Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Nos. 20113 20034 21100 21086 21113 21088 21103 21221 21223 21224 21085 20045 20093 20110 21066 21031 21139 21006 20155 21112 21122 21185 20088 20089 20106 20107 21052 21089 21213 20167 21098 21174 21057 21101 21116 21117 21118 21120 21091 21092 21037 21076 21077 21200 21203 21209 20149 21163 21172 21171 21173 20094 21115 20047 21194 20193 21211 21212 21186 21188 21094 21039 21040 21108 20035 20185 21061 21065 21205 21095 21148 21189 21140 21175 21023 21001 21127 21123 21013 20159 21160
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems controls may (depending on the values being processed) initiate a correction of the tube stand position, resulting in an unexpected movement.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.

Product Codes/Lot Numbers:

Serial Nos. 20113 20034 21100 21086 21113 21088 21103 21221 21223 21224 21085 20045 20093 20110 21066 21031 21139 21006 20155 21112 21122 21185 20088 20089 20106 20107 21052 21089 21213 20167 21098 21174 21057 21101 21116 21117 21118 21120 21091 21092 21037 21076 21077 21200 21203 21209 20149 21163 21172 21171 21173 20094 21115 20047 21194 20193 21211 21212 21186 21188 21094 21039 21040 21108 20035 20185 21061 21065 21205 21095 21148 21189 21140 21175 21023 21001 21127 21123 21013 20159 21160

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1501-2020

Related Recalls

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →
Class I - Dangerous

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA

Class I - Dangerous

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Mar 12, 2026 Other Medical Devices View Details →