Achieva 1.5T, Magnetic Resonance System.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model No. 781196, 781296, 781343; UDI-DI n/a; Serial No. 32508, 21935, 32084, 22167, 21313, 21681, 32550, 32717, 32083, 22136, 22319, 32095, 32481, 22001, 21237, 21968, 32474, 32500, 21715, 22245, 21693, 21798, 21572, 22025, 32966, 21045, 22140, 21183, 21962, 32034, 21428, 21192, 21957, 21176, 32171, 22215, 21232, 32085, 22207, 22004, 22066, 21666, 21851, 32390, 21104, 21574, 21272, 21111, 32381, 21598, 21351, 21432, 21965, 21178, 21812, 21805, 21688, 21073, 21120, 21117, 21128, 21085, 21086, 21185, 21162, 21125, 21250, 21153, 21144, 21243, 21235, 21315, 21360, 21265, 21559, 21263, 21244, 21160, 21317, 21342, 21312, 21165, 21274, 21215, 21416, 21136, 21382, 21443, 21454, 21329, 21348, 21226, 21344, 21225, 21297, 21460, 21414, 21281, 21530, 21306, 21444, 21459, 21413, 21430, 21457, 21449, 21052, 21592, 21427, 21431, 21238, 21489, 21488, 21570, 21896, 21625, 21465, 21401, 21578, 21398, 21518, 21295, 21013, 21397, 21134, 21503, 21509, 21624, 21463, 21658, 21588, 21585, 21646, 21647, 21689, 21627, 21711, 21696, 21412, 21699, 21548, 21378, 21654, 21698, 21492, 21707, 21338, 21708, 21690, 21677, 21782, 21787, 21729, 21738, 21731, 21760, 21686, 21914, 22047, 21903, 21750, 21724, 21911, 21794, 21808, 21875, 21876, 21877, 21730, 21815, 21951, 21900, 21100, 21186, 21593, 21655, 21704, 21747, 21751, 21795, 21802, 21984, 21950, 21961, 21919, 21757, 22019, 22233, 21989, 22039, 22032, 22050, 21955, 22037, 22046, 21985, 22002, 22054, 22065, 22028, 22033, 22000, 22062, 21943, 22108, 22038, 21790, 22058, 22068, 21920, 21983, 22006, 22022, 22082, 22078, 22072, 22075, 22076, 22024, 22080, 22023, 22079, 22090, 22016, 21938, 21719, 21753, 22104, 22099, 22262, 22052, 22034, 22029, 22117, 21986, 21496, 22119, 22073, 22126, 22135, 22142, 22056, 21807, 22148, 22150, 22151, 22161, 22178, 22163, 22069, 22125, 22173, 22176, 22179, 22180, 22185, 22187, 22192, 22193, 22195, 22198, 22084, 22162, 22203, 22209, 22153, 21557, 22225, 22186, 22234, 22201, 22239, 22260, 22246, 22250, 22085, 22265, 22261, 22259, 22270, 21626, 22274, 22275, 22277, 22278, 22279, 22293, 22281, 22283, 22291, 22304, 22300, 22302, 22312, 22308, 22309, 22318, 22305, 22311, 22263, 22327, 22326, 22350, 22325, 22330, 22243, 32038, 32006, 22355, 32075, 32089, 21667, 21713, 32108, 21603, 22264, 32100, 32118, 32068, 22183, 22149, 32112, 32011, 32141, 32139, 22204, 32121, 32097, 32156, 32063, 32037, 32160, 32124, 32036, 32149, 32166, 32150, 32093, 32179, 32198, 21609, 32104, 32146, 32136, 32177, 32079, 32067, 32155, 32193, 32255, 32209, 32211, 32222, 32123, 32204, 32205, 32190, 32228, 32231, 32259, 21504, 32216, 32212, 32242, 32248, 32247, 32249, 32263, 32213, 32214, 32265, 32220, 32047, 32274, 32273, 32128, 32275, 32226, 32061, 32246, 32289, 32188, 32192, 32290, 32291, 32294, 32300, 32309, 32243, 21954, 32215, 22071, 21499, 21799, 22021, 21008, 21001, 21083, 21276, 22020, 21897, 21763, 21525, 22287, 32116, 32315, 32317, 32318, 32230, 32306, 32319, 32329, 32321, 32326, 32325, 32332, 32328, 32298, 32241, 32343, 32208, 32345, 32276, 32302, 32354, 32356, 32357, 32288, 32096, 32362, 32365, 32369, 32371, 32304, 32379, 32303, 32372, 32293, 32197, 32380, 32382, 32383, 32386, 32389, 32301, 32367, 32394, 32395, 32396, 32399, 32404, 32400, 32401, 32402, 32715, 32407, 32408, 32387, 32105, 32409, 32415, 32413, 32414, 32416, 32417, 32181, 32420, 32421, 32427, 32425, 32447, 32428, 32429, 