Nuclear Gamma Cameras (ADAC Vertex Classic, ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60) Philips Medical Systems, Cleveland, OH
Class I - DangerousWhat Should You Do?
- Check if you have this product: Vertex Classic; Model #: 890122; Serial #s: 9404100, 9601238, 9609084, 0407V, 0682V, 0795V, 0822VP2, 103, 20, 500, 5405V, 64, 9308007, 9309020, 9310032, 93-10660-1292, 9401059, 9401061, 9403076, 9403083, 9404106, 9407122, 9409148, 9410002, 9411006, 9411007, 9411016, 9501049, 9501210, 9502067, 9502086, 9503003, 9503079, 9503109, 9504019, 9505021, 9505124, 9506140, 9508169, 9508171, 9508184, 9508193, 9509198, 9509206, 9510222, 9510226, 9601146, 9602249, 9608042, 9901004, B0106154, B02120033, M@10229AMT, M@207JD, M@4404V, M@4431V, M@4432V, M@4480V, M@4623V, M@4630V, M@4639V, M@4686V, M@4692V, M@4858V, M@4893V, M@4959V, M@5032V, M@5033V, M@5038V, M@5105V, M@5145V, M@5158V, M@5189V, M@5280V, M@5289V, M@5338V, M@5360V, M@5373V, M@5398V, M@5407V, M@5412V, M@5431V, M@5439V, M@5467V, M@5476V, M@5497V, M@5498V, M@5515V, M@5526V, M@5590V, M@5617V, M@5627V, M@5661V, M@5662V, M@5663V, M@5665V, M@5681V, M@5688V, M@5689V, M@5751V, M@5752V, M@5782V, M@5811V, M@5835V, M@5869V, M@5900V, M@5913V, M@5915V, M@5960V, M@5970V, M@6004V, M@6007V, M@6010V, M@6056V, M@6900V, M@9304RV, M@9645RV, M@MVS5236V, M@MVS5852V, M@O1325, UNK-AMT1238A. VERTEX PLUS/SOLUS/CARDIO: Model #: 88297; Serial #s: 9603015, 9605036, 9610076, 9610077, 9703171, 9706192, 9710103, 9711277, 9802304, 9803109, 9807121, 9808122, 9809056, 9809124, 9809415, 9809424, 9912461, _0002032, ~9608052, 0007001M, 0008009B, 0011002B, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 1006, 1008, 10252, 1027, 1037, 10630, 1092, 1097, 11, 130234-01NVERTEX, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 5934V, 6052, 6261, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 8002, 802080322, 85247, 9054VM/V, 9056VM/V, 9073152, 9130VM, 9205345, 940384, 9408136, 9503084, 9504096, 9506016, 9506145, 9507163-CN, 9511123, 9512008, 9512009, 9512018, 9601015, 9601036, 9601240-CN, 9602161, 9602245, 9602246, 9602254, 9603336, 9604016, 9604021, 9605026, 9605037, 9606211, 9607015, 9607024, 9608046, 9608049, 9608050, 9608052, 9608064, 9609063, 9609081, 9610058, 9611003, 9611077, 9611082, 9612072, 9701127, 9702061-CN, 9702095, 9702150, 9703154, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708068-CN, 9708069, 9708076, 9708078, 9708084, 9708227, 9709022, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711078, 9711261, 9711271, 9801047, 9801227, 9801232, 9801297, 9801299-CN, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803312, 9803316, 9803322, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9806006, 9806014, 9806363, 9806377, 9806388, 98066002, 9807002, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809062 (-CN), 9809065-CN, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810428ADAC#7973VM, 9810442, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905012, 9905659, 9905667, 9906670, 9907016, 9907017, 9907019, 9907029, 9909161, 9910031, 9911458, 9911698, 9911702, 9912181, 9912463, AD1261, AMT00848VM, B0006129, B0007273, B0008003, B0008012, B0008021, B0008030, B001200212, B0105138, B0106169, B01070191, B01070192, B0107179, B01080204, B01080207, B01090227, B01100231, B01100233, B01100238, B01110251, B02050071, B02050262, B02060167, B02060270, B02060279, B0206027S, B02070290, B02070300, B0207296, B02080318, B0208312, B02090327, B02090333, B02090334, B02100345, B03010355, B03010359, B03020048, B03080380, C0008070, C01080152, C01080154, C03050230, C03060193, M@10458, M@10573, M@10582, M@10589, M@10772, M@10837, M@10850, M@10927, M@10967, M@11020, M@11036, M@11109, M@11140, M@11178, M@11498, M@11525, M@11571, M@11590, M@11881, M@11949, M@11957, M@11972, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13665, M@13688, M@13691, M@13748, M@13813, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@5961K, M@6002J, M@6012J, M@6075VM, M@6108VA, M@6123VA, M@6160VM, M@6183VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6368J, M@6370VA, M@6375J, M@6376J, M@6460VM, M@6464VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6946VM, M@6947J, M@6985J, M@7003VA, M@7004VA, M@7005VA, M@7017VM, M@7024VM, M@7034K, M@7042VM, M@7053VM, M@7073VM, M@7085K, M@7086J, M@7095VM, M@7100VM, M@7101VM, M@7102VM, M@7151VM, M@7153VM, M@7159VM, M@7166K, M@7171VM, M@7172K, M@7181J, M@7190VM, M@7198VA, M@7215K, M@7216VM, M@7217VM, M@7221VA, M@7222VA, M@7229VM, M@7233VM, M@7237VM, M@7245J, M@7246VM, M@7247VM, M@7250VM, M@7300J, M@7308VM, M@7376VM, M@7380VM, M@7384VM, M@7390VM, M@7393VM, M@7394VM, M@7410VM, M@7420VM, M@7464VA, M@7470VM, M@7490VM, M@7494K, M@7501VM, M@7504VM, M@7505VM1, M@7514VM, M@7515VM, M@7536VM, M@7537VM, M@7538VM, M@7556VM, M@7557VM, M@7569VM, M@7572VM, M@7591VM, M@7592VM, M@7603K, M@7605VM, M@7606VM, M@7623K, M@7626K, M@7627K, M@7628VA, M@7638VM, M@7643VM, M@7654VM, M@7664VM, M@7665K, M@7673VM, M@7674K, M@7679VM, M@7682K, M@7685VM, M@7691VM, M@7695VM, M@7697VM, M@7723VM, M@7729VM, M@7733VM, M@7739K, M@7760VM, M@7761VM, M@7770VM, M@7776VM, M@7789K, M@7801VM, M@7810VM, M@7811VM, M@7812VM, M@7819VA, M@7823K, M@7824VM, M@7830K, M@7855J, M@7857VM, M@7860VM, M@7861VM, M@7874VM, M@7877VM, M@7918K, M@7924K, M@7925K, M@7929VM, M@7930VM, M@7932VM, M@7946J, M@7957VM, M@7961VM, M@7962VM, M@7965VM, M@7968VM, M@7975K, M@7987VM, M@7995VM, M@7997VM, M@8004VM, M@8013VM, M@8018J, M@8021VM, M@8044VM, M@8060VM, M@8064VM, M@8076K, M@8089VM, M@8096VM, M@8105VM, M@8124VA, M@8124VA, M@8146VM, M@8155VM, M@8178VM, M@8203VM, M@8219VA, M@8220K, M@8240K, M@8241VM, M@8248VM, M@8252VM, M@8284K, M@8301K, M@8302VM, M@8306K, M@8337VM, M@8356K, M@8380VM, M@8399VM, M@8484VM, M@8528VM, M@8537VM, M@8551VM, M@8561K, M@8606K, M@8609VM, M@8651VM, M@8683VM, M@8686K, M@8693VM, M@8726VM, M@8738K, M@8759K, M@8771VM, M@8781VM, M@8797K, M@8799RV, M@8800VM, M@8803K, M@8812VM, M@8814VM, M@8882K, M@8888K, M@8897K, M@8903K, M@8906RK, M@8920VM, M@8922VM, M@8935K, M@8948RV, M@8955K, M@8990K, M@8996VM, M@9032K, M@9046K, M@9049VM, M@9081K, M@9091VM, M@9107RV, M@9166VM, M@9185K, M@9194VM, M@9221VM, M@9222VM, M@9247RV, M@9252VM, M@9267VM, M@9309RV, M@9316VM, M@9327K, M@9338VM, M@9340VM, M@9360VM, M@9369VM, M@9418VM, M@9426K, M@9438K, M@9471RV, M@9515VM, M@9532VM, M@9542VM, M@9543K, M@9583VM, M@9607VM, M@9653VM, M@9672VM, M@9724VM, M@9730K, M@9816