SW Bari Lift & Transfer, Model Number 38060000

Class I - Dangerous
🏥 Medical Devices Recalled: January 25, 2024 Raye's Other Medical Devices Nationwide

What Should You Do?

  1. Check if you have this product:
    UDI/DI 00845699001040, All serial numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Raye's Inc.
Reason for Recall:
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

SW Bari Lift & Transfer, Model Number 38060000

Product Codes/Lot Numbers:

UDI/DI 00845699001040, All serial numbers

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1595-2024