Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a software package that will add additional features to the Spectrum Infusion Pump. The Master Drug Library will permit electronic communications with the Spectrum pump and other external peripheral devices. The intended use of the Spectrum pump includes common drug error prevention, through the stand alone settings features of the pump. This includes drug parameter limits and associated drug name identification. With the Master Drug Library, the intended use is to reduce user errors associated with drug selection, drug dose rates, drug dose concentrations, and patient identification associated with the prescribed drug.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial numbers ranging from 700000 through 794213: 700000, 700001, 700002, 700003, 700004, 700005, 700006, 700007, 700010, 700012, 700013, 700014, 700017, 700018, 700020, 700021, 700022, 700023, 700024, 700029, 700030, 700031, 700032, 700033, 700034, 700035, 700037, 700038, 700039, 700040, 700041, 700042, 700043, 700045, 700046, 700047, 700048, 700049, 700050, 700051, 700052, 700053, 700054, 700055, 700057, 700058, 700059, 700060, 700061, 700062, 700063, 700064, 700065, 700066, 700067, 700069, 700071, 700073, 700075, 700076, 700077, 700080, 700081, 700082, 700085, 700086, 700088, 700089, 700090, 700093, 700094, 700095, 700097, 700098, 700100, 700101, 700103, 700104, 700106, 700108, 700109, 700110, 700111, 700112, 700114, 700117, 700118, 700119, 700120, 700121, 700123, 700126, 700127, 700129, 700130, 700132, 700133, 700134, 700135, 700137, 700138, 700139, 700141, 700142, 700143, 700144, 700145, 700147, 700149, 700151, 700152, 700153, 700156, 700157, 700158, 700159, 700161, 700162, 700163, 700165, 700166, 700168, 700169, 700170, 700172, 700173, 700174, 700175, 700176, 700177, 700178, 700179, 700180, 700183, 700184, 700187, 700189, 700193, 700194, 700195, 700197, 700201, 700202, 700203, 700204, 700205, 700206, 700207, 700208, 700209, 700210, 700211, 700212, 700213, 700214, 700215, 700216, 700217, 700218, 700219, 700220, 700222, 700223, 700224, 700225, 700226, 700229, 700231, 700232, 700233, 700234, 700235, 700236, 700237, 700238, 700239, 700240, 700241, 700242, 700243, 700244, 700245, 700246, 700248, 700249, 700250, 700251, 700252, 700253, 700255, 700256, 700257, 700258, 700259, 700260, 700263, 700265, 700267, 700268, 700269, 700271, 700272, 700273, 700274, 700275, 700277, 700278, 700280, 700282, 700283, 700284, 700285, 700286, 700287, 700288, 700289, 700292, 700293, 700294, 700295, 700297, 700298, 700299, 700301, 700302, 700303, 700305, 700306, 700307, 700308, 700309, 700310, 700311, 700312, 700314, 700316, 700317, 700318, 700319, 700320, 700321, 700322, 700323, 700324, 700326, 700327, 700328, 700329, 700330, 700331, 700332, 700333, 700334, 700335, 700338, 700339, 700340, 700341, 700343, 700344, 700345, 700346, 700348, 700349, 700350, 700351, 700352, 700353, 700354, 700355, 700357, 700358, 700359, 700360, 700361, 700362, 700363, 700364, 700365, 700366, 700367, 700368, 700369, 700370, 700371, 700372, 700373, 700374, 700375, 700376, 700377, 700378, 700379, 700380, 700381, 700382, 700383, 700384, 700385, 700386, 700387, 700388, 700389, 700390, 700392, 700395, 700396, 700398, 700399, 700400, 700401, 700402, 700403, 700404, 700405, 700406, 700407, 700408, 700409, 700410, 700411, 700412, 700413, 700414, 700415, 700416, 700417, 700419, 700420, 700421, 700422, 700423, 700424, 700425, 700426, 700427, 700428, 700429, 700433, 700434, 700436, 700437, 700438, 700439, 700440, 700441, 700442, 700443, 700444, 700446, 700447, 700448, 700449, 700450, 700451, 700452, 700453, 700454, 700455, 700456, 700457, 700458, 700459, 700460, 700461, 700462, 700463, 700464, 700466, 700468, 700470, 700471, 700472, 700473, 700474, 700475, 700476, 700478, 700479, 700480, 700481, 700483, 700484, 700485, 700489, 700490, 700491, 700494, 700495, 700496, 700497, 700502, 700503, 700504, 700505, 700506, 700507, 700508, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700519, 700520, 700524, 700525, 700526, 700527, 700528, 700530, 700531, 700532, 700533, 700534, 700536, 700537, 700538, 700539, 700540, 700542, 700548, 700552, 700553, 700555, 700556, 700557, 700558, 700560, 700561, 700562, 700563, 700564, 700565, 700566, 700567, 700568, 700569, 700570, 700571, 700572, 700573, 700574, 700575, 700576, 700577, 700578, 700579, 700580, 700581, 700582, 700583, 700584, 700585, 700586, 700587, 700588, 700589, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700598, 700599, 700600, 700601, 700602, 700603, 700604, 700605, 700606, 700607, 700608, 700611, 700612, 700613, 700614, 700616, 700617, 700618, 700619, 700620, 700621, 700622, 700623, 700624, 700625, 700626, 700627, 700629, 700631, 700632, 700633, 700634, 700635, 700636, 700638, 700639, 700641, 700642, 700643, 700646, 700647, 700648, 700649, 700650, 700652, 700654, 700655, 700656, 700657, 700659, 700661, 700662, 700663, 700664, 700665, 700666, 700667, 700668, 700669, 700670, 700671, 700672, 700673, 700674, 700676, 700677, 700681, 700682, 700683, 700684, 700685, 700687, 700689, 700690, 700691, 700693, 700694, 700695, 700696, 700697, 700698, 700699, 700700, 700701, 700702, 700703, 700704, 700707, 700708, 700709, 700710, 700711, 700712, 700713, 700715, 700716, 700717, 700718, 700721, 700722, 700723, 700724, 700725, 700726, 700727, 700729, 700731, 700732, 700733, 700734, 700735, 700736, 700737, 700738, 700741, 700742, 700743, 700744, 700747, 700748, 700749, 700750, 700751, 700752, 700753, 700754, 700755, 700756, 700758, 700759, 700761, 700762, 700763, 700765, 700766, 700767, 700768, 700769, 700770, 700771, 700773, 700774, 700775, 700776, 700777, 700778, 700780, 700781, 700782, 700783, 700784, 700785, 700790, 700791, 700792, 700794, 700797, 700798, 700799, 700800, 700803, 700805, 700807, 700808, 700809, 700810, 700811, 700813, 700815, 700817, 700818, 700819, 700821, 700822, 700823, 700826, 700828, 700829, 700830, 700831, 700832, 700833, 700834, 700835, 700836, 700838, 700839, 700840, 700844, 700845, 700846, 700849, 700850, 700851, 700852, 700854, 700856, 700857, 700858, 700859, 700860, 700861, 700862, 700863, 700864, 700865, 700866, 700867, 700870, 700871, 700873, 700874, 700875, 700877, 700878, 700880, 700881, 700882, 700883, 700884, 700885, 700887, 700888, 700889, 700890, 700891, 700892, 700893, 700894, 700896, 700897, 700898, 700900, 700902, 700904, 700905, 700908, 700909, 700910, 700911, 700913, 700914, 700919, 700920, 700921, 700925, 700926, 700927, 700928, 700929, 700930, 700931, 700932, 700934, 700935, 700936, 700937, 700938, 700939, 700940, 700941, 700942, 700943, 700944, 700945, 700946, 700947, 700948, 700949, 700950, 700951, 700952, 700953, 700954, 700955, 700956, 700958, 700959, 700960, 700961, 700962, 700963, 700964, 700966, 700967, 700968, 700969, 700970, 700971, 700972, 700973, 700974, 700975, 700976, 700977, 700978, 700979, 700980, 700981, 700982, 700983, 700984, 700985, 700986, 700987, 700988, 700989, 700990, 700991, 700992, 700993, 700994, 700995, 700997, 700998, 700999, 701000, 701002, 701003, 701004, 701005, 701007, 701008, 701009, 701010, 701011, 701012, 701013, 701014, 701015, 701016, 701017, 701018, 701019, 701020, 701022, 701025, 701027, 701028, 701029, 701030, 701031, 701032, 701033, 701034, 701036, 701037, 701039, 701040, 701041, 701042, 701043, 701045, 701046, 701047, 701048, 701049, 701050, 701051, 701052, 701053, 701054, 701056, 701057, 701058, 701059, 701060, 701061, 701063, 701064, 701065, 701066, 701067, 701068, 701069, 701070, 701071, 701072, 701073, 701074, 701075, 701076, 701077, 701078, 701079, 701080, 701081, 701082, 701083, 701084, 701085, 701086, 701087, 701088, 701089, 701092, 701093, 701094, 701095, 701096, 701097, 701098, 701099, 701100, 701101, 701102, 701105, 701106, 701107, 701108, 701109, 701110, 701111, 701112, 701113, 701114, 701115, 701116, 701117, 701118, 701119, 701120, 701121, 701122, 701123, 701124, 701125, 701126, 701127, 701128, 701129, 701130, 701131, 701132, 701133, 701134, 701135, 701136, 701137, 701138, 701139, 701140, 701141, 701142, 701143, 701144, 701145, 701146, 701147, 701148, 701149, 701150, 701151, 701152, 701153, 701154, 701155, 701156, 701157, 701158, 701159, 701160, 701161, 701162, 701163, 701164, 701165, 701166, 701167, 701168, 701169, 701170, 701171, 701173, 701174, 701175, 701176, 701177, 701178, 701179, 701180, 701181, 701182, 701183, 701184, 701185, 701186, 701187, 701188, 701189, 701190, 701191, 701192, 701193, 701194, 701195, 701196, 701197, 701198, 701199, 701201, 701202, 701203, 701204, 701205, 701206, 701207, 701208, 701209, 701210, 701211, 701212, 701214, 701215, 701216, 701217, 701218, 701219, 701220, 701221, 701222, 701223, 701224, 701225, 701226, 701227, 701228, 701229, 701230, 701231, 701232, 701233, 701234, 701235, 701236, 701237, 701238, 701239, 701240, 701241, 701242, 701243, 701244, 701245, 701246, 701247, 701248, 701249, 701250, 701251, 701252, 701254, 701255, 701256, 701257, 701259, 701260, 701261, 701262, 701263, 701264, 701265, 701266, 701267, 701268, 701269, 701270, 701271, 701272, 701273, 701274, 701276, 701277, 701278, 701279, 701280, 701281, 701283, 701284, 701285, 701286, 701287, 701288, 701289, 701290, 701291, 701292, 701293, 701294, 701295, 701296, 701297, 701298, 701299, 701301, 701302, 701304, 701305, 701306, 701307, 701308, 701309, 701310, 701311, 701313, 701314, 701315, 701317, 701318, 701319, 701320, 701321, 701323, 701324, 701325, 701326, 701327, 701329, 701330, 701331, 701332, 701333, 701334, 701336, 701337, 701338, 701339, 701340, 701341, 701342, 701344, 701345, 701346, 701347, 701348, 701349, 701350, 701351, 701352, 701353, 701354, 701355, 701356, 701357, 701358, 701359, 701360, 701361, 701362, 701363, 701364, 701366, 701367, 701368, 701369, 701370, 701371, 701372, 701373, 701374, 701375, 701376, 701378, 701379, 701380, 701381, 701382, 701383, 701385, 701386, 701387, 701388, 701389, 701390, 701391, 701392, 701393, 701394, 701395, 701396, 701397, 701399, 701400, 701401, 701403, 701404, 701405, 701406, 701407, 701408, 701409, 701410, 701411, 701412, 701413, 701414, 701415, 701417, 701418, 701419, 701420, 701421, 701422, 701423, 701424, 701425, 701426, 701427, 701428, 701429, 701430, 701431, 701432, 701433, 701434, 701435, 701436, 701437, 701438, 701439, 701440, 701441, 701443, 701445, 701446, 701447, 701448, 701449, 701450, 701457, 701462, 701465, 701467, 701472, 701475, 701482, 701483, 701484, 701487, 701488, 701492, 701494, 701495, 701501, 701505, 701507, 701508, 701509, 701511, 701512, 701514, 701515, 701516, 701518, 701520, 701523, 701524, 701525, 701529, 701530, 701532, 701533, 701534, 701535, 701536, 701537, 701538, 701539, 701540, 701541, 701542, 701543, 701544, 701545, 701546, 701547, 701548, 701549, 701550, 701551, 701552, 701554, 701555, 701556, 701557, 701558, 701559, 701563, 701564, 701565, 701566, 701567, 701568, 701569, 701570, 701571, 701573, 701574, 701575, 701576, 701577, 701578, 701579, 701580, 701582, 701583, 701585, 701586, 701588, 701589, 701590, 701591, 701592, 701593, 701594, 701595, 701596, 701597, 701598, 701600, 701601, 701602, 701603, 701604, 701605, 701606, 701607, 701609, 701610, 701611, 701612, 701613, 701614, 701615, 701617, 701618, 701620, 701621, 701622, 701624, 701625, 701626, 701627, 701628, 701629, 701631, 701632, 701633, 701634, 701635, 701636, 701637, 701638, 701639, 701640, 701641, 701642, 701644, 701645, 701646, 701647, 701648, 701649, 701650, 701651, 701652, 701653, 701654, 701655, 701656, 701657, 701658, 701659, 701660, 701661, 701662, 701664, 701665, 701666, 701667, 701668, 701669, 701670, 701671, 701672, 701673, 701674, 701676, 701677, 701678, 701679, 701680, 701681, 701682, 701683, 701684, 701685, 701686, 701687, 701688, 701689, 701690, 701691, 701692, 701693, 701694, 701696, 701697, 701698, 701699, 701700, 701701, 701703, 701704, 701705, 701706, 701707, 701708, 701709, 701711, 701712, 701713, 701714, 701715, 701716, 701717, 701718, 701719, 701720, 701721, 701722, 701724, 701725, 701726, 701727, 701729, 701732, 701733, 701734, 701735, 701737, 701739, 701740, 701741, 701742, 701743, 701744, 701745, 701746, 701747, 701748, 701749, 701750, 701751, 701754, 701755, 701756, 701758, 701759, 701761, 701762, 701763, 701764, 701766, 701767, 701768, 701769, 701770, 701771, 701772, 701773, 701774, 701775, 701776, 701778, 701780, 701781, 701782, 701784, 701787, 701788, 701789, 701790, 701792, 701794, 701796, 701797, 701798, 701799, 701800, 701801, 701802, 701803, 701804, 701805, 701806, 701808, 701809, 701813, 701814, 701818, 701819, 701820, 701821, 701822, 701823, 701825, 701827, 701829, 701830, 701831, 701832, 701833, 701834, 701835, 701836, 701837, 701838, 701839, 701840, 701843, 701844, 701845, 701846, 701848, 701850, 701851, 701852, 701853, 701854, 701856, 701857, 701858, 701859, 701860, 701861, 701863, 701865, 701866, 701867, 701868, 701869, 701870, 701871, 701872, 701873, 701874, 701875, 701876, 701878, 701879, 701880, 701881, 701882, 701883, 701885, 701886, 701888, 701889, 701890, 701891, 701892, 701896, 701897, 701898, 701899, 701900, 701901, 701902, 701903, 701904, 701905, 701906, 701907, 701908, 701912, 701913, 701916, 701917, 701919, 701920, 701921, 701923, 701925, 701927, 701928, 701929, 701931, 701932, 701933, 701935, 701936, 701937, 701938, 701939, 701940, 701941, 701942, 701943, 701944, 701945, 701947, 701949, 701950, 701951, 701953, 701954, 701955, 701956, 701957, 701958, 701959, 701960, 701961, 701963, 701965, 701967, 701969, 701970, 701971, 701972, 701973, 701975, 701976, 701978, 701979, 701980, 701982, 701983, 701984, 701986, 701987, 701988, 701990, 701991, 701993, 701996, 701997, 701998, 702000, 702001, 702002, 702003, 702005, 702006, 702007, 702008, 702011, 702012, 702013, 702014, 702015, 702016, 702018, 702020, 702021, 702022, 702023, 702024, 702025, 702027, 702028, 702029, 702030, 702031, 702032, 702033, 702034, 702035, 702036, 702037, 702038, 702039, 702040, 702041, 702044, 702046, 702047, 702049, 702050, 702051, 702052, 702053, 702054, 702055, 702056, 702057, 702058, 702059, 702060, 702064, 702065, 702066, 702068, 702069, 702070, 702071, 702072, 702073, 702074, 702075, 702076, 702078, 702080, 702083, 702085, 702087, 702088, 702089, 702090, 702092, 702095, 702096, 702098, 702099, 702100, 702106, 702107, 702108, 702109, 702111, 702112, 702113, 702115, 702117, 702118, 702119, 702122, 702123, 702124, 702125, 702128, 702130, 702131, 702132, 702134, 702135, 702136, 702138, 702140, 702141, 702143, 702144, 702146, 702148, 702149, 702152, 702153, 702154, 702155, 702156, 702157, 702158, 702160, 702161, 702162, 702163, 702164, 702165, 702166, 702167, 702168, 702169, 702171, 702172, 702173, 702174, 702175, 702176, 702177, 702178, 702180, 702181, 702182, 702184, 702186, 702187, 702188, 702189, 702190, 702191, 702192, 702194, 702195, 702196, 702197, 702198, 702199, 702200, 702202, 702203, 702204, 702206, 702207, 702210, 702211, 702212, 702214, 702215, 702216, 702218, 702219, 702220, 702221, 702222, 702223, 702224, 702225, 702226, 702227, 702228, 702229, 702230, 702233, 702234, 702235, 702236, 702237, 702238, 702240, 702241, 702242, 702243, 702244, 702245, 702246, 702247, 702248, 702249, 702250, 702251, 702252, 702253, 702254, 702255, 702256, 702257, 702258, 702259, 702260, 702261, 702262, 702263, 702264, 702267, 702268, 702269, 702270, 702271, 702273, 702274, 702275, 702276, 702277, 702278, 702279, 702280, 702281, 702284, 702285, 702286, 702287, 702289, 702290, 702291, 702292, 702293, 702294, 702295, 702296, 702297, 702298, 702299, 702300, 702301, 702302, 702303, 702304, 702305, 702306, 702307, 702308, 702309, 702310, 702311, 702312, 702313, 702314, 702315, 702316, 702317, 702318, 702319, 702320, 702321, 702322, 702323, 702324, 702325, 702326, 702328, 702329, 702330, 702331, 702332, 702333, 702334, 702335, 702336, 702337, 702338, 702339, 702340, 702341, 702343, 702344, 702345, 702346, 702347, 702348, 702349, 702350, 702351, 702352, 702353, 702354, 702355, 702357, 702358, 702360, 702361, 702362, 702364, 702365, 702366, 702368, 702369, 702370, 702371, 702372, 702373, 702374, 702375, 702376, 702377, 702378, 702379, 702380, 702381, 702382, 702383, 702384, 702386, 702387, 702388, 702390, 702391, 702392, 702393, 702394, 702395, 702396, 702397, 702398, 702400, 702401, 702402, 702403, 702405, 702406, 702407, 702408, 702410, 702411, 702412, 702413, 702414, 702415, 702416, 702417, 702418, 702419, 702420, 702421, 702422, 702423, 702424, 702425, 702426, 702427, 702428, 702429, 702430, 702431, 702433, 702434, 702436, 702438, 702439, 702440, 702441, 702443, 702444, 702445, 702446, 702447, 702448, 702449, 702450, 702451, 702452, 702453, 702454, 702455, 702456, 702457, 702458, 702459, 702460, 702461, 702462, 702464, 702465, 702466, 702467, 702468, 702469, 702470, 702471, 702472, 702473, 702474, 702475, 702476, 702477, 702478, 702479, 702480, 702481, 702482, 702483, 702484, 702485, 702486, 702487, 702488, 702490, 702491, 702492, 702493, 702494, 702495, 702496, 702497, 702498, 702499, 702500, 702501, 702502, 702503, 702504, 702505, 702506, 