Ultraview SL Command Modules, Model 91496-M, upgraded with the Masimo SpO2 Option. Product Usage: The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use. Physiological parameters that may be monitored include ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (Sp02) and cardiac output. Acquired data may then be communicated to all information network for display, recording, editing and analysis.
Class I - DangerousWhat Should You Do?
- Check if you have this product: SERIAL NUMBERS IN THE US: 1496-000473, 1496-000590, 1496-001126, 1496-001189, 1496-001214, 1496-001299, 1496-001300, 1496-001301, 1496-001302, 1496-001303, 1496-001304, 1496-001305, 1496-001306, 1496-001307, 1496-001309, 1496-001328, 1496-001329, 1496-001330, 1496-001332, 1496-001333, 1496-001335, 1496-001336, 1496-001337, 1496-001712, 1496-001713, 1496-001714, 1496-002442, 1496-002568, 1496-002691, 1496-002743, 1496-002832, 1496-002833, 1496-002834, 1496-002836, 1496-002837, 1496-002838, 1496-002839, 1496-002840, 1496-002841, 1496-002917, 1496-002918, 1496-002919, 1496-002920, 1496-002998, 1496-002999, 1496-003000, 1496-003001, 1496-003002, 1496-003003, 1496-003005, 1496-003006, 1496-003007, 1496-003021, 1496-003022, 1496-003025, 1496-003035, 1496-003036, 1496-003037, 1496-003042, 1496-003043, 1496-003044, 1496-003046, 1496-003049, 1496-003051, 1496-003055, 1496-003158, 1496-003161, 1496-003162, 1496-003163, 1496-003164, 1496-003179, 1496-003187, 1496-003190, 1496-003191, 1496-003192, 1496-003193, 1496-003210, 1496-003528, 1496-003602, 1496-003796, 1496-003797, 1496-003819, 1496-004075, 1496-004081, 1496-004082, 1496-004088, 1496-004095, 1496-004098, 1496-004100, 1496-004103, 1496-004104, 1496-004110, 1496-004174, 1496-004175, 1496-004222, 1496-004302, 1496-004382, 1496-004385, 1496-004386, 1496-004387, 1496-004390, 1496-004392, 1496-004394, 1496-004622, 1496-004667, 1496-004674, 1496-004675, 1496-004676, 1496-004737, 1496-004762, 1496-004805, 1496-004806, 1496-004886, 1496-004892, 1496-004951, 1496-004980, 1496-005204, 1496-005291, 1496-005292, 1496-005293, 1496-005294, 1496-005342, 1496-005343, 1496-005358, 1496-005359, 1496-005371, 1496-005374, 1496-005380, 1496-005611, 1496-005779, 1496-005780, 1496-005781, 1496-005782, 1496-005785, 1496-005786, 1496-005787, 1496-005889, 1496-005890, 1496-005891, 1496-005892, 1496-005894, 1496-005896, 1496-005897, 1496-005898, 1496-005899, 1496-005900, 1496-005901, 1496-005903, 1496-005904, 1496-005905, 1496-006038, 1496-006058, 1496-006078, 1496-006089, 1496-006108, 1496-006109, 1496-006110, 1496-006140, 1496-006141, 1496-006142, 1496-006143, 1496-006144, 1496-006145, 1496-006249, 1496-006250, 1496-006251, 1496-006252, 1496-006253, 1496-006254, 1496-006255, 1496-006256, 1496-006257, 1496-006258, 1496-006259, 1496-006260, 1496-006261, 1496-006262, 1496-006266, 1496-006267, 1496-006268, 1496-006269, 1496-006270, 1496-006271, 1496-006272, 1496-006273, 1496-006327, 1496-006328, 1496-006329, 1496-006330, 1496-006331, 1496-006332, 1496-006333, 1496-006334, 1496-006336, 1496-006337, 1496-006338, 1496-006341, 1496-006342, 1496-006343, 1496-006457, 1496-006472, 1496-006523, 1496-006544, 1496-006593, 1496-006594, 1496-006595, 1496-006596, 1496-006597, 1496-006877, 1496-006972, 1496-006973, 1496-006981, 1496-006982, 1496-006983, 1496-006984, 1496-006985, 1496-006986, 1496-006994, 1496-006995, 1496-006996, 1496-006997, 1496-006998, 1496-007132, 1496-007137, 1496-007318, 1496-007319, 1496-007320, 1496-007321, 1496-007322, 1496-007323, 1496-007324, 1496-007325, 1496-007326, 1496-007327, 1496-007328, 1496-007329, 1496-007330, 1496-007331, 1496-007332, 1496-007333, 1496-007334, 1496-007335, 1496-007336, 1496-007337, 1496-007338, 1496-007339, 1496-007340, 1496-007376, 1496-007587, 1496-007589, 1496-007590, 1496-007593, 1496-007594, 1496-007595, 1496-007596, 1496-007631, 1496-007647, 1496-007650, 1496-007651, 1496-007652, 1496-007653, 1496-007654, 1496-007655, 1496-007656, 1496-007657, 1496-007658, 1496-007745, 1496-007750, 1496-007753, 1496-007755, 1496-007756, 1496-007757, 1496-007969, 1496-007970, 1496-008009, 1496-008010, 1496-008011, 1496-008012, 1496-008013, 1496-008014, 1496-008015, 1496-008046, 1496-008047, 1496-008048, 1496-008161, 1496-008162, 1496-008163, 1496-008250, 1496-008319, 1496-008383, 1496-008450, 1496-008451, 1496-008452, 1496-008453, 1496-008454, 1496-008455, 1496-008456, 1496-008580, 1496-008584, 1496-008587, 1496-008873, 1496-009392, 1496-009393, 1496-009394, 1496-009395, 1496-009396, 1496-009397, 1496-009398, 1496-009399, 1496-009400, 1496-009401, 1496-009524, 1496-009584, 1496-009585, 1496-009586, 1496-009587, 1496-009588, 1496-009589, 1496-009590, 1496-009591, 1496-009592, 1496-009593, 1496-009594, 1496-009595, 1496-009596, 1496-009597, 1496-009598, 1496-009599, 1496-009600, 1496-009601, 1496-009602, 1496-009603, 1496-009604, 1496-009605, 1496-009606, 1496-009607, 1496-009608, 1496-009715, 1496-009733, 1496-009734, 1496-009737, 1496-009738, 1496-009760, 1496-009761, 1496-009762, 1496-009763, 1496-009764, 1496-009765, 1496-009766, 1496-009767, 1496-009778, 1496-009779, 1496-009781, 1496-009782, 1496-009783, 1496-009784, 1496-009785, 1496-009786, 1496-009787, 1496-009788, 1496-009856, 1496-009861, 1496-009862, 1496-009867, 1496-009874, 1496-009877, 1496-009879, 1496-009963, 1496-009968, 1496-010008, 1496-010009, 1496-010010, 1496-010011, 1496-010012, 1496-010013, 1496-010017, 1496-010018, 1496-010019, 1496-010068, 1496-010113, 1496-010252, 1496-010260, 1496-010273, 1496-010278, 1496-010287, 1496-010288, 1496-010290, 1496-010291, 1496-010292, 1496-010293, 1496-010294, 1496-010295, 1496-010296, 1496-010297, 1496-010298, 1496-010299, 1496-010300, 1496-010301, 1496-010302, 1496-010469, 1496-010470, 1496-010471, 1496-010472, 1496-010473, 1496-010474, 1496-010475, 1496-010476, 1496-010477, 1496-010478, 1496-010523, 1496-010524, 1496-010526, 1496-010527, 1496-010528, 1496-010529, 1496-010530, 1496-010531, 1496-010532, 1496-010533, 1496-010534, 1496-010535, 1496-010537, 1496-010539, 1496-010540, 1496-010541, 1496-010542, 1496-010543, 1496-010545, 1496-010546, 1496-010547, 1496-010548, 1496-010550, 1496-010691, 1496-010692, 1496-010809, 1496-010830, 1496-010927, 1496-010933, 1496-010952, 1496-010959, 1496-010960, 1496-011059, 1496-011060, 1496-011069, 1496-011070, 1496-011210, 1496-011211, 1496-011246, 1496-011247, 1496-011248, 1496-011249, 1496-011250, 1496-011252, 1496-011253, 1496-011256, 1496-011259, 1496-011306, 1496-011307, 1496-011308, 1496-011309, 1496-011310, 1496-011311, 1496-011312, 1496-011313, 1496-011314, 1496-011315, 1496-011317, 1496-011552, 1496-011555, 1496-011716, 1496-011894, 1496-011902, 1496-011908, 1496-011923, 1496-012278, 1496-012280, 1496-012307, 1496-012320, 1496-012343, 1496-012426, 1496-012429, 1496-012431, 1496-012432, 1496-012434, 1496-012470, 1496-012478, 