32772, 32431, 32342, 32432, 32433, 32434, 32442, 32435, 32437, 32440, 33625, 32552, 32439, 32448, 32451, 32453, 32459, 32460, 32338, 32232, 32466, 32471, 32472, 32475, 21435, 32476, 32464, 32480, 32482, 32418, 32483, 32485, 32486, 32491, 32495, 32183, 32497, 32499, 32377, 32501, 32503, 21190, 32504, 21191, 32519, 32509, 32510, 32514, 32518, 32520, 32521, 32522, 32523, 32524, 32525, 32526, 32527, 32528, 32529, 32490, 32530, 32531, 32532, 32533, 32534, 32535, 32536, 32537, 32538, 32540, 32541, 32051, 32544, 32545, 32546, 32548, 32549, 32339, 32551, 32358, 32462, 22214, 22251, 21188, 32560, 32144, 21810, 21467, 21924, 32101, 32039, 22112, 21425, 32261, 22063, 32187, 32436, 22165, 22170, 21422, 32392, 22269, 32161, 22145, 21956, 32153, 32310, 21594, 21487, 32224, 32199, 21196, 21571, 22226, 21779, 32282, 22232, 22128, 22335, 32558, 32559, 32515, 32561, 32562, 32411, 32563, 32555, 32567, 32487, 32512, 32572, 32502, 32571, 32494, 32576, 32674, 32554, 32307, 32578, 32505, 32582, 32584, 32585, 32569, 32277, 32602, 32624, 32591, 32592, 32574, 32596, 32597, 32604, 32600, 32601, 32603, 32605, 32606, 32609, 32484, 32669, 32611, 32612, 32613, 32614, 32615, 32617, 32618, 32989, 32493, 32608, 32853, 32619, 32620, 21500, 32625, 32627, 32626, 32628, 32632, 32629, 32565, 32630, 32750, 32631, 32633, 32637, 32635, 32638, 32639, 32640, 32642, 22242, 32645, 32646, 32647, 32648, 32649, 32650, 32570, 32651, 32652, 32654, 32658, 32656, 32657, 32660, 32598, 32662, 32664, 32667, 32671, 32677, 32679, 32681, 32683, 32684, 32686, 32688, 32690, 21988, 32691, 32692, 32693, 32694, 32695, 32697, 32700, 32701, 32702, 32706, 32711, 32703, 32704, 32705, 32708, 32746, 32713, 32714, 32659, 32720, 32721, 32723, 32731, 32740, 32699, 32736, 32728, 32727, 32733, 32737, 32729, 32641, 32748, 32749, 32751, 32758, 32786, 32759, 32768, 32764, 32763, 32765, 32820, 32783, 32808, 32767, 32766, 32824, 32769, 32770, 32159, 32634, 32773, 32587, 32774, 32776, 32777, 32779, 32780, 32781, 32782, 32734, 32785, 32787, 32788, 32789, 32790, 32793, 32795, 32794, 32655, 32797, 32799, 32800, 32801, 32802, 32806, 32809, 32810, 32811, 32813, 32817, 32812, 32814, 32829, 32815, 32775, 32819, 32818, 32821, 32747, 32823, 32756, 32826, 32831, 32832, 32834, 32835, 32836, 32863, 32837, 32852, 32843, 32860, 32842, 32846, 32847, 32848, 32850, 32855, 32856, 32861, 32865, 32866, 32867, 32868, 32869, 32870, 21096, 32871, 32872, 32873, 32874, 32877, 32878, 32879, 32887, 32880, 32888, 21197, 21569, 32881, 32882, 32229, 32885, 32886, 32889, 32890, 32892, 32753, 32893, 32894, 32897, 32964, 32901, 32862, 32898, 32899, 32974, 32900, 22210, 32902, 32903, 32905, 32908, 32909, 32517, 32911, 32912, 32913, 32914, 32916, 32917, 32922, 32921, 32925, 32931, 32926, 32927, 32928, 32933, 32934, 32936, 32940, 32876, 32595, 32944, 32959, 32945, 32948, 32950, 32951, 32952, 32955, 32954, 32957, 33067, 32962, 32965, 32967, 32370, 32968, 32969, 32970, 32971, 32975, 32976, 32977, 32979, 32980, 32983, 32981, 32982, 32986, 32987, 21508, 32938, 32991, 32932, 32988, 32943, 32992, 32997, 32995, 32998, 33000, 33001, 33002, 33004, 33006, 33007, 33010, 33008, 33009, 33012, 33011, 32825, 33024, 33015, 21993, 33016, 33017, 33018, 33020, 33021, 33027, 33029, 32376, 32003, 33031, 33032, 33030, 33036, 21224, 33040, 33041, 33042, 33046, 33049, 33013, 33057, 33055, 33056, 33064, 33063, 33061, 33060, 32996, 33065, 33066, 21168, 33068, 33070, 33071, 33087, 33073, 33076, 33078, 33081, 33088, 33123, 33142, 21589, 33096, 32738, 33274, 33101, 33102, 32741, 32985, 33107, 32956, 33110, 33112, 33120, 33115, 33047, 33026, 33116, 33117, 33118, 33119, 33099, 33053, 32930, 33108, 33109, 32816, 33128, 33127, 33105, 33129, 33132, 33043, 33133, 33134, 33135, 33136, 33097, 32716, 33158, 33146, 33148, 33111, 33083, 33161, 33091, 33163, 33124, 33164, 33140, 33166, 33167, 33168, 33170, 33094, 33157, 33144, 33172, 33173, 33174, 33171, 33143, 33153, 33152, 33181, 33150, 33147, 33160, 33182, 33155, 32994, 33090, 33156, 33141, 33187, 33188, 33189, 33190, 33191, 33192, 33194, 33195, 33196, 33197, 33200, 33201, 33202, 33203, 33204, 33175, 33214, 33216, 33222, 33223, 33186, 33224, 33165, 33225, 33221, 33226, 33227, 33183, 33228, 33231, 33229, 33079, 32993, 33232, 33233, 33205, 33178, 33236, 33237, 33238, 33239, 33241, 33125, 33242, 33245, 33248, 33249, 33149, 33218, 33254, 33255, 33256, 33139, 33151, 33259, 33260, 33261, 33263, 33265, 33266, 33267, 33268, 33272, 33185, 21545, 33275, 33278, 33291, 33292, 33093, 33282, 33206, 33283, 33207, 33284, 33289, 33286, 33287, 33244, 33209, 33253, 33293, 33294, 33296, 33297, 33299, 33300, 33302, 33301, 33303, 33304, 33219, 33180, 33305, 33306, 33177, 33240, 33311, 33258, 33313, 33154, 33314, 33251, 32875, 33234, 33235, 33318, 33176, 33089, 33084, 33329, 33330, 33335, 33351, 33338, 33269, 33336, 33339, 33341, 33342, 33346, 33343, 33344, 33345, 33279, 33348, 33347, 33349, 33252, 33369, 33360, 33361, 33363, 33312, 33362, 33364, 33365, 33367, 33368, 32973, 33370, 33372, 33392, 33471, 33375, 33374, 33376, 33328, 33377, 33379, 33380, 33382, 33383, 33386, 33387, 33388, 33452, 33389, 33058, 33098, 33394, 33393, 33080, 33398, 33397, 33396, 33399, 33402, 33406, 33408, 33409, 33410, 33412, 33413, 33414, 33416, 33454, 33456, 33417, 33418, 33419, 33421, 33423, 33422, 33426, 33432, 33504, 33428, 33430, 33425, 33433, 33435, 33434, 33462, 33627, 33438, 33925, 33439, 33441, 33443, 33444, 33445, 33446, 33447, 33449, 33453, 33455, 33458, 33463, 33464, 33465, 33470, 33407, 33468, 33474, 33476, 33475, 33130, 33378, 33478, 33479, 33480, 33483, 33491, 33482, 33484, 32978, 33485, 33488, 33490, 33492, 33493, 33494, 33086, 33497, 33499, 33500, 33333, 33050, 32424, 33501, 33502, 33503, 33505, 33506, 33507, 33508, 33509, 33510, 33512, 33515, 33514, 33516, 33250, 33095, 33517, 33518, 33519, 33521, 33523, 33524, 33525, 33526, 33527, 33298, 33528, 33530, 33531, 33532, 33533, 33537, 21868, 33539, 33541, 33540, 33542, 33544, 33543, 33545, 33547, 33548, 33549, 33550, 33551, 33552, 22130, 33553, 33554, 33555, 33535, 33557, 33556, 33560, 33558, 33562, 33561, 33563, 33564, 33566, 33568, 33569, 33570, 33571, 33459, 33575, 33576, 33577, 33322, 33578, 33579, 33581, 33582, 33584, 33585, 33586, 33588, 33589, 33590, 33587, 33591, 33595, 33594, 33596, 33597, 33605, 33603, 33606, 33608, 33607, 33609, 33610, 33611, 33340, 33632, 33626, 33633, 33628, 33285, 21982, 33634, 33636, 33638, 33637, 33668, 33644, 33642, 33643, 33645, 33646, 33647, 33648, 33649, 33650, 33651, 33670, 33652, 33655, 33654, 33656, 33658, 33669, 33660, 33661, 21639, 33662, 33664, 33665, 33666, 33684, 33673, 33674, 33675, 33676, 33680, 72029, 33681, 32644, 33683, 72010, 33688, 33687, 33689, 33691, 33690, 33692, 33696, 33695, 33131, 72003, 21211, 21550, 72007, 33052, 33700, 33701, 33702, 72005, 33704, 21264, 33705, 33706, 33851, 33710, 72016, 72015, 72001, 33713, 33714, 33715, 33717, 33722, 33723, 33724, 33726, 33727, 32792, 33728, 33730, 33731, 33736, 33699, 33534, 33740, 