Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The Aveir leadless pacemaker system supports the implantation and use of an Aveir Leadless Pacemaker (LP) within the targeted chamber(s) of the heart for monitoring a patient s heart rate and providing rate responsive bradycardia pacing therapy to regulate heart rate.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number: LSP202V UDI-DI code: 05415067040725 Serial Numbers: 1357831, 1357869, 1357885, 1357894, 1357895 Model Number: LSP112V UDI-DI code: 05415067034472 Serial Number: 1357956
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Medical
- Reason for Recall:
- Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The Aveir leadless pacemaker system supports the implantation and use of an Aveir Leadless Pacemaker (LP) within the targeted chamber(s) of the heart for monitoring a patient s heart rate and providing rate responsive bradycardia pacing therapy to regulate heart rate.
Product Codes/Lot Numbers:
Model Number: LSP202V UDI-DI code: 05415067040725 Serial Numbers: 1357831, 1357869, 1357885, 1357894, 1357895 Model Number: LSP112V UDI-DI code: 05415067034472 Serial Number: 1357956
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1557-2025
Related Recalls
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).
Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.