Maquet Getinge-T 8029 PRESSURE PIGTAIL 6" Material: 701055995
Class I - DangerousWhat Should You Do?
- Check if you have this product: Batch Numbers: 3000058045 3000069467 3000072507 3000080702
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Maquet Cardiovascular, LLC
- Reason for Recall:
- Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Maquet Getinge-T 8029 PRESSURE PIGTAIL 6" Material: 701055995
Product Codes/Lot Numbers:
Batch Numbers: 3000058045 3000069467 3000072507 3000080702
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1545-2019
Related Recalls
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Heartstring III Proximal Seal System. Intravascular anastomosis occluder.
Maquet Cardiovascular
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
Maquet Cardiovascular
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.