33741, 33742, 33746, 33747, 33748, 33749, 33751, 33750, 33752, 33754, 33756, 33755, 72012, 72018, 72020, 72019, 72160, 72022, 72023, 72025, 72247, 72154, 72038, 72157, 72032, 72036, 72034, 33764, 33766, 33767, 33768, 72040, 33774, 72153, 33771, 33772, 33773, 72159, 72156, 72136, 72047, 72048, 72049, 72052, 33777, 33778, 33779, 33781, 72053, 33782, 33854, 33085, 33784, 33787, 33788, 33257, 72249, 33718, 33806, 33849, 33815, 33850, 72145, 33794, 33797, 33072, 22159, 72134, 33823, 33805, 33825, 33828, 33830, 33852, 33836, 33837, 72111, 72113, 72112, 33838, 33839, 33848, 33842, 33846, 33841, 72109, 33807, 72114, 72118, 72120, 72119, 72121, 72122, 72124, 72123, 72126, 72128, 72129, 33853, 72137, 72138, 72140, 72139, 72141, 72143, 72142, 72144, 72147, 72148, 72150, 72151, 72152, 33855, 33856, 33858, 33859, 72168, 72170, 33865, 33864, 72174, 72171, 72251, 33866, 72173, 33867, 36001, 33883, 33872, 33873, 33874, 33875, 33942, 33880, 33881, 33884, 33885, 72175, 72177, 72176, 32272, 33887, 33888, 72183, 72013, 21796, 21996, 22041, 33138, 32143, 33895, 33896, 33897, 33898, 33894, 33900, 33902, 33972, 33903, 33906, 33909, 33910, 72185, 33928, 33911, 72194, 33916, 33947, 33955, 33919, 33920, 72186, 72188, 72189, 33941, 33921, 33922, 72190, 21925, 72192, 72193, 72195, 72196, 72199, 72200, 33924, 72201, 72202, 72203, 72204, 32148, 33932, 33933, 33934, 33939, 72207, 72217, 72255, 72253, 72260, 72257, 72256, 72259, 72254, 72258, 33938, 33940, 72221, 33943, 32346, 21269, 33935, 33936, 72223, 72180, 33945, 72227, 33958, 33948, 33949, 72264, 33983, 33950, 33952, 32233, 36005, 32351, 72228, 33954, 33961, 72230, 72231, 33957, 33963, 33959, 72232, 33960, 33962, 72237, 33964, 72236, 72233, 33956, 33965, 33981, 33966, 72235, 72234, 33967, 72238, 33968, 72240, 33973, 33977, 33974, 72309, 33975, 33976, 33978, 72244, 33979, 32201, 33984, 32210, 72265, 72266, 33982, 72267, 72269, 33985, 33986, 36002, 22217, 33988, 72270, 72271, 72273, 33989, 36000, 33991, 33994, 33992, 36009, 33996, 33995, 21556, 36004, 36006, 36007, 36008, 72279, 72224, 36013, 36015, 36016, 36017, 36018, 36010, 21630, 72280, 72281, 36019, 72287, 36020, 32689, 72288, 36023, 32712, 72290, 72292, 36021, 36022, 22095, 36024, 36027, 36028, 36025, 72294, 36026, 36030, 36029, 36032, 36031, 36033, 72304, 72301, 72302, 36034, 36035, 36036, 36037, 32581, 72303, 36038, 36041, 36042, 36040, 36056, 36043, 72278, 36045, 36046, 36047, 36048, 72306, 36049, 36050, 36057, 36051, 36062, 36054, 36065, 32907, 36058, 72308, 22315, 36061, 36069, 72310, 36070, 36063, 36071, 72312, 72311, 72313, 36077, 36078, 72316, 36079, 22223, 33122, 36081, 72318, 72317, 36082, 22110, 36083, 36093, 36084, 36091, 36085, 36086, 36092, 36089, 36087, 36088, 72321, 72320, 36095, 72319, 36103, 36108, 36109, 36110, 72226, 36107, 36106, 13051, 36105, 72324, 72325, 21776, 21284, 72326, 72327, 21754, 32001, 32616, 21674, 72116, 72117, 22129, 32668, 21246, 21252, 21474, 21675, 21651, 21495, 32739, 21455, 21076, 21405, 32463, 36099, 22155, 21240, 21320, 21879, 21479, 32080, 21475, 21066, 32151, 21251, 33678, 33679, 33538, 33677, 33686, 32254, 32757, 32430, 21678, 21558, 21113, 32314, 21321, 32002, 21334, 22286, 21009, 21077, 21112, 21254, 32122, 32320, 21494, 32388.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America Llc
- Reason for Recall:
- Patient support table floor plate may be incorrectly installed.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Achieva 1.5T, Magnetic Resonance System.
