Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Panorama Central Station's CPU Unit, Tower (p/n 0998-00-0709-01) Panorama Central Station's CPU Unit, 2U case (p/n 0998-00-0708-01) Panorama Centrall Station's eGateway Tower (p/n 0998-00-0709-03) Panorama Central Station's eGateway 2U case (p/n 0998-00-0708-03)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Mindray DS USA, Inc. dba Mindray North America
- Reason for Recall:
- Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.
Product Codes/Lot Numbers:
Panorama Central Station's CPU Unit, Tower (p/n 0998-00-0709-01) Panorama Central Station's CPU Unit, 2U case (p/n 0998-00-0708-01) Panorama Centrall Station's eGateway Tower (p/n 0998-00-0709-03) Panorama Central Station's eGateway 2U case (p/n 0998-00-0708-03)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1575-2016
Related Recalls
Potential for anesthesia leakage.
Potential for anesthesia leakage.