702507, 702508, 702509, 702510, 702511, 702513, 702514, 702515, 702516, 702517, 702520, 702521, 702522, 702523, 702524, 702525, 702526, 702527, 702528, 702529, 702530, 702531, 702532, 702533, 702534, 702535, 702537, 702538, 702540, 702542, 702544, 702545, 702547, 702548, 702550, 702551, 702552, 702553, 702554, 702555, 702556, 702557, 702558, 702559, 702561, 702562, 702563, 702564, 702565, 702566, 702567, 702568, 702570, 702572, 702573, 702574, 702575, 702577, 702579, 702580, 702582, 702583, 702584, 702585, 702586, 702587, 702588, 702589, 702590, 702591, 702592, 702593, 702594, 702595, 702597, 702599, 702601, 702602, 702603, 702604, 702605, 702606, 702607, 702608, 702609, 702610, 702611, 702612, 702613, 702614, 702615, 702616, 702617, 702618, 702619, 702621, 702625, 702626, 702628, 702629, 702630, 702631, 702632, 702633, 702634, 702635, 702636, 702637, 702638, 702639, 702640, 702641, 702642, 702644, 702647, 702648, 702650, 702651, 702652, 702653, 702654, 702656, 702657, 702659, 702660, 702662, 702663, 702666, 702667, 702668, 702670, 702671, 702673, 702674, 702678, 702681, 702682, 702683, 702684, 702685, 702686, 702687, 702689, 702690, 702694, 702695, 702696, 702698, 702699, 702700, 702705, 702706, 702707, 702708, 702709, 702710, 702711, 702712, 702714, 702715, 702716, 702717, 702718, 702719, 702720, 702721, 702722, 702724, 702725, 702726, 702727, 702729, 702730, 702731, 702732, 702733, 702734, 702735, 702736, 702737, 702738, 702740, 702741, 702742, 702743, 702745, 702746, 702747, 702748, 702749, 702750, 702751, 702752, 702753, 702754, 702755, 702757, 702758, 702759, 702760, 702761, 702762, 702763, 702764, 702765, 702767, 702768, 702769, 702770, 702771, 702772, 702773, 702774, 702775, 702776, 702777, 702778, 702779, 702780, 702781, 702782, 702784, 702787, 702788, 702790, 702792, 702793, 702794, 702795, 702796, 702797, 702798, 702799, 702800, 702802, 702804, 702805, 702806, 702807, 702808, 702809, 702810, 702811, 702812, 702813, 702814, 702815, 702816, 702817, 702818, 702819, 702820, 702821, 702823, 702824, 702825, 702826, 702827, 702828, 702829, 702830, 702831, 702832, 702834, 702835, 702836, 702837, 702838, 702839, 702840, 702841, 702843, 702844, 702845, 702846, 702847, 702848, 702849, 702850, 702851, 702852, 702853, 702854, 702855, 702856, 702857, 702858, 702859, 702860, 702861, 702862, 702863, 702865, 702866, 702867, 702868, 702869, 702870, 702872, 702873, 702874, 702875, 702877, 702878, 702879, 702880, 702881, 702882, 702883, 702884, 702885, 702886, 702887, 702888, 702889, 702890, 702892, 702895, 702896, 702898, 702899, 702900, 702901, 702904, 702905, 702906, 702907, 702908, 702909, 702910, 702911, 702913, 702915, 702916, 702917, 702918, 702919, 702920, 702921, 702922, 702923, 702924, 702925, 702926, 702927, 702929, 702930, 702932, 702934, 702935, 702936, 702937, 702938, 702939, 702940, 702941, 702942, 702943, 702944, 702945, 702946, 702947, 702948, 702949, 702951, 702952, 702953, 702954, 702956, 702958, 702959, 702960, 702961, 702962, 702963, 702965, 702967, 702968, 702969, 702970, 702971, 702972, 702973, 702974, 702975, 702976, 702977, 702978, 702979, 702980, 702981, 702982, 702983, 702984, 702987, 702988, 702989, 702990, 702991, 702993, 702994, 702996, 702997, 702998, 703000, 703001, 703002, 703003, 703004, 703005, 703006, 703007, 703008, 703009, 703010, 703011, 703012, 703013, 703014, 703015, 703016, 703017, 703018, 703019, 703020, 703021, 703022, 703023, 703024, 703025, 703027, 703029, 703030, 703031, 703033, 703034, 703035, 703036, 703037, 703038, 703039, 703040, 703041, 703042, 703044, 703045, 703046, 703047, 703048, 703049, 703050, 703051, 703052, 703053, 703054, 703055, 703056, 703057, 703058, 703059, 703060, 703061, 703062, 703063, 703065, 703066, 703067, 703068, 703069, 703070, 703071, 703073, 703074, 703075, 703076, 703077, 703078, 703079, 703080, 703081, 703082, 703083, 703084, 703085, 703086, 703087, 703088, 703089, 703090, 703091, 703092, 703093, 703094, 703095, 703096, 703098, 703099, 703100, 703101, 703102, 703103, 703104, 703105, 703106, 703107, 703108, 703109, 703110, 703111, 703113, 703115, 703116, 703117, 703119, 703120, 703121, 703122, 703123, 703124, 703125, 703126, 703129, 703130, 703131, 703132, 703133, 703134, 703135, 703137, 703138, 703139, 703140, 703141, 703142, 703143, 703144, 703145, 703147, 703149, 703150, 703152, 703153, 703155, 703156, 703157, 703158, 703159, 703161, 703162, 703163, 703164, 703165, 703167, 703168, 703169, 703170, 703171, 703172, 703173, 703174, 703175, 703176, 703177, 703178, 703179, 703180, 703181, 703182, 703183, 703184, 703185, 703186, 703188, 703189, 703190, 703191, 703192, 703193, 703195, 703196, 703197, 703199, 703200, 703201, 703202, 703203, 703204, 703205, 703206, 703207, 703208, 703210, 703211, 703212, 703213, 703214, 703217, 703219, 703220, 703221, 703222, 703223, 703224, 703225, 703226, 703228, 703229, 703230, 703231, 703232, 703233, 703234, 703235, 703236, 703238, 703239, 703240, 703241, 703242, 703243, 703245, 703246, 703247, 703248, 703251, 703252, 703253, 703254, 703255, 703256, 703257, 703258, 703259, 703260, 703262, 703263, 703264, 703265, 703266, 703268, 703269, 703270, 703271, 703272, 703273, 703275, 703276, 703277, 703278, 703279, 703280, 703281, 703282, 703283, 703284, 703285, 703287, 703289, 703290, 703291, 703292, 703293, 703294, 703295, 703296, 703297, 703298, 703299, 703300, 703301, 703304, 703305, 703306, 703307, 703309, 703310, 703311, 703312, 703313, 703314, 703315, 703318, 703319, 703320, 703321, 703323, 703325, 703327, 703328, 703330, 703331, 703332, 703334, 703335, 703336, 703337, 703338, 703339, 703340, 703341, 703342, 703343, 703344, 703345, 703346, 703347, 703348, 703349, 703351, 703352, 703353, 703354, 703356, 703357, 703358, 703359, 703360, 703361, 703362, 703363, 703364, 703365, 703366, 703367, 703368, 703370, 703371, 703372, 703373, 703375, 703376, 703377, 703378, 703379, 703380, 703381, 703382, 703383, 703384, 703385, 703386, 703387, 703389, 703390, 703391, 703393, 703394, 703395, 703396, 703397, 703398, 703399, 703400, 703401, 703402, 703403, 703404, 703405, 703406, 703408, 703410, 703412, 703414, 703415, 703417, 703418, 703419, 703420, 703421, 703422, 703425, 703426, 703427, 703428, 703429, 703430, 703432, 703433, 703434, 703435, 703437, 703438, 703439, 703440, 703441, 703442, 703443, 703444, 703445, 703447, 703448, 703450, 703451, 703453, 703454, 703455, 703456, 703457, 703458, 703459, 703460, 703461, 703462, 703463, 703466, 703467, 703468, 703469, 703470, 703471, 703472, 703473, 703476, 703477, 703480, 703481, 703482, 703484, 703485, 703486, 703487, 703489, 703490, 703491, 703492, 703493, 703494, 703495, 703496, 703498, 703499, 703501, 703502, 703503, 703504, 703505, 703506, 703507, 703508, 703509, 703510, 703511, 703512, 703513, 703514, 703515, 703516, 703517, 703519, 703520, 703522, 703523, 703524, 703525, 703526, 703527, 703529, 703530, 703531, 703533, 703534, 703535, 703536, 703537, 703538, 703539, 703540, 703541, 703542, 703544, 703545, 703546, 703547, 703548, 703549, 703550, 703551, 703552, 703553, 703554, 703555, 703556, 703557, 703558, 703559, 703560, 703561, 703562, 703563, 703564, 703565, 703566, 703567, 703568, 703569, 703570, 703571, 703572, 703574, 703575, 703576, 703578, 703580, 703581, 703582, 703583, 703584, 703586, 703587, 703588, 703589, 703590, 703591, 703592, 703593, 703594, 703595, 703596, 703597, 703599, 703600, 703601, 703602, 703603, 703604, 703605, 703606, 703607, 703608, 703609, 703610, 703611, 703612, 703614, 703615, 703616, 703617, 703618, 703619, 703621, 703622, 703623, 703624, 703625, 703626, 703627, 703628, 703629, 703631, 703632, 703633, 703634, 703635, 703636, 703637, 703639, 703640, 703642, 703643, 703644, 703645, 703646, 703647, 703648, 703649, 703650, 703651, 703652, 703654, 703655, 703656, 703657, 703658, 703660, 703661, 703662, 703663, 703664, 703665, 703666, 703667, 703668, 703669, 703670, 703671, 703672, 703673, 703674, 703675, 703676, 703677, 703678, 703679, 703680, 703681, 703682, 703683, 703684, 703686, 703687, 703689, 703690, 703691, 703692, 703693, 703694, 703695, 703696, 703697, 703698, 703700, 703701, 703702, 703703, 703704, 703705, 703706, 703707, 703708, 703709, 703710, 703711, 703712, 703713, 703714, 703715, 703716, 703718, 703719, 703721, 703722, 703723, 703724, 703726, 703727, 703729, 703731, 703732, 703733, 703734, 703736, 703738, 703739, 703740, 703741, 703743, 703744, 703746, 703747, 703748, 703749, 703750, 703751, 703752, 703753, 703754, 703755, 703756, 703759, 703760, 703761, 703762, 703763, 