IWS Infant Warmer System-Infant radiant warmers provide infrared heat, in a controlled manner, to neonates who are unable to thermoregulate based on their own physiology or necessitate external heat to smoothen the transition from the uterus to the external environment. They fulfill the same purpose of neonatal incubators but they may be preferred over incubators when total access to the patient is needed or desirable (e.g., surgical procedures, extracorporeal membrane oxygenation, frequent resuscitation, etc.). Most infant radiant warmers can be used in two operating modes: 1. Manual: The clinician sets the appropriate heater output for maintaining the desired patient temperature. The heater output is initially selected based on the clinician s training and experience and then is adjusted based on the patients needs and clinical status. 2. Servo: The clinician sets the desired patient temperature. A skin temperature probe senses the patient temperature and feeds this information to the controller of the infant radiant warmer. The controller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Infant radiant warmers have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Infant radiant warmers may incorporate other features, such as phototherapy, observation light, tilting of the bed, and data output to remote monitors or nurse call systems
Class I - DangerousWhat Should You Do?
- Check if you have this product: See Attached Spreadsheat
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ohmeda Medical
- Reason for Recall:
- Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
IWS Infant Warmer System-Infant radiant warmers provide infrared heat, in a controlled manner, to neonates who are unable to thermoregulate based on their own physiology or necessitate external heat to smoothen the transition from the uterus to the external environment. They fulfill the same purpose of neonatal incubators but they may be preferred over incubators when total access to the patient is needed or desirable (e.g., surgical procedures, extracorporeal membrane oxygenation, frequent resuscitation, etc.). Most infant radiant warmers can be used in two operating modes: 1. Manual: The clinician sets the appropriate heater output for maintaining the desired patient temperature. The heater output is initially selected based on the clinician s training and experience and then is adjusted based on the patients needs and clinical status. 2. Servo: The clinician sets the desired patient temperature. A skin temperature probe senses the patient temperature and feeds this information to the controller of the infant radiant warmer. The controller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Infant radiant warmers have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Infant radiant warmers may incorporate other features, such as phototherapy, observation light, tilting of the bed, and data output to remote monitors or nurse call systems
Product Codes/Lot Numbers:
See Attached Spreadsheat
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1511-2017
Related Recalls
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.
There is the potential for the Rotating IV Pole to fall. A fall could result in an injury to a bystander holding the patient or to the operator.
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.