Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D, Model Numbers: DTPA2QQ, DTPA2Q1

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a. Model Numbers: DTPA2D4: GTIN 00763000178130, Serial Numbers: RTG600136S, RTG600270S, RTG601323S, RTG601338S, RTG601367S, RTG601433S, RTG601484S, RTG601489S, RTG601575S, RTG601583S, RTG601595S, RTG601632S, RTG601643S, RTG601653S, RTG601657S, RTG601669S, RTG601775S, RTG601790S, RTG601802S, RTG601805S, RTG601820S, RTG601847S, RTG601892S, RTG601904S, RTG601968S, RTG602017S, RTG602019S, RTG602125S, RTG602147S, RTG602179S, RTG602182S, RTG602191S, RTG602207S, RTG602209S, RTG602221S, RTG602226S, RTG602238S, RTG602239S, RTG602360S, RTG602439S, RTG602477S, RTG602480S, RTG602490S, RTG602506S, RTG602512S, RTG602515S, RTG602516S, RTG602524S, RTG602550S, RTG602580S, RTG602584S; GTIN 00763000178154, Serial Numbers: RTG600028S, RTG600029S, RTG600030S, RTG600033S, RTG600036S, RTG600037S, RTG600040S, RTG600048S, RTG600049S, RTG600054S, RTG600055S, RTG600056S, RTG600059S, RTG600060S, RTG600061S, RTG600062S, RTG600063S, RTG600064S, RTG600065S, RTG600066S, RTG600067S, RTG600068S, RTG600069S, RTG600070S, RTG600071S, RTG600072S, RTG600073S, RTG600074S, RTG600075S, RTG600076S, RTG600077S, RTG600079S, RTG600081S, RTG600082S, RTG600083S, RTG600085S, RTG600086S, RTG600087S, RTG600089S, RTG600090S, RTG600091S, RTG600092S, RTG600095S, RTG600096S, RTG600097S, RTG600098S, RTG600099S, RTG600100S, RTG600101S, RTG600104S, RTG600105S, RTG600106S, RTG600107S, RTG600108S, RTG600110S, RTG600111S, RTG600112S, RTG600113S, RTG600114S, RTG600115S, RTG600116S, RTG600117S, RTG600118S, RTG600119S, RTG600120S, RTG600124S, RTG600126S, RTG600146S, RTG600256S, RTG600268S, RTG600294S, RTG600310S, RTG600341S, RTG600345S, RTG600346S, RTG600348S, RTG600352S, RTG600356S, RTG600363S, RTG600366S, RTG600367S, RTG600375S, RTG600377S, RTG600378S, RTG600388S, RTG600390S, RTG600393S, RTG600395S, RTG600405S, RTG600415S, RTG600455S, RTG600462S, RTG600468S, RTG600469S, RTG600474S, RTG600486S, RTG600491S, RTG600523S, RTG600525S, RTG600531S, RTG600534S, RTG600537S, RTG600538S, RTG600540S, RTG600546S, RTG600602S, RTG600670S, RTG600671S, RTG600672S, RTG600673S, RTG600674S, RTG600675S, RTG600676S, RTG600677S, RTG600678S, RTG600679S, RTG600757S, RTG600760S, RTG600772S, RTG600773S, RTG600774S, RTG600780S, RTG600782S, RTG600783S, RTG600784S, RTG600787S, RTG600794S, RTG600795S, RTG600796S, RTG600798S, RTG600800S, RTG600801S, RTG600803S, RTG600805S, RTG600808S, RTG600809S, RTG600810S, RTG600811S, RTG600812S, RTG600813S, RTG600817S, RTG600819S, RTG600821S, RTG600834S, RTG600837S, RTG600838S, RTG600839S, RTG600840S, RTG600842S, RTG600843S, RTG600844S, RTG600846S, RTG600848S, RTG600849S, RTG600850S, RTG600851S, RTG600853S, RTG600854S, RTG600855S, RTG600856S, RTG600857S, RTG600858S, RTG600859S, RTG600860S, RTG600862S, RTG600863S, RTG600864S, RTG600865S, RTG600866S, RTG600867S, RTG600869S, RTG600871S, RTG600872S, RTG600873S, RTG600874S, RTG600876S, RTG600877S, RTG600878S, RTG600879S, RTG600880S, RTG600881S, RTG600882S, RTG600883S, RTG600884S, RTG600885S, RTG600887S, RTG600888S, RTG600889S, RTG600890S, RTG600891S, RTG600892S, RTG600893S, RTG600894S, RTG600895S, RTG600896S, RTG600897S, RTG600898S, RTG600900S, RTG600901S, RTG600902S, RTG600903S, RTG600904S, RTG600905S, RTG600906S, RTG600907S, RTG600908S, RTG600909S, RTG600910S, RTG600911S, RTG600912S, RTG600913S, RTG600914S, RTG600915S, RTG600916S, RTG600917S, RTG600918S, RTG600919S, RTG600920S, RTG600921S, RTG600922S, RTG600923S, RTG600924S, RTG600925S, RTG600926S, RTG600927S, RTG600928S, RTG600929S, RTG600930S, RTG600932S, RTG600933S, RTG600936S, RTG600939S, RTG600940S, RTG600941S, RTG600942S, RTG600943S, RTG600944S, RTG600945S, RTG600946S, RTG600947S, RTG600948S, RTG600949S, RTG600951S, RTG600952S, RTG600953S, RTG600955S, RTG600956S, RTG600957S, RTG600958S, RTG600959S, RTG600960S, RTG600961S, RTG600962S, RTG600963S, RTG600964S, RTG600965S, RTG600966S, RTG600967S, RTG600968S, RTG600969S, RTG600970S, RTG600971S, RTG600972S, RTG600973S, RTG600974S, RTG600975S, RTG600976S, RTG600977S, RTG600978S, RTG600979S, RTG600980S, RTG600981S, RTG600982S, RTG600983S, RTG600984S, RTG600985S, RTG600986S, RTG600987S, RTG600988S, RTG600989S, RTG600990S, RTG600991S, RTG600993S, RTG600996S, RTG600997S, RTG601000S, RTG601002S, RTG601003S, RTG601004S, RTG601005S, RTG601006S, RTG601010S, RTG601011S, RTG601014S, RTG601015S, RTG601016S, RTG601017S, RTG601018S, RTG601019S, RTG601020S, RTG601021S, RTG601022S, RTG601023S, RTG601024S, RTG601026S, RTG601027S, RTG601028S, RTG601033S, RTG601035S, RTG601036S, RTG601037S, RTG601038S, RTG601039S, RTG601041S, RTG601042S, RTG601044S, RTG601045S, RTG601047S, RTG601048S, RTG601049S, RTG601050S, RTG601051S, RTG601052S, RTG601053S, RTG601054S, RTG601055S, RTG601056S, RTG601057S, RTG601058S, RTG601059S, RTG601060S, RTG601061S, RTG601062S, RTG601063S, RTG601064S, RTG601066S, RTG601067S, RTG601068S, RTG601069S, RTG601070S, RTG601071S, RTG601072S, RTG601074S, RTG601075S, RTG601076S, RTG601077S, RTG601078S, RTG601080S, RTG601081S, RTG601083S, RTG601084S, RTG601087S, RTG601088S, RTG601108S, RTG601112S, RTG601129S, RTG601134S, RTG601135S, RTG601136S, RTG601137S, RTG601138S, RTG601139S, RTG601140S, RTG601141S, RTG601142S, RTG601143S, RTG601144S, RTG601145S, RTG601146S, RTG601147S, RTG601148S, RTG601149S, RTG601150S, RTG601151S, RTG601152S, RTG601153S, RTG601154S, RTG601155S, RTG601156S, RTG601157S, RTG601158S, RTG601159S, RTG601160S, RTG601163S, RTG601164S, RTG601165S, RTG601166S, RTG601167S, RTG601168S, RTG601169S, RTG601170S, RTG601171S, RTG601172S, RTG601173S, RTG601175S, RTG601176S, RTG601177S, RTG601178S, RTG601179S, RTG601180S, RTG601181S, RTG601182S, RTG601183S, RTG601184S, RTG601185S, RTG601186S, RTG601187S, RTG601189S, RTG601190S, RTG601191S, RTG601192S, RTG601193S, RTG601194S, RTG601195S, RTG601196S, RTG601197S, RTG601198S, RTG601199S, RTG601200S, RTG601201S, RTG601202S, RTG601203S, RTG601204S, RTG601205S, RTG601206S, RTG601207S, RTG601209S, RTG601215S, RTG601216S, RTG601217S, RTG601218S, RTG601219S, RTG601221S, RTG601222S, RTG601223S, RTG601224S, RTG601225S, RTG601226S, RTG601227S, RTG601228S, RTG601229S, RTG601231S, RTG601234S, RTG601235S, RTG601236S, RTG601238S, RTG601240S, RTG601242S, RTG601243S, RTG601244S, RTG601245S, RTG601246S, RTG601247S, RTG601248S, RTG601249S, RTG601252S, RTG601253S, RTG601255S, RTG601256S, RTG601257S, RTG601258S, RTG601260S, RTG601262S, RTG601263S, RTG601265S, RTG601268S, RTG601269S, RTG601270S, RTG601271S, RTG601272S, RTG601274S, RTG601275S, RTG601276S, RTG601277S, RTG601284S, RTG601286S, RTG601287S, RTG601288S, RTG601289S, RTG601292S, RTG601293S, RTG601294S, RTG601295S, RTG601296S, RTG601298S, RTG601300S, RTG601302S, RTG601303S, RTG601304S, RTG601306S, RTG601307S, RTG601308S, RTG601309S, RTG601310S, RTG601312S, RTG601313S, RTG601314S, RTG601315S, RTG601316S, RTG601317S, RTG601318S, RTG601319S, RTG601320S, RTG601321S, RTG601322S, RTG601328S, RTG601329S, RTG601344S, RTG601347S, RTG601348S, RTG601350S, RTG601351S, RTG601352S, RTG601353S, RTG601354S, RTG601356S, RTG601357S, RTG601359S, RTG601360S, RTG601361S, RTG601381S, RTG601384S, RTG601402S, RTG601404S, RTG601408S, RTG601416S, RTG601425S, RTG601429S, RTG601431S, RTG601434S, RTG601440S, RTG601443S, RTG601450S, RTG601451S, RTG601461S, RTG601464S, RTG601466S, RTG601467S, RTG601469S, RTG601470S, RTG601472S, RTG601473S, RTG601476S, RTG601477S, RTG601478S, RTG601480S, RTG601481S, RTG601483S, RTG601485S, RTG601486S, RTG601490S, RTG601492S, RTG601493S, RTG601496S, RTG601497S, RTG601499S, RTG601501S, RTG601505S, RTG601507S, RTG601516S, RTG601518S, RTG601527S, RTG601531S, RTG601535S, RTG601538S, RTG601543S, RTG601544S, RTG601545S, RTG601546S, RTG601547S, RTG601548S, RTG601549S, RTG601550S, RTG601552S, RTG601554S, RTG601555S, RTG601556S, RTG601557S, RTG601558S, RTG601559S, RTG601561S, RTG601566S, RTG601577S, RTG601578S, RTG601579S, RTG601582S, RTG601584S, RTG601587S, RTG601588S, RTG601590S, RTG601591S, RTG601592S, RTG601598S, RTG601609S, RTG601667S, RTG601686S, RTG601701S, RTG601702S, RTG601703S, RTG601704S, RTG601705S, RTG601707S, RTG601708S, RTG601709S, RTG601710S, RTG601711S, RTG601712S, RTG601713S, RTG601714S, RTG601715S, RTG601716S, RTG601717S, RTG601718S, RTG601719S, RTG601720S, RTG601721S, RTG601722S, RTG601723S, RTG601724S, RTG601725S, RTG601726S, RTG601727S, RTG601728S, RTG601729S, RTG601730S, RTG601731S, RTG601732S, RTG601733S, RTG601734S, RTG601735S, RTG601736S, RTG601737S, RTG601738S, RTG601739S, RTG601740S, RTG601741S, RTG601742S, RTG601743S, RTG601744S, RTG601745S, RTG601746S, RTG601747S, RTG601748S, RTG601749S, RTG601750S, RTG601751S, RTG601752S, RTG601753S, RTG601754S, RTG601755S, RTG601756S, RTG601757S, RTG601758S, RTG601759S, RTG601760S, RTG601766S, RTG601769S, RTG601770S, RTG601772S, RTG601774S, RTG601777S, RTG601779S, RTG601780S, RTG601781S, RTG601785S, RTG601788S, RTG601789S, RTG601792S, RTG601793S, RTG601795S, RTG601799S, RTG601800S, RTG601803S, RTG601804S, RTG601811S, RTG601815S, RTG601819S, RTG601831S, RTG601833S, RTG601836S, RTG601918S, RTG601924S, RTG601925S, RTG601935S, RTG601936S, RTG601939S, RTG601940S, RTG601941S, RTG601952S, RTG601954S, RTG601962S, RTG601985S, RTG601987S, RTG601988S, RTG601990S, RTG602000S, RTG602001S, RTG602002S, RTG602004S, RTG602005S, RTG602007S, RTG602013S, RTG602021S, RTG602022S, RTG602023S, RTG602026S, RTG602029S, RTG602030S, RTG602032S, RTG602038S, RTG602041S, RTG602043S, RTG602045S, RTG602046S, RTG602048S, RTG602052S, RTG602055S, RTG602059S, RTG602065S, RTG602066S, RTG602067S, RTG602071S, RTG602072S, RTG602074S, RTG602075S, RTG602078S, RTG602079S, RTG602080S, RTG602081S, RTG602082S, RTG602085S, RTG602086S, RTG602088S, RTG602089S, RTG602090S, RTG602094S, RTG602095S, RTG602099S, RTG602102S, RTG602103S, RTG602104S, RTG602105S, RTG602106S, RTG602107S, RTG602109S, RTG602111S, RTG602113S, RTG602114S, RTG602115S, RTG602116S, RTG602119S, RTG602120S, RTG602121S, RTG602122S, RTG602124S, RTG602129S, RTG602133S, RTG602169S, RTG602243S, RTG602245S, RTG602246S, RTG602247S, RTG602248S, RTG602250S, RTG602251S, RTG602254S, RTG602257S, RTG602259S, RTG602261S, RTG602263S, RTG602264S, RTG602265S, RTG602267S, RTG602268S, RTG602269S, RTG602270S, RTG