Hitachi Oasis MRI System
Class I - DangerousWhat Should You Do?
- Check if you have this product: M001, M002, M003, M004, M005, M006, M007, M008, M009, M010, M011, M012, M013, M014, M015, M016, M017, M018, M019, M020, M021, M022, M023, M024, M025, M026, M027, M028, M030, M031, M032, M033, M034, M035, M036, M037, M038, M039, M040, M041, M042, M043, M044, M045, M046, M047, M048, M049, M050, M051, M052, M053, M054, M055, M056, M057, M058, M059, M060, M061, M062, M063, M064, M065, M066, M067, M068, M069, M070, M071, M072, M073, M074, M075, M076, M077, M078, M079, M080, M081, M082, M083, M084, M085, M086, M087, M088, M089, M090, M091, M092, M093, M094, M095, M096, M097, M098, M099, M100, M101, M102, M103, M104, M105, M106, M107, M108, M110, M111, M112, M113, M114, M115, M116, M117, M118, M119, M120, M121, M122, M123, M124, M125, M126, M127, M128, M129, M130, M131, M132, M133, M134, M135, M136, M137, M138, M139, M140, M141, M142, M143, M144, M145, M146, M147, M148, M149, M150, M151, M152, M153, M154, M155, M156, M157, M158, M159, M160, M161, M162, M163, M164, M165, M166, M167, M168, M169, M170, M171, M172, M173, M174, M175, M176, M177, M178, M179, M180, M181, M182, M183, M184, M185, M186, M187, M188, M189, M190, M191, M192, M193, M194, M195, M196, M197, M198, M199, M200, M201, M202, M203, M204, M205, M206, M207, M208, M209, M210, M951, M952
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Hitachi Medical Systems America Inc
- Reason for Recall:
- Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Hitachi Oasis MRI System
Product Codes/Lot Numbers:
M001, M002, M003, M004, M005, M006, M007, M008, M009, M010, M011, M012, M013, M014, M015, M016, M017, M018, M019, M020, M021, M022, M023, M024, M025, M026, M027, M028, M030, M031, M032, M033, M034, M035, M036, M037, M038, M039, M040, M041, M042, M043, M044, M045, M046, M047, M048, M049, M050, M051, M052, M053, M054, M055, M056, M057, M058, M059, M060, M061, M062, M063, M064, M065, M066, M067, M068, M069, M070, M071, M072, M073, M074, M075, M076, M077, M078, M079, M080, M081, M082, M083, M084, M085, M086, M087, M088, M089, M090, M091, M092, M093, M094, M095, M096, M097, M098, M099, M100, M101, M102, M103, M104, M105, M106, M107, M108, M110, M111, M112, M113, M114, M115, M116, M117, M118, M119, M120, M121, M122, M123, M124, M125, M126, M127, M128, M129, M130, M131, M132, M133, M134, M135, M136, M137, M138, M139, M140, M141, M142, M143, M144, M145, M146, M147, M148, M149, M150, M151, M152, M153, M154, M155, M156, M157, M158, M159, M160, M161, M162, M163, M164, M165, M166, M167, M168, M169, M170, M171, M172, M173, M174, M175, M176, M177, M178, M179, M180, M181, M182, M183, M184, M185, M186, M187, M188, M189, M190, M191, M192, M193, M194, M195, M196, M197, M198, M199, M200, M201, M202, M203, M204, M205, M206, M207, M208, M209, M210, M951, M952
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1540-2017
Related Recalls
Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic ultrasound imaging or fluid flow analysis of the human body as Continuous-wave Doppler (CWD) for Cardiac Adult, Cardiac Pediatric, and Peripheral Vessel.
Hitachi Medical Systems America
These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.
Hitachi Scenaria Whole-body X-ray CT System
Hitachi Medical Systems America
There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.