K, M@9831VM, M@AMT1258, M@MCD6416VM, M@MCD7036VM, M@R372820, M@S5195VA, M@S5365-V, M@S5820VM, M@S5907V11, M@S6155VA, M@S7156VM, M@S7223K, M@S7559VA, M@S7571VM, M@S7856K, M@S7864K, M@S7904VM, M@S7941VA, M@S8687K, M0101023, R03010363, S0011003, S6235VM, STOR_QUO_47521360, UNK-6597VM, UNK-7590VM, UNK-8533VM1, UNK-8860VM, UNK-9868K, V0007070, V0010090, V0012099, V0104109, V0104110, V0105112, V0106121, V0106123, V0106124, V01070133, V01090146, V01100149, V01100151, V01100156, V02050166, V02070174, V02070179, V02080190, V02080194, V02080199, V02080200, V02090206, V03060224, V03070229, V03091241. VERTEX V60 Model #: 882030; Serial #s: 9601243, 9602253, 9603008, 9603013, 9604028, 9604041, 9605033, 9606038, 9610256, 9704175, 9704177, 9801284, 9803324, 9807089, 9810434, 9904005, 970001-9610187, M0007018, V0009076, V0011096, V0106122, V0106125, V02050163, V02050165, V02070181, 9705007, 9711262, 00559VM, 00634V, 00657VA, 00970VA, 01190VM, 01697VM, 01894VM, 0371V, 10588, 1189-17, 12001, 1273-06, 1274-06, 9601004, 9607063, 9707206, 9710247, 9712001, 9712008, 9712009, 9803003, 9806019, 9808021, 9902027, 9909039, M0007001-CN, V0005062, V01090141, V01090142, V03080235, V03080236, V2070177, VETEJ, X0005002, X0010004, X0101006, X0101007, X03030012, Z00657VA, Z01798VM.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems (Cleveland) Inc
- Reason for Recall:
- During clinical use of a Vertex Plus Gamma Camera, as the detector heads were in the relative 180 degree position, Detector #1 drifted down and made contact with the patients chin during a whole body study acquisition.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Nuclear Gamma Cameras (ADAC Vertex Classic, ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60) Philips Medical Systems, Cleveland, OH
Product Codes/Lot Numbers:
Vertex Classic; Model #: 890122; Serial #s: 9404100, 9601238, 9609084, 0407V, 0682V, 0795V, 0822VP2, 103, 20, 500, 5405V, 64, 9308007, 9309020, 9310032, 93-10660-1292, 9401059, 9401061, 9403076, 9403083, 9404106, 9407122, 9409148, 9410002, 9411006, 9411007, 9411016, 9501049, 9501210, 9502067, 9502086, 9503003, 9503079, 9503109, 9504019, 9505021, 9505124, 9506140, 9508169, 9508171, 9508184, 9508193, 9509198, 9509206, 9510222, 9510226, 9601146, 9602249, 9608042, 9901004, B0106154, B02120033, M@10229AMT, M@207JD, M@4404V, M@4431V, M@4432V, M@4480V, M@4623V, M@4630V, M@4639V, M@4686V, M@4692V, M@4858V, M@4893V, M@4959V, M@5032V, M@5033V, M@5038V, M@5105V, M@5145V, M@5158V, M@5189V, M@5280V, M@5289V, M@5338V, M@5360V, M@5373V, M@5398V, M@5407V, M@5412V, M@5431V, M@5439V, M@5467V, M@5476V, M@5497V, M@5498V, M@5515V, M@5526V, M@5590V, M@5617V, M@5627V, M@5661V, M@5662V, M@5663V, M@5665V, M@5681V, M@5688V, M@5689V, M@5751V, M@5752V, M@5782V, M@5811V, M@5835V, M@5869V, M@5900V, M@5913V, M@5915V, M@5960V, M@5970V, M@6004V, M@6007V, M@6010V, M@6056V, M@6900V, M@9304RV, M@9645RV, M@MVS5236V, M@MVS5852V, M@O1325, UNK-AMT1238A. VERTEX PLUS/SOLUS/CARDIO: Model #: 88297; Serial #s: 9603015, 9605036, 9610076, 9610077, 9703171, 9706192, 9710103, 9711277, 9802304, 9803109, 9807121, 9808122, 9809056, 9809124, 9809415, 9809424, 9912461, _0002032, ~9608052, 0007001M, 0008009B, 