV02771 AJV02744 BDV01316 BCV01397 BBV03241 BBV03239 BBV03238 BBV03237 AEV01829 BAV06518 BAV05177 BCV01391 ADV04115 BAV01139 BEV01964 BDV01862 AEV02170 AEV02169 AEV02148 AEV02051 AEV01866 AJV02948 BDV01374 BBV01612 BAV06800 BAV06797 BAV06794 BAV06793 BAV06788 BAV06787 BAV06786 BAV06785 BAV06776 BAV06095 BAV06094 AJV04187 AHV03024 AEV02229 AEV02158 AEV02147 AEV02126 AEV02095 AEV01675 ADV06223 ADV06140 ADV02130 ADV02081 AEV02160 BDV01454 BBV03122 BCV02855 BCV02606 BCV02345 BCV02268 BCV02267 AJV02829 AKV01224 AKV01180 ADV02134 BDV01342 BDV01331 BDV01319 BDV01304 BDV01173 BDV01471 BDV01465 BDV01358 BDV01355 BDV01354 BDV01350 BDV01348 BDV01321 BAV06795 BAV06792 BAV06790 BAV06789 BAV06779 BAV06006 BAV05298 AEV02121 BAV06806 BAV06799 BAV05300 BEV01521 BEV01516 BEV01194 BEV01189 BEV01187 BEV01115 BEV01110 BDV01757 BDV01755 BDV01537 BDV01536 BDV01532 BDV01527 BDV01458 AJV02827 BBV01614 BAV06456 BAV04059 BAV04056 AKV01925 AKV01923 AKV01912 AKV01910 AKV01909 AKV01908 AKV01904 BEV02203 BEV02202 BEV02201 BEV02200 BEV02199 BEV02198 BEV02196 BEV02195 BEV02194 BEV02192 BEV02190 AJV01182 AJV02598 AHV03106 AHV03104 AFV02030 BDV01363 AKV01443 AKV01434 AKV01417 AJV02872 ADV01827 BAV01170 BAV01167 BAV01127 BDV01507 BDV01499 BDV01493 BDV01492 BDV01486 BDV01484 BDV01483 BDV01475 BDV01473 BDV01468 BDV01282 BCV02703 BCV02702 BCV02679 BAV01168 AJV02600 AEV02704 BBV06103 AJV01106 BCV01138 BCV01250 BAV05171 BAV04333 AKV04125 BAV04339 AJV02585 AJV02583 AJV02581 AJV02548 AJV01107 AHV01151 AHV01145 AJV04209 BBV03549 BBV03548 BBV03547 BBV03546 BBV03545 BBV03543 BBV03542 BBV03535 BBV03514 BBV03513 BBV03512 BBV03511 BBV03409 BBV03408 BBV03407 BBV03406 BBV03405 BBV03404 BBV03401 BBV03400 BBV03399 BBV03397 BBV03396 BBV03395 BBV03392 BBV03391 BBV03390 BBV03387 BBV03386 BBV03383 BBV03382 BBV03381 BBV03380 BBV03249 BBV03248 BBV03247 BDV02016 BDV02015 BDV02014 BDV02003 BDV01997 BDV01985 BDV01979 BDV01972 BDV01875 BDV01813 BBV03240 BBV03130 ADV01645 BCV02879 BCV02876 BCV02859 BAV06693 BAV06464 AKV04484 AKV04482 BAV06772 BAV06771 BAV06040 BAV05159 BAV05158 BAV05157 BAV05154 BAV05152 BAV05151 BAV05148 BAV05146 BAV05145 BAV05144 BAV05143 BAV05142 BAV05141 BAV05140 BAV05137 BAV05136 BAV05135 BAV05134 BAV05133 BAV05132 BAV05130 AKV02195 AKV02194 AKV02178 AKV02169 AKV02135 AHV03331 AHV03327 AHV03302 BAV05147 BBV01007 AJV01502 BAV04152 BAV04151 BAV04150 BAV04144 BAV04140 BAV04131 BAV04128 BBV03387 BAV04838 BAV04811 BAV04145 AHV02016 AEV01765 AEV01753 ADV04100 ADV04008 BAV06763 BAV06761 BAV06759 ADV06327 ADV06309 ADV06190 BAV06773 AGV02291 AEV01703 BBV01026 BBV01024 AKV01780 AKV01373 AKV01385 BDV01610 BDV01597 BDV01505 BDV01497 BDV01298 BDV01271 BDV01270 BDV01266 BDV01265 BDV01250 BCV02707 BCV02680 BCV02662 BCV02650 BCV01261 BCV01260 BCV01259 BCV01258 BCV01254 BCV01252 BCV01251 BCV01149 BCV01148 BCV01147 BCV01146 BCV01145 BCV01144 BCV01143 BCV01142 BCV01141 BCV01140 BCV01136 BCV01111 BBV06111 BBV06074 BBV01059 BBV01058 BBV01057 BBV01055 BBV01054 BBV01003 BBV01056 BAV05168 BAV05166 BAV05165 BAV04340 BAV04338 BAV04337 BAV04335 BAV04275 AJV02490 AJV02486 AHV01166 AGV02304 AGV02226 AGV02193 AGV02081 AGV02217 AEV02266 AEV02214 AEV01831 BBV01053 BAV04721 AEV01822 AHV02140 BDV01356 BDV01345 BDV01312 BEV01464 BAV06197 AJV02391 AJV01693 AJV01690 AJV01569 AJV01207 AKV04346 BCV01709 BCV01621 BCV01618 BCV01615 BCV01610 BCV01604 AKV02547 BAV05100 BAV04825 BAV04798 AKV01474 AKV01473 AKV01467 BAV04814 AKV01455 AKV01454 AKV01458 AKV01427 AKV01318 AKV01314 AKV01967 AKV01460 AKV01461 AKV01464 AKV01414 AKV01457 AKV01344 AKV01292 BCV02870 BCV02868 BCV02866 BCV02864 BCV02857 BDV01777 BDV01764 BDV01533 BDV01530 AKV02721 AKV02720 AKV02706 AKV02705 ADV01522 AKV01300 BDV01977 BDV01814 BEV01109 BEV01106 BEV01105 BEV01102 BEV01100 BDV01861 BDV01858 BDV01857 BDV01854 BDV01513 BDV01511 BDV01496 BDV01464 BDV01455 BDV01389 BDV01386 BDV01352 BDV01346 BDV01332 BDV01498 BAV01108 BAV01109 BAV01106 BAV01107 BBV06116 BBV06119 AHV01395 AHV01116 AHV01018 AHV01002 BAV06475 ADV03113 BDV01414 BDV01404 BBV03138 AKV01930 AGV01809 BBV06142 BBV06141 BBV06140 BBV06139 BBV06137 BBV06136 BBV06135 BBV06134 BBV06133 BBV06122 BBV06121 BBV06118 BBV06115 BBV06114 BBV06110 BBV06087 BBV01049 BBV01022 BBV01021 BAV06728 BAV06715 BAV06714 BAV06711 BAV06707 BAV06705 BAV06704 BAV06699 BAV06698 BAV06697 BAV06696 AHV01482 AHV01404 AHV01394 AHV01357 AFV01523 AFV01002 BAV06710 ADV06406 ADV04031 ADV03055 ADV05003 BBV06138 BAV06701 AJV02501 AJV02179 AJV02164 AJV02159 BCV01690 BCV01626 BBV03591 BAV05119 BDV01654 BDV01652 BDV01651 BDV01447 BDV01441 BDV01439 AKV01489 AEV02030 AEV01948 AEV01714 AEV01682 AEV01610 AEV02187 BDV01988 BDV01596 AKV02597 BEV02167 BEV02164 BEV02158 BEV02156 BEV02152 BEV02151 BEV02002 BEV01496 BEV01351 BEV01350 BDV01216 BDV01213 BDV01204 BCV1009 BCV02251 BBV03246 BBV03245 BBV03146 BBV03123 BBV03001 BBV01635 BBV01634 BBV01630 BAV05233 BAV04812 AKV04074 AKV04073 AKV01670 AKV01668 AKV01662 AKV01295 AJV01742 AJV01135 AJV01104 AJV01064 AHV01308 AGV01618 AGV01535 AGV01532 AGV01464 AGV01410 BBV03002 BAV06429 AGV01569 AGV01463 BDV01555 BDV01615 BDV01595 BDV01584 BDV01323 BDV01322 BCV02630 BBV03544 BBV03534 BBV03533 BBV03532 BBV03531 BAV04593 AHV01910 AFV02473 AFV02332 AHV03304 BAV04592 BCV02305 BAV01115 AGV01871 AKV01931 AJV02769 AJV02735 AJV02734 