1496-012495, 1496-012496, 1496-012497, 1496-012615, 1496-012616, 1496-012617, 1496-012618, 1496-012620, 1496-012621, 1496-012623, 1496-012624, 1496-012625, 1496-012626, 1496-012628, 1496-012629, 1496-012630, 1496-012631, 1496-012632, 1496-012633, 1496-012634, 1496-012635, 1496-012636, 1496-012637, 1496-012638, 1496-012639, 1496-012640, 1496-012642, 1496-013314, 1496-013383, 1496-013563, 1496-013564, 1496-013565, 1496-013566, 1496-013567, 1496-013568, 1496-013569, 1496-013570, 1496-013571, 1496-013572, 1496-013573, 1496-013574, 1496-013575, 1496-013578, 1496-013579, 1496-013580, 1496-013583, 1496-013584, 1496-013585, 1496-013586, 1496-013587, 1496-013588, 1496-013590, 1496-013591, 1496-013683, 1496-013705, 1496-013706, 1496-013709, 1496-013710, 1496-013711, 1496-013712, 1496-013713, 1496-013732, 1496-013733, 1496-013734, 1496-013735, 1496-013998, 1496-013999, 1496-014000, 1496-014001, 1496-014002, 1496-014003, 1496-014004, 1496-014005, 1496-014006, 1496-014007, 1496-014008, 1496-014009, 1496-014010, 1496-014011, 1496-014012, 1496-014015, 1496-014016, 1496-014017, 1496-014018, 1496-014019, 1496-014020, 1496-014065, 1496-014135, 1496-014136, 1496-014172, 1496-014173, 1496-014174, 1496-014175, 1496-014177, 1496-014178, 1496-014179, 1496-014180, 1496-014181, 1496-014339, 1496-014340, 1496-014341, 1496-014342, 1496-014343, 1496-014344, 1496-014345, 1496-014346, 1496-014348, 1496-014349, 1496-014350, 1496-014351, 1496-014352, 1496-014353, 1496-014354, 1496-014355, 1496-014356, 1496-014357, 1496-014358, 1496-014359, 1496-014360, 1496-014361, 1496-014362, 1496-014445, 1496-014446, 1496-014447, 1496-014448, 1496-014449, 1496-014451, 1496-014462, 1496-014484, 1496-014485, 1496-014486, 1496-014487, 1496-014506, 1496-014517, 1496-014536, 1496-014537, 1496-014542, 1496-014552, 1496-014554, 1496-014555, 1496-014835, 1496-014964, 1496-014965, 1496-014966, 1496-014967, 1496-014968, 1496-014970, 1496-014971, 1496-014972, 1496-014973, 1496-014974, 1496-014975, 1496-014976, 1496-014978, 1496-014979, 1496-014980, 1496-014982, 1496-014983, 1496-014984, 1496-014985, 1496-014986, 1496-014988, 1496-015059, 1496-015082, 1496-015084, 1496-015181, 1496-015183, 1496-015184, 1496-015185, 1496-015217, 1496-015218, 1496-015219, 1496-015220, 1496-015221, 1496-015222, 1496-015223, 1496-015224, 1496-015225, 1496-015226, 1496-015227, 1496-015228, 1496-015229, 1496-015230, 1496-015231, 1496-015232, 1496-015233, 1496-015234, 1496-015235, 1496-015236, 1496-015237, 1496-015257, 1496-015258, 1496-015259, 1496-015260, 1496-015261, 1496-015263, 1496-015264, 1496-015265, 1496-015266, 1496-015267, 1496-015268, 1496-015269, 1496-015270, 1496-015272, 1496-015273, 1496-015274, 1496-015276, 1496-015277, 1496-015278, 1496-015279, 1496-015280, 1496-015281, 1496-015282, 1496-015283, 1496-015284, 1496-015285, 1496-015286, 1496-015288, 1496-015289, 1496-015290, 1496-015292, 1496-015293, 1496-015294, 1496-015295, 1496-015296, 1496-015297, 1496-015298, 1496-015299, 1496-015300, 1496-015301, 1496-015302, 1496-015303, 1496-015304, 1496-015312, 1496-016051, 1496-016052, 1496-016053, 1496-016056, 1496-016249, 1496-016250, 1496-016499, 1496-016509, 1496-016526, 1496-016527, 1496-016530, 1496-016804, 1496-016870, 1496-016938, 1496-016946, 1496-016956, 1496-016959, 1496-016960, 1496-016965, 1496-016966, 1496-016967, 1496-016968, 1496-016969, 1496-016970, 1496-016972, 1496-016974, 1496-016975, 1496-016977, 1496-016979, 1496-016980, 1496-016981, 1496-016983, 1496-016984, 1496-016985, 1496-016986, 1496-016987, 1496-017379, 1496-017408, 1496-017409, 1496-017410, 1496-017411, 1496-017412, 1496-017413, 1496-017414, 1496-017415, 1496-017416, 1496-017417, 1496-017418, 1496-017419, 1496-017420, 1496-017421, 1496-017422, 1496-017423, 1496-017424, 1496-017425, 1496-017426, 1496-017427, 1496-017428, 1496-017429, 1496-017430, 1496-017431, 1496-017432, 1496-017433, 1496-017434, 1496-017435, 1496-017436, 1496-017437, 1496-017438, 1496-017439, 1496-017440, 1496-017441, 1496-017442, 1496-017443, 1496-017444, 1496-017445, 1496-017446, 1496-017447, 1496-017448, 1496-017449, 1496-017491, 1496-017492, 1496-017493, 1496-017494, 1496-017547, 1496-017548, 1496-017549, 1496-017550, 1496-017551, 1496-017552, 1496-017555, 1496-017556, 1496-017557, 1496-017558, 1496-017559, 1496-017560, 1496-017561, 1496-017562, 1496-017563, 1496-017564, 1496-017599, 1496-017659, 1496-017660, 1496-017661, 1496-017663, 1496-017664, 1496-017665, 1496-017666, 1496-017667, 1496-017882, 1496-017883, 1496-017884, 1496-017885, 1496-017886, 1496-017888, 1496-017889, 1496-017890, 1496-017891, 1496-017892, 1496-017893, 1496-017895, 1496-017896, 1496-017897, 1496-017898, 1496-017899, 1496-017900, 1496-017901, 1496-017902, 1496-017903, 1496-017904, 1496-017905, 1496-018140, 1496-018277, 1496-018498, 1496-018499, 1496-018501, 1496-018506, 1496-018511, 1496-018512, 1496-018513, 1496-018657, 1496-018658, 1496-018659, 1496-018660, 1496-018661, 1496-018662, 1496-018663, 1496-018664, 1496-018665, 1496-018666, 1496-018667, 1496-018668, 1496-018669, 1496-018670, 1496-018671, 1496-018672, 1496-018673, 1496-018674, 1496-018675, 1496-018676, 1496-018715, 1496-018716, 1496-018717, 1496-018719, 1496-018720, 1496-018721, 1496-018722, 1496-019217, 1496-019221, 1496-019491, 1496-019492, 1496-019584, 1496-020047, 1496-020049, 1496-020073, 1496-020074, 1496-020075, 1496-020076, 1496-020077, 1496-020078, 1496-020079, 1496-020080, 1496-020081, 1496-020082, 1496-020083, 1496-020084, 1496-020085, 1496-020086, 1496-020087, 1496-020088, 1496-020089, 1496-020090, 1496-020091, 1496-020178, 1496-020278, 1496-020279, 1496-020280, 1496-020281, 1496-020282, 1496-020283, 1496-020284, 1496-020293, 1496-020294, 1496-020295, 1496-020296, 1496-020297, 1496-020298, 1496-020308, 1496-020311, 1496-020348, 1496-020349, 1496-020350, 1496-020807, 1496-021310, 1496-021443, 1496-021457, 1496-021458, 1496-021459, 1496-021460, 1496-021461, 1496-021462, 1496-021463, 1496-021464, 1496-021465, 1496-021466, 1496-021467, 1496-021468, 1496-021469, 1496-021470, 1496-021471, 1496-021472, 1496-021473, 1496-021474, 1496-021475, 1496-021476, 1496-021477, 1496-021478, 1496-021479, 1496-021480, 1496-021481, 1496-021482, 1496-021483, 1496-021484, 1496-021485, 1496-021486, 1496-021487, 1496-021488, 1496-021504, 1496-021505, 1496-021506, 1496-021507, 1496-021508, 1496-021509, 1496-021510, 1496-021511, 1496-021512, 1496-021513, 1496-021520, 1496-021677, 1496-021715, 1496-021723, 1496-021743, 1496-021758, 1496-023510, 1496-023558, 1496-023559, 1496-023560, 1496-023561, 1496-023563, 1496-023564, 1496-023565, 1496-023566, 1496-023567, 1496-023858, 1496-024742, 1496-024804, 1496-024832, 1496-025119, 1496-025120, 1496-025121, 1496-025122, 1496-025123, 1496-025124, 1496-025126, 1496-025197, 1496-025199, 1496-025202, 1496-025203, 1496-025204, 1496-025205, 1496-025206, 1496-025207, 