Product Codes/Lot Numbers:
Model No. 781196, 781296, 781343; UDI-DI n/a; Serial No. 32508, 21935, 32084, 22167, 21313, 21681, 32550, 32717, 32083, 22136, 22319, 32095, 32481, 22001, 21237, 21968, 32474, 32500, 21715, 22245, 21693, 21798, 21572, 22025, 32966, 21045, 22140, 21183, 21962, 32034, 21428, 21192, 21957, 21176, 32171, 22215, 21232, 32085, 22207, 22004, 22066, 21666, 21851, 32390, 21104, 21574, 21272, 21111, 32381, 21598, 21351, 21432, 21965, 21178, 21812, 21805, 21688, 21073, 21120, 21117, 21128, 21085, 21086, 21185, 21162, 21125, 21250, 21153, 21144, 21243, 21235, 21315, 21360, 21265, 21559, 21263, 21244, 21160, 21317, 21342, 21312, 21165, 21274, 21215, 21416, 21136, 21382, 21443, 21454, 21329, 21348, 21226, 21344, 21225, 21297, 21460, 21414, 21281, 21530, 21306, 21444, 21459, 21413, 21430, 21457, 21449, 21052, 21592, 21427, 21431, 21238, 21489, 21488, 21570, 21896, 21625, 21465, 21401, 21578, 21398, 21518, 21295, 21013, 21397, 21134, 21503, 21509, 21624, 21463, 21658, 21588, 21585, 21646, 21647, 21689, 21627, 21711, 21696, 21412, 21699, 21548, 21378, 21654, 21698, 21492, 21707, 21338, 21708, 21690, 21677, 21782, 21787, 21729, 21738, 21731, 21760, 21686, 21914, 22047, 21903, 21750, 21724, 21911, 21794, 21808, 21875, 21876, 21877, 21730, 21815, 21951, 21900, 21100, 21186, 21593, 21655, 21704, 21747, 21751, 21795, 21802, 21984, 21950, 21961, 21919, 21757, 22019, 22233, 21989, 22039, 22032, 22050, 21955, 22037, 22046, 21985, 22002, 22054, 22065, 22028, 22033, 22000, 22062, 21943, 22108, 22038, 21790, 22058, 22068, 21920, 21983, 22006, 22022, 22082, 22078, 22072, 22075, 22076, 22024, 22080, 22023, 22079, 22090, 22016, 21938, 21719, 21753, 22104, 22099, 22262, 22052, 22034, 22029, 22117, 21986, 21496, 22119, 22073, 22126, 22135, 22142, 22056, 21807, 22148, 22150, 22151, 22161, 22178, 22163, 22069, 22125, 22173, 22176, 22179, 22180, 22185, 22187, 22192, 22193, 22195, 22198, 22084, 22162, 22203, 22209, 22153, 21557, 22225, 22186, 22234, 22201, 22239, 22260, 22246, 22250, 22085, 22265, 22261, 22259, 22270, 21626, 22274, 22275, 22277, 22278, 22279, 22293, 22281, 22283, 22291, 22304, 22300, 22302, 22312, 22308, 22309, 22318, 22305, 22311, 22263, 22327, 22326, 22350, 22325, 22330, 22243, 32038, 32006, 22355, 32075, 32089, 21667, 21713, 32108, 21603, 22264, 32100, 32118, 32068, 22183, 22149, 32112, 32011, 32141, 32139, 22204, 32121, 32097, 32156, 32063, 32037, 32160, 32124, 32036, 32149, 32166, 32150, 32093, 32179, 32198, 21609, 32104, 32146, 32136, 32177, 32079, 32067, 32155, 32193, 32255, 32209, 32211, 32222, 32123, 32204, 32205, 32190, 32228, 32231, 32259, 21504, 32216, 32212, 32242, 32248, 32247, 32249, 32263, 32213, 32214, 32265, 32220, 32047, 32274, 32273, 32128, 32275, 32226, 32061, 32246, 32289, 32188, 32192, 32290, 32291, 32294, 32300, 32309, 32243, 21954, 32215, 22071, 21499, 21799, 22021, 21008, 21001, 21083, 21276, 22020, 21897, 21763, 21525, 22287, 32116, 32315, 32317, 32318, 32230, 32306, 32319, 32329, 32321, 32326, 32325, 32332, 32328, 32298, 32241, 32343, 32208, 32345, 32276, 32302, 32354, 32356, 32357, 32288, 32096, 32362, 32365, 32369, 32371, 32304, 32379, 32303, 32372, 32293, 32197, 32380, 32382, 32383, 32386, 32389, 32301, 32367, 32394, 32395, 32396, 32399, 32404, 32400, 32401, 32402, 32715, 32407, 32408, 32387, 32105, 32409, 32415, 32413, 32414, 32416, 32417, 32181, 32420, 32421, 32427, 32425, 32447, 32428, 32429, 32772, 32431, 32342, 32432, 32433, 32434, 32442, 32435, 32437, 32440, 33625, 32552, 