AHV03158 AGV01968 AGV01873 AGV01870 AFV01499 AFV01471 AFV01468 AFV01467 AFV01441 AFV01430 AFV01395 ADV02154 AFV01435 AJV01700 AEV02747 AEV02617 AJV01248 BDV02009 BDV02008 BDV02000 BDV01999 AJV01196 AJV01222 AHV01060 AGV02662 AJV01197 aev01792 BEV01250 BCV01003 BCV01000 BAV06660 AKV01525 AJV02915 AJV01151 AEV01536 AJV02815 BCV01001 BBV03835 BBV01205 AGV01526 AGV01524 AKV01698 AKV01710 AKV01709 AKV01699 BDV01642 AEV01586 BEV02140 BBV06106 AJV02376 AGV02282 AFV01217 aev01921 BAV05098 BAV05094 BAV05099 AKV01530 AGV02089 ADV06381 AFV02500 ADV01553 BEV01764 BEV01760 AJV03041 AJV03026 AJV03007 AJV02989 AJV02907 AJV02895 AJV02666 AGV01846 AEV01764 BCV02163 BCV02161 BCV02843 BCV02842 BBV03129 BBV03108 AKV01223 AKV01174 AKV01533 AKV01528 AKV01527 AKV01523 AKV01522 AKV01520 BBV01280 AGV02564 AFV02061 AEV02092 AEV01619 AEV01617 BBV01204 BEV01266 BEV01265 BEV01261 BEV01253 BEV01252 BEV01249 BEV01248 BEV01247 BEV01246 BEV01243 BEV01239 BEV01237 BEV01236 BEV01234 BEV01229 BEV01223 BEV01216 BEV01204 BDV01768 BDV01706 BDV01641 BDV01638 BDV01637 BDV01636 BDV01634 BDV01630 BDV01617 BDV01580 BCV02025 BCV02024 BCV02023 BCV02001 BCV02000 BCV01998 BCV01270 BCV01268 BCV01267 BBV01011 BAV06048 BAV06433 BAV06426 BAV06052 BAV06051 BAV06049 AKV01771 AKV01743 AKV01218 AKV01217 AKV01216 AKV01214 AKV01194 AKV01163 AKV01158 AKV01149 AKV01144 AHV03310 AHV03213 AKV01215 BAV06428 BEV01165 BEV01162 BEV01160 BEV01145 BEV01119 BCV02608 BCV02607 BCV02604 BCV02598 BCV02590 BCV02567 BBV03712 BBV03711 BBV03710 BBV03659 BBV03658 BBV03657 BBV03656 BBV03655 BBV03653 BBV03652 BBV03649 BBV03648 BBV03647 BBV03646 BBV03645 BBV03644 BBV03643 BBV03642 BBV03641 BBV03640 BBV03639 BBV03638 BBV03637 BBV03636 BBV03635 BBV03634 BBV03633 BBV03632 BBV03631 BBV03630 BBV03629 BBV03628 BBV03627 BBV03555 BBV03366 BBV03361 BBV03357 BBV03356 BBV03354 BBV03353 BBV03347 BBV01271 BBV01270 BBV01269 BBV01268 BBV01267 BBV01266 BBV01265 BBV01264 BBV01263 BBV01262 BAV04803 BAV04802 BAV04801 BAV04799 BAV04626 BAV04625 BAV04624 BAV04623 BAV04621 BAV04620 BAV04619 AKV01944 AKV01920 AKV01362 AKV01358 AKV01355 AKV01351 AKV01348 AKV01345 AKV01336 AKV01334 AKV01333 AKV01329 AKV01328 AKV01326 AKV01325 AKV01324 AKV01323 AKV01313 AKV01312 AKV01311 AKV01310 AKV01309 AKV01308 AKV01307 AKV01306 AKV01305 AKV01304 AKV01303 AKV01272 AKV01257 AJV02342 AGV01024 AJV02325 BAV04629 AKV01327 BAV04800 AJV02326 BBV03654 BAV04618 BAV04617 AKV01347 AKV01350 AKV01339 AKV01338 AKV01337 BBV06102 AJV01605 BDV01644 BAV01268 AJV01692 BDV02012 BDV02007 BDV01983 BDV01978 BDV01976 BDV01649 AHV01161 AFV02497 AFV01981 AFV01907 AFV01893 AFV01870 AFV01835 AFV01819 BDV01812 BBV01276 BCV02193 