1496-025208, 1496-025209, 1496-025234, 1496-025235, 1496-025236, 1496-025237, 1496-025238, 1496-025506, 1496-025512, 1496-025513, 1496-025514, 1496-025515, 1496-025516, 1496-025517, 1496-025518, 1496-025618, 1496-025619, 1496-025654, 1496-025657, 1496-025658, 1496-025660, 1496-025661, 1496-025662, 1496-025663, 1496-025664, 1496-025665, 1496-025666, 1496-025667, 1496-025668, 1496-025669, 1496-025670, 1496-025671, 1496-025672, 1496-025673, 1496-025674, 1496-025675, 1496-025676, 1496-025677, 1496-025678, 1496-025679, 1496-025680, 1496-025681, 1496-025682, 1496-025683, 1496-025684, 1496-025685, 1496-025686, 1496-025687, 1496-025688, 1496-025689, 1496-025690, 1496-025691, 1496-025692, 1496-025693, 1496-025694, 1496-025695, 1496-025696, 1496-025697, 1496-025698, 1496-025699, 1496-025701, 1496-025702, 1496-025704, 1496-025705, 1496-025706, 1496-025707, 1496-025708, 1496-025709, 1496-025710, 1496-025769, 1496-025770, 1496-025771, 1496-025941, 1496-025942, 1496-026067, 1496-026068, 1496-026070, 1496-026072, 1496-026075, 1496-026076, 1496-026077, 1496-026078, 1496-026079, 1496-026080, 1496-026081, 1496-026082, 1496-026083, 1496-026309, 1496-026310, 1496-026311, 1496-026362, 1496-026363, 1496-026364, 1496-026365, 1496-026366, 1496-026367, 1496-026368, 1496-026369, 1496-026370, 1496-026373, 1496-026374, 1496-026429, 1496-027022, 1496-027070, 1496-027071, 1496-027103, 1496-027111, 1496-027112, 1496-027409, 1496-027410, 1496-027411, 1496-027412, 1496-027413, 1496-027414, 1496-027415, 1496-027416, 1496-027472, 1496-027578, 1496-027579, 1496-027580, 1496-027581, 1496-028419, 1496-028442, 1496-028459, 1496-028461, 1496-028462, 1496-028465, 1496-028466, 1496-028467, 1496-028665, 1496-028710, 1496-028711, 1496-028822, 1496-028824, 1496-028825, 1496-028826, 1496-028827, 1496-028829, 1496-028831, 1496-028832, 1496-028833, 1496-028834, 1496-028836, 1496-028837, 1496-028838, 1496-028839, 1496-028840, 1496-028841, 1496-028842, 1496-028843, 1496-028844, 1496-028845, 1496-028846, 1496-028847, 1496-028848, 1496-028850, 1496-028851, 1496-028852, 1496-028853, 1496-028854, 1496-028855, 1496-029028, 1496-029051, 1496-029575, 1496-029595, 1496-029723, 1496-029724, 1496-029725, 1496-029726, 1496-029727, 1496-029728, 1496-029985, 1496-030010, 1496-030932, 1496-030933, 1496-030934, 1496-031144, 1496-031145, 1496-031190, 1496-031784, 1496-031785, 1496-031786, 1496-031787, 1496-031788, 1496-031789, 1496-031790, 1496-031791, 1496-031792, 1496-031793, 1496-031794, 1496-031795, 1496-031796, 1496-031798, 1496-031799, 1496-031800, 1496-031802, 1496-031803, 1496-031804, 1496-031805, 1496-031806, 1496-031807, 1496-031809, 1496-031810, 1496-031811, 1496-031812, 1496-031814, 1496-031815, 1496-031816, 1496-032015, 1496-032016, 1496-032375, 1496-032414, 1496-032418, 1496-032660, 1496-032661, 1496-032662, 1496-032663, 1496-032721, 1496-032792, 1496-032793, 1496-033426, 1496-033785, 1496-034033, 1496-034251, 1496-034259, 1496-034260, 1496-034261, 1496-034262, 1496-034263, 1496-034264, 1496-034265, 1496-034266, 1496-034267, 1496-034268, 1496-034540, 1496-034553, 1496-034560, 1496-034583, 1496-034711, 1496-034712, 1496-034713, 1496-034925, 1496-035276, 1496-035296, 1496-035417, 1496-036293, 1496-036294, 1496-036295, 1496-036424, 1496-037006, 1496-037007, 1496-037008, 1496-037009, 1496-037010, 1496-037011, 1496-037012, 1496-037013, 1496-037014, 1496-037015, 1496-037157, 1496-037158, 1496-037633, 1496-037684, 1496-037685, 1496-037686, 1496-037687, 1496-037688, 1496-037689, 1496-045292, 1496-045293, 1496-045294, 1496-045295, 1496-045296, 1496-045297, 1496-045310, 1496-045626, 1496-045627, 1496-045628, 1496-045909, 1496-045911, 1496-045913, 1496-045914, 1496-045915, 1496-045916, 1496-046010, 1496-046011, 1496-046012, 1496-046013, 1496-046761, 1496-046763, 1496-046764, 1496-046765, 1496-046766, 1496-046767, 1496-046768, 1496-046769, 1496-046770, 1496-046771, 1496-046772, 1496-046773, 1496-046774, 1496-046775, 1496-046776, 1496-046777, 1496-046779, 1496-046780, 1496-048077, 1496-048078, 1496-048079, 1496-048081, 1496-048082, 1496-048083, 1496-048175, 1496-048176, 1496-049930, 1496-100886, 1496-100887, 1496-100888, 1496-100889, 1496-100890, 1496-100891, 1496-100892, 1496-100893, 1496-100894, 1496-100895, 1496-100896, 1496-100897, 1496-100898, 1496-100899, 1496-101134, 1496-101140, 1496-101141, 1496-101148, 1496-101151, 1496-101189, 1496-101190, 1496-101191, 1496-101192, 1496-101193, 1496-101194, 1496-101195, 1496-101196, 1496-101197, 1496-101198, 1496-101199, 1496-101200, 1496-101202, 1496-101203, 1496-101204, 1496-101205, 1496-102018, 1496-102019, 1496-102020, 1496-102021, 1496-102022, 1496-102023, 1496-102024, 1496-102025, 1496-102026, 1496-102027, 1496-102045, 1496-102054, 1496-102055, 1496-102056, 1496-102057, 1496-102058, 1496-102158, 1496-102159, 1496-102160, 1496-102161, 1496-102163, 1496-102164, 1496-102165, 1496-102216, 1496-102549, 1496-102655, 1496-102656, 1496-102657, 1496-102658, 1496-102659, 1496-102787, 1496-102788, 1496-102789, 1496-102790, 1496-102791, 1496-103435, 1496-103436, 1496-103437, 1496-103438, 1496-103439, 1496-103440, 1496-103441, 1496-103610, 1496-103611, 1496-103639, 1496-103710, 1496-103711, 1496-103712, 1496-103713, 1496-103714, 1496-103715, 1496-103716, 1496-103717, 1496-103718, 1496-103719, 1496-104419, 1496-104424, 1496-104425, 1496-104426, 1496-104427, 1496-104428, 1496-104547, 1496-104548, 1496-104549, 1496-104550, 1496-104551, 1496-104552, 1496-104553, 1496-104554, 1496-104555, 1496-104556, 1496-104557, 1496-104558, 1496-104559, 1496-104560, 1496-104561, 1496-104563, 1496-104564, 1496-104565, 1496-104566, 1496-104567, 1496-104568, 1496-105356, 1496-105357, 1496-105358, 1496-105360, 1496-105373, 1496-105382, 1496-105383, 1496-105688, 1496-105747, 1496-105748, 1496-105749, 1496-105750, 1496-105751, 1496-105752, 1496-105753, 1496-105754, 1496-105755, 1496-105756, 1496-105757, 1496-105758, 1496-105759, 1496-105760, 1496-105761, 1496-105762, 1496-105763, 1496-105764, 1496-105766, 1496-105767, 1496-105768, 1496-105769, 1496-105770, 1496-105771, 1496-105772, 1496-105773, 1496-105774, 1496-105775, 1496-105776, 1496-105777, 1496-105778, 1496-105779, 1496-105780, 1496-105781, 1496-105782, 1496-105783, 1496-105784, 1496-105785, 1496-105786, 1496-105787, 1496-105788, 1496-105789, 1496-105790, 1496-105791, 1496-105792, 1496-105793, 1496-105794, 1496-105795, 1496-105796, 1496-105797, 1496-105798, 1496-105799, 1496-105800, 1496-105801, 1496-105802, 1496-105803, 1496-105804, 1496-105805, 1496-105806, 1496-105807, 1496-105808, 1496-105809, 1496-105810, 1496-105811, 1496-105812, 1496-105813, 1496-105814, 1496-105815, 1496-105816, 1496-105817, 1496-105818, 1496-105819, 1496-105821, 1496-105822, 1496-105823, 1496-105824, 1496-105825, 1496-105826, 1496-105827, 1496-105828, 1496-105829, 1496-105830, 1496-105831, 1496-105832, 1496-105833, 1496-105834, 1496-105835, 1496-105836, 1496-105837, 1496-105838, 1496-105839, 1496-105840, 1496-105841, 1496-105842, 1496-105843, 1496-105844, 1496-105845, 1496-105846, 1496-105847, 1496-105848, 1496-105849, 1496-105850, 1496-105851, 1496-105852, 1496-105853, 1496-105854, 1496-105855, 1496-105856, 1496-105857, 1496-105858, 1496-105859, 1496-105860, 1496-105861, 1496-105862, 1496-105863, 1496-105864, 1496-105865, 1496-105866, 1496-105867, 1496-105868, 1496-105869, 1496-105870, 1496-105871, 1496-105872, 1496-105873, 1496-105874, 1496-105875, 1496-105876, 1496-105877, 1496-105878, 1496-105879, 1496-105880, 1496-105881, 1496-105882, 1496-105883, 1496-105884, 1496-105885, 1496-105886, 1496-105887, 1496-105888, 1496-105889, 1496-105890, 1496-105891, 1496-105892, 1496-105893, 1496-105894, 1496-105895, 1496-105896, 1496-105897, 1496-105898, 1496-105899, 1496-105900, 1496-105901, 1496-105902, 1496-105903, 1496-105904, 1496-105905, 1496-105906, 1496-105907, 1496-105908, 1496-105909, 1496-105910, 1496-105911, 1496-105912, 1496-105913, 1496-105914, 1496-105915, 1496-105916, 1496-105917, 1496-105918, 1496-105919, 1496-105920, 1496-105921, 1496-105922, 1496-105923, 1496-105924, 1496-105925, 1496-105926, 1496-105927, 1496-105928, 1496-105929, 1496-105930, 1496-105931, 1496-105932, 1496-105933, 1496-105934, 1496-105935, 1496-105936, 1496-105937, 1496-105938, 1496-105939, 1496-105940, 1496-105941, 1496-105942, 1496-105943, 1496-105944, 1496-105945, 1496-105946, 1496-105947, 1496-105948, 1496-105949, 1496-105950, 1496-105951, 1496-105952, 1496-105953, 1496-105954, 1496-105955, 1496-105956, 1496-105957, 1496-105958, 1496-105959, 1496-105960, 1496-105961, 1496-105962, 1496-105963, 1496-105964, 1496-105965, 1496-105966, 1496-105967, 1496-105968, 1496-105969, 1496-105970, 1496-105971, 1496-105972, 1496-105973, 1496-105974, 1496-105975, 1496-105976, 1496-105977, 1496-105978, 1496-105979, 1496-105980, 1496-105981, 1496-105982, 1496-105983, 1496-105984, 1496-105985, 1496-105986, 1496-105987, 1496-105988, 1496-105989, 1496-105990, 1496-105991, 1496-105992, 1496-105993, 1496-105994, 1496-105995, 1496-105996, 1496-105997, 1496-105998, 1496-105999, 1496-106000, 1496-106001, 1496-106002, 1496-106003, 1496-106004, 1496-106005, 1496-106006, 1496-106007, 1496-106008, 1496-106009, 1496-106010, 1496-106011, 1496-106012, 1496-106013, 1496-106014, 1496-106015, 1496-106016, 1496-106017, 1496-106018, 1496-106019, 1496-106020, 1496-106021, 1496-106022, 1496-106023, 1496-106024, 1496-106025, 1496-106026, 1496-106027, 1496-106028, 1496-106029, 1496-106030, 1496-106031, 1496-106032, 1496-106033, 1496-106034, 1496-106035, 1496-106036, 1496-106037, 1496-106038, 1496-106039, 1496-106040, 1496-106041, 1496-106042, 1496-106043, 1496-106044, 1496-106045, 1496-106046, 1496-106047, 1496-106048, 1496-106049, 1496-106050, 1496-106051, 1496-106052, 1496-106053, 1496-106054, 1496-106055, 1496-106056, 1496-106057, 1496-106058, 1496-106059, 1496-106060, 1496-106061, 1496-106062, 1496-106063, 1496-106064, 1496-106065, 1496-106066, 1496-106067, 1496-106068, 1496-106069, 1496-106070, 1496-106071, 1496-106072, 1496-106073, 1496-106074, 1496-106286, 1496-106287, 1496-106288, 1496-106289, 1496-106290, 1496-106291, 1496-106292, 1496-106293, 1496-106294, 1496-106295, 1496-106296, 1496-106297, 1496-106298, 1496-106299, 1496-106300, 1496-106301, 1496-106302, 1496-106303, 1496-106304, 1496-106305, 1496-106306, 1496-106307, 1496-106308, 1496-106309, 1496-106310, 1496-106311, 1496-106458, 1496-106459, 1496-106460, 1496-106461, 1496-106462, 1496-106463, 1496-106464, 1496-106465, 1496-106466, 1496-106467, 1496-106468, 1496-106469, 1496-106470, 1496-106471, 1496-106472, 1496-106498, 1496-106500, 1496-106501, 1496-106502, 1496-106503, 1496-106504, 1496-106505, 1496-106506, 1496-106507, 1496-106508, 1496-106509, 1496-106510, 1496-106511, 1496-106512, 1496-106513, 1496-106514, 1496-106515, 1496-106516, 1496-106517, 1496-106518, 1496-106519, 1496-106520, 1496-106521, 1496-106522, 1496-106523, 1496-106524, 1496-106525, 1496-106526, 1496-106527, 1496-106528, 1496-106529, 1496-107878, 1496-107882, 1496-107884, 1496-107903, 1496-107904, 1496-107905, 1496-107906, 1496-109257, 1496-109258, 1496-109259, 1496-109260, 1496-109261, 1496-109263, 1496-109264, 1496-109265, 1496-109266, 1496-109267, 1496-109298, 1496-109299, 1496-109300, 1496-109301, 1496-109302, 1496-109303, 1496-109304, 1496-109305, 1496-109306, 1496-109322, 1496-109323, 1496-109324, 1496-109325, 1496-112374, 1496-112632, 1496-117845, 1496-120087, 1496-120088, 1496-120089, 1496-120090, 1496-120091, 1496-120605, 1496-120725, 1496-120726, 1496-120727, 1496-120728, 1496-120729, 1496-120730, 1496-120731, 1496-120732, 1496-120733, 1496-120734, 1496-120757, 1496-120758, 1496-120759, 1496-120760, 1496-120761, 1496-120762, 1496-120763, 1496-120764, 1496-120765, 1496-120766, 1496-120767, 1496-120768, 1496-120769, 1496-121121, 1496-121973, 1496-121974, 1496-121975, 1496-121976, 1496-121977, 1496-121978, 1496-121979, 1496-121980, 1496-121981, 1496-121982, 1496-121983, 1496-121984, 1496-121985, 1496-121986, 1496-121987, 1496-121988, 1496-121989, 1496-121990, 1496-121991, and 1496-121992. **********SERIAL NUMBERS INTERNATIONAL********************************* 1496-001139, 1496-001925, 1496-002364, 1496-002372, 1496-002410, 1496-002411, 1496-002483, 1496-002486, 1496-002501, 1496-002509, 1496-002521, 1496-002542, 1496-002543, 1496-003059, 1496-006322, 1496-006323, 1496-006324, 1496-006702, 1496-009407, 1496-009518, 1496-009550, 1496-009739, 1496-009953, 1496-009979, 1496-010250, 1496-010251, 1496-010257, 1496-010259, 1496-010261, 1496-010779, 1496-011011, 1496-011322, 1496-011553, 1496-011927, 1496-012393, 1496-012715, 1496-012729, 1496-014969, 1496-016724, 1496-017329, 1496-019070, 1496-033510, 1496-038108, 1496-113895, 1496-113896, 1496-113911, 1496-113935, 1496-113936, 1496-113938, 1496-113939, 1496-113940, 1496-113941, 1496-113944, 1496-113945, 1496-113946, 1496-113947, 1496-113948, 1496-113949, 1496-113950, 1496-113951, 1496-113952, 1496-113953, 1496-113954, 1496-113956, 1496-113957, 1496-113958, and 1496-113959.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Spacelabs Healthcare Inc
- Reason for Recall:
- Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters will be suspended. Upon completion of this sequence, alarms will be reset to their default values.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Ultraview SL Command Modules, Model 91496-M, upgraded with the Masimo SpO2 Option. Product Usage: The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use. Physiological parameters that may be monitored include ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (Sp02) and cardiac output. Acquired data may then be communicated to all information network for display, recording, editing and analysis.