0011002B, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 1006, 1008, 10252, 1027, 1037, 10630, 1092, 1097, 11, 130234-01NVERTEX, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 5934V, 6052, 6261, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 8002, 802080322, 85247, 9054VM/V, 9056VM/V, 9073152, 9130VM, 9205345, 940384, 9408136, 9503084, 9504096, 9506016, 9506145, 9507163-CN, 9511123, 9512008, 9512009, 9512018, 9601015, 9601036, 9601240-CN, 9602161, 9602245, 9602246, 9602254, 9603336, 9604016, 9604021, 9605026, 9605037, 9606211, 9607015, 9607024, 9608046, 9608049, 9608050, 9608052, 9608064, 9609063, 9609081, 9610058, 9611003, 9611077, 9611082, 9612072, 9701127, 9702061-CN, 9702095, 9702150, 9703154, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708068-CN, 9708069, 9708076, 9708078, 9708084, 9708227, 9709022, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711078, 9711261, 9711271, 9801047, 9801227, 9801232, 9801297, 9801299-CN, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803312, 9803316, 9803322, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9806006, 9806014, 9806363, 9806377, 9806388, 98066002, 9807002, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809062 (-CN), 9809065-CN, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810428ADAC#7973VM, 9810442, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905012, 9905659, 9905667, 9906670, 9907016, 9907017, 9907019, 9907029, 9909161, 9910031, 9911458, 9911698, 9911702, 9912181, 9912463, AD1261, AMT00848VM, B0006129, B0007273, B0008003, B0008012, B0008021, B0008030, B001200212, B0105138, B0106169, B01070191, B01070192, B0107179, B01080204, B01080207, B01090227, B01100231, B01100233, B01100238, B01110251, B02050071, B02050262, B02060167, B02060270, B02060279, B0206027S, B02070290, B02070300, B0207296, B02080318, B0208312, B02090327, B02090333, B02090334, B02100345, B03010355, B03010359, B03020048, B03080380, C0008070, C01080152, C01080154, C03050230, C03060193, M@10458, M@10573, M@10582, M@10589, M@10772, M@10837, M@10850, M@10927, M@10967, M@11020, M@11036, M@11109, M@11140, M@11178, M@11498, M@11525, M@11571, M@11590, M@11881, M@11949, M@11957, M@11972, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13665, M@13688, M@13691, M@13748, M@13813, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@5961K, M@6002J, M@6012J, M@6075VM, M@6108VA, M@6123VA, M@6160VM, M@6183VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6368J, M@6370VA, M@6375J, M@6376J, M@6460VM, M@6464VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6946VM, M@6947J, M@6985J, M@7003VA, M@7004VA, M@7005VA, M@7017VM, M@7024VM, M@7034K, M@7042VM, M@7053VM, M@7073VM, M@7085K, M@7086J, M@7095VM, M@7100VM, M@7101VM, M@7102VM, M@7151VM, M@7153VM, M@7159VM, M@7166K, M@7171VM, M@7172K, M@7181J, M@7190VM, M@7198VA, M@7215K, M@7216VM, M@7217VM, M@7221VA, M@7222VA, M@7229VM, M@7233VM, M@7237VM, M@7245J, M@7246VM, M@7247VM, M@7250VM, M@7300J, M@7308VM, M@7376VM, M@7380VM, M@7384VM, M@7390VM, M@7393VM, M@7394VM, M@7410VM, M@7420VM, M@7464VA, M@7470VM, M@7490VM, M@7494K, M@7501VM, M@7504VM, M@7505VM1, M@7514VM, M@7515VM, M@7536VM, M@7537VM, M@7538VM, M@7556VM, M@7