BCV02185 BCV01834 BCV01833 BCV01265 BCV01264 AKV01789 AKV01787 AKV01786 AKV01784 AKV01782 AKV01779 AHV02014 AGV01812 AGV01782 AGV01764 AGV01757 AGV01780 BAV06512 AKV04312 BEV01807 BEV01802 BCV01513 BAV04224 BAV04051 BAV04050 BAV04052 BAV04053 ADV05172 AHV01082 AHV01013 AHV01009 AGV02717 AJV01674 AJV01656 AJV01682 BCV01849 BAV06513 BAV06511 AKV04313 AFV02053 AJV02434 BCV01517 AFV01267 AFV01284 AFV01275 AFV01271 AFV01262 AFV01241 AEV02351 AHV01296 AHV03336 AHV01123 AHV01061 AHV01006 AGV02708 BAV04268 AHV03271 BAV05170 BAV05163 AKV01079 AKV01062 AJV01325 AFV02530 AFV02083 AEV02332 AEV02288 AFV02329 BDV01296 BDV01272 BDV01269 BDV01268 BDV01267 AFV02260 BAV04702 BAV04467 AJV02957 AJV02880 AJV01038 AFV02416 AFV02264 AFV02201 AEV02387 BEV01477 BEV01465 BEV01254 BEV01251 BEV01242 BEV01231 BEV01655 BEV01652 BEV01650 BEV01649 BEV01645 BEV01177 BDV01850 BDV01788 BDV01360 BDV01343 BCV01583 BCV01582 BCV01581 BCV01580 BCV01579 BCV01577 BCV01574 BCV01566 BCV01506 BCV01395 BCV01386 BAV04783 AKV01792 AKV01790 AKV01733 AKV01729 AJV04075 AJV04071 AJV04065 AHV01036 AGV02400 AGV02324 BAV06534 BAV06626 BAV06535 BAV06769 BAV06765 BAV06768 BAV06767 AKV01419 AKV01343 AKV01331 AKV01330 BBV03216 BBV03215 BEV01752 BEV01747 BEV01679 BEV01340 BEV01339 BEV01338 BEV01336 BEV01335 BEV01332 BEV01331 BEV01330 BEV01302 BEV01297 AFV02357 BAV06042 AKV04322 AGV01020 AKV04323 AKV04321 BCV02309 BCV02308 BCV02295 BCV02292 BCV02286 BCV01314 BCV01311 BCV01310 BCV01302 BCV01049 BCV01032 BCV01026 BCV01018 BCV01017 BBV03811 BBV03707 BBV03705 BBV03702 BBV03700 BBV03502 BBV03501 BBV03500 BBV03491 BBV03490 BBV03489 BBV03488 BBV03487 BBV03482 BBV03473 AKV04497 AKV04496 AKV04495 AKV04493 AKV01973 AJV01621 AJV01618 AHV03262 AHV03248 AHV03245 BCV01030 BAV06659 AEV01597 BAV06658 BAV06656 BAV06653 BAV06655 BAV06754 BCV01781 BCV01780 BCV01779 BCV01764 BCV01778 BCV01949 BCV01948 BBV01513 BBV01512 BBV01488 BBV01485 BBV01327 BBV01509 AKV01058 AHV02064 AHV02040 AHV01849 AHV01837 BEV01502 BDV01834 BDV01829 BDV01828 BDV01794 BCV02609 BCV02602 BCV02592 BBV03236 BBV03234 BBV03233 BBV03230 BCV02391 BCV02376 BCV02367 BCV02362 BCV01744 BCV01695 BCV01687 BCV01686 BCV01643 BCV01629 BCV01082 BDV01643 AJV02374 AJV01115 BDV02004 BBV06042 BBV06058 BAV04336 BAV04315 BAV04314 BAV04311 BAV04307 BAV04303 BAV04298 BAV04297 BAV04296 BAV04284 BDV01989 BDV01973 BDV01817 BCV02716 BCV02688 BCV02686 BCV01698 BCV01697 BCV01688 BBV03594 BBV03593 BBV03592 BBV03141 BBV01610 BAV06766 AKV02227 AKV01086 AJV04222 AHV02124 AHV01137 AGV
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1609-2015
Related Recalls
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).