Product Codes/Lot Numbers:
SERIAL NUMBERS IN THE US: 1496-000473, 1496-000590, 1496-001126, 1496-001189, 1496-001214, 1496-001299, 1496-001300, 1496-001301, 1496-001302, 1496-001303, 1496-001304, 1496-001305, 1496-001306, 1496-001307, 1496-001309, 1496-001328, 1496-001329, 1496-001330, 1496-001332, 1496-001333, 1496-001335, 1496-001336, 1496-001337, 1496-001712, 1496-001713, 1496-001714, 1496-002442, 1496-002568, 1496-002691, 1496-002743, 1496-002832, 1496-002833, 1496-002834, 1496-002836, 1496-002837, 1496-002838, 1496-002839, 1496-002840, 1496-002841, 1496-002917, 1496-002918, 1496-002919, 1496-002920, 1496-002998, 1496-002999, 1496-003000, 1496-003001, 1496-003002, 1496-003003, 1496-003005, 1496-003006, 1496-003007, 1496-003021, 1496-003022, 1496-003025, 1496-003035, 1496-003036, 1496-003037, 1496-003042, 1496-003043, 1496-003044, 1496-003046, 1496-003049, 1496-003051, 1496-003055, 1496-003158, 1496-003161, 1496-003162, 1496-003163, 1496-003164, 1496-003179, 1496-003187, 1496-003190, 1496-003191, 1496-003192, 1496-003193, 1496-003210, 1496-003528, 1496-003602, 1496-003796, 1496-003797, 1496-003819, 1496-004075, 1496-004081, 1496-004082, 1496-004088, 1496-004095, 1496-004098, 1496-004100, 1496-004103, 1496-004104, 1496-004110, 1496-004174, 1496-004175, 1496-004222, 1496-004302, 1496-004382, 1496-004385, 1496-004386, 1496-004387, 1496-004390, 1496-004392, 1496-004394, 1496-004622, 1496-004667, 1496-004674, 1496-004675, 1496-004676, 1496-004737, 1496-004762, 1496-004805, 1496-004806, 1496-004886, 1496-004892, 1496-004951, 1496-004980, 1496-005204, 1496-005291, 1496-005292, 1496-005293, 1496-005294, 1496-005342, 1496-005343, 1496-005358, 1496-005359, 1496-005371, 1496-005374, 1496-005380, 1496-005611, 1496-005779, 1496-005780, 1496-005781, 1496-005782, 1496-005785, 1496-005786, 1496-005787, 1496-005889, 1496-005890, 1496-005891, 1496-005892, 1496-005894, 1496-005896, 1496-005897, 1496-005898, 1496-005899, 1496-005900, 1496-005901, 1496-005903, 1496-005904, 1496-005905, 1496-006038, 1496-006058, 1496-006078, 1496-006089, 1496-006108, 1496-006109, 1496-006110, 1496-006140, 1496-006141, 1496-006142, 1496-006143, 1496-006144, 1496-006145, 1496-006249, 1496-006250, 1496-006251, 1496-006252, 1496-006253, 1496-006254, 1496-006255, 1496-006256, 1496-006257, 1496-006258, 1496-006259, 1496-006260, 1496-006261, 1496-006262, 1496-006266, 1496-006267, 1496-006268, 1496-006269, 1496-006270, 1496-006271, 1496-006272, 1496-006273, 1496-006327, 1496-006328, 1496-006329, 1496-006330, 1496-006331, 1496-006332, 1496-006333, 1496-006334, 1496-006336, 1496-006337, 1496-006338, 1496-006341, 1496-006342, 1496-006343, 1496-006457, 1496-006472, 1496-006523, 1496-006544, 1496-006593, 1496-006594, 1496-006595, 1496-006596, 1496-006597, 1496-006877, 1496-006972, 1496-006973, 1496-006981, 1496-006982, 1496-006983, 1496-006984, 1496-006985, 1496-006986, 1496-006994, 1496-006995, 1496-006996, 1496-006997, 1496-006998, 1496-007132, 1496-007137, 1496-007318, 1496-007319, 1496-007320, 1496-007321, 1496-007322, 1496-007323, 1496-007324, 1496-007325, 1496-007326, 1496-007327, 1496-007328, 1496-007329, 1496-007330, 1496-007331, 1496-007332, 1496-007333, 1496-007334, 1496-007335, 1496-007336, 1496-007337, 1496-007338, 1496-007339, 1496-007340, 1496-007376, 1496-007587, 1496-007589, 1496-007590, 1496-007593, 1496-007594, 1496-007595, 1496-007596, 1496-007631, 1496-007647, 1496-007650, 1496-007651, 1496-007652, 1496-007653, 1496-007654, 1496-007655, 1496-007656, 1496-007657, 1496-007658, 1496-007745, 1496-007750, 1496-007753, 1496-007755, 1496-007756, 1496-007757, 1496-007969, 1496-007970, 1496-008009, 1496-008010, 1496-008011, 1496-008012, 1496-008013, 1496-008014, 1496-008015, 1496-008046, 1496-008047, 1496-008048, 1496-008161, 1496-008162, 1496-008163, 1496-008250, 1496-008319, 1496-008383, 1496-008450, 1496-008451, 1496-008452, 1496-008453, 1496-008454, 1496-008455, 1496-008456, 1496-008580, 1496-008584, 1496-008587, 1496-008873, 1496-009392, 1496-009393, 1496-009394, 1496-009395, 1496-009396, 1496-009397, 1496-009398, 1496-009399, 1496-009400, 1496-009401, 1496-009524, 1496-009584, 1496-009585, 1496-009586, 1496-009587, 1496-009588, 1496-009589, 1496-009590, 1496-009591, 1496-009592, 1496-009593, 1496-009594, 1496-009595, 1496-009596, 1496-009597, 1496-009598, 1496-009599, 1496-009600, 1496-009601, 1496-009602, 1496-009603, 1496-009604, 1496-009605, 1496-009606, 1496-009607, 1496-009608, 1496-009715, 1496-009733, 1496-009734, 1496-009737, 1496-009738, 1496-009760, 1496-009761, 1496-009762, 1496-009763, 1496-009764, 1496-009765, 1496-009766, 1496-009767, 1496-009778, 1496-009779, 1496-009781, 1496-009782, 1496-009783, 1496-009784, 1496-009785, 1496-009786, 1496-009787, 1496-009788, 1496-009856, 1496-009861, 1496-009862, 1496-009867, 1496-009874, 1496-009877, 1496-009879, 1496-009963, 1496-009968, 1496-010008, 1496-010009, 1496-010010, 1496-010011, 1496-010012, 1496-010013, 1496-010017, 1496-010018, 1496-010019, 1496-010068, 1496-010113, 1496-010252, 1496-010260, 1496-010273, 1496-010278, 1496-010287, 1496-010288, 1496-010290, 1496-010291, 1496-010292, 1496-010293, 1496-010294, 1496-010295, 1496-010296, 1496-010297, 1496-010298, 1496-010299, 1496-010300, 1496-010301, 1496-010302, 1496-010469, 1496-010470, 1496-010471, 1496-010472, 1496-010473, 1496-010474, 1496-010475, 1496-010476, 1496-010477, 1496-010478, 1496-010523, 1496-010524, 1496-010526, 1496-010527, 1496-010528, 1496-010529, 1496-010530, 1496-010531, 1496-010532, 1496-010533, 1496-010534, 1496-010535, 1496-010537, 1496-010539, 1496-010540, 1496-010541, 1496-010542, 1496-010543, 1496-010545, 1496-010546, 1496-010547, 1496-010548, 1496-010550, 1496-010691, 1496-010692, 1496-010809, 1496-010830, 1496-010927, 1496-010933, 1496-010952, 1496-010959, 1496-010960, 1496-011059, 1496-011060, 1496-011069, 1496-011070, 1496-011210, 1496-011211, 1496-011246, 1496-011247, 1496-011248, 1496-011249, 1496-011250, 1496-011252, 1496-011253, 1496-011256, 1496-011259, 1496-011306, 1496-011307, 1496-011308, 1496-011309, 1496-011310, 1496-011311, 1496-011312, 1496-011313, 1496-011314, 1496-011315, 1496-011317, 1496-011552, 1496-011555, 1496-011716, 1496-011894, 1496-011902, 1496-011908, 1496-011923, 1496-012278, 1496-012280, 1496-012307, 1496-012320, 1496-012343, 1496-012426, 1496-012429, 1496-012431, 1496-012432, 1496-012434, 1496-012470, 1496-012478, 1496-012495, 1496-012496, 1496-012497, 1496-012615, 1496-012616, 1496-012617, 1496-012618, 1496-012620, 1496-012621, 1496-012623, 1496-012624, 1496-012625, 1496-012626, 1496-012628, 1496-012629, 1496-012630, 1496-012631, 1496-012632, 1496-012633, 1496-012634, 1496-012635, 1496-012636, 1496-012637, 1496-012638, 1496-012639, 1496-012640, 1496-012642, 1496-013314, 1496-013383, 1496-013563, 1496-013564, 1496-013565, 1496-013566, 1496-013567, 1496-013568, 1496-013569, 1496-013570, 1496-013571, 1496-013572, 1496-013573, 1496-013574, 1496-013575, 1496-013578, 1496-013579, 1496-013580, 1496-013583, 1496-013584, 1496-013585, 1496-013586, 1496-013587, 1496-013588, 1496-013590, 1496-013591, 1496-013683, 1496-013705, 1496-013706, 1496-013709, 1496-013710, 1496-013711, 1496-013712, 1496-013713, 1496-013732, 1496-013733, 1496-013734, 1496-013735, 1496-013998, 1496-013999, 1496-014000, 1496-014001, 1496-014002, 1496-014003, 1496-014004, 1496-014005, 1496-014006, 1496-014007, 1496-014008, 1496-014009, 1496-014010, 1496-014011, 1496-014012, 1496-014015, 1496-014016, 1496-014017, 1496-014018, 1496-014019, 1496-014020, 1496-014065, 1496-014135, 1496-014136, 1496-014172, 1496-014173, 1496-014174, 1496-014175, 1496-014177, 1496-014178, 1496-014179, 1496-014180, 1496-014181, 1496-014339, 1496-014340, 1496-014341, 1496-014342, 1496-014343, 1496-014344, 1496-014345, 1496-014346, 1496-014348, 1496-014349, 1496-014350, 1496-014351, 1496-014352, 1496-014353, 1496-014354, 1496-014355, 1496-014356, 1496-014357, 1496-014358, 1496-014359, 1496-014360, 1496-014361, 1496-014362, 1496-014445, 1496-014446, 1496-014447, 1496-014448, 1496-014449, 1496-014451, 1496-014462, 1496-014484, 1496-014485, 1496-014486, 1496-014487, 1496-014506, 1496-014517, 1496-014536, 1496-014537, 1496-014542, 1496-014552, 1496-014554, 1496-014555, 1496-014835, 1496-014964, 1496-014965, 1496-014966, 1496-014967, 1496-014968, 1496-014970, 1496-014971, 1496-014972, 1496-014973, 1496-014974, 1496-014975, 1496-014976, 1496-014978, 1496-014979, 1496-014980, 1496-014982, 1496-014983, 1496-014984, 1496-014985, 1496-014986, 1496-014988, 1496-015059, 1496-015082, 1496-015084, 1496-015181, 1496-015183, 1496-015184, 1496-015185, 1496-015217, 1496-015218, 1496-015219, 1496-015220, 1496-015221, 1496-015222, 1496-015223, 1496-015224, 1496-015225, 1496-015226, 1496-015227, 1496-015228, 1496-015229, 1496-015230, 1496-015231, 1496-015232, 1496-015233, 1496-015234, 1496-015235, 1496-015236, 1496-015237, 1496-015257, 1496-015258, 1496-015259, 1496-015260, 1496-015261, 1496-015263, 1496-015264, 1496-015265, 1496-015266, 1496-015267, 1496-015268, 1496-015269, 1496-015270, 1496-015272, 1496-015273, 1496-015274, 1496-015276, 1496-015277, 1496-015278, 1496-015279, 1496-015280, 1496-015281, 1496-015282, 1496-015283, 1496-015284, 1496-015285, 1496-015286, 1496-015288, 1496-015289, 1496-015290, 1496-015292, 1496-015293, 1496-015294, 1496-015295, 1496-015296, 1496-015297, 1496-015298, 1496-015299, 1496-015300, 1496-015301, 1496-015302, 1496-015303, 1496-015304, 1496-015312, 1496-016051, 1496-016052, 1496-016053, 1496-016056, 1496-016249, 1496-016250, 1496-016499, 1496-016509, 1496-016526, 1496-016527, 1496-016530, 1496-016804, 1496-016870, 1496-016938, 1496-016946, 1496-016956, 1496-016959, 1496-016960, 1496-016965, 1496-016966, 1496-016967, 1496-016968, 1496-016969, 1496-016970, 1496-016972, 1496-016974, 1496-016975, 1496-016977, 1496-016979, 1496-016980, 1496-016981, 1496-016983, 1496-016984, 1496-016985, 1496-016986, 1496-016987, 1496-017379, 1496-017408, 1496-017409, 1496-017410, 1496-017411, 1496-017412, 1496-017413, 1496-017414, 1496-017415, 1496-017416, 1496-017417, 1496-017418, 1496-017419, 1496-017420, 1496-017421, 1496-017422, 1496-017423, 1496-017424, 1496-017425, 1496-017426, 1496-017427, 1496-017428, 1496-017429, 1496-017430, 1496-017431, 1496-017432, 1496-017433, 1496-017434, 1496-017435, 1496-017436, 1496-017437, 1496-017438, 1496-017439, 1496-017440, 1496-017441, 1496-017442, 1496-017443, 1496-017444, 1496-017445, 1496-017446, 1496-017447, 1496-017448, 1496-017449, 1496-017491, 1496-017492, 1496-017493, 1496-017494, 1496-017547, 1496-017548, 1496-017549, 1496-017550, 1496-017551, 1496-017552, 1496-017555, 1496-017556, 1496-017557, 1496-017558, 1496-017559, 1496-017560, 1496-017561, 1496-017562, 1496-017563, 1496-017564, 1496-017599, 1496-017659, 1496-017660, 1496-017661, 1496-017663, 1496-017664, 1496-017665, 1496-017666, 1496-017667, 1496-017882, 1496-017883, 1496-017884, 1496-017885, 1496-017886, 1496-017888, 1496-017889, 1496-017890, 1496-017891, 1496-017892, 1496-017893, 1496-017895, 1496-017896, 1496-017897, 1496-017898, 1496-017899, 1496-017900, 1496-017901, 1496-017902, 1496-017903, 1496-017904, 1496-017905, 1496-018140, 1496-018277, 1496-018498, 1496-018499, 1496-018501, 1496-018506, 1496-018511, 1496-018512, 1496-018513, 1496-018657, 1496-018658, 1496-018659, 1496-018660, 1496-018661, 1496-018662, 1496-018663, 1496-018664, 1496-018665, 1496-018666, 1496-018667, 1496-018668, 1496-018669, 1496-018670, 1496-018671, 1496-018672, 1496-018673, 1496-018674, 1496-018675, 1496-018676, 1496-018715, 1496-018716, 1496-018717, 1496-018719, 1496-018720, 1496-018721, 1496-018722, 1496-019217, 1496-019221, 1496-019491, 1496-019492, 1496-019584, 1496-020047, 1496-020049, 1496-020073, 1496-020074, 1496-020075, 1496-020076, 1496-020077, 1496-020078, 1496-020079, 1496-020080, 1496-020081, 1496-020082, 1496-020083, 1496-020084, 1496-020085, 1496-020086, 1496-020087, 1496-020088, 1496-020089, 1496-020090, 1496-020091, 1496-020178, 1496-020278, 1496-020279, 1496-020280, 1496-020281, 1496-020282, 1496-020283, 1496-020284, 1496-020293, 1496-020294, 1496-020295, 1496-020296, 1496-020297, 1496-020298, 1496-020308, 1496-020311, 1496-020348, 1496-020349, 1496-020350, 1496-020807, 1496-021310, 1496-021443, 1496-021457, 1496-021458, 1496-021459, 1496-021460, 1496-021461, 1496-021462, 1496-021463, 1496-021464, 1496-021465, 1496-021466, 1496-021467, 1496-021468, 1496-021469, 1496-021470, 1496-021471, 1496-021472, 1496-021473, 1496-021474, 1496-021475, 1496-021476, 1496-021477, 1496-021478, 1496-021479, 1496-021480, 1496-021481, 1496-021482, 1496-021483, 1496-021484, 1496-021485, 1496-021486, 1496-021487, 1496-021488, 1496-021504, 1496-021505, 1496-021506, 1496-021507, 1496-021508, 1496-021509, 1496-021510, 1496-021511, 1496-021512, 1496-021513, 1496-021520, 1496-021677, 1496-021715, 1496-021723, 1496-021743, 1496-021758, 1496-023510, 1496-023558, 1496-023559, 1496-023560, 1496-023561, 1496-023563, 1496-023564, 1496-023565, 1496-023566, 1496-023567, 1496-023858, 1496-024742, 1496-024804, 1496-024832, 1496-025119, 1496-025120, 1496-025121, 1496-025122, 1496-025123, 1496-025124, 1496-025126, 1496-025197, 1496-025199, 1496-025202, 1496-025203, 1496-025204, 1496-025205, 1496-025206, 1496-025207, 1496-025208, 1496-025209, 1496-025234, 1496-025235, 1496-025236, 1496-025237, 1496-025238, 1496-025506, 1496-025512, 1496-025513, 1496-025514, 1496-025515, 1496-025516, 1496-025517, 1496-025518, 1496-025618, 1496-025619, 1496-025654, 1496-025657, 1496-025658, 1496-025660, 1496-025661, 1496-025662, 1496-025663, 1496-025664, 1496-025665, 1496-025666, 1496-025667, 1496-025668, 1496-025669, 1496-025670, 1496-025671, 1496-025672, 1496-025673, 1496-025674, 1496-025675, 1496-025676, 1496-025677, 1496-025678, 1496-025679, 1496-025680, 1496-025681, 1496-025682, 1496-025683, 1496-025684, 1496-025685, 1496-025686, 1496-025687, 1496-025688, 1496-025689, 1496-025690, 1496-025691, 1496-025692, 1496-025693, 1496-025694, 1496-025695, 1496-025696, 1496-025697, 1496-025698, 1496-025699, 1496-025701, 1496-025702, 1496-025704, 1496-025705, 1496-025706, 1496-025707, 1496-025708, 1496-025709, 1496-025710, 1496-025769, 1496-025770, 1496-025771, 1496-025941, 1496-025942, 1496-026067, 1496-026068, 1496-026070, 1496-026072, 1496-026075, 1496-026076, 1496-026077, 1496-026078, 1496-026079, 1496-026080, 1496-026081, 1496-026082, 1496-026083, 1496-026309, 1496-026310, 1496-026311, 1496-026362, 1496-026363, 1496-026364, 1496-026365, 1496-026366, 1496-026367, 1496-026368, 1496-026369, 1496-026370, 1496-026373, 1496-026374, 1496-026429, 1496-027022, 1496-027070, 1496-027071, 1496-027103, 1496-027111, 1496-027112, 1496-027409, 1496-027410, 1496-027411, 1496-027412, 1496-027413, 1496-027414, 1496-027415, 1496-027416, 1496-027472, 1496-027578, 1496-027579, 1496-027580, 1496-027581, 1496-028419, 1496-028442, 1496-028459, 1496-028461, 1496-028462, 1496-028465, 1496-028466, 1496-028467, 1496-028665, 1496-028710, 1496-028711, 1496-028822, 1496-028824, 1496-028825, 1496-028826, 1496-028827, 1496-028829, 1496-028831, 1496-028832, 1496-028833, 1496-028834, 1496-028836, 1496-028837, 1496-028838, 1496-028839, 1496-028840, 1496-028841, 1496-028842, 1496-028843, 1496-028844, 1496-028845, 1496-028846, 1496-028847, 1496-028848, 1496-028850, 1496-028851, 1496-028852, 1496-028853, 1496-028854, 1496-028855, 1496-029028, 1496-029051, 1496-029575, 1496-029595, 1496-029723, 1496-029724, 1496-029725, 1496-029726, 1496-029727, 1496-029728, 1496-029985, 1496-030010, 1496-030932, 1496-030933, 1496-030934, 1496-031144, 1496-031145, 1496-031190, 1496-031784, 1496-031785, 1496-031786, 1496-031787, 1496-031788, 1496-031789, 1496-031790, 1496-031791, 1496-031792, 1496-031793, 1496-031794, 1496-031795, 1496-031796, 1496-031798, 1496-031799, 1496-031800, 1496-031802, 1496-031803, 1496-031804, 1496-031805, 1496-031806, 1496-031807, 1496-031809, 1496-031810, 1496-031811, 1496-031812, 1496-031814, 1496-031815, 1496-031816, 1496-032015, 1496-032016, 1496-032375, 1496-032414, 1496-032418, 1496-032660, 1496-032661, 1496-032662, 1496-032663, 1496-032721, 1496-032792, 1496-032793, 1496-033426, 1496-033785, 1496-034033, 1496-034251, 1496-034259, 1496-034260, 1496-034261, 1496-034262, 1496-034263, 1496-034264, 1496-034265, 1496-034266, 1496-034267, 1496-034268, 1496-034540, 1496-034553, 1496-034560, 1496-034583, 1496-034711, 1496-034712, 1496-034713, 1496-034925, 1496-035276, 1496-035296, 1496-035417, 1496-036293, 1496-036294, 1496-036295, 1496-036424, 1496-037006, 1496-037007, 1496-037008, 1496-037009, 1496-037010, 1496-037011, 1496-037012, 1496-037013, 1496-037014, 1496-037015, 1496-037157, 1496-037158, 1496-037633, 1496-037684, 1496-037685, 1496-037686, 1496-037687, 1496-037688, 1496-037689, 1496-045292, 1496-045293, 1496-045294, 1496-045295, 1496-045296, 1496-045297, 1496-045310, 1496-045626, 1496-045627, 1496-045628, 1496-045909, 1496-045911, 1496-045913, 1496-045914, 1496-045915, 1496-045916, 1496-046010, 1496-046011, 1496-046012, 1496-046013, 1496-046761, 1496-046763, 1496-046764, 1496-046765, 1496-046766, 1496-046767, 1496-046768, 1496-046769, 1496-046770, 1496-046771, 1496-046772, 1496-046773, 1496-046774, 1496-046775, 1496-046776, 1496-046777, 1496-046779, 1496-046780, 1496-048077, 1496-048078, 1496-048079, 1496-048081, 1496-048082, 1496-048083, 1496-048175, 1496-048176, 1496-049930, 1496-100886, 1496-100887, 1496-100888, 1496-100889, 1496-100890, 1496-100891, 1496-100892, 1496-100893, 1496-100894, 1496-100895, 1496-100896, 1496-100897, 1496-100898, 1496-100899, 1496-101134, 1496-101140, 1496-101141, 1496-101148, 1496-101151, 1496-101189, 1496-101190, 1496-101191, 1496-101192, 1496-101193, 1496-101194, 1496-101195, 1496-101196, 1496-101197, 1496-101198, 1496-101199, 1496-101200, 1496-101202, 1496-101203, 1496-101204, 1496-101205, 1496-102018, 1496-102019, 1496-102020, 1496-102021, 1496-102022, 1496-102023, 1496-102024, 1496-102025, 1496-102026, 1496-102027, 1496-102045, 1496-102054, 1496-102055, 1496-102056, 1496-102057, 1496-102058, 1496-102158, 1496-102159, 1496-102160, 1496-102161, 1496-102163, 1496-102164, 1496-102165, 1496-102216, 1496-102549, 1496-102655, 1496-102656, 1496-102657, 1496-102658, 1496-102659, 1496-102787, 1496-102788, 1496-102789, 1496-102790, 1496-102791, 1496-103435, 1496-103436, 1496-103437, 1496-103438, 1496-103439, 1496-103440, 1496-103441, 1496-103610, 1496-103611, 1496-103639, 1496-103710, 1496-103711, 1496-103712, 1496-103713, 1496-103714, 1496-103715, 1496-103716, 1496-103717, 1496-103718, 1496-103719, 1496-104419, 1496-104424, 1496-104425, 1496-104426, 1496-104427, 1496-104428, 1496-104547, 1496-104548, 1496-104549, 1496-104550, 1496-104551, 1496-104552, 1496-104553, 1496-104554, 1496-104555, 1496-104556, 1496-104557, 1496-104558, 1496-104559, 1496-104560, 1496-104561, 1496-104563, 1496-104564, 1496-104565, 1496-104566, 1496-104567, 1496-104568, 1496-105356, 1496-105357, 1496-105358, 1496-105360, 1496-105373, 1496-105382, 1496-105383, 1496-105688, 1496-105747, 1496-105748, 1496-105749, 1496-105750, 1496-105751, 1496-105752, 1496-105753, 1496-105754, 1496-105755, 1496-105756, 1496-105757, 1496-105758, 1496-105759, 1496-105760, 1496-105761, 1496-105762, 1496-105763, 1496-105764, 1496-105766, 1496-105767, 1496-105768, 1496-105769, 1496-105770, 1496-105771, 1496-105772, 1496-105773, 1496-105774, 1496-105775, 1496-105776, 1496-105777, 1496-105778, 1496-105779, 1496-105780, 1496-105781, 1496-105782, 1496-105783, 1496-105784, 1496-105785, 1496-105786, 1496-105787, 1496-105788, 1496-105789, 1496-105790, 1496-105791, 1496-105792, 1496-105793, 1496-105794, 1496-105795, 1496-105796, 1496-105797, 1496-105798, 1496-105799, 1496-105800, 1496-105801, 1496-105802, 1496-105803, 1496-105804, 1496-105805, 1496-105806, 