Ultraview SL (UVSL) Command Module, Model 91496
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software version 2.03.06 and earlier and software version 2.04 and later.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Spacelabs Healthcare, Inc.
- Reason for Recall:
- Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Ultraview SL (UVSL) Command Module, Model 91496
Product Codes/Lot Numbers:
Software version 2.03.06 and earlier and software version 2.04 and later.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1503-2023
Related Recalls
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.