32439, 32448, 32451, 32453, 32459, 32460, 32338, 32232, 32466, 32471, 32472, 32475, 21435, 32476, 32464, 32480, 32482, 32418, 32483, 32485, 32486, 32491, 32495, 32183, 32497, 32499, 32377, 32501, 32503, 21190, 32504, 21191, 32519, 32509, 32510, 32514, 32518, 32520, 32521, 32522, 32523, 32524, 32525, 32526, 32527, 32528, 32529, 32490, 32530, 32531, 32532, 32533, 32534, 32535, 32536, 32537, 32538, 32540, 32541, 32051, 32544, 32545, 32546, 32548, 32549, 32339, 32551, 32358, 32462, 22214, 22251, 21188, 32560, 32144, 21810, 21467, 21924, 32101, 32039, 22112, 21425, 32261, 22063, 32187, 32436, 22165, 22170, 21422, 32392, 22269, 32161, 22145, 21956, 32153, 32310, 21594, 21487, 32224, 32199, 21196, 21571, 22226, 21779, 32282, 22232, 22128, 22335, 32558, 32559, 32515, 32561, 32562, 32411, 32563, 32555, 32567, 32487, 32512, 32572, 32502, 32571, 32494, 32576, 32674, 32554, 32307, 32578, 32505, 32582, 32584, 32585, 32569, 32277, 32602, 32624, 32591, 32592, 32574, 32596, 32597, 32604, 32600, 32601, 32603, 32605, 32606, 32609, 32484, 32669, 32611, 32612, 32613, 32614, 32615, 32617, 32618, 32989, 32493, 32608, 32853, 32619, 32620, 21500, 32625, 32627, 32626, 32628, 32632, 32629, 32565, 32630, 32750, 32631, 32633, 32637, 32635, 32638, 32639, 32640, 32642, 22242, 32645, 32646, 32647, 32648, 32649, 32650, 32570, 32651, 32652, 32654, 32658, 32656, 32657, 32660, 32598, 32662, 32664, 32667, 32671, 32677, 32679, 32681, 32683, 32684, 32686, 32688, 32690, 21988, 32691, 32692, 32693, 32694, 32695, 32697, 32700, 32701, 32702, 32706, 32711, 32703, 32704, 32705, 32708, 32746, 32713, 32714, 32659, 32720, 32721, 32723, 32731, 32740, 32699, 32736, 32728, 32727, 32733, 32737, 32729, 32641, 32748, 32749, 32751, 32758, 32786, 32759, 32768, 32764, 32763, 32765, 32820, 32783, 32808, 32767, 32766, 32824, 32769, 32770, 32159, 32634, 32773, 32587, 32774, 32776, 32777, 32779, 32780, 32781, 32782, 32734, 32785, 32787, 32788, 32789, 32790, 32793, 32795, 32794, 32655, 32797, 32799, 32800, 32801, 32802, 32806, 32809, 32810, 32811, 32813, 32817, 32812, 32814, 32829, 32815, 32775, 32819, 32818, 32821, 32747, 32823, 32756, 32826, 32831, 32832, 32834, 32835, 32836, 32863, 32837, 32852, 32843, 32860, 32842, 32846, 32847, 32848, 32850, 32855, 32856, 32861, 32865, 32866, 32867, 32868, 32869, 32870, 21096, 32871, 32872, 32873, 32874, 32877, 32878, 32879, 32887, 32880, 32888, 21197, 21569, 32881, 32882, 32229, 32885, 32886, 32889, 32890, 32892, 32753, 32893, 32894, 32897, 32964, 32901, 32862, 32898, 32899, 32974, 32900, 22210, 32902, 32903, 32905, 32908, 32909, 32517, 32911, 32912, 32913, 32914, 32916, 32917, 32922, 32921, 32925, 32931, 32926, 32927, 32928, 32933, 32934, 32936, 32940, 32876, 32595, 32944, 32959, 32945, 32948, 32950, 32951, 32952, 32955, 32954, 32957, 33067, 32962, 32965, 32967, 32370, 32968, 32969, 32970, 32971, 32975, 32976, 32977, 32979, 32980, 32983, 32981, 32982, 32986, 32987, 21508, 32938, 32991, 32932, 32988, 32943, 32992, 32997, 32995, 32998, 33000, 33001, 33002, 33004, 33006, 33007, 33010, 33008, 33009, 33012, 33011, 32825, 33024, 33015, 21993, 33016, 33017, 33018, 33020, 33021, 33027, 33029, 32376, 32003, 33031, 33032, 33030, 33036, 21224, 33040, 33041, 33042, 33046, 33049, 33013, 33057, 33055, 33056, 33064, 33063, 33061, 33060, 32996, 33065, 33066, 21168, 33068, 33070, 33071, 33087, 33073, 33076, 33078, 33081, 33088, 33123, 33142, 21589, 33096, 32738, 33274, 33101, 33102, 32741, 32985, 33107, 32956, 33110, 33112, 33120, 33115, 33047, 33026, 33116, 