ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    REF/UDI-DI: 11711750/04056869995809, and 11711779/04056869995830. Serial Numbers: 700127 700125 700109 700123 700124 700126 700131 700077 700081 700070 700071 700073 700075 700076 700072 700074 700096 700097 700098 700100 700103 700082 700030 700036 700037 700028 700029 700031 700033 700022 700023 700024 700025 700035 700121 700111 700132 700065 700058 700026 700019 700020 900581 900582 900583 900585 700034 700032 700041 700040 700039 700052 700053 700057 700038 700063 700062 700060 700064 700068 700059 700084 700083 700107 700114 700115 700135.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc.
Reason for Recall:
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A

Product Codes/Lot Numbers:

REF/UDI-DI: 11711750/04056869995809, and 11711779/04056869995830. Serial Numbers: 700127 700125 700109 700123 700124 700126 700131 700077 700081 700070 700071 700073 700075 700076 700072 700074 700096 700097 700098 700100 700103 700082 700030 700036 700037 700028 700029 700031 700033 700022 700023 700024 700025 700035 700121 700111 700132 700065 700058 700026 700019 700020 900581 900582 900583 900585 700034 700032 700041 700040 700039 700052 700053 700057 700038 700063 700062 700060 700064 700068 700059 700084 700083 700107 700114 700115 700135.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1601-2024

Related Recalls

MAMMOMAT Inspiration;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Fusion;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Revelation;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →