LINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 8 MM X 50 MM Item Number: 15-2030/14
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 04026575436712 All lot numbers within the labeled expiration date
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Linkbio Corp.
- Reason for Recall:
- Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
LINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 8 MM X 50 MM Item Number: 15-2030/14
Product Codes/Lot Numbers:
UDI-DI: 04026575436712 All lot numbers within the labeled expiration date
Distribution:
Distributed in: NY, NV
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1537-2023
Related Recalls
Contains the incorrect implant in the packaging. The packaging incorrectly contains the opposite side of the implant (Lateral-Right/Medial-Left instead of Medial-Right/Lateral-Left).
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
LINK SLED Knee System POREX ENDO MODEL SLED TIBIAL COMP 11 MM X 55 MM Item Number: 15-2230/11
Linkbio
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected