NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Numbers: N16E090002, N16E090003, N16E090007, N16E090008, N16E090010, N16E100023, N16E110023, N16E110026, N16E120016, N16E120023, N16E120052, N16E120053, N16E130012, N16E130038, N16E130045, N16E140008, N16E140007, N16E140012, N16E140017, N16E140020, N16E140029, N16E140056, N16E140057, N16E140058, N16E150004, N16E150014, N16E150013, N16E150015, N16E150025, N16E150024, N16E150027, N16E150038, N16E150039, N16E150040, N16E100017, N16E140030 and N16E140031
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips and Neusoft Medical Systems Co., Ltd.
Reason for Recall:
The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Product Codes/Lot Numbers:

Serial Numbers: N16E090002, N16E090003, N16E090007, N16E090008, N16E090010, N16E100023, N16E110023, N16E110026, N16E120016, N16E120023, N16E120052, N16E120053, N16E130012, N16E130038, N16E130045, N16E140008, N16E140007, N16E140012, N16E140017, N16E140020, N16E140029, N16E140056, N16E140057, N16E140058, N16E150004, N16E150014, N16E150013, N16E150015, N16E150025, N16E150024, N16E150027, N16E150038, N16E150039, N16E150040, N16E100017, N16E140030 and N16E140031

Distribution:

Distributed in: NC, OH, NE, SC, TX, LA, PR, MO, FL, CT

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1573-2016

Related Recalls

User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.

Mar 4, 2013 Diagnostic Equipment Nationwide View Details →

User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.

Mar 4, 2013 Diagnostic Equipment Nationwide View Details →

The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.

Oct 18, 2012 Diagnostic Equipment Nationwide View Details →