HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Number: 20IB0 Exp. Date: 2025/09/04
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Guangdong Haiou Medical Apparatus Co., Ltd
- Reason for Recall:
- Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25
Product Codes/Lot Numbers:
Lot Number: 20IB0 Exp. Date: 2025/09/04
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1506-2022
Related Recalls
Haiou Needle Retractable Safety Syringe with Needle-1ML 23G x 1 (0.6x25mm). 200pcs/Box Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-23G25
Guangdong Haiou Medical Apparatus Co.
Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user