602272S, RTG602274S, RTG602278S, RTG602279S, RTG602281S, RTG602283S, RTG602286S, RTG602287S, RTG602288S, RTG602289S, RTG602291S, RTG602292S, RTG602294S, RTG602297S, RTG602299S, RTG602300S, RTG602301S, RTG602307S, RTG602311S, RTG602328S, RTG602356S, RTG602362S, RTG602363S, RTG602366S, RTG602367S, RTG602368S, RTG602370S, RTG602371S, RTG602372S, RTG602374S, RTG602375S, RTG602376S, RTG602377S, RTG602378S, RTG602379S, RTG602380S, RTG602381S, RTG602382S, RTG602383S, RTG602384S, RTG602386S, RTG602387S, RTG602388S, RTG602389S, RTG602390S, RTG602391S, RTG602392S, RTG602393S, RTG602394S, RTG602395S, RTG602396S, RTG602397S, RTG602398S, RTG602399S, RTG602400S, RTG602401S, RTG602403S, RTG602404S, RTG602405S, RTG602406S, RTG602407S, RTG602408S, RTG602409S, RTG602410S, RTG602411S, RTG602412S, RTG602413S, RTG602414S, RTG602415S, RTG602416S, RTG602417S, RTG602418S, RTG602419S, RTG602420S, RTG602424S, RTG602429S, RTG602452S, RTG602518S, RTG602547S, RTG602558S, RTG602560S, RTG602564S, RTG602567S, RTG602568S, RTG602572S, RTG602574S, RTG602575S, RTG602589S, RTG602591S, RTG602592S, RTG602593S, RTG602596S, RTG602598S, RTG602599S, RTG602603S, RTG602604S, RTG602607S, RTG602608S, RTG602609S, RTG602610S, RTG602611S, RTG602613S, RTG602614S, RTG602621S, RTG602624S, RTG602626S, RTG602628S, RTG602632S, RTG602633S, RTG602635S, RTG602636S, RTG602637S, RTG602638S, RTG602639S, RTG602640S, RTG602641S, RTG602642S, RTG602643S, RTG602644S, RTG602645S, RTG602646S, RTG602647S, RTG602648S, RTG602649S, RTG602650S, RTG602651S, RTG602652S, RTG602653S, RTG602655S, RTG602656S, RTG602657S, RTG602658S, RTG602659S, RTG602660S, RTG602661S, RTG602664S, RTG602668S, RTG602670S, RTG602672S, RTG602673S, RTG602674S, RTG602676S, RTG602677S, RTG602678S, RTG602679S, RTG602680S, RTG602681S, RTG602682S, RTG602683S, RTG602684S, RTG602685S, RTG602688S, RTG602689S, RTG602690S, RTG602691S, RTG602694S, RTG602695S, RTG602696S, RTG602699S, RTG602701S, RTG602703S, RTG602705S, RTG602707S, RTG602708S; GTIN 00763000402211, Serial Numbers: RTG600302S, RTG600337S, RTG600344S, RTG600353S, RTG600355S, RTG600357S, RTG600362S, RTG600365S, RTG600371S, RTG600376S, RTG600379S, RTG600382S, RTG600387S, RTG600389S, RTG600391S, RTG600392S, RTG600396S, RTG600397S, RTG600407S, RTG600409S, RTG600410S, RTG600413S, RTG600414S, RTG600416S, RTG600417S, RTG600419S, RTG600420S, RTG600421S, RTG600422S, RTG600425S, RTG600427S, RTG600429S, RTG600431S, RTG600432S, RTG600433S, RTG600439S, RTG600442S, RTG600775S, RTG600776S, RTG600777S, RTG600778S, RTG600785S, RTG600789S, RTG600790S, RTG600797S, RTG600802S, RTG600804S, RTG600807S, RTG600814S, RTG600815S, RTG600818S, RTG600822S, RTG600823S, RTG600824S, RTG600825S, RTG600826S, RTG600827S, RTG600828S, RTG600830S, RTG600832S, RTG600833S, RTG600835S, RTG600836S, RTG600847S, RTG600861S, RTG600868S, RTG600870S, RTG600875S, RTG600992S, RTG600994S, RTG600995S, RTG601007S, RTG601009S, RTG601012S, RTG601025S, RTG601029S, RTG601031S, RTG601032S, RTG601040S, RTG601091S, RTG601093S, RTG601094S, RTG601095S, RTG601097S, RTG601099S, RTG601100S, RTG601101S, RTG601104S, RTG601105S, RTG601107S, RTG601110S, RTG601113S, RTG601114S, RTG601115S, RTG601116S, RTG601117S, RTG601118S, RTG601119S, RTG601120S, RTG601121S, RTG601122S, RTG601123S, RTG601124S, RTG601125S, RTG601126S, RTG601128S, RTG601131S, RTG601132S, RTG601133S, RTG601188S, RTG601208S, RTG601214S, RTG601220S, RTG601233S, RTG601250S, RTG601251S, RTG601254S, RTG601259S, RTG601261S, RTG601266S, RTG601267S, RTG601278S, RTG601279S, RTG601280S, RTG601282S, RTG601290S, RTG601291S, RTG601297S, RTG601305S, RTG601311S, RTG601569S, RTG601823S, RTG601825S, RTG601835S, RTG601838S, RTG601946S, RTG601947S, RTG602003S, RTG602009S, RTG602010S, RTG602044S, RTG602050S, RTG602056S, RTG602062S, RTG602063S, RTG602068S, RTG602076S, RTG602084S, RTG602087S, RTG602091S, RTG602097S, RTG602098S, RTG602100S, RTG602101S, RTG602110S, RTG602252S, RTG602255S, RTG602256S, RTG602262S, RTG602276S, RTG602277S, RTG602280S, RTG602285S, RTG602293S, RTG602298S, RTG602323S, RTG602337S, RTG602343S, RTG602344S, RTG602402S, RTG602541S, RTG602570S, RTG602578S, RTG602597S, RTG602606S, RTG602612S, RTG602618S, RTG602629S, RTG602667S, RTG602715S, RTG602717S, RTG602727S, RTG602729S; GTIN none, Serial Numbers: RTG601829S, RTG602061S, RTG602258S, RTG602669S, RTG602724S; DTPA2D1: GTIN 00763000178147, Serial Numbers: RTH600598S, RTH601437S, RTH601544S, RTH601557S, RTH601590S, RTH601762S, RTH601782S, RTH601832S, RTH601844S, RTH602085S, RTH602090S, RTH602094S, RTH602104S, RTH602105S, RTH602150S, RTH602168S, RTH602172S, RTH602198S, RTH602220S, RTH602255S, RTH602261S, RTH602265S, RTH602267S, RTH602282S, RTH602312S, RTH602320S; GTIN 00763000178161, Serial Numbers: RTH600013S, RTH600014S, RTH600015S, RTH600016S, RTH600021S, RTH600030S, RTH600031S, RTH600032S, RTH600033S, RTH600034S, RTH600035S, RTH600036S, RTH600037S, RTH600039S, RTH600041S, RTH600042S, RTH600044S, RTH600046S, RTH600048S, RTH600049S, RTH600051S, RTH600052S, RTH600054S, RTH600055S, RTH600056S, RTH600058S, RTH600059S, RTH600061S, RTH600062S, RTH600063S, RTH600064S, RTH600065S, RTH600066S, RTH600071S, RTH600072S, RTH600073S, RTH600075S, RTH600078S, RTH600079S, RTH600080S, RTH600083S, RTH600085S, RTH600086S, RTH600087S, RTH600090S, RTH600276S, RTH600283S, RTH600299S, RTH600328S, RTH600344S, RTH600374S, RTH600442S, RTH600526S, RTH600531S, RTH600539S, RTH600571S, RTH600575S, RTH600691S, RTH600693S, RTH600695S, RTH600720S, RTH600721S, RTH600727S, RTH600730S, RTH600733S, RTH600739S, RTH600740S, RTH600741S, RTH600745S, RTH600750S, RTH600759S, RTH600760S, RTH600761S, RTH600762S, RTH600763S, RTH600764S, RTH600765S, RTH600771S, RTH600772S, RTH600773S, RTH600774S, RTH600775S, RTH600778S, RTH600779S, RTH600780S, RTH600782S, RTH600785S, RTH600788S, RTH600789S, RTH600790S, RTH600791S, RTH600792S, RTH600793S, RTH600795S, RTH600796S, RTH600797S, RTH600813S, RTH600848S, RTH600858S, RTH600861S, RTH600864S, RTH600865S, RTH600866S, RTH600867S, RTH600871S, RTH600874S, RTH600880S, RTH600886S, RTH600890S, RTH600892S, RTH600895S, RTH600896S, RTH600899S, RTH600900S, RTH600901S, RTH600902S, RTH600903S, RTH600904S, RTH600907S, RTH600908S, RTH600909S, RTH600910S, RTH600911S, RTH600912S, RTH600913S, RTH600914S, RTH600915S, RTH600916S, RTH600917S, RTH600919S, RTH600920S, RTH600921S, RTH600924S, RTH600925S, RTH600927S, RTH600928S, RTH600929S, RTH600930S, RTH600931S, RTH600932S, RTH600933S, RTH600934S, RTH600936S, RTH600938S, RTH600939S, RTH600941S, RTH600942S, RTH600943S, RTH600946S, RTH600948S, RTH600950S, RTH600951S, RTH600952S, RTH600955S, RTH600957S, RTH600958S, RTH600959S, RTH600960S, RTH600962S, RTH600963S, RTH600964S, RTH600967S, RTH600968S, RTH600970S, RTH600971S, RTH600972S, RTH600973S, RTH600974S, RTH600975S, RTH600976S, RTH600977S, RTH600978S, RTH600980S, RTH600981S, RTH600984S, RTH600986S, RTH600988S, RTH600989S, RTH600993S, RTH600994S, RTH600996S, RTH600998S, RTH601000S, RTH601002S, RTH601004S, RTH601005S, RTH601006S, RTH601007S, RTH601008S, RTH601011S, RTH601012S, RTH601013S, RTH601014S, RTH601015S, RTH601016S, RTH601017S, RTH601018S, RTH601019S, RTH601020S, RTH601021S, RTH601022S, RTH601023S, RTH601026S, RTH601028S, RTH601031S, RTH601032S, RTH601033S, RTH601034S, RTH601035S, RTH601036S, RTH601038S, RTH601039S, RTH601040S, RTH601041S, RTH601042S, RTH601043S, RTH601044S, RTH601045S, RTH601046S, RTH601047S, RTH601048S, RTH601049S, RTH601050S, RTH601051S, RTH601052S, RTH601054S, RTH601058S, RTH601059S, RTH601060S, RTH601062S, RTH601063S, RTH601065S, RTH601066S, RTH601067S, RTH601068S, RTH601069S, RTH601070S, RTH601071S, RTH601073S, RTH601074S, RTH601075S, RTH601076S, RTH601077S, RTH601080S, RTH601081S, RTH601082S, RTH601083S, RTH601084S, RTH601085S, RTH601086S, RTH601087S, RTH601088S, RTH601089S, RTH601090S, RTH601091S, RTH601092S, RTH601094S, RTH601096S, RTH601097S, RTH601098S, RTH601099S, RTH601100S, RTH601101S, RTH601102S, RTH601103S, RTH601105S, RTH601106S, RTH601107S, RTH601108S, RTH601109S, RTH601110S, RTH601111S, RTH601112S, RTH601113S, RTH601114S, RTH601115S, RTH601116S, RTH601117S, RTH601118S, RTH601119S, RTH601120S, RTH601121S, RTH601122S, RTH601123S, RTH601124S, RTH601126S, RTH601127S, RTH601131S, RTH601133S, RTH601135S, RTH601136S, RTH601137S, RTH601138S, RTH601140S, RTH601141S, RTH601143S, RTH601164S, RTH601165S, RTH601166S, RTH601169S, RTH601171S, RTH601172S, RTH601173S, RTH601174S, RTH601176S, RTH601178S, RTH601179S, RTH601180S, RTH601181S, RTH601183S, RTH601184S, RTH601186S, RTH601187S, RTH601188S, RTH601189S, RTH601190S, RTH601191S, RTH601194S, RTH601195S, RTH601196S, RTH601197S, RTH601198S, RTH601199S, RTH601200S, RTH601201S, RTH601202S, RTH601203S, RTH601204S, RTH601215S, RTH601276S, RTH601279S, RTH601280S, RTH601281S, RTH601283S, RTH601284S, RTH601285S, RTH601286S, RTH601287S, RTH601288S, RTH601289S, RTH601290S, RTH601291S, RTH601292S, RTH601293S, RTH601314S, RTH601315S, RTH601320S, RTH601324S, RTH601328S, RTH601335S, RTH601336S, RTH601338S, RTH601345S, RTH601349S, RTH601351S, RTH601358S, RTH601359S, RTH601360S, RTH601362S, RTH601363S, RTH601384S, RTH601461S, RTH601462S, RTH601464S, RTH601465S, RTH601467S, RTH601468S, RTH601469S, RTH601471S, RTH601472S, RTH601473S, RTH601474S, RTH601476S, RTH601477S, RTH601478S, RTH601481S, RTH601488S, RTH601489S, RTH601490S, RTH601491S, RTH601492S, RTH601493S, RTH601495S, RTH601496S, RTH601497S, RTH601498S, RTH601499S, RTH601500S, RTH601501S, RTH601503S, RTH601505S, RTH601506S, RTH601507S, RTH601512S, RTH601513S, RTH601515S, RTH601519S, RTH601523S, RTH601524S, RTH601527S, RTH601528S, RTH601529S, RTH601530S, RTH601531S, RTH601532S, RTH601535S, RTH601536S, RTH601537S, RTH601539S, RTH601540S, RTH601592S, RTH601593S, RTH601594S, RTH601595S, RTH601596S, RTH601597S, RTH601598S, RTH601599S, RTH601600S, RTH601601S, RTH601602S, RTH601603S, RTH601604S, RTH601605S, RTH601606S, RTH601607S, RTH601608S, RTH601609S, RTH601612S, RTH601628S, RTH601630S, RTH601639S, RTH601650S, RTH601654S, RTH601659S, RTH601660S, RTH601668S, RTH601672S, RTH601674S, RTH601676S, RTH601678S, RTH601680S, RTH601681S, RTH601682S, RTH601683S, RTH601684S, RTH601685S, RTH601689S, RTH601690S, RTH601691S, RTH601692S, RTH601693S, RTH601698S, RTH601700S, RTH601709S, RTH601717S, RTH601720S, RTH601