ADVIA Chemistry Total Bilirubin_2 assay- SMN: 10341115 (40 mL) & 10341113 (70 mL). Product Usage: ADVIA TBIL_2 is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA¿ Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Class I - DangerousWhat Should You Do?
- Check if you have this product: ADVIA CH TBIL_2 (40 mL) 492094 (01)00630414513072(10)492094(17)20210828 ADVIA CH TBIL_2 (40 mL) 486525 (01)00630414513072(10)486525(17)20210728 ADVIA CH TBIL_2 (40 mL) 478149 (01)00630414513072(10)478149(17)20210428 ADVIA CH TBIL_2 (40 mL) 472677 (01)00630414513072(10)472677(17)20210328 ADVIA CH TBIL_2 (40 mL) 469083 (01)00630414513072(10)469083(17)20210128 ADVIA CH TBIL_2 (40 mL) 462464 (01)00630414513072(10)462464(17)20201128 ADVIA CH TBIL_2 (40 mL) 455486 (01)00630414513072(10)455486(17)20200828 ADVIA CH TBIL_2 (40 mL) 452851 (01)00630414513072(10)452851(17)20200728 ADVIA CH TBIL_2 (40 mL) 447317 (01)00630414513072(10)447317(17)20200628 ADVIA CH TBIL_2 (40 mL) 436498 (01)00630414513072(10)436498(17)20200328 ADVIA CH TBIL_2 (70 mL) 492102 (01)00630414513058(10)492102(17)20210828 ADVIA CH TBIL_2 (70 mL) 492111 (01)00630414513058(10)492111(17)20210828 ADVIA CH TBIL_2 (70 mL) 486533 (01)00630414513058(10)486533(17)20210728 ADVIA CH TBIL_2 (70 mL) 478154 (01)00630414513058(10)478154(17)20210428 ADVIA CH TBIL_2 (70 mL) 478159 (01)00630414513058(10)478159(17)20210428 ADVIA CH TBIL_2 (70 mL) 493710 (01)00630414513058(10)493710(17)20210428 ADVIA CH TBIL_2 (70 mL) 490263 (01)00630414513058(10)490263(17)20210328 ADVIA CH TBIL_2 (70 mL) 490264 (01)00630414513058(10)490264(17)20210328 ADVIA CH TBIL_2 (70 mL) 469088 (01)00630414513058(10)469088(17)20210128 ADVIA CH TBIL_2 (70 mL) 469094 (01)00630414513058(10)469094(17)20210128 ADVIA CH TBIL_2 (70 mL) 462470 (01)00630414513058(10)462470(17)20201128 ADVIA CH TBIL_2 (70 mL) 462477 (01)00630414513058(10)462477(17)20201128 ADVIA CH TBIL_2 (70 mL) 455496 (01)00630414513058(10)455496(17)20200828 ADVIA CH TBIL_2 (70 mL) 455501 (01)00630414513058(10)455501(17)20200828 ADVIA CH TBIL_2 (70 mL) 452858 (01)00630414513058(10)452858(17)20200728 ADVIA CH TBIL_2 (70 mL) 452863 (01)00630414513058(10)452863(17)20200728 ADVIA CH TBIL_2 (70 mL) 452870 (01)00630414513058(10)452870(17)20200728 ADVIA CH TBIL_2 (70 mL) 447325 (01)00630414513058(10)447325(17)20200628 ADVIA CH TBIL_2 (70 mL) 447332 (01)00630414513058(10)447332(17)20200628 ADVIA CH TBIL_2 (70 mL) 436510 (01)00630414513058(10)436510(17)20200328 ADVIA CH TBIL_2 (70 mL) 436515 (01)00630414513058(10)436515(17)20200328
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ADVIA Chemistry Total Bilirubin_2 assay- SMN: 10341115 (40 mL) & 10341113 (70 mL). Product Usage: ADVIA TBIL_2 is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA¿ Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Product Codes/Lot Numbers:
ADVIA CH TBIL_2 (40 mL) 492094 (01)00630414513072(10)492094(17)20210828 ADVIA CH TBIL_2 (40 mL) 486525 (01)00630414513072(10)486525(17)20210728 ADVIA CH TBIL_2 (40 mL) 478149 (01)00630414513072(10)478149(17)20210428 ADVIA CH TBIL_2 (40 mL) 472677 (01)00630414513072(10)472677(17)20210328 ADVIA CH TBIL_2 (40 mL) 469083 (01)00630414513072(10)469083(17)20210128 ADVIA CH TBIL_2 (40 mL) 462464 (01)00630414513072(10)462464(17)20201128 ADVIA CH TBIL_2 (40 mL) 455486 (01)00630414513072(10)455486(17)20200828 ADVIA CH TBIL_2 (40 mL) 452851 (01)00630414513072(10)452851(17)20200728 ADVIA CH TBIL_2 (40 mL) 447317 (01)00630414513072(10)447317(17)20200628 ADVIA CH TBIL_2 (40 mL) 436498 (01)00630414513072(10)436498(17)20200328 ADVIA CH TBIL_2 (70 mL) 492102 (01)00630414513058(10)492102(17)20210828 ADVIA CH TBIL_2 (70 mL) 492111 (01)00630414513058(10)492111(17)20210828 ADVIA CH TBIL_2 (70 mL) 486533 (01)00630414513058(10)486533(17)20210728 ADVIA CH TBIL_2 (70 mL) 478154 (01)00630414513058(10)478154(17)20210428 ADVIA CH TBIL_2 (70 mL) 478159 (01)00630414513058(10)478159(17)20210428 ADVIA CH TBIL_2 (70 mL) 493710 (01)00630414513058(10)493710(17)20210428 ADVIA CH TBIL_2 (70 mL) 490263 (01)00630414513058(10)490263(17)20210328 ADVIA CH TBIL_2 (70 mL) 490264 (01)00630414513058(10)490264(17)20210328 ADVIA CH TBIL_2 (70 mL) 469088 (01)00630414513058(10)469088(17)20210128 ADVIA CH TBIL_2 (70 mL) 469094 (01)00630414513058(10)469094(17)20210128 ADVIA CH TBIL_2 (70 mL) 462470 (01)00630414513058(10)462470(17)20201128 ADVIA CH TBIL_2 (70 mL) 462477 (01)00630414513058(10)462477(17)20201128 ADVIA CH TBIL_2 (70 mL) 455496 (01)00630414513058(10)455496(17)20200828 ADVIA CH TBIL_2 (70 mL) 455501 (01)00630414513058(10)455501(17)20200828 ADVIA CH TBIL_2 (70 mL) 452858 (01)00630414513058(10)452858(17)20200728 ADVIA CH TBIL_2 (70 mL) 452863 (01)00630414513058(10)452863(17)20200728 ADVIA CH TBIL_2 (70 mL) 452870 (01)00630414513058(10)452870(17)20200728 ADVIA CH TBIL_2 (70 mL) 447325 (01)00630414513058(10)447325(17)20200628 ADVIA CH TBIL_2 (70 mL) 447332 (01)00630414513058(10)447332(17)20200628 ADVIA CH TBIL_2 (70 mL) 436510 (01)00630414513058(10)436510(17)20200328 ADVIA CH TBIL_2 (70 mL) 436515 (01)00630414513058(10)436515(17)20200328
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1539-2020
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.