703764, 703765, 703767, 703768, 703769, 703770, 703771, 703772, 703774, 703775, 703776, 703777, 703778, 703779, 703780, 703781, 703783, 703784, 703785, 703786, 703787, 703788, 703789, 703790, 703791, 703792, 703793, 703794, 703795, 703796, 703797, 703798, 703799, 703800, 703801, 703802, 703803, 703804, 703805, 703806, 703808, 703809, 703810, 703811, 703812, 703813, 703814, 703815, 703816, 703817, 703818, 703819, 703820, 703821, 703822, 703823, 703825, 703826, 703827, 703828, 703829, 703830, 703832, 703833, 703834, 703835, 703836, 703837, 703838, 703839, 703840, 703841, 703842, 703843, 703844, 703845, 703846, 703847, 703848, 703849, 703850, 703851, 703852, 703853, 703854, 703855, 703856, 703857, 703858, 703859, 703860, 703861, 703862, 703863, 703864, 703865, 703866, 703867, 703868, 703870, 703871, 703872, 703873, 703874, 703876, 703877, 703878, 703879, 703881, 703882, 703883, 703885, 703886, 703887, 703890, 703892, 703893, 703894, 703896, 703898, 703899, 703900, 703901, 703903, 703904, 703905, 703906, 703907, 703908, 703909, 703910, 703911, 703912, 703913, 703914, 703916, 703917, 703918, 703919, 703920, 703921, 703922, 703923, 703924, 703925, 703926, 703927, 703928, 703929, 703930, 703932, 703933, 703934, 703937, 703938, 703939, 703940, 703941, 703942, 703944, 703945, 703946, 703947, 703948, 703949, 703950, 703951, 703952, 703953, 703954, 703955, 703956, 703957, 703958, 703959, 703960, 703961, 703962, 703963, 703964, 703965, 703966, 703967, 703968, 703969, 703970, 703973, 703974, 703976, 703977, 703978, 703980, 703981, 703982, 703983, 703984, 703986, 703987, 703988, 703989, 703990, 703991, 703992, 703993, 703995, 703996, 703997, 703998, 703999, 704000, 704001, 704002, 704005, 704006, 704007, 704008, 704009, 704010, 704013, 704014, 704015, 704018, 704019, 704020, 704021, 704022, 704023, 704024, 704025, 704026, 704027, 704028, 704029, 704033, 704034, 704035, 704036, 704037, 704038, 704039, 704040, 704041, 704042, 704043, 704044, 704045, 704046, 704047, 704048, 704049, 704050, 704051, 704052, 704053, 704054, 704056, 704057, 704058, 704059, 704060, 704061, 704062, 704063, 704065, 704066, 704067, 704068, 704069, 704070, 704071, 704072, 704073, 704074, 704075, 704076, 704077, 704078, 704079, 704081, 704083, 704084, 704085, 704087, 704088, 704089, 704090, 704091, 704092, 704093, 704094, 704095, 704096, 704098, 704100, 704101, 704103, 704104, 704105, 704106, 704107, 704108, 704109, 704110, 704111, 704112, 704113, 704115, 704116, 704117, 704120, 704121, 704123, 704124, 704125, 704128, 704129, 704130, 704131, 704132, 704133, 704134, 704135, 704136, 704137, 704138, 704139, 704140, 704141, 704142, 704144, 704146, 704147, 704148, 704149, 704150, 704151, 704152, 704153, 704154, 704155, 704156, 704157, 704158, 704159, 704160, 704161, 704162, 704163, 704164, 704165, 704166, 704167, 704168, 704169, 704170, 704171, 704172, 704173, 704174, 704176, 704177, 704178, 704179, 704181, 704182, 704183, 704184, 704186, 704187, 704189, 704191, 704192, 704193, 704194, 704195, 704196, 704198, 704199, 704200, 704201, 704203, 704204, 704205, 704207, 704208, 704210, 704214, 704215, 704216, 704217, 704220, 704221, 704222, 704223, 704224, 704225, 704226, 704227, 704228, 704230, 704231, 704232, 704233, 704234, 704236, 704237, 704238, 704239, 704240, 704241, 704242, 704243, 704244, 704245, 704246, 704247, 704248, 704249, 704250, 704251, 704252, 704253, 704256, 704257, 704258, 704259, 704261, 704262, 704263, 704264, 704265, 704267, 704268, 704269, 704270, 704272, 704273, 704276, 704277, 704278, 704279, 704280, 704281, 704282, 704283, 704284, 704285, 704286, 704287, 704288, 704289, 704290, 704291, 704292, 704293, 704294, 704295, 704296, 704297, 704298, 704299, 704300, 704301, 704302, 704303, 704304, 704305, 704307, 704308, 704309, 704310, 704311, 704312, 704313, 704314, 704315, 704316, 704317, 704318, 704319, 704320, 704321, 704322, 704323, 704325, 704326, 704327, 704328, 704329, 704330, 704331, 704332, 704333, 704334, 704335, 704337, 704338, 704339, 704340, 704341, 704342, 704345, 704346, 704347, 704348, 704349, 704351, 704352, 704353, 704354, 704355, 704356, 704357, 704359, 704360, 704361, 704362, 704363, 704364, 704365, 704366, 704367, 704368, 704369, 704370, 704371, 704373, 704374, 704375, 704376, 704377, 704378, 704379, 704380, 704381, 704382, 704383, 704384, 704385, 704386, 704387, 704389, 704390, 704392, 704393, 704394, 704395, 704396, 704397, 704398, 704399, 704400, 704401, 704402, 704403, 704404, 704405, 704406, 704408, 704409, 704411, 704414, 704415, 704417, 704418, 704420, 704421, 704422, 704423, 704424, 704425, 704426, 704427, 704429, 704430, 704431, 704432, 704433, 704434, 704435, 704436, 704437, 704438, 704439, 704440, 704441, 704442, 704443, 704445, 704446, 704447, 704448, 704449, 704450, 704451, 704452, 704453, 704454, 704455, 704457, 704458, 704459, 704460, 704461, 704462, 704463, 704464, 704465, 704466, 704467, 704468, 704469, 704470, 704471, 704472, 704473, 704474, 704475, 704477, 704478, 704479, 704480, 704481, 704482, 704483, 704484, 704485, 704487, 704488, 704489, 704490, 704491, 704492, 704493, 704494, 704495, 704496, 704497, 704498, 704499, 704500, 704501, 704502, 704503, 704504, 704505, 704507, 704509, 704510, 704511, 704512, 704513, 704514, 704515, 704517, 704518, 704519, 704520, 704521, 704522, 704523, 704524, 704525, 704526, 704527, 704528, 704529, 704530, 704531, 704532, 704534, 704535, 704536, 704537, 704538, 704539, 704540, 704541, 704542, 704543, 704544, 704545, 704546, 704547, 704548, 704549, 704550, 704551, 704552, 704553, 704554, 704555, 704557, 704558, 704560, 704561, 704562, 704563, 704564, 704566, 704567, 704569, 704570, 704571, 704572, 704573, 704574, 704576, 704577, 704578, 704579, 704580, 704581, 704582, 704583, 704584, 704585, 704586, 704587, 704588, 704589, 704590, 704591, 704593, 704594, 704596, 704597, 704598, 704599, 704600, 704601, 704602, 704603, 704604, 704605, 704606, 704607, 704608, 704609, 704610, 704611, 704613, 704614, 704615, 704616, 704617, 704618, 704619, 704621, 704622, 704623, 704624, 704625, 704627, 704628, 704629, 704630, 704631, 704632, 704633, 704634, 704635, 704636, 704637, 704640, 704641, 704642, 704643, 704644, 704645, 704646, 704649, 704650, 704652, 704653, 704654, 704658, 704659, 704661, 704662, 704663, 704664, 704665, 704667, 704668, 704670, 704672, 704673, 704674, 704675, 704676, 704677, 704678, 704679, 704680, 704682, 704683, 704684, 704685, 704687, 704688, 704690, 704691, 704692, 704693. 704694, 704695, 704697, 704698, 704699, 704700, 704701, 704702, 704703, 704704, 704705, 704706, 704708, 704710, 704711, 704712, 704713, 704714, 704716, 704718, 704719, 704720, 704721, 704722, 704724, 704725, 704726, 704728, 704729, 704731, 704733, 704735, 704737, 704738, 704739, 704740, 704742, 704743, 704744, 704747, 704748, 704749, 704750, 704751, 704752, 704753, 704754, 704757, 704758, 704761, 704762, 704763, 704764, 704765, 704766, 704767, 704768, 704769, 704770, 704773, 704774, 704776, 704779, 704780, 704781, 704782, 704783, 704784, 704786, 704788, 704791, 704792, 704793, 704794, 704795, 704796, 704797, 704798, 704800, 704801, 704803, 704804, 704807, 704808, 704809, 704810, 704811, 704814, 704815, 704816, 704817, 704818, 704819, 704820, 704821, 704822, 704823, 704824, 704825, 704827, 704829, 704830, 704831, 704832, 704833, 704834, 704835, 704837, 704838, 704839, 704840, 704842, 704843, 704844, 704845, 704847, 704848, 704849, 704850, 704851, 704852, 704853, 704854, 704857, 704858, 704859, 704860, 704861, 704862, 704864, 704865, 704866, 704867, 704868, 704869, 704870, 704871, 704872, 704873, 704874, 704875, 704876, 704877, 704878, 704879, 704880, 704881, 704882, 704883, 704884, 704885, 704886, 704887, 704888, 704889, 704890, 704891, 704892, 704894, 704895,
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Baxter Healthcare Corporation
Reason for Recall:
Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a software package that will add additional features to the Spectrum Infusion Pump. The Master Drug Library will permit electronic communications with the Spectrum pump and other external peripheral devices. The intended use of the Spectrum pump includes common drug error prevention, through the stand alone settings features of the pump. This includes drug parameter limits and associated drug name identification. With the Master Drug Library, the intended use is to reduce user errors associated with drug selection, drug dose rates, drug dose concentrations, and patient identification associated with the prescribed drug.