722S, RTH601723S, RTH601775S, RTH601824S, RTH601834S, RTH601841S, RTH601847S, RTH601850S, RTH601855S, RTH601860S, RTH601867S, RTH601869S, RTH601870S, RTH601881S, RTH601882S, RTH601887S, RTH601893S, RTH601898S, RTH601900S, RTH601901S, RTH601903S, RTH601914S, RTH601917S, RTH601918S, RTH601919S, RTH601920S, RTH601921S, RTH601922S, RTH601924S, RTH601925S, RTH601926S, RTH601927S, RTH601928S, RTH601931S, RTH601933S, RTH601935S, RTH601940S, RTH601941S, RTH601943S, RTH601945S, RTH601946S, RTH601947S, RTH601953S, RTH601955S, RTH601957S, RTH601960S, RTH601961S, RTH601968S, RTH601969S, RTH601971S, RTH601972S, RTH601980S, RTH601983S, RTH601984S, RTH601985S, RTH601987S, RTH601989S, RTH601990S, RTH602001S, RTH602004S, RTH602006S, RTH602007S, RTH602010S, RTH602011S, RTH602013S, RTH602120S, RTH602124S, RTH602125S, RTH602132S, RTH602145S, RTH602157S, RTH602158S, RTH602161S, RTH602163S, RTH602166S, RTH602184S, RTH602253S, RTH602273S, RTH602274S, RTH602283S, RTH602285S, RTH602286S, RTH602287S, RTH602290S, RTH602291S, RTH602292S, RTH602293S, RTH602294S, RTH602295S, RTH602296S, RTH602297S, RTH602299S, RTH602300S, RTH602304S, RTH602307S, RTH602308S, RTH602309S, RTH602310S, RTH602311S, RTH602313S, RTH602314S, RTH602315S, RTH602316S, RTH602318S, RTH602322S, RTH602323S, RTH602324S, RTH602325S, RTH602326S, RTH602328S, RTH602329S, RTH602330S, RTH602331S, RTH602332S, RTH602333S, RTH602334S, RTH602335S, RTH602336S, RTH602337S, RTH602338S, RTH602339S, RTH602340S, RTH602341S, RTH602342S, RTH602343S, RTH602344S, RTH602345S, RTH602346S, RTH602347S, RTH602348S, RTH602349S, RTH602350S, RTH602351S, RTH602358S, RTH602359S, RTH602362S, RTH602364S, RTH602367S, RTH602370S, RTH602371S, RTH602373S, RTH602376S, RTH602378S, RTH602379S, RTH602382S, RTH602384S, RTH602386S, RTH602393S, RTH602398S, RTH602400S, RTH602403S, RTH602404S, RTH602405S, RTH602406S, RTH602410S, RTH602411S, RTH602416S, RTH602422S, RTH602423S, RTH602424S, RTH602425S, RTH602426S, RTH602432S, RTH602433S, RTH602435S, RTH602436S, RTH602438S, RTH602440S; GTIN 00763000402204, Serial Numbers: RTH600724S, RTH600725S, RTH600726S, RTH600729S, RTH600731S, RTH600732S, RTH600735S, RTH600736S, RTH600737S, RTH600738S, RTH600742S, RTH600743S, RTH600744S, RTH600746S, RTH600747S, RTH600749S, RTH600751S, RTH600752S, RTH600755S, RTH600756S, RTH600757S, RTH600758S, RTH600766S, RTH600767S, RTH600769S, RTH600777S, RTH600783S, RTH600786S, RTH600794S, RTH600798S, RTH600799S, RTH600800S, RTH600801S, RTH600802S, RTH600803S, RTH600804S, RTH600805S, RTH600807S, RTH600808S, RTH600809S, RTH600810S, RTH600812S, RTH600814S, RTH600815S, RTH600816S, RTH600817S, RTH600818S, RTH600819S, RTH600820S, RTH600821S, RTH600822S, RTH600824S, RTH600825S, RTH600826S, RTH600827S, RTH600828S, RTH600829S, RTH600830S, RTH600831S, RTH600832S, RTH600833S, RTH600834S, RTH600837S, RTH600838S, RTH600839S, RTH600840S, RTH600841S, RTH600842S, RTH600843S, RTH600844S, RTH600846S, RTH600849S, RTH600850S, RTH600851S, RTH600853S, RTH600854S, RTH600855S, RTH600857S, RTH600859S, RTH600860S, RTH600862S, RTH600863S, RTH600868S, RTH600869S, RTH600870S, RTH600872S, RTH600873S, RTH600875S, RTH600876S, RTH600877S, RTH600878S, RTH600879S, RTH600881S, RTH600884S, RTH600887S, RTH600888S, RTH600889S, RTH600893S, RTH600897S, RTH600898S, RTH600905S, RTH600906S, RTH600918S, RTH600922S, RTH600923S, RTH600926S, RTH600937S, RTH600940S, RTH600944S, RTH600945S, RTH600947S, RTH600953S, RTH600956S, RTH600961S, RTH600965S, RTH600966S, RTH600969S, RTH600985S, RTH600990S, RTH600992S, RTH600995S, RTH601001S, RTH601027S, RTH601029S, RTH601030S, RTH601037S, RTH601053S, RTH601061S, RTH601095S, RTH601129S, RTH601132S, RTH601139S, RTH601142S, RTH601205S, RTH601206S, RTH601207S, RTH601208S, RTH601209S, RTH601210S, RTH601211S, RTH601212S, RTH601214S, RTH601216S, RTH601217S, RTH601218S, RTH601219S, RTH601220S, RTH601221S, RTH601222S, RTH601223S, RTH601224S, RTH601225S, RTH601226S, RTH601227S, RTH601228S, RTH601229S, RTH601230S, RTH601231S, RTH601232S, RTH601233S, RTH601234S, RTH601235S, RTH601236S, RTH601237S, RTH601238S, RTH601239S, RTH601240S, RTH601241S, RTH601242S, RTH601243S, RTH601244S, RTH601245S, RTH601246S, RTH601247S, RTH601248S, RTH601249S, RTH601250S, RTH601251S, RTH601252S, RTH601305S, RTH601308S, RTH601311S, RTH601312S, RTH601326S, RTH601327S, RTH601329S, RTH601332S, RTH601339S, RTH601346S, RTH601347S, RTH601348S, RTH601352S, RTH601355S, RTH601381S, RTH601385S, RTH601389S, RTH601958S, RTH601962S, RTH601964S, RTH601970S, RTH601974S, RTH601976S, RTH601979S, RTH601982S, RTH601986S, RTH601993S, RTH601996S, RTH602008S, RTH602009S, RTH602114S, RTH602115S, RTH602117S, RTH602119S, RTH602121S, RTH602123S, RTH602126S, RTH602127S, RTH602128S, RTH602130S, RTH602131S, RTH602289S, RTH602352S, RTH602360S, RTH602361S, RTH602363S, RTH602366S, RTH602368S, RTH602369S, RTH602372S, RTH602374S, RTH602375S, RTH602377S, RTH602380S, RTH602381S, RTH602387S, RTH602391S, RTH602392S, RTH602394S, RTH602395S, RTH602399S, RTH602402S, RTH602408S, RTH602415S, RTH602417S, RTH602419S; GTIN none, Serial Numbers: RTH602129S, RTH602327S, RTH602409S; b. Model Numbers: DTPA2QQ: GTIN 