ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: 04056869046877 S/N: Serial # 164140 164159 164138 164332 164331 164033 164287 164282 164210 164184 164183 164311 164142 164312 164071 164328 164034 164391 164318 164244 164306 164155 164340 164189 164277 164173 164292 164231 164095 164057 164055 164054 164329 164072 164190 164048 164025 164026 164027 164338 164094 164241 164326 164267 164186 164068 164121 164157 164019 164152 164161 164013 164069 164213 164073 164293 164263 164298 164060 164083 164113 164021 164066 164398 164362 164124 164125 164366 164279 164075 164098 164028 164153 164330 164389 164394 164172 164086 164084 164074 164295 164353 164234 164321 164242 164275 164270 164315 164344 164163 164158 164053 164171 164297 164134 164253 164333 164188 164377 164403 164063 164064 164397 164300 164299 164105 164160 164024 164080 164049 164240 164359 164227 164228 164018 164022 164250 164061 164187 164115 164399 164011 164211 164216 164204 164205 164212 164100 164097 164269 164303 164322 164355 164400 164196 164352 164168 164319 164201 164202 164371
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the C-Arm will reach the target position with an inaccuracy of 5-10 mm. As a result, the message Endposition reached will be displayed .Leading to unintended direction of the movement causing crushing of a patient, staff member, operator, or equipmen Live images may not match the previously stored reference images. Overlay images may be shown inaccurate on anatomy (e.g., DSA Roadmap workflow does not match real anatomy). This may cause e.g., a vessel perforation in DSA roadmap.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Product Codes/Lot Numbers:
UDI: 04056869046877 S/N: Serial # 164140 164159 164138 164332 164331 164033 164287 164282 164210 164184 164183 164311 164142 164312 164071 164328 164034 164391 164318 164244 164306 164155 164340 164189 164277 164173 164292 164231 164095 164057 164055 164054 164329 164072 164190 164048 164025 164026 164027 164338 164094 164241 164326 164267 164186 164068 164121 164157 164019 164152 164161 164013 164069 164213 164073 164293 164263 164298 164060 164083 164113 164021 164066 164398 164362 164124 164125 164366 164279 164075 164098 164028 164153 164330 164389 164394 164172 164086 164084 164074 164295 164353 164234 164321 164242 164275 164270 164315 164344 164163 164158 164053 164171 164297 164134 164253 164333 164188 164377 164403 164063 164064 164397 164300 164299 164105 164160 164024 164080 164049 164240 164359 164227 164228 164018 164022 164250 164061 164187 164115 164399 164011 164211 164216 164204 164205 164212 164100 164097 164269 164303 164322 164355 164400 164196 164352 164168 164319 164201 164202 164371
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1498-2023
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.