33117, 33118, 33119, 33099, 33053, 32930, 33108, 33109, 32816, 33128, 33127, 33105, 33129, 33132, 33043, 33133, 33134, 33135, 33136, 33097, 32716, 33158, 33146, 33148, 33111, 33083, 33161, 33091, 33163, 33124, 33164, 33140, 33166, 33167, 33168, 33170, 33094, 33157, 33144, 33172, 33173, 33174, 33171, 33143, 33153, 33152, 33181, 33150, 33147, 33160, 33182, 33155, 32994, 33090, 33156, 33141, 33187, 33188, 33189, 33190, 33191, 33192, 33194, 33195, 33196, 33197, 33200, 33201, 33202, 33203, 33204, 33175, 33214, 33216, 33222, 33223, 33186, 33224, 33165, 33225, 33221, 33226, 33227, 33183, 33228, 33231, 33229, 33079, 32993, 33232, 33233, 33205, 33178, 33236, 33237, 33238, 33239, 33241, 33125, 33242, 33245, 33248, 33249, 33149, 33218, 33254, 33255, 33256, 33139, 33151, 33259, 33260, 33261, 33263, 33265, 33266, 33267, 33268, 33272, 33185, 21545, 33275, 33278, 33291, 33292, 33093, 33282, 33206, 33283, 33207, 33284, 33289, 33286, 33287, 33244, 33209, 33253, 33293, 33294, 33296, 33297, 33299, 33300, 33302, 33301, 33303, 33304, 33219, 33180, 33305, 33306, 33177, 33240, 33311, 33258, 33313, 33154, 33314, 33251, 32875, 33234, 33235, 33318, 33176, 33089, 33084, 33329, 33330, 33335, 33351, 33338, 33269, 33336, 33339, 33341, 33342, 33346, 33343, 33344, 33345, 33279, 33348, 33347, 33349, 33252, 33369, 33360, 33361, 33363, 33312, 33362, 33364, 33365, 33367, 33368, 32973, 33370, 33372, 33392, 33471, 33375, 33374, 33376, 33328, 33377, 33379, 33380, 33382, 33383, 33386, 33387, 33388, 33452, 33389, 33058, 33098, 33394, 33393, 33080, 33398, 33397, 33396, 33399, 33402, 33406, 33408, 33409, 33410, 33412, 33413, 33414, 33416, 33454, 33456, 33417, 33418, 33419, 33421, 33423, 33422, 33426, 33432, 33504, 33428, 33430, 33425, 33433, 33435, 33434, 33462, 33627, 33438, 33925, 33439, 33441, 33443, 33444, 33445, 33446, 33447, 33449, 33453, 33455, 33458, 33463, 33464, 33465, 33470, 33407, 33468, 33474, 33476, 33475, 33130, 33378, 33478, 33479, 33480, 33483, 33491, 33482, 33484, 32978, 33485, 33488, 33490, 33492, 33493, 33494, 33086, 33497, 33499, 33500, 33333, 33050, 32424, 33501, 33502, 33503, 33505, 33506, 33507, 33508, 33509, 33510, 33512, 33515, 33514, 33516, 33250, 33095, 33517, 33518, 33519, 33521, 33523, 33524, 33525, 33526, 33527, 33298, 33528, 33530, 33531, 33532, 33533, 33537, 21868, 33539, 33541, 33540, 33542, 33544, 33543, 33545, 33547, 33548, 33549, 33550, 33551, 33552, 22130, 33553, 33554, 33555, 33535, 33557, 33556, 33560, 33558, 33562, 33561, 33563, 33564, 33566, 33568, 33569, 33570, 33571, 33459, 33575, 33576, 33577, 33322, 33578, 33579, 33581, 33582, 33584, 33585, 33586, 33588, 33589, 33590, 33587, 33591, 33595, 33594, 33596, 33597, 33605, 33603, 33606, 33608, 33607, 33609, 33610, 33611, 33340, 33632, 33626, 33633, 33628, 33285, 21982, 33634, 33636, 33638, 33637, 33668, 33644, 33642, 33643, 33645, 33646, 33647, 33648, 33649, 33650, 33651, 33670, 33652, 33655, 33654, 33656, 33658, 33669, 33660, 33661, 21639, 33662, 33664, 33665, 33666, 33684, 33673, 33674, 33675, 33676, 33680, 72029, 33681, 32644, 33683, 72010, 33688, 33687, 33689, 33691, 33690, 33692, 33696, 33695, 33131, 72003, 21211, 21550, 72007, 33052, 33700, 33701, 33702, 72005, 33704, 21264, 33705, 33706, 33851, 33710, 72016, 72015, 72001, 33713, 33714, 33715, 33717, 33722, 33723, 33724, 33726, 33727, 32792, 33728, 33730, 33731, 33736, 33699, 33534, 33740, 33741, 33742, 33746, 33747, 33748, 33749, 33751, 33750, 33752, 33754, 33756, 33755, 72012, 72018, 72020, 72019, 72160, 72022, 