00763000178093, Serial Numbers: RTC600339S, RTC600388S, RTC600657S, RTC600779S, RTC602354S, RTC605066S, RTC605335S, RTC605388S, RTC605468S, RTC606108S, RTC606304S, RTC606480S, RTC606523S, RTC606525S, RTC606531S, RTC606532S, RTC607084S, RTC607516S, RTC607842S, RTC607849S, RTC607859S, RTC608327S, RTC608331S, RTC608332S, RTC609193S, RTC609471S, RTC609652S, RTC609694S, RTC609704S, RTC610036S, RTC610579S, RTC610587S, RTC610594S, RTC610597S, RTC610612S, RTC610616S, RTC610620S, RTC610627S, RTC610632S, RTC610634S, RTC610640S, RTC610642S, RTC610644S, RTC610663S, RTC610674S, RTC610678S, RTC610744S, RTC610783S, RTC610910S, RTC610914S, RTC610963S, RTC610992S, RTC611007S, RTC611013S, RTC611037S, RTC611044S, RTC611058S, RTC611214S, RTC611215S, RTC611268S, RTC611287S, RTC611350S, RTC611710S, RTC611717S, RTC611748S; GTIN 00763000178116, Serial Numbers: RTC600048S, RTC600050S, RTC600053S, RTC600055S, RTC600056S, RTC600057S, RTC600058S, RTC600059S, RTC600060S, RTC600061S, RTC600062S, RTC600063S, RTC600064S, RTC600065S, RTC600066S, RTC600067S, RTC600068S, RTC600069S, RTC600070S, RTC600071S, RTC600072S, RTC600073S, RTC600074S, RTC600075S, RTC600076S, RTC600077S, RTC600078S, RTC600079S, RTC600080S, RTC600081S, RTC600082S, RTC600083S, RTC600085S, RTC600087S, RTC600093S, RTC600094S, RTC600095S, RTC600096S, RTC600098S, RTC600099S, RTC600100S, RTC600101S, RTC600102S, RTC600103S, RTC600104S, RTC600105S, RTC600106S, RTC600107S, RTC600108S, RTC600109S, RTC600110S, RTC600111S, RTC600112S, RTC600113S, RTC600114S, RTC600115S, RTC600116S, RTC600117S, RTC600118S, RTC600119S, RTC600120S, RTC600121S, RTC600122S, RTC600123S, RTC600124S, RTC600125S, RTC600126S, RTC600127S, RTC600128S, RTC600129S, RTC600130S, RTC600131S, RTC600132S, RTC600134S, RTC600135S, RTC600136S, RTC600137S, RTC600138S, RTC600139S, RTC600140S, RTC600141S, RTC600142S, RTC600143S, RTC600144S, RTC600145S, RTC600146S, RTC600147S, RTC600148S, RTC600149S, RTC600150S, RTC600151S, RTC600152S, RTC600153S, RTC600154S, RTC600155S, RTC600156S, RTC600157S, RTC600158S, RTC600159S, RTC600160S, RTC600161S, RTC600162S, RTC600163S, RTC600164S, RTC600168S, RTC600169S, RTC600170S, RTC600171S, RTC600172S, RTC600173S, RTC600174S, RTC600175S, RTC600176S, RTC600177S, RTC600179S, RTC600180S, RTC600181S, RTC600182S, RTC600183S, RTC600184S, RTC600185S, RTC600186S, RTC600187S, RTC600188S, RTC600189S, RTC600200S, RTC600313S, RTC600403S, RTC600544S, RTC600550S, RTC600557S, RTC600558S, RTC600559S, RTC600560S, RTC600561S, RTC600563S, RTC600564S, RTC600565S, RTC600566S, RTC600568S, RTC600569S, RTC600572S, RTC600575S, RTC600584S, RTC600585S, RTC600589S, RTC600590S, RTC600596S, RTC600599S, RTC600600S, RTC600603S, RTC600606S, RTC600608S, RTC600609S, RTC600611S, RTC600612S, RTC600616S, RTC600618S, RTC600622S, RTC600626S, RTC600627S, RTC600629S, RTC600630S, RTC600631S, RTC600632S, RTC600634S, RTC600635S, RTC600636S, RTC600637S, RTC600640S, RTC600641S, RTC600644S, RTC600645S, RTC600646S, RTC600651S, RTC600652S, RTC600654S, RTC600661S, RTC600665S, RTC600681S, RTC600733S, RTC600734S, RTC600735S, RTC600737S, RTC600739S, RTC600747S, RTC600748S, RTC600749S, RTC600752S, RTC600753S, RTC600754S, RTC600758S, RTC600760S, RTC600764S, RTC600766S, RTC600770S, RTC600771S, RTC600772S, RTC600775S, RTC600777S, RTC600803S, RTC600804S, RTC600805S, RTC600806S, RTC600807S, RTC600808S, RTC600809S, RTC600810S, RTC600811S, RTC600812S, RTC600813S, RTC600814S, RTC600815S, RTC600816S, RTC600817S, RTC600818S, RTC600819S, RTC600820S, RTC600821S, RTC600822S, RTC600823S, RTC600824S, RTC600825S, RTC600826S, RTC600827S, RTC600828S, RTC600913S, RTC600918S, RTC600919S, RTC600922S, RTC600927S, RTC600931S, RTC600932S, RTC600933S, RTC600934S, RTC600935S, RTC600936S, RTC600937S, RTC600938S, RTC600939S, RTC600940S, RTC600941S, RTC600942S, RTC600943S, RTC600944S, RTC600945S, RTC600946S, RTC600947S, RTC600948S, RTC600949S, RTC600950S, RTC600951S, RTC600952S, RTC600953S, RTC600954S, RTC600955S, RTC600956S, RTC600957S, RTC600958S, RTC600959S, RTC600960S, RTC600961S, RTC600962S, RTC600963S, RTC600964S, RTC600965S, RTC600966S, RTC600967S, RTC600968S, RTC600969S, RTC600970S, RTC600971S, RTC600972S, RTC600973S, RTC600974S, RTC600975S, RTC600976S, RTC600977S, RTC600978S, RTC600979S, RTC600981S, RTC600982S, RTC600983S, RTC600984S, RTC600985S, RTC600986S, RTC600987S, RTC600988S, RTC600989S, RTC600990S, RTC600991S, RTC600992S, RTC600993S, RTC600994S, RTC600995S, RTC600996S, RTC601003S, RTC601009S, RTC601010S, RTC601011S, RTC601012S, RTC601013S, RTC601014S, RTC601015S, RTC601016S, RTC601017S, RTC601018S, RTC601019S, RTC601020S, RTC601021S, RTC601022S, RTC601023S, RTC601024S, RTC601025S, RTC601026S, RTC601027S, RTC601028S, RTC601029S, RTC601030S, RTC601031S, RTC601032S, RTC601033S, RTC601034S, RTC601035S, RTC601036S, RTC601037S, RTC601038S, RTC601039S, RTC601040S, RTC601041S, RTC601042S, RTC601043S, RTC601044S, RTC601045S, RTC601046S, RTC601047S, RTC601048S, RTC601049S, RTC601050S, RTC601051S, RTC601052S, RTC601053S, RTC601054S, RTC601055S, RTC601056S, RTC601057S, RTC601058S, RTC601059S, RTC601060S, RTC601061S, RTC601063S, RTC601064S, RTC601065S, RTC601