Product Codes/Lot Numbers:

Serial numbers ranging from 700000 through 794213: 700000, 700001, 700002, 700003, 700004, 700005, 700006, 700007, 700010, 700012, 700013, 700014, 700017, 700018, 700020, 700021, 700022, 700023, 700024, 700029, 700030, 700031, 700032, 700033, 700034, 700035, 700037, 700038, 700039, 700040, 700041, 700042, 700043, 700045, 700046, 700047, 700048, 700049, 700050, 700051, 700052, 700053, 700054, 700055, 700057, 700058, 700059, 700060, 700061, 700062, 700063, 700064, 700065, 700066, 700067, 700069, 700071, 700073, 700075, 700076, 700077, 700080, 700081, 700082, 700085, 700086, 700088, 700089, 700090, 700093, 700094, 700095, 700097, 700098, 700100, 700101, 700103, 700104, 700106, 700108, 700109, 700110, 700111, 700112, 700114, 700117, 700118, 700119, 700120, 700121, 700123, 700126, 700127, 700129, 700130, 700132, 700133, 700134, 700135, 700137, 700138, 700139, 700141, 700142, 700143, 700144, 700145, 700147, 700149, 700151, 700152, 700153, 700156, 700157, 700158, 700159, 700161, 700162, 700163, 700165, 700166, 700168, 700169, 700170, 700172, 700173, 700174, 700175, 700176, 700177, 700178, 700179, 700180, 700183, 700184, 700187, 700189, 700193, 700194, 700195, 700197, 700201, 700202, 700203, 700204, 700205, 700206, 700207, 700208, 700209, 700210, 700211, 700212, 700213, 700214, 700215, 700216, 700217, 700218, 700219, 700220, 700222, 700223, 700224, 700225, 700226, 700229, 700231, 700232, 700233, 700234, 700235, 700236, 700237, 700238, 700239, 700240, 700241, 700242, 700243, 700244, 700245, 700246, 700248, 700249, 700250, 700251, 700252, 700253, 700255, 700256, 700257, 700258, 700259, 700260, 700263, 700265, 700267, 700268, 700269, 700271, 700272, 700273, 700274, 700275, 700277, 700278, 700280, 700282, 700283, 700284, 700285, 700286, 700287, 700288, 700289, 700292, 700293, 700294, 700295, 700297, 700298, 700299, 700301, 700302, 700303, 700305, 700306, 700307, 700308, 700309, 700310, 700311, 700312, 700314, 700316, 700317, 700318, 700319, 700320, 700321, 700322, 700323, 700324, 700326, 700327, 700328, 700329, 700330, 700331, 700332, 700333, 700334, 700335, 700338, 700339, 700340, 700341, 700343, 700344, 700345, 700346, 700348, 700349, 700350, 700351, 700352, 700353, 700354, 700355, 700357, 700358, 700359, 700360, 700361, 700362, 700363, 700364, 700365, 700366, 700367, 700368, 700369, 700370, 700371, 700372, 700373, 700374, 700375, 700376, 700377, 700378, 700379, 700380, 700381, 700382, 700383, 700384, 700385, 700386, 700387, 700388, 700389, 700390, 700392, 700395, 700396, 700398, 700399, 700400, 700401, 700402, 700403, 700404, 700405, 700406, 700407, 700408, 700409, 700410, 700411, 700412, 700413, 700414, 700415, 700416, 700417, 700419, 700420, 700421, 700422, 700423, 700424, 700425, 700426, 700427, 700428, 700429, 700433, 700434, 700436, 700437, 700438, 700439, 700440, 700441, 700442, 700443, 700444, 700446, 700447, 700448, 700449, 700450, 700451, 700452, 700453, 700454, 700455, 700456, 700457, 700458, 700459, 700460, 700461, 700462, 700463, 700464, 700466, 700468, 700470, 700471, 700472, 700473, 700474, 700475, 700476, 700478, 700479, 700480, 700481, 700483, 700484, 700485, 700489, 700490, 700491, 700494, 700495, 700496, 700497, 700502, 700503, 700504, 700505, 700506, 700507, 700508, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700519, 700520, 700524, 700525, 700526, 700527, 700528, 700530, 700531, 700532, 700533, 700534, 700536, 700537, 700538, 700539, 700540, 700542, 700548, 700552, 700553, 700555, 700556, 700557, 700558, 700560, 700561, 700562, 700563, 700564, 700565, 700566, 700567, 700568, 700569, 700570, 700571, 700572, 700573, 700574, 700575, 700576, 700577, 700578, 700579, 700580, 700581, 700582, 700583, 700584, 700585, 700586, 700587, 700588, 700589, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700598, 700599, 700600, 700601, 700602, 700603, 700604, 700605, 700606, 700607, 700608, 700611, 700612, 700613, 700614, 700616, 700617, 700618, 700619, 700620, 700621, 700622, 700623, 700624, 700625, 700626, 700627, 700629, 700631, 700632, 700633, 700634, 700635, 700636, 700638, 700639, 700641, 700642, 700643, 700646, 700647, 700648, 700649, 700650, 700652, 700654, 700655, 700656, 700657, 700659, 700661, 700662, 700663, 700664, 700665, 700666, 700667, 700668, 700669, 700670, 700671, 700672, 700673, 700674, 700676, 700677, 700681, 700682, 700683, 700684, 700685, 700687, 700689, 700690, 700691, 700693, 700694, 700695, 700696, 700697, 700698, 700699, 700700, 700701, 700702, 700703, 700704, 700707, 700708, 700709, 700710, 700711, 700712, 700713, 700715, 700716, 700717, 700718, 700721, 700722, 700723, 700724, 700725, 700726, 700727, 700729, 700731, 700732, 700733, 700734, 700735, 700736, 700737, 700738, 700741, 700742, 700743, 700744, 700747, 700748, 700749, 700750, 700751, 700752, 700753, 700754, 700755, 700756, 700758, 700759, 700761, 700762, 700763, 700765, 700766, 700767, 700768, 700769, 700770, 700771, 700773, 700774, 700775, 700776, 700777, 700778, 700780, 700781, 700782, 700783, 700784, 700785, 700790, 700791, 700792, 700794, 700797, 700798, 700799, 700800, 700803, 700805, 700807, 700808, 700809, 700810, 700811, 700813, 700815, 700817, 700818, 700819, 700821, 700822, 700823, 700826, 700828, 700829, 700830, 700831, 700832, 700833, 700834, 700835, 700836, 700838, 700839, 700840, 700844, 700845, 700846, 700849, 700850, 700851, 700852, 700854, 700856, 700857, 700858, 700859, 700860, 700861, 700862, 700863, 700864, 700865, 700866, 700867, 700870, 700871, 700873, 700874, 700875, 700877, 700878, 700880, 700881, 700882, 700883, 700884, 700885, 700887, 700888, 700889, 700890, 700891, 700892, 700893, 700894, 700896, 700897, 700898, 700900, 700902, 700904, 700905, 700908, 700909, 700910, 700911, 700913, 700914, 700919, 700920, 700921, 700925, 700926, 700927, 700928, 700929, 700930, 700931, 700932, 700934, 700935, 700936, 700937, 700938, 700939, 700940, 700941, 700942, 700943, 700944, 700945, 700946, 700947, 700948, 700949, 700950, 700951, 700952, 700953, 700954, 700955, 700956, 700958, 700959, 700960, 700961, 700962, 700963, 700964, 700966, 700967, 700968, 700969, 700970, 700971, 700972, 700973, 700974, 700975, 700976, 700977, 700978, 700979, 700980, 700981, 700982, 700983, 700984, 700985, 700986, 700987, 700988, 700989, 700990, 700991, 700992, 700993, 700994, 700995, 700997, 700998, 700999, 701000, 701002, 701003, 701004, 701005, 701007, 701008, 701009, 701010, 701011, 701012, 701013, 701014, 701015, 701016, 701017, 701018, 701019, 701020, 701022, 701025, 701027, 701028, 701029, 701030, 701031, 701032, 701033, 701034, 701036, 701037, 701039, 701040, 701041, 701042, 701043, 701045, 701046, 701047, 701048, 701049, 701050, 701051, 701052, 701053, 701054, 701056, 701057, 701058, 701059, 701060, 701061, 701063, 701064, 701065, 701066, 701067, 701068, 701069, 701070, 701071, 701072, 701073, 701074, 701075, 701076, 701077, 701078, 701079, 701080, 701081, 701082, 701083, 701084, 701085, 701086, 701087, 701088, 701089, 701092, 701093, 701094, 701095, 701096, 701097, 701098, 701099, 701100, 701101, 701102, 701105, 701106, 701107, 701108, 701109, 701110, 701111, 701112, 701113, 701114, 701115, 701116, 701117, 701118, 701119, 701120, 701121, 701122, 701123, 701124, 701125, 701126, 701127, 701128, 701129, 701130, 701131, 701132, 701133, 701134, 701135, 701136, 701137, 701138, 701139, 701140, 701141, 701142, 701143, 701144, 701145, 701146, 701147, 701148, 701149, 701150, 701151, 701152, 701153, 701154, 701155, 701156, 701157, 701158, 701159, 701160, 701161, 701162, 701163, 701164, 701165, 701166, 701167, 701168, 701169, 701170, 701171, 701173, 701174, 701175, 701176, 701177, 701178, 701179, 701180, 701181, 701182, 701183, 701184, 701185, 701186, 701187, 701188, 701189, 701190, 701191, 701192, 701193, 701194, 701195, 701196, 701197, 701198, 701199, 701201, 701202, 701203, 701204, 701205, 701206, 701207, 701208, 701209, 701210, 701211, 701212, 701214, 701215, 701216, 701217, 701218, 701219, 701220, 701221, 701222, 701223, 701224, 701225, 701226, 701227, 701228, 701229, 701230, 701231, 701232, 701233, 701234, 701235, 701236, 701237, 701238, 701239, 701240, 701241, 701242, 701243, 701244, 701245, 701246, 701247, 701248, 