Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems. Produce cross-sectional images of the body by computer reconstruction of x-ray transmission.
Class I - DangerousWhat Should You Do?
- Check if you have this product: The following codes are affected by this recall:System Code #728231, Serial #: 9654, 91003, 95130, 95157, 95504 System Code #728326, Serial #: 32000,300003,300004,300005,300010,300011,300012,300013, 300014,300015,300018, 300027,300041,300053,300054, 300059, 300066, 300070,300071, 300073, 300074, 30077, 300078, 300079, 300080,300082,300084,300086,300091,300092, 300095, 300101, 300102,300112, 300119, 300127, 300128, 300129, 300131, 300132, 310064, 320030 System Code #728321, Serial #: 310096, 310070 System Code #728323, Serial #: 320033
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems (Cleveland) Inc
- Reason for Recall:
- If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with the mouse. Also, it is possible to begin a normal clinical scan while the anti-virus scan is running. However, if the user is running a bol
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems. Produce cross-sectional images of the body by computer reconstruction of x-ray transmission.
Product Codes/Lot Numbers:
The following codes are affected by this recall:System Code #728231, Serial #: 9654, 91003, 95130, 95157, 95504 System Code #728326, Serial #: 32000,300003,300004,300005,300010,300011,300012,300013, 300014,300015,300018, 300027,300041,300053,300054, 300059, 300066, 300070,300071, 300073, 300074, 30077, 300078, 300079, 300080,300082,300084,300086,300091,300092, 300095, 300101, 300102,300112, 300119, 300127, 300128, 300129, 300131, 300132, 310064, 320030 System Code #728321, Serial #: 310096, 310070 System Code #728323, Serial #: 320033
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1588-2013
Related Recalls
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Philips Medical Systems (Cleveland)
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Philips Medical Systems (Cleveland)
When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.