DEVILBISS HEALTHCARE 1025 SERIES OXYGEN CONCENTRATOR, Model Numbers 1025DS, 1025UK, 1025KS, and 1025KS-AR

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model 1025DS - UDI-DI: 00885304000846; Serial Numbers A175230001DS to A232160185DS; Model 1025UK - UDI-DI: 00885304022480; Serial Numbers A179210001US to A232060432US; Model 1025KS - UDI-DI: 00853040224734; Serial Numbers A176060001KS to A232220085KS; Model 1025KS-AR - UDI-DI: 00853040224734; Serial Numbers A216240003KS to A229210010KR
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
DeVilbiss Healthcare LLC
Reason for Recall:
The instruction manual is being replaced with an updated version that is in alignment with labeling requirements specified in the IEC-60101-1 standard.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

DEVILBISS HEALTHCARE 1025 SERIES OXYGEN CONCENTRATOR, Model Numbers 1025DS, 1025UK, 1025KS, and 1025KS-AR

Product Codes/Lot Numbers:

Model 1025DS - UDI-DI: 00885304000846; Serial Numbers A175230001DS to A232160185DS; Model 1025UK - UDI-DI: 00885304022480; Serial Numbers A179210001US to A232060432US; Model 1025KS - UDI-DI: 00853040224734; Serial Numbers A176060001KS to A232220085KS; Model 1025KS-AR - UDI-DI: 00853040224734; Serial Numbers A216240003KS to A229210010KR

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1558-2023

Related Recalls

The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.

Nov 18, 2022 Other Medical Devices Nationwide View Details →