Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Batch Numbers: 3000041542 3000046483 3000049065 3000055328 3000057682 3000060483 3000063712
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Maquet Cardiovascular, LLC
Reason for Recall:
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551

Product Codes/Lot Numbers:

Batch Numbers: 3000041542 3000046483 3000049065 3000055328 3000057682 3000060483 3000063712

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1579-2019

Related Recalls

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.