CROSSCHECK(R) Plating system, Utility Plate, 2-Hole, REF CCP-UTN2. Orthopedic use.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: 3504927, 35049-27A, 35049-27A-1, 35049-28, 362801, 36280-1-1, 8752-010381
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Wright Medical Technology, Inc.
- Reason for Recall:
- The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CROSSCHECK(R) Plating system, Utility Plate, 2-Hole, REF CCP-UTN2. Orthopedic use.
Product Codes/Lot Numbers:
Lot Numbers: 3504927, 35049-27A, 35049-27A-1, 35049-28, 362801, 36280-1-1, 8752-010381
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1494-2015
Related Recalls
stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty
Wright Medical Technology
Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.
Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.
Wright Medical Technology
The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A221RT, bone stabilization and fixation
Wright Medical Technology
Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.