701249, 701250, 701251, 701252, 701254, 701255, 701256, 701257, 701259, 701260, 701261, 701262, 701263, 701264, 701265, 701266, 701267, 701268, 701269, 701270, 701271, 701272, 701273, 701274, 701276, 701277, 701278, 701279, 701280, 701281, 701283, 701284, 701285, 701286, 701287, 701288, 701289, 701290, 701291, 701292, 701293, 701294, 701295, 701296, 701297, 701298, 701299, 701301, 701302, 701304, 701305, 701306, 701307, 701308, 701309, 701310, 701311, 701313, 701314, 701315, 701317, 701318, 701319, 701320, 701321, 701323, 701324, 701325, 701326, 701327, 701329, 701330, 701331, 701332, 701333, 701334, 701336, 701337, 701338, 701339, 701340, 701341, 701342, 701344, 701345, 701346, 701347, 701348, 701349, 701350, 701351, 701352, 701353, 701354, 701355, 701356, 701357, 701358, 701359, 701360, 701361, 701362, 701363, 701364, 701366, 701367, 701368, 701369, 701370, 701371, 701372, 701373, 701374, 701375, 701376, 701378, 701379, 701380, 701381, 701382, 701383, 701385, 701386, 701387, 701388, 701389, 701390, 701391, 701392, 701393, 701394, 701395, 701396, 701397, 701399, 701400, 701401, 701403, 701404, 701405, 701406, 701407, 701408, 701409, 701410, 701411, 701412, 701413, 701414, 701415, 701417, 701418, 701419, 701420, 701421, 701422, 701423, 701424, 701425, 701426, 701427, 701428, 701429, 701430, 701431, 701432, 701433, 701434, 701435, 701436, 701437, 701438, 701439, 701440, 701441, 701443, 701445, 701446, 701447, 701448, 701449, 701450, 701457, 701462, 701465, 701467, 701472, 701475, 701482, 701483, 701484, 701487, 701488, 701492, 701494, 701495, 701501, 701505, 701507, 701508, 701509, 701511, 701512, 701514, 701515, 701516, 701518, 701520, 701523, 701524, 701525, 701529, 701530, 701532, 701533, 701534, 701535, 701536, 701537, 701538, 701539, 701540, 701541, 701542, 701543, 701544, 701545, 701546, 701547, 701548, 701549, 701550, 701551, 701552, 701554, 701555, 701556, 701557, 701558, 701559, 701563, 701564, 701565, 701566, 701567, 701568, 701569, 701570, 701571, 701573, 701574, 701575, 701576, 701577, 701578, 701579, 701580, 701582, 701583, 701585, 701586, 701588, 701589, 701590, 701591, 701592, 701593, 701594, 701595, 701596, 701597, 701598, 701600, 701601, 701602, 701603, 701604, 701605, 701606, 701607, 701609, 701610, 701611, 701612, 701613, 701614, 701615, 701617, 701618, 701620, 701621, 701622, 701624, 701625, 701626, 701627, 701628, 701629, 701631, 701632, 701633, 701634, 701635, 701636, 701637, 701638, 701639, 701640, 701641, 701642, 701644, 701645, 701646, 701647, 701648, 701649, 701650, 701651, 701652, 701653, 701654, 701655, 701656, 701657, 701658, 701659, 701660, 701661, 701662, 701664, 701665, 701666, 701667, 701668, 701669, 701670, 701671, 701672, 701673, 701674, 701676, 701677, 701678, 701679, 701680, 701681, 701682, 701683, 701684, 701685, 701686, 701687, 701688, 701689, 701690, 701691, 701692, 701693, 701694, 701696, 701697, 701698, 701699, 701700, 701701, 701703, 701704, 701705, 701706, 701707, 701708, 701709, 701711, 701712, 701713, 701714, 701715, 701716, 701717, 701718, 701719, 701720, 701721, 701722, 701724, 701725, 701726, 701727, 701729, 701732, 701733, 701734, 701735, 701737, 701739, 701740, 701741, 701742, 701743, 701744, 701745, 701746, 701747, 701748, 701749, 701750, 701751, 701754, 701755, 701756, 701758, 701759, 701761, 701762, 701763, 701764, 701766, 701767, 701768, 701769, 701770, 701771, 701772, 701773, 701774, 701775, 701776, 701778, 701780, 701781, 701782, 701784, 701787, 701788, 701789, 701790, 701792, 701794, 701796, 701797, 701798, 701799, 701800, 701801, 701802, 701803, 701804, 701805, 701806, 701808, 701809, 701813, 701814, 701818, 701819, 701820, 701821, 701822, 701823, 701825, 701827, 701829, 701830, 701831, 701832, 701833, 701834, 701835, 701836, 701837, 701838, 701839, 701840, 701843, 701844, 701845, 701846, 701848, 701850, 701851, 701852, 701853, 701854, 701856, 701857, 701858, 701859, 701860, 701861, 701863, 701865, 701866, 701867, 701868, 701869, 701870, 701871, 701872, 701873, 701874, 701875, 701876, 701878, 701879, 701880, 701881, 701882, 701883, 701885, 701886, 701888, 701889, 701890, 701891, 701892, 701896, 701897, 701898, 701899, 701900, 701901, 701902, 701903, 701904, 701905, 701906, 701907, 701908, 701912, 701913, 701916, 701917, 701919, 701920, 701921, 701923, 701925, 701927, 701928, 701929, 701931, 701932, 701933, 701935, 701936, 701937, 701938, 701939, 701940, 701941, 701942, 701943, 701944, 701945, 701947, 701949, 701950, 701951, 701953, 701954, 701955, 701956, 701957, 701958, 701959, 701960, 701961, 701963, 701965, 701967, 701969, 701970, 701971, 701972, 701973, 701975, 701976, 701978, 701979, 701980, 701982, 701983, 701984, 701986, 701987, 701988, 701990, 701991, 701993, 701996, 701997, 701998, 702000, 702001, 702002, 702003, 702005, 702006, 702007, 702008, 702011, 702012, 702013, 702014, 702015, 702016, 702018, 702020, 702021, 702022, 702023, 702024, 702025, 702027, 702028, 702029, 702030, 702031, 702032, 702033, 702034, 702035, 702036, 702037, 702038, 702039, 702040, 702041, 702044, 702046, 702047, 702049, 702050, 702051, 702052, 702053, 702054, 702055, 702056, 702057, 702058, 702059, 702060, 702064, 702065, 702066, 702068, 702069, 702070, 702071, 702072, 702073, 702074, 702075, 702076, 702078, 702080, 702083, 702085, 702087, 702088, 702089, 702090, 702092, 702095, 702096, 702098, 702099, 702100, 702106, 702107, 702108, 702109, 702111, 702112, 702113, 702115, 702117, 702118, 702119, 702122, 702123, 702124, 702125, 702128, 702130, 702131, 702132, 702134, 702135, 702136, 702138, 702140, 702141, 702143, 702144, 702146, 702148, 702149, 702152, 702153, 702154, 702155, 702156, 702157, 702158, 702160, 702161, 702162, 702163, 702164, 702165, 702166, 702167, 702168, 702169, 702171, 702172, 702173, 702174, 702175, 702176, 702177, 702178, 702180, 702181, 702182, 702184, 702186, 702187, 702188, 702189, 702190, 702191, 702192, 702194, 702195, 702196, 702197, 702198, 702199, 702200, 702202, 702203, 702204, 702206, 702207, 702210, 702211, 702212, 702214, 702215, 702216, 702218, 702219, 702220, 702221, 702222, 702223, 702224, 702225, 702226, 702227, 702228, 702229, 702230, 702233, 702234, 702235, 702236, 702237, 702238, 702240, 702241, 702242, 702243, 702244, 702245, 702246, 702247, 702248, 702249, 702250, 702251, 702252, 702253, 702254, 702255, 702256, 702257, 702258, 702259, 702260, 702261, 702262, 702263, 702264, 702267, 702268, 702269, 702270, 702271, 702273, 702274, 702275, 702276, 702277, 702278, 702279, 702280, 702281, 702284, 702285, 702286, 702287, 702289, 702290, 702291, 702292, 702293, 702294, 702295, 702296, 702297, 702298, 702299, 702300, 702301, 702302, 702303, 702304, 702305, 702306, 702307, 702308, 702309, 702310, 702311, 702312, 702313, 702314, 702315, 702316, 702317, 702318, 702319, 702320, 702321, 702322, 702323, 702324, 702325, 702326, 702328, 702329, 702330, 702331, 702332, 702333, 702334, 702335, 702336, 702337, 702338, 702339, 702340, 702341, 702343, 702344, 702345, 702346, 702347, 702348, 702349, 702350, 702351, 702352, 702353, 702354, 702355, 702357, 702358, 702360, 702361, 702362, 702364, 702365, 702366, 702368, 702369, 702370, 702371, 702372, 702373, 702374, 702375, 702376, 702377, 702378, 702379, 702380, 702381, 702382, 702383, 702384, 702386, 702387, 702388, 702390, 702391, 702392, 702393, 702394, 702395, 702396, 702397, 702398, 702400, 702401, 702402, 702403, 702405, 702406, 702407, 702408, 702410, 702411, 702412, 702413, 702414, 702415, 702416, 702417, 702418, 702419, 702420, 702421, 702422, 702423, 702424, 702425, 702426, 702427, 702428, 702429, 702430, 702431, 702433, 702434, 702436, 702438, 702439, 702440, 702441, 702443, 702444, 702445, 702446, 702447, 702448, 702449, 702450, 702451, 702452, 702453, 702454, 702455, 702456, 702457, 702458, 702459, 702460, 702461, 702462, 702464, 702465, 702466, 702467, 702468, 702469, 702470, 702471, 702472, 702473, 702474, 702475, 702476, 702477, 702478, 702479, 702480, 702481, 702482, 702483, 702484, 702485, 702486, 702487, 702488, 702490, 702491, 702492, 702493, 702494, 702495, 702496, 702497, 702498, 702499, 702500, 702501, 702502, 702503, 702504, 702505, 702506, 702507, 702508, 702509, 702510, 702511, 702513, 