1. PowerPort M.R.I. Implantable Port with Attachable 9.6 F Open-Ended Single-Lumen Venous Catheter with Suture Plugs, REF: 1809600, UDI: (01)00801741027130 ; 2. PowerPort M.R.I. Implantable Port with Attachable 9.6 F Open-Ended Single-Lumen Venous Catheter Without Suture Plugs, REF: 1809601, UDI: (01)00801741027147; 3. PowerPort M.R.I. Implantable Port With Pre-Attached 9.6 F Open-Ended Single-Lumen Venous Catheter with Suture Plugs, REF: 1859600, UDI: (01)00801741027208 Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1. Lot #s 1809600- RECR1434, RECT0091 2. Lot #s 1809601- RECT0189, 3. Lot #s 1859600- RECR2072, RECS1118
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Bard Peripheral Vascular Inc
Reason for Recall:
Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

1. PowerPort M.R.I. Implantable Port with Attachable 9.6 F Open-Ended Single-Lumen Venous Catheter with Suture Plugs, REF: 1809600, UDI: (01)00801741027130 ; 2. PowerPort M.R.I. Implantable Port with Attachable 9.6 F Open-Ended Single-Lumen Venous Catheter Without Suture Plugs, REF: 1809601, UDI: (01)00801741027147; 3. PowerPort M.R.I. Implantable Port With Pre-Attached 9.6 F Open-Ended Single-Lumen Venous Catheter with Suture Plugs, REF: 1859600, UDI: (01)00801741027208 Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.

Product Codes/Lot Numbers:

1. Lot #s 1809600- RECR1434, RECT0091 2. Lot #s 1809601- RECT0189, 3. Lot #s 1859600- RECR2072, RECS1118

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1571-2020

Related Recalls

Class I - Dangerous

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

Nov 6, 2025 Surgical Instruments Nationwide View Details →

Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

Aug 21, 2025 Diagnostic Equipment Nationwide View Details →