Tecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run previously defined scripts and processes on a Freedom EVO platform (Touch Tools Suite is an optional software add-on for Freedom EVOware (the base operational software for the Freedom EVO 2 instrument.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Instrument Freedom EVO 100 Base Unit: Material number: 10641100, Serial numbers: 1211001569, 1212000203 and 1212000205; Instrument Freedom EVO 100 MCA: Material number: 30020010, Serial number: 1301007643; Instrument Freedom EVO 150 Base Unit: Material number10641150, Serial number: 1301006708; Instrument Freedom EVO 150 MCA: Material number: 30020015, Serial number: 1211007442; Instrument Freedom EVO 150 MCA384: Material number: 30032015, Serial number: 1211011594 and 1109003270; Instrument Freedom EVO 200 Base Unit: Material number: 1641200, Serial number: 1106003923 and 1211011578 and Instrument Freedom EVO 200 MCA: Material number: 30020020, Serial number: 1208007288 and 1208007284.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Tecan US, Inc.
- Reason for Recall:
- WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without warning). This could lead to the instrument moving unexpectedly and potentially cause an injury to the operator if they are physically inside t
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Tecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run previously defined scripts and processes on a Freedom EVO platform (Touch Tools Suite is an optional software add-on for Freedom EVOware (the base operational software for the Freedom EVO 2 instrument.
Product Codes/Lot Numbers:
Instrument Freedom EVO 100 Base Unit: Material number: 10641100, Serial numbers: 1211001569, 1212000203 and 1212000205; Instrument Freedom EVO 100 MCA: Material number: 30020010, Serial number: 1301007643; Instrument Freedom EVO 150 Base Unit: Material number10641150, Serial number: 1301006708; Instrument Freedom EVO 150 MCA: Material number: 30020015, Serial number: 1211007442; Instrument Freedom EVO 150 MCA384: Material number: 30032015, Serial number: 1211011594 and 1109003270; Instrument Freedom EVO 200 Base Unit: Material number: 1641200, Serial number: 1106003923 and 1211011578 and Instrument Freedom EVO 200 MCA: Material number: 30020020, Serial number: 1208007288 and 1208007284.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1560-2013
Related Recalls
Gas springs not always replaced within the interval specified.
Freedom EV0-2 with any version of EVOware and EVOware Sample Tracking Add-On V1.0 or higher
Tecan US
There is a software issue that can potentially impact sample identification and traceability under specific conditions when using EVOware Add On Software: Sample Tracking -Multi Instrument V1.0 or higher, Part Number 30019971.
Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.