Merlin PCS 3650 programmer Model 3330 software

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Software Model 3330 v26.0.1 rev 1 or earlier; UDI/DI 05414734509725
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
St. Jude Medical, Cardiac Rhythm Management Division
Reason for Recall:
There is a potential for programmer software (Merlin PCS) and remote monitoring software application (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Merlin PCS 3650 programmer Model 3330 software

Product Codes/Lot Numbers:

Software Model 3330 v26.0.1 rev 1 or earlier; UDI/DI 05414734509725

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1495-2022

Related Recalls

Merlin 2 PCS MER37000 programmer Model MER3400 software

St. Jude Medical, Cardiac Rhythm Management Division

Class I - Dangerous

The Merlin" 2 PCS model MER3700 is a portable, dedicated programming system designed to interrogate, program, display data from, and test Abbott Medical implantable medical devices during implant and follow-up.

Jun 17, 2022 Other Medical Devices Nationwide View Details →

Merlin.net model MN5000 Software

St. Jude Medical, Cardiac Rhythm Management Division

Class I - Dangerous

The Merlin" 2 PCS model MER3700 is a portable, dedicated programming system designed to interrogate, program, display data from, and test Abbott Medical implantable medical devices during implant and follow-up.

Jun 17, 2022 Other Medical Devices Nationwide View Details →

Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q

St. Jude Medical, Cardiac Rhythm Management Division

Class I - Dangerous

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Mar 10, 2022 Implants & Prosthetics Nationwide View Details →