Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The TRD is intended to distribute patient case, appropriate parts information, track changes, and transform test requests from LIS or manual entries to connected Dako systems.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    N/A Serial Number: 1311004352, 1308004862, 1307000537, 1405002266, 1410008844, 1307006995, 1311001075, 1402002186, 1401010308, 1402005995, 1306001697, 1311007317, 1410004066, 1404006200, 1410004067, 1405002263, 1404007225, 1410002144, 1311007321, 1404004226, 1401005960, 1305006010, 1401000919, 1405000371, 1401005963, 1403005465, 1304008205, 1401003711, 1401001048, 1310010425, 1401003715, 1311004354, 1311004355
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Dako North America Inc.
Reason for Recall:
If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The TRD is intended to distribute patient case, appropriate parts information, track changes, and transform test requests from LIS or manual entries to connected Dako systems.

Product Codes/Lot Numbers:

N/A Serial Number: 1311004352, 1308004862, 1307000537, 1405002266, 1410008844, 1307006995, 1311001075, 1402002186, 1401010308, 1402005995, 1306001697, 1311007317, 1410004066, 1404006200, 1410004067, 1405002263, 1404007225, 1410002144, 1311007321, 1404004226, 1401005960, 1305006010, 1401000919, 1405000371, 1401005963, 1403005465, 1304008205, 1401003711, 1401001048, 1310010425, 1401003715, 1311004354, 1311004355

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1548-2015

Related Recalls

An incorrect HER2 probe concentration was used when producing the probe component (Vial 3) in this kit lot. This has resulted in the HER2 probe concentration being too low.

Jun 10, 2016 Diagnostic Equipment Nationwide View Details →

There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.

Sep 4, 2015 Diagnostic Equipment Nationwide View Details →