72023, 72025, 72247, 72154, 72038, 72157, 72032, 72036, 72034, 33764, 33766, 33767, 33768, 72040, 33774, 72153, 33771, 33772, 33773, 72159, 72156, 72136, 72047, 72048, 72049, 72052, 33777, 33778, 33779, 33781, 72053, 33782, 33854, 33085, 33784, 33787, 33788, 33257, 72249, 33718, 33806, 33849, 33815, 33850, 72145, 33794, 33797, 33072, 22159, 72134, 33823, 33805, 33825, 33828, 33830, 33852, 33836, 33837, 72111, 72113, 72112, 33838, 33839, 33848, 33842, 33846, 33841, 72109, 33807, 72114, 72118, 72120, 72119, 72121, 72122, 72124, 72123, 72126, 72128, 72129, 33853, 72137, 72138, 72140, 72139, 72141, 72143, 72142, 72144, 72147, 72148, 72150, 72151, 72152, 33855, 33856, 33858, 33859, 72168, 72170, 33865, 33864, 72174, 72171, 72251, 33866, 72173, 33867, 36001, 33883, 33872, 33873, 33874, 33875, 33942, 33880, 33881, 33884, 33885, 72175, 72177, 72176, 32272, 33887, 33888, 72183, 72013, 21796, 21996, 22041, 33138, 32143, 33895, 33896, 33897, 33898, 33894, 33900, 33902, 33972, 33903, 33906, 33909, 33910, 72185, 33928, 33911, 72194, 33916, 33947, 33955, 33919, 33920, 72186, 72188, 72189, 33941, 33921, 33922, 72190, 21925, 72192, 72193, 72195, 72196, 72199, 72200, 33924, 72201, 72202, 72203, 72204, 32148, 33932, 33933, 33934, 33939, 72207, 72217, 72255, 72253, 72260, 72257, 72256, 72259, 72254, 72258, 33938, 33940, 72221, 33943, 32346, 21269, 33935, 33936, 72223, 72180, 33945, 72227, 33958, 33948, 33949, 72264, 33983, 33950, 33952, 32233, 36005, 32351, 72228, 33954, 33961, 72230, 72231, 33957, 33963, 33959, 72232, 33960, 33962, 72237, 33964, 72236, 72233, 33956, 33965, 33981, 33966, 72235, 72234, 33967, 72238, 33968, 72240, 33973, 33977, 33974, 72309, 33975, 33976, 33978, 72244, 33979, 32201, 33984, 32210, 72265, 72266, 33982, 72267, 72269, 33985, 33986, 36002, 22217, 33988, 72270, 72271, 72273, 33989, 36000, 33991, 33994, 33992, 36009, 33996, 33995, 21556, 36004, 36006, 36007, 36008, 72279, 72224, 36013, 36015, 36016, 36017, 36018, 36010, 21630, 72280, 72281, 36019, 72287, 36020, 32689, 72288, 36023, 32712, 72290, 72292, 36021, 36022, 22095, 36024, 36027, 36028, 36025, 72294, 36026, 36030, 36029, 36032, 36031, 36033, 72304, 72301, 72302, 36034, 36035, 36036, 36037, 32581, 72303, 36038, 36041, 36042, 36040, 36056, 36043, 72278, 36045, 36046, 36047, 36048, 72306, 36049, 36050, 36057, 36051, 36062, 36054, 36065, 32907, 36058, 72308, 22315, 36061, 36069, 72310, 36070, 36063, 36071, 72312, 72311, 72313, 36077, 36078, 72316, 36079, 22223, 33122, 36081, 72318, 72317, 36082, 22110, 36083, 36093, 36084, 36091, 36085, 36086, 36092, 36089, 36087, 36088, 72321, 72320, 36095, 72319, 36103, 36108, 36109, 36110, 72226, 36107, 36106, 13051, 36105, 72324, 72325, 21776, 21284, 72326, 72327, 21754, 32001, 32616, 21674, 72116, 72117, 22129, 32668, 21246, 21252, 21474, 21675, 21651, 21495, 32739, 21455, 21076, 21405, 32463, 36099, 22155, 21240, 21320, 21879, 21479, 32080, 21475, 21066, 32151, 21251, 33678, 33679, 33538, 33677, 33686, 32254, 32757, 32430, 21678, 21558, 21113, 32314, 21321, 32002, 21334, 22286, 21009, 21077, 21112, 21254, 32122, 32320, 21494, 32388.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1550-2024
Related Recalls
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Philips North America
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
IntelliVue MP90. Product Number: M8010A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.
IntelliVue MP30. Product Number: M8002A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.