557VM, M@7569VM, M@7572VM, M@7591VM, M@7592VM, M@7603K, M@7605VM, M@7606VM, M@7623K, M@7626K, M@7627K, M@7628VA, M@7638VM, M@7643VM, M@7654VM, M@7664VM, M@7665K, M@7673VM, M@7674K, M@7679VM, M@7682K, M@7685VM, M@7691VM, M@7695VM, M@7697VM, M@7723VM, M@7729VM, M@7733VM, M@7739K, M@7760VM, M@7761VM, M@7770VM, M@7776VM, M@7789K, M@7801VM, M@7810VM, M@7811VM, M@7812VM, M@7819VA, M@7823K, M@7824VM, M@7830K, M@7855J, M@7857VM, M@7860VM, M@7861VM, M@7874VM, M@7877VM, M@7918K, M@7924K, M@7925K, M@7929VM, M@7930VM, M@7932VM, M@7946J, M@7957VM, M@7961VM, M@7962VM, M@7965VM, M@7968VM, M@7975K, M@7987VM, M@7995VM, M@7997VM, M@8004VM, M@8013VM, M@8018J, M@8021VM, M@8044VM, M@8060VM, M@8064VM, M@8076K, M@8089VM, M@8096VM, M@8105VM, M@8124VA, M@8124VA, M@8146VM, M@8155VM, M@8178VM, M@8203VM, M@8219VA, M@8220K, M@8240K, M@8241VM, M@8248VM, M@8252VM, M@8284K, M@8301K, M@8302VM, M@8306K, M@8337VM, M@8356K, M@8380VM, M@8399VM, M@8484VM, M@8528VM, M@8537VM, M@8551VM, M@8561K, M@8606K, M@8609VM, M@8651VM, M@8683VM, M@8686K, M@8693VM, M@8726VM, M@8738K, M@8759K, M@8771VM, M@8781VM, M@8797K, M@8799RV, M@8800VM, M@8803K, M@8812VM, M@8814VM, M@8882K, M@8888K, M@8897K, M@8903K, M@8906RK, M@8920VM, M@8922VM, M@8935K, M@8948RV, M@8955K, M@8990K, M@8996VM, M@9032K, M@9046K, M@9049VM, M@9081K, M@9091VM, M@9107RV, M@9166VM, M@9185K, M@9194VM, M@9221VM, M@9222VM, M@9247RV, M@9252VM, M@9267VM, M@9309RV, M@9316VM, M@9327K, M@9338VM, M@9340VM, M@9360VM, M@9369VM, M@9418VM, M@9426K, M@9438K, M@9471RV, M@9515VM, M@9532VM, M@9542VM, M@9543K, M@9583VM, M@9607VM, M@9653VM, M@9672VM, M@9724VM, M@9730K, M@9816K, M@9831VM, M@AMT1258, M@MCD6416VM, M@MCD7036VM, M@R372820, M@S5195VA, M@S5365-V, M@S5820VM, M@S5907V11, M@S6155VA, M@S7156VM, M@S7223K, M@S7559VA, M@S7571VM, M@S7856K, M@S7864K, M@S7904VM, M@S7941VA, M@S8687K, M0101023, R03010363, S0011003, S6235VM, STOR_QUO_47521360, UNK-6597VM, UNK-7590VM, UNK-8533VM1, UNK-8860VM, UNK-9868K, V0007070, V0010090, V0012099, V0104109, V0104110, V0105112, V0106121, V0106123, V0106124, V01070133, V01090146, V01100149, V01100151, V01100156, V02050166, V02070174, V02070179, V02080190, V02080194, V02080199, V02080200, V02090206, V03060224, V03070229, V03091241. VERTEX V60 Model #: 882030; Serial #s: 9601243, 9602253, 9603008, 9603013, 9604028, 9604041, 9605033, 9606038, 9610256, 9704175, 9704177, 9801284, 9803324, 9807089, 9810434, 9904005, 970001-9610187, M0007018, V0009076, V0011096, V0106122, V0106125, V02050163, V02050165, V02070181, 9705007, 9711262, 00559VM, 00634V, 00657VA, 00970VA, 01190VM, 01697VM, 01894VM, 0371V, 10588, 1189-17, 12001, 1273-06, 1274-06, 9601004, 9607063, 9707206, 9710247, 9712001, 9712008, 9712009, 9803003, 9806019, 9808021, 9902027, 9909039, M0007001-CN, V0005062, V01090141, V01090142, V03080235, V03080236, V2070177, VETEJ, X0005002, X0010004, X0101006, X0101007, X03030012, Z00657VA, Z01798VM.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1598-2015
Related Recalls
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Philips Medical Systems (Cleveland)
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Philips Medical Systems (Cleveland)
When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.