702514, 702515, 702516, 702517, 702520, 702521, 702522, 702523, 702524, 702525, 702526, 702527, 702528, 702529, 702530, 702531, 702532, 702533, 702534, 702535, 702537, 702538, 702540, 702542, 702544, 702545, 702547, 702548, 702550, 702551, 702552, 702553, 702554, 702555, 702556, 702557, 702558, 702559, 702561, 702562, 702563, 702564, 702565, 702566, 702567, 702568, 702570, 702572, 702573, 702574, 702575, 702577, 702579, 702580, 702582, 702583, 702584, 702585, 702586, 702587, 702588, 702589, 702590, 702591, 702592, 702593, 702594, 702595, 702597, 702599, 702601, 702602, 702603, 702604, 702605, 702606, 702607, 702608, 702609, 702610, 702611, 702612, 702613, 702614, 702615, 702616, 702617, 702618, 702619, 702621, 702625, 702626, 702628, 702629, 702630, 702631, 702632, 702633, 702634, 702635, 702636, 702637, 702638, 702639, 702640, 702641, 702642, 702644, 702647, 702648, 702650, 702651, 702652, 702653, 702654, 702656, 702657, 702659, 702660, 702662, 702663, 702666, 702667, 702668, 702670, 702671, 702673, 702674, 702678, 702681, 702682, 702683, 702684, 702685, 702686, 702687, 702689, 702690, 702694, 702695, 702696, 702698, 702699, 702700, 702705, 702706, 702707, 702708, 702709, 702710, 702711, 702712, 702714, 702715, 702716, 702717, 702718, 702719, 702720, 702721, 702722, 702724, 702725, 702726, 702727, 702729, 702730, 702731, 702732, 702733, 702734, 702735, 702736, 702737, 702738, 702740, 702741, 702742, 702743, 702745, 702746, 702747, 702748, 702749, 702750, 702751, 702752, 702753, 702754, 702755, 702757, 702758, 702759, 702760, 702761, 702762, 702763, 702764, 702765, 702767, 702768, 702769, 702770, 702771, 702772, 702773, 702774, 702775, 702776, 702777, 702778, 702779, 702780, 702781, 702782, 702784, 702787, 702788, 702790, 702792, 702793, 702794, 702795, 702796, 702797, 702798, 702799, 702800, 702802, 702804, 702805, 702806, 702807, 702808, 702809, 702810, 702811, 702812, 702813, 702814, 702815, 702816, 702817, 702818, 702819, 702820, 702821, 702823, 702824, 702825, 702826, 702827, 702828, 702829, 702830, 702831, 702832, 702834, 702835, 702836, 702837, 702838, 702839, 702840, 702841, 702843, 702844, 702845, 702846, 702847, 702848, 702849, 702850, 702851, 702852, 702853, 702854, 702855, 702856, 702857, 702858, 702859, 702860, 702861, 702862, 702863, 702865, 702866, 702867, 702868, 702869, 702870, 702872, 702873, 702874, 702875, 702877, 702878, 702879, 702880, 702881, 702882, 702883, 702884, 702885, 702886, 702887, 702888, 702889, 702890, 702892, 702895, 702896, 702898, 702899, 702900, 702901, 702904, 702905, 702906, 702907, 702908, 702909, 702910, 702911, 702913, 702915, 702916, 702917, 702918, 702919, 702920, 702921, 702922, 702923, 702924, 702925, 702926, 702927, 702929, 702930, 702932, 702934, 702935, 702936, 702937, 702938, 702939, 702940, 702941, 702942, 702943, 702944, 702945, 702946, 702947, 702948, 702949, 702951, 702952, 702953, 702954, 702956, 702958, 702959, 702960, 702961, 702962, 702963, 702965, 702967, 702968, 702969, 702970, 702971, 702972, 702973, 702974, 702975, 702976, 702977, 702978, 702979, 702980, 702981, 702982, 702983, 702984, 702987, 702988, 702989, 702990, 702991, 702993, 702994, 702996, 702997, 702998, 703000, 703001, 703002, 703003, 703004, 703005, 703006, 703007, 703008, 703009, 703010, 703011, 703012, 703013, 703014, 703015, 703016, 703017, 703018, 703019, 703020, 703021, 703022, 703023, 703024, 703025, 703027, 703029, 703030, 703031, 703033, 703034, 703035, 703036, 703037, 703038, 703039, 703040, 703041, 703042, 703044, 703045, 703046, 703047, 703048, 703049, 703050, 703051, 703052, 703053, 703054, 703055, 703056, 703057, 703058, 703059, 703060, 703061, 703062, 703063, 703065, 703066, 703067, 703068, 703069, 703070, 703071, 703073, 703074, 703075, 703076, 703077, 703078, 703079, 703080, 703081, 703082, 703083, 703084, 703085, 703086, 703087, 703088, 703089, 703090, 703091, 703092, 703093, 703094, 703095, 703096, 703098, 703099, 703100, 703101, 703102, 703103, 703104, 703105, 703106, 703107, 703108, 703109, 703110, 703111, 703113, 703115, 703116, 703117, 703119, 703120, 703121, 703122, 703123, 703124, 703125, 703126, 703129, 703130, 703131, 703132, 703133, 703134, 703135, 703137, 703138, 703139, 703140, 703141, 703142, 703143, 703144, 703145, 703147, 703149, 703150, 703152, 703153, 703155, 703156, 703157, 703158, 703159, 703161, 703162, 703163, 703164, 703165, 703167, 703168, 703169, 703170, 703171, 703172, 703173, 703174, 703175, 703176, 703177, 703178, 703179, 703180, 703181, 703182, 703183, 703184, 703185, 703186, 703188, 703189, 703190, 703191, 703192, 703193, 703195, 703196, 703197, 703199, 703200, 703201, 703202, 703203, 703204, 703205, 703206, 703207, 703208, 703210, 703211, 703212, 703213, 703214, 703217, 703219, 703220, 703221, 703222, 703223, 703224, 703225, 703226, 703228, 703229, 703230, 703231, 703232, 703233, 703234, 703235, 703236, 703238, 703239, 703240, 703241, 703242, 703243, 703245, 703246, 703247, 703248, 703251, 703252, 703253, 703254, 703255, 703256, 703257, 703258, 703259, 703260, 703262, 703263, 703264, 703265, 703266, 703268, 703269, 703270, 703271, 703272, 703273, 703275, 703276, 703277, 703278, 703279, 703280, 703281, 703282, 703283, 703284, 703285, 703287, 703289, 703290, 703291, 703292, 703293, 703294, 703295, 703296, 703297, 703298, 703299, 703300, 703301, 703304, 703305, 703306, 703307, 703309, 703310, 703311, 703312, 703313, 703314, 703315, 703318, 703319, 703320, 703321, 703323, 703325, 703327, 703328, 703330, 703331, 703332, 703334, 703335, 703336, 703337, 703338, 703339, 703340, 703341, 703342, 703343, 703344, 703345, 703346, 703347, 703348, 703349, 703351, 703352, 703353, 703354, 703356, 703357, 703358, 703359, 703360, 703361, 703362, 703363, 703364, 703365, 703366, 703367, 703368, 703370, 703371, 703372, 703373, 703375, 703376, 703377, 703378, 703379, 703380, 703381, 703382, 703383, 703384, 703385, 703386, 703387, 703389, 703390, 703391, 703393, 703394, 703395, 703396, 703397, 703398, 703399, 703400, 703401, 703402, 703403, 703404, 703405, 703406, 703408, 703410, 703412, 703414, 703415, 703417, 703418, 703419, 703420, 703421, 703422, 703425, 703426, 703427, 703428, 703429, 703430, 703432, 703433, 703434, 703435, 703437, 703438, 703439, 703440, 703441, 703442, 703443, 703444, 703445, 703447, 703448, 703450, 703451, 703453, 703454, 703455, 703456, 703457, 703458, 703459, 703460, 703461, 703462, 703463, 703466, 703467, 703468, 703469, 703470, 703471, 703472, 703473, 703476, 703477, 703480, 703481, 703482, 703484, 703485, 703486, 703487, 703489, 703490, 703491, 703492, 703493, 703494, 703495, 703496, 703498, 703499, 703501, 703502, 703503, 703504, 703505, 703506, 703507, 703508, 703509, 703510, 703511, 703512, 703513, 703514, 703515, 703516, 703517, 703519, 703520, 703522, 703523, 703524, 703525, 703526, 703527, 703529, 703530, 703531, 703533, 703534, 703535, 703536, 703537, 703538, 703539, 703540, 703541, 703542, 703544, 703545, 703546, 703547, 703548, 703549, 703550, 703551, 703552, 703553, 703554, 703555, 703556, 703557, 703558, 703559, 703560, 703561, 703562, 703563, 703564, 703565, 703566, 703567, 703568, 703569, 703570, 703571, 703572, 703574, 703575, 703576, 703578, 703580, 703581, 703582, 703583, 703584, 703586, 703587, 703588, 703589, 703590, 703591, 703592, 703593, 703594, 703595, 703596, 703597, 703599, 703600, 703601, 703602, 703603, 703604, 703605, 703606, 703607, 703608, 703609, 703610, 703611, 703612, 703614, 703615, 703616, 703617, 703618, 703619, 703621, 703622, 703623, 703624, 703625, 703626, 703627, 703628, 703629, 703631, 703632, 703633, 703634, 703635, 703636, 703637, 703639, 703640, 703642, 703643, 703644, 703645, 703646, 703647, 703648, 703649, 703650, 703651, 703652, 703654, 703655, 703656, 703657, 703658, 703660, 703661, 703662, 703663, 703664, 703665, 703666, 703667, 703668, 703669, 703670, 703671, 703672, 703673, 703674, 703675, 703676, 703677, 703678, 703679, 703680, 703681, 703682, 703683, 703684, 703686, 703687, 703689, 703690, 703691, 703692, 703693, 703694, 703695, 703696, 703697, 703698, 703700, 703701, 703702, 703703, 703704, 703705, 703706, 703707, 703708, 703709, 703710, 703711, 703712, 703713, 703714, 703715, 703716, 703718, 703719, 703721, 703722, 703723, 703724, 703726, 703727, 703729, 703731, 703732, 703733, 703734, 703736, 703738, 703739, 703740, 703741, 703743, 703744, 703746, 703747, 703748, 703749, 703750, 703751, 703752, 703753, 703754, 703755, 703756, 703759, 703760, 703761, 703762, 703763, 703764, 703765, 703767, 703768, 703769, 703770, 703771, 703772, 703774, 703775, 703776, 703777, 703778, 703779, 703780, 703781, 703783, 703784, 703785, 703786, 703787, 703788, 703789, 703790, 703791, 703792, 703793, 703794, 703795, 703796, 703797, 703798, 703799, 703800, 703801, 703802, 703803, 703804, 703805, 703806, 703808, 703809, 703810, 703811, 703812, 703813, 703814, 703815, 703816, 703817, 703818, 703819, 703820, 703821, 703822, 703823, 703825, 703826, 703827, 703828, 703829, 703830, 703832, 703833, 703834, 703835, 703836, 703837, 703838, 703839, 703840, 703841, 703842, 703843, 703844, 703845, 703846, 703847, 703848, 703849, 703850, 703851, 703852, 703853, 703854, 703855, 703856, 703857, 703858, 703859, 703860, 703861, 703862, 703863, 703864, 703865, 703866, 703867, 703868, 703870, 703871, 703872, 703873, 703874, 703876, 703877, 703878, 703879, 703881, 703882, 703883, 703885, 703886, 703887, 703890, 703892, 703893, 703894, 703896, 703898, 703899, 703900, 703901, 703903, 703904, 703905, 703906, 703907, 703908, 703909, 703910, 703911, 703912, 703913, 703914, 703916, 703917, 703918, 703919, 703920, 703921, 703922, 703923, 703924, 703925, 703926, 703927, 703928, 703929, 703930, 703932, 703933, 703934, 703937, 703938, 703939, 703940, 703941, 703942, 703944, 703945, 703946, 703947, 703948, 703949, 703950, 703951, 703952, 703953, 703954, 703955, 703956, 703957, 703958, 703959, 703960, 703961, 703962, 703963, 703964, 703965, 703966, 703967, 703968, 703969, 703970, 703973, 703974, 703976, 703977, 703978, 703980, 703981, 703982, 703983, 703984, 703986, 703987, 703988, 703989, 703990, 703991, 703992, 703993, 703995, 703996, 703997, 703998, 703999, 704000, 704001, 704002, 704005, 704006, 704007, 704008, 704009, 704010, 704013, 704014, 704015, 704018, 704019, 704020, 704021, 704022, 704023, 704024, 704025, 704026, 704027, 704028, 704029, 704033, 704034, 704035, 704036, 704037, 704038, 704039, 704040, 704041, 704042, 704043, 704044, 704045, 704046, 704047, 704048, 704049, 704050, 704051, 704052, 704053, 704054, 704056, 704057, 704058, 704059, 704060, 704061, 704062, 704063, 704065, 704066, 704067, 704068, 704069, 704070, 704071, 704072, 704073, 704074, 704075, 704076, 704077, 704078, 704079, 704081, 704083, 704084, 704085, 704087, 704088, 704089, 704090, 704091, 704092, 704093, 704094, 704095, 704096, 704098, 704100, 704101, 704103, 704104, 704105, 704106, 704107, 704108, 704109, 704110, 704111, 704112, 704113, 704115, 704116, 704117, 704120, 704121, 704123, 704124, 704125, 704128, 704129, 704130, 704131, 704132, 704133, 704134, 704135, 704136, 704137, 704138, 704139, 704140, 704141, 704142, 704144, 704146, 704147, 704148, 704149, 704150, 704151, 704152, 704153, 704154, 704155, 704156, 704157, 704158, 704159, 704160, 704161, 704162, 704163, 704164, 704165, 704166, 704167, 704168, 704169, 704170, 704171, 704172, 704173, 704174, 704176, 704177, 704178, 704179, 704181, 704182, 704183, 704184, 704186, 704187, 704189, 704191, 704192, 704193, 704194, 704195, 704196, 704198, 704199, 704200, 704201, 704203, 704204, 704205, 704207, 704208, 704210, 704214, 704215, 704216, 704217, 704220, 704221, 704222, 704223, 704224, 704225, 704226, 704227, 704228, 704230, 704231, 704232, 704233, 704234, 704236, 704237, 704238, 704239, 704240, 704241, 704242, 704243, 704244, 704245, 704246, 704247, 704248, 704249, 704250, 704251, 704252, 704253, 704256, 704257, 704258, 704259, 704261, 704262, 704263, 704264, 704265, 704267, 704268, 704269, 704270, 704272, 704273, 704276, 704277, 704278, 704279, 704280, 704281, 704282, 704283, 704284, 704285, 704286, 704287, 704288, 704289, 704290, 704291, 704292, 704293, 704294, 704295, 704296, 704297, 704298, 704299, 704300, 704301, 704302, 704303, 704304, 704305, 704307, 704308, 704309, 704310, 704311, 704312, 704313, 704314, 704315, 704316, 704317, 704318, 704319, 704320, 704321, 704322, 704323, 704325, 704326, 704327, 704328, 704329, 704330, 704331, 704332, 704333, 704334, 704335, 704337, 704338, 704339, 704340, 704341, 704342, 704345, 704346, 704347, 704348, 704349, 704351, 704352, 704353, 704354, 704355, 704356, 704357, 704359, 704360, 704361, 704362, 704363, 704364, 704365, 704366, 704367, 704368, 704369, 704370, 704371, 704373, 704374, 704375, 704376, 704377, 704378, 704379, 704380, 704381, 704382, 704383, 704384, 704385, 704386, 704387, 704389, 704390, 704392, 704393, 704394, 704395, 704396, 704397, 704398, 704399, 704400, 704401, 704402, 704403, 704404, 704405, 704406, 704408, 704409, 704411, 704414, 704415, 704417, 704418, 704420, 704421, 704422, 704423, 704424, 704425, 704426, 704427, 704429, 704430, 704431, 704432, 704433, 704434, 704435, 704436, 704437, 704438, 704439, 704440, 704441, 704442, 704443, 704445, 704446, 704447, 704448, 704449, 704450, 704451, 704452, 704453, 704454, 704455, 704457, 704458, 704459, 704460, 704461, 704462, 704463, 704464, 704465, 704466, 704467, 704468, 704469, 704470, 704471, 704472, 704473, 704474, 704475, 704477, 704478, 704479, 704480, 704481, 704482, 704483, 704484, 704485, 704487, 704488, 704489, 704490, 704491, 704492, 704493, 704494, 704495, 704496, 704497, 704498, 704499, 704500, 704501, 704502, 704503, 704504, 704505, 704507, 704509, 704510, 704511, 704512, 704513, 704514, 704515, 704517, 704518, 704519, 704520, 704521, 704522, 704523, 704524, 704525, 704526, 704527, 704528, 704529, 704530, 704531, 704532, 704534, 704535, 704536, 704537, 704538, 704539, 704540, 704541, 704542, 704543, 704544, 704545, 704546, 704547, 704548, 704549, 704550, 704551, 704552, 704553, 704554, 704555, 704557, 704558, 704560, 704561, 704562, 704563, 704564, 704566, 704567, 704569, 704570, 704571, 704572, 704573, 704574, 704576, 704577, 704578, 704579, 704580, 704581, 704582, 704583, 704584, 704585, 704586, 704587, 704588, 704589, 704590, 704591, 704593, 704594, 704596, 704597, 704598, 704599, 704600, 704601, 704602, 704603, 704604, 704605, 704606, 704607, 704608, 704609, 704610, 704611, 704613, 704614, 704615, 704616, 704617, 704618, 704619, 704621, 704622, 704623, 704624, 704625, 704627, 704628, 704629, 704630, 704631, 704632, 704633, 704634, 704635, 704636, 704637, 704640, 704641, 704642, 704643, 704644, 704645, 704646, 704649, 704650, 704652, 704653, 704654, 704658, 704659, 704661, 704662, 704663, 704664, 704665, 704667, 704668, 704670, 704672, 704673, 704674, 704675, 704676, 704677, 704678, 704679, 704680, 704682, 704683, 704684, 704685, 704687, 704688, 704690, 704691, 704692, 704693. 704694, 704695, 704697, 704698, 704699, 704700, 704701, 704702, 704703, 704704, 704705, 704706, 704708, 704710, 704711, 704712, 704713, 704714, 704716, 704718, 704719, 704720, 704721, 704722, 704724, 704725, 704726, 704728, 704729, 704731, 704733, 704735, 704737, 704738, 704739, 704740, 704742, 704743, 704744, 704747, 704748, 704749, 704750, 704751, 704752, 704753, 704754, 704757, 704758, 704761, 704762, 704763, 704764, 704765, 704766, 704767, 704768, 704769, 704770, 704773, 704774, 704776, 704779, 704780, 704781, 704782, 704783, 704784, 704786, 704788, 704791, 704792, 704793, 704794, 704795, 704796, 704797, 704798, 704800, 704801, 704803, 704804, 704807, 704808, 704809, 704810, 704811, 704814, 704815, 704816, 704817, 704818, 704819, 704820, 704821, 704822, 704823, 704824, 704825, 704827, 704829, 704830, 704831, 704832, 704833, 704834, 704835, 704837, 704838, 704839, 704840, 704842, 704843, 704844, 704845, 704847, 704848, 704849, 704850, 704851, 704852, 704853, 704854, 704857, 704858, 704859, 704860, 704861, 704862, 704864, 704865, 704866, 704867, 704868, 704869, 704870, 704871, 704872, 704873, 704874, 704875, 704876, 704877, 704878, 704879, 704880, 704881, 704882, 704883, 704884, 704885, 704886, 704887, 704888, 704889, 704890, 704891, 704892, 704894, 704895,

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1490-2012

Related Recalls

Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.

Nov 28, 2025 Infusion Pumps Nationwide View Details →