1496-105807, 1496-105808, 1496-105809, 1496-105810, 1496-105811, 1496-105812, 1496-105813, 1496-105814, 1496-105815, 1496-105816, 1496-105817, 1496-105818, 1496-105819, 1496-105821, 1496-105822, 1496-105823, 1496-105824, 1496-105825, 1496-105826, 1496-105827, 1496-105828, 1496-105829, 1496-105830, 1496-105831, 1496-105832, 1496-105833, 1496-105834, 1496-105835, 1496-105836, 1496-105837, 1496-105838, 1496-105839, 1496-105840, 1496-105841, 1496-105842, 1496-105843, 1496-105844, 1496-105845, 1496-105846, 1496-105847, 1496-105848, 1496-105849, 1496-105850, 1496-105851, 1496-105852, 1496-105853, 1496-105854, 1496-105855, 1496-105856, 1496-105857, 1496-105858, 1496-105859, 1496-105860, 1496-105861, 1496-105862, 1496-105863, 1496-105864, 1496-105865, 1496-105866, 1496-105867, 1496-105868, 1496-105869, 1496-105870, 1496-105871, 1496-105872, 1496-105873, 1496-105874, 1496-105875, 1496-105876, 1496-105877, 1496-105878, 1496-105879, 1496-105880, 1496-105881, 1496-105882, 1496-105883, 1496-105884, 1496-105885, 1496-105886, 1496-105887, 1496-105888, 1496-105889, 1496-105890, 1496-105891, 1496-105892, 1496-105893, 1496-105894, 1496-105895, 1496-105896, 1496-105897, 1496-105898, 1496-105899, 1496-105900, 1496-105901, 1496-105902, 1496-105903, 1496-105904, 1496-105905, 1496-105906, 1496-105907, 1496-105908, 1496-105909, 1496-105910, 1496-105911, 1496-105912, 1496-105913, 1496-105914, 1496-105915, 1496-105916, 1496-105917, 1496-105918, 1496-105919, 1496-105920, 1496-105921, 1496-105922, 1496-105923, 1496-105924, 1496-105925, 1496-105926, 1496-105927, 1496-105928, 1496-105929, 1496-105930, 1496-105931, 1496-105932, 1496-105933, 1496-105934, 1496-105935, 1496-105936, 1496-105937, 1496-105938, 1496-105939, 1496-105940, 1496-105941, 1496-105942, 1496-105943, 1496-105944, 1496-105945, 1496-105946, 1496-105947, 1496-105948, 1496-105949, 1496-105950, 1496-105951, 1496-105952, 1496-105953, 1496-105954, 1496-105955, 1496-105956, 1496-105957, 1496-105958, 1496-105959, 1496-105960, 1496-105961, 1496-105962, 1496-105963, 1496-105964, 1496-105965, 1496-105966, 1496-105967, 1496-105968, 1496-105969, 1496-105970, 1496-105971, 1496-105972, 1496-105973, 1496-105974, 1496-105975, 1496-105976, 1496-105977, 1496-105978, 1496-105979, 1496-105980, 1496-105981, 1496-105982, 1496-105983, 1496-105984, 1496-105985, 1496-105986, 1496-105987, 1496-105988, 1496-105989, 1496-105990, 1496-105991, 1496-105992, 1496-105993, 1496-105994, 1496-105995, 1496-105996, 1496-105997, 1496-105998, 1496-105999, 1496-106000, 1496-106001, 1496-106002, 1496-106003, 1496-106004, 1496-106005, 1496-106006, 1496-106007, 1496-106008, 1496-106009, 1496-106010, 1496-106011, 1496-106012, 1496-106013, 1496-106014, 1496-106015, 1496-106016, 1496-106017, 1496-106018, 1496-106019, 1496-106020, 1496-106021, 1496-106022, 1496-106023, 1496-106024, 1496-106025, 1496-106026, 1496-106027, 1496-106028, 1496-106029, 1496-106030, 1496-106031, 1496-106032, 1496-106033, 1496-106034, 1496-106035, 1496-106036, 1496-106037, 1496-106038, 1496-106039, 1496-106040, 1496-106041, 1496-106042, 1496-106043, 1496-106044, 1496-106045, 1496-106046, 1496-106047, 1496-106048, 1496-106049, 1496-106050, 1496-106051, 1496-106052, 1496-106053, 1496-106054, 1496-106055, 1496-106056, 1496-106057, 1496-106058, 1496-106059, 1496-106060, 1496-106061, 1496-106062, 1496-106063, 1496-106064, 1496-106065, 1496-106066, 1496-106067, 1496-106068, 1496-106069, 1496-106070, 1496-106071, 1496-106072, 1496-106073, 1496-106074, 1496-106286, 1496-106287, 1496-106288, 1496-106289, 1496-106290, 1496-106291, 1496-106292, 1496-106293, 1496-106294, 1496-106295, 1496-106296, 1496-106297, 1496-106298, 1496-106299, 1496-106300, 1496-106301, 1496-106302, 1496-106303, 1496-106304, 1496-106305, 1496-106306, 1496-106307, 1496-106308, 1496-106309, 1496-106310, 1496-106311, 1496-106458, 1496-106459, 1496-106460, 1496-106461, 1496-106462, 1496-106463, 1496-106464, 1496-106465, 1496-106466, 1496-106467, 1496-106468, 1496-106469, 1496-106470, 1496-106471, 1496-106472, 1496-106498, 1496-106500, 1496-106501, 1496-106502, 1496-106503, 1496-106504, 1496-106505, 1496-106506, 1496-106507, 1496-106508, 1496-106509, 1496-106510, 1496-106511, 1496-106512, 1496-106513, 1496-106514, 1496-106515, 1496-106516, 1496-106517, 1496-106518, 1496-106519, 1496-106520, 1496-106521, 1496-106522, 1496-106523, 1496-106524, 1496-106525, 1496-106526, 1496-106527, 1496-106528, 1496-106529, 1496-107878, 1496-107882, 1496-107884, 1496-107903, 1496-107904, 1496-107905, 1496-107906, 1496-109257, 1496-109258, 1496-109259, 1496-109260, 1496-109261, 1496-109263, 1496-109264, 1496-109265, 1496-109266, 1496-109267, 1496-109298, 1496-109299, 1496-109300, 1496-109301, 1496-109302, 1496-109303, 1496-109304, 1496-109305, 1496-109306, 1496-109322, 1496-109323, 1496-109324, 1496-109325, 1496-112374, 1496-112632, 1496-117845, 1496-120087, 1496-120088, 1496-120089, 1496-120090, 1496-120091, 1496-120605, 1496-120725, 1496-120726, 1496-120727, 1496-120728, 1496-120729, 1496-120730, 1496-120731, 1496-120732, 1496-120733, 1496-120734, 1496-120757, 1496-120758, 1496-120759, 1496-120760, 1496-120761, 1496-120762, 1496-120763, 1496-120764, 1496-120765, 1496-120766, 1496-120767, 1496-120768, 1496-120769, 1496-121121, 1496-121973, 1496-121974, 1496-121975, 1496-121976, 1496-121977, 1496-121978, 1496-121979, 1496-121980, 1496-121981, 1496-121982, 1496-121983, 1496-121984, 1496-121985, 1496-121986, 1496-121987, 1496-121988, 1496-121989, 1496-121990, 1496-121991, and 1496-121992. **********SERIAL NUMBERS INTERNATIONAL********************************* 1496-001139, 1496-001925, 1496-002364, 1496-002372, 1496-002410, 1496-002411, 1496-002483, 1496-002486, 1496-002501, 1496-002509, 1496-002521, 1496-002542, 1496-002543, 1496-003059, 1496-006322, 1496-006323, 1496-006324, 1496-006702, 1496-009407, 1496-009518, 1496-009550, 1496-009739, 1496-009953, 1496-009979, 1496-010250, 1496-010251, 1496-010257, 1496-010259, 1496-010261, 1496-010779, 1496-011011, 1496-011322, 1496-011553, 1496-011927, 1496-012393, 1496-012715, 1496-012729, 1496-014969, 1496-016724, 1496-017329, 1496-019070, 1496-033510, 1496-038108, 1496-113895, 1496-113896, 1496-113911, 1496-113935, 1496-113936, 1496-113938, 1496-113939, 1496-113940, 1496-113941, 1496-113944, 1496-113945, 1496-113946, 1496-113947, 1496-113948, 1496-113949, 1496-113950, 1496-113951, 1496-113952, 1496-113953, 1496-113954, 1496-113956, 1496-113957, 1496-113958, and 1496-113959.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1542-2015
Related Recalls
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.