066S, RTC601067S, RTC601068S, RTC601069S, RTC601070S, RTC601071S, RTC601072S, RTC601073S, RTC601074S, RTC601075S, RTC601076S, RTC601077S, RTC601078S, RTC601079S, RTC601080S, RTC601081S, RTC601084S, RTC601091S, RTC601092S, RTC601095S, RTC601096S, RTC601098S, RTC601105S, RTC601120S, RTC601122S, RTC601133S, RTC601136S, RTC601139S, RTC601140S, RTC601169S, RTC601220S, RTC601230S, RTC601238S, RTC601286S, RTC601289S, RTC601294S, RTC601295S, RTC601296S, RTC601300S, RTC601302S, RTC601306S, RTC601307S, RTC601315S, RTC601321S, RTC601325S, RTC601335S, RTC601347S, RTC601364S, RTC601365S, RTC601369S, RTC601374S, RTC601377S, RTC601378S, RTC601380S, RTC601382S, RTC601406S, RTC601411S, RTC601415S, RTC601437S, RTC601441S, RTC601462S, RTC601468S, RTC601480S, RTC601481S, RTC601492S, RTC601499S, RTC601500S, RTC601501S, RTC601502S, RTC601503S, RTC601504S, RTC601505S, RTC601507S, RTC601508S, RTC601509S, RTC601510S, RTC601511S, RTC601512S, RTC601513S, RTC601514S, RTC601515S, RTC601516S, RTC601517S, RTC601518S, RTC601519S, RTC601520S, RTC601521S, RTC601522S, RTC601523S, RTC601524S, RTC601525S, RTC601526S, RTC601527S, RTC601528S, RTC601529S, RTC601530S, RTC601531S, RTC601532S, RTC601533S, RTC601534S, RTC601535S, RTC601536S, RTC601537S, RTC601538S, RTC601539S, RTC601540S, RTC601541S, RTC601542S, RTC601543S, RTC601544S, RTC601545S, RTC601546S, RTC601547S, RTC601548S, RTC601549S, RTC601550S, RTC601551S, RTC601552S, RTC601553S, RTC601554S, RTC601555S, RTC601556S, RTC601557S, RTC601558S, RTC601559S, RTC601560S, RTC601561S, RTC601562S, RTC601563S, RTC601564S, RTC601565S, RTC601566S, RTC601567S, RTC601568S, RTC601569S, RTC601570S, RTC601571S, RTC601572S, RTC601573S, RTC601574S, RTC601575S, RTC601578S, RTC601579S, RTC601580S, RTC601581S, RTC601582S, RTC601583S, RTC601584S, RTC601585S, RTC601586S, RTC601587S, RTC601588S, RTC601589S, RTC601590S, RTC601591S, RTC601592S, RTC601593S, RTC601594S, RTC601595S, RTC601596S, RTC601597S, RTC601598S, RTC601599S, RTC601601S, RTC601602S, RTC601603S, RTC601604S, RTC601605S, RTC601606S, RTC601607S, RTC601608S, RTC601609S, RTC601610S, RTC601611S, RTC601612S, RTC601613S, RTC601614S, RTC601615S, RTC601616S, RTC601617S, RTC601618S, RTC601619S, RTC601620
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Reason for Recall:
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D, Model Numbers: DTPA2QQ, DTPA2Q1

Product Codes/Lot Numbers:

a. Model Numbers: DTPA2D4: GTIN 00763000178130, Serial Numbers: RTG600136S, RTG600270S, RTG601323S, RTG601338S, RTG601367S, RTG601433S, RTG601484S, RTG601489S, RTG601575S, RTG601583S, RTG601595S, RTG601632S, RTG601643S, RTG601653S, RTG601657S, RTG601669S, RTG601775S, RTG601790S, RTG601802S, RTG601805S, RTG601820S, RTG601847S, RTG601892S, RTG601904S, RTG601968S, RTG602017S, RTG602019S, RTG602125S, RTG602147S, RTG602179S, RTG602182S, RTG602191S, RTG602207S, RTG602209S, RTG602221S, RTG602226S, RTG602238S, RTG602239S, RTG602360S, RTG602439S, RTG602477S, RTG602480S, RTG602490S, RTG602506S, RTG602512S, RTG602515S, RTG602516S, RTG602524S, RTG602550S, RTG602580S, RTG602584S; GTIN 00763000178154, Serial Numbers: RTG600028S, RTG600029S, RTG600030S, RTG600033S, RTG600036S, RTG600037S, RTG600040S, RTG600048S, RTG600049S, RTG600054S, RTG600055S, RTG600056S, RTG600059S, RTG600060S, RTG600061S, RTG600062S, RTG600063S, RTG600064S, RTG600065S, RTG600066S, RTG600067S, RTG600068S, RTG600069S, RTG600070S, RTG600071S, RTG600072S, RTG600073S, RTG600074S, RTG600075S, RTG600076S, RTG600077S, RTG600079S, RTG600081S, RTG600082S, RTG600083S, RTG600085S, RTG600086S, RTG600087S, RTG600089S, RTG600090S, RTG600091S, RTG600092S, RTG600095S, RTG600096S, RTG600097S, RTG600098S, RTG600099S, RTG600100S, RTG600101S, RTG600104S, RTG600105S, RTG600106S, RTG600107S, RTG600108S, RTG600110S, RTG600111S, RTG600112S, RTG600113S, RTG600114S, RTG600115S, RTG600116S, RTG600117S, RTG600118S, RTG600119S, RTG600120S, RTG600124S, RTG600126S, RTG600146S, RTG600256S, RTG600268S, RTG600294S, RTG600310S, RTG600341S, RTG600345S, RTG600346S, RTG600348S, RTG600352S, RTG600356S, RTG600363S, RTG600366S, RTG600367S, RTG600375S, RTG600377S, RTG600378S, RTG600388S, RTG600390S, RTG600393S, RTG600395S, RTG600405S, RTG600415S, RTG600455S, RTG600462S, RTG600468S, RTG600469S, RTG600474S, RTG600486S, RTG600491S, RTG600523S, RTG600525S, RTG600531S, RTG600534S, RTG600537S, RTG600538S, RTG600540S, RTG600546S, RTG600602S, RTG600670S, RTG600671S, RTG600672S, RTG600673S, RTG600674S, RTG600675S, RTG600676S, RTG600677S, RTG600678S, RTG600679S, RTG600757S, RTG600760S, RTG600772S, RTG600773S, RTG600774S, RTG600780S, RTG600782S, RTG600783S, RTG600784S, RTG600787S, RTG600794S, RTG600795S, RTG600796S, RTG600798S, RTG600800S, RTG600801S, RTG600803S, RTG600805S, RTG600808S, RTG600809S, RTG600810S, RTG600811S, RTG600812S, RTG600813S, RTG600817S, RTG600819S, RTG600821S, RTG600834S, RTG600837S, RTG600838S, RTG600839S, RTG600840S, RTG600842S, RTG600843S, RTG600844S, RTG600846S, RTG600848S, RTG600849S, RTG600850S, RTG600851S, RTG600853S, RTG600854S, RTG600855S, RTG600856S, RTG600857S, RTG600858S, RTG600859S, RTG600860S, RTG600862S, RTG600863S, RTG600864S, RTG600865S, RTG600866S, RTG600867S, RTG600869S, RTG600871S, RTG600872S, RTG600873S, RTG600874S, RTG600876S, RTG600877S, RTG600878S, RTG600879S, RTG600880S, RTG600881S, RTG600882S, RTG600883S, RTG600884S, RTG600885S, RTG600887S, RTG600888S, RTG600889S, RTG600890S, RTG600891S, RTG600892S, RTG600893S, RTG600894S, RTG600895S, RTG600896S, RTG600897S, RTG600898S, RTG600900S, RTG600901S, RTG600902S, RTG600903S, RTG600904S, RTG600905S, RTG600906S, RTG600907S, RTG600908S, RTG600909S, RTG600910S, RTG600911S, RTG600912S, RTG600913S, RTG600914S, RTG600915S, RTG600916S, RTG600917S, RTG600918S, RTG600919S, RTG600920S, RTG600921S, RTG600922S, RTG600923S, RTG600924S, RTG600925S, RTG600926S, RTG600927S, RTG600928S, RTG600929S, RTG600930S, RTG600932S, RTG600933S, RTG600936S, RTG600939S, RTG600940S, RTG600941S, RTG600942S, RTG600943S, RTG600944S, RTG600945S, RTG600946S, RTG600947S, RTG600948S, RTG600949S, RTG600951S, RTG600952S, RTG600953S, RTG600955S, RTG600956S, RTG600957S, RTG600958S, RTG600959S, RTG600960S, RTG600961S, RTG600962S, RTG600963S, RTG600964S, RTG600965S, RTG600966S, RTG600967S, RTG600968S, RTG600969S, RTG600970S, RTG600971S, RTG600972S, RTG600973S, RTG600974S, RTG600975S, RTG600976S, RTG600977S, RTG600978S, RTG600979S, RTG600980S, RTG600981S, RTG600982S, RTG600983S, RTG600984S, RTG600985S, RTG600986S, RTG600987S, RTG600988S, RTG600989S, RTG600990S, RTG600991S, RTG600993S, RTG600996S, RTG600997S, RTG601000S, RTG601002S, RTG601003S, RTG601004S, RTG601005S, RTG601006S, RTG601010S, RTG601011S, RTG601014S, RTG601015S, RTG601016S, RTG601017S, RTG601018S, RTG601019S, RTG601020S, RTG601021S, RTG601022S, RTG601023S, RTG601024S, RTG601026S, RTG601027S, RTG601028S, RTG601033S, RTG601035S, RTG601036S, RTG601037S, RTG601038S, RTG601039S, RTG601041S, RTG601042S, RTG601044S, RTG601045S, RTG601047S, RTG601048S, RTG601049S, RTG601050S, RTG601051S, RTG601052S, RTG601053S, RTG601054S, RTG601055S, RTG601056S, RTG601057S, RTG601058S, RTG601059S, RTG601060S, RTG601061S, RTG601062S, RTG601063S, RTG601064S, RTG601066S, RTG601067S, RTG601068S, RTG601069S, RTG601070S, RTG601071S, RTG601072S, RTG601074S, RTG601075S, RTG601076S, RTG601077S, RTG601078S, RTG601080S, RTG601081S, RTG601083S, RTG601084S, RTG601087S, RTG601088S, RTG601108S, RTG601112S, RTG601129S, RTG601134S, RTG601135S, RTG601136S, RTG601137S, RTG601138S, RTG601139S, RTG601140S, RTG601141S, RTG601142S, RTG601143S, RTG601144S, RTG601145S, RTG601146S, RTG601147S, RTG601148S, RTG601149S, RTG601150S, RTG601151S, RTG601152S, RTG601153S, RTG601154S, RTG601155S, RTG601156S, RTG601157S, RTG601158S, RTG601159S, RTG601160S, RTG601163S, RTG601164S, RTG601165S, RTG601166S, RTG601167S, RTG601168S, RTG601169S, RTG601170S, RTG601171S, RTG601172S, RTG601173S, RTG601175S, RTG601176S, RTG601177S, RTG601178S, RTG601179S, RTG601180S, RTG601181S, RTG601182S, RTG601183S, RTG601184S, RTG601185S, RTG601186S, RTG601187S, RTG601189S, RTG601190S, RTG601191S, RTG601192S, RTG601193S, RTG601194S, RTG601195S, RTG601196S, RTG601197S, RTG601198S, RTG601199S, RTG601200S, RTG601201S, RTG601202S, RTG601203S, RTG601204S, RTG601205S, RTG601206S, RTG601207S, RTG601209S, RTG601215S, RTG601216S, RTG601217S, RTG601218S, RTG601219S, RTG601221S, RTG601222S, RTG601223S, RTG601224S, RTG601225S, RTG601226S, RTG601227S, RTG601228S, RTG601229S, RTG601231S, RTG601234S, RTG601235S, RTG601236S, RTG601238S, RTG601240S, RTG601242S, RTG601243S, RTG601244S, RTG601245S, RTG601246S, RTG601247S, RTG601248S, RTG601249S, RTG601252S, RTG601253S, RTG601255S, RTG601256S, RTG601257S, RTG601258S, RTG601260S, RTG601262S, RTG601263S, RTG601265S, RTG601268S, RTG601269S, RTG601270S, RTG601271S, RTG601272S, RTG601274S, RTG601275S, RTG601276S, RTG601277S, RTG601284S, RTG601286S, RTG601287S, RTG601288S, RTG601289S, RTG601292S, RTG601293S, RTG601294S, RTG601295S, RTG601296S, RTG601298S, RTG601300S, RTG601302S, RTG601303S, RTG601304S, RTG601306S, RTG601307S, RTG601308S, RTG601309S, RTG601310S, RTG601312S, RTG601313S, RTG601314S, RTG601315S, RTG601316S, RTG601317S, RTG601318S, RTG601319S, RTG601320S, RTG601321S, RTG601322S, RTG601328S, RTG601329S, RTG601344S, RTG601347S, RTG601348S, RTG601350S, RTG601351S, RTG601352S, RTG601353S, RTG601354S, RTG601356S, RTG601357S, RTG601359S, RTG601360S, RTG601361S, RTG601381S, RTG601384S, RTG601402S, RTG601404S, RTG601408S, RTG601416S, RTG601425S, RTG601429S, RTG601431S, RTG601434S, RTG601440S, RTG601443S, RTG601450S, RTG601451S, RTG601461S, RTG601464S, RTG601466S, RTG601467S, RTG601469S, RTG601470S, RTG601472S, RTG601473S, RTG601476S, RTG601477S, RTG601478S, RTG601480S, RTG601481S, RTG601483S, RTG601485S, RTG601486S, RTG601490S, RTG601492S, RTG601493S, RTG601496S, RTG601497S, RTG601499S, RTG601501S, RTG601505S, RTG601507S, RTG601516S, RTG601518S, RTG601527S, RTG601531S, RTG601535S, RTG601538S, RTG601543S, RTG601544S, RTG601545S, RTG601546S, RTG601547S, RTG601548S, RTG601549S, RTG601550S, RTG601552S, RTG601554S, RTG601555S, RTG601556S, RTG601557S, RTG601558S, RTG601559S, RTG601561S, RTG601566S, RTG601577S, RTG601578S, RTG601579S, RTG601582S, RTG601584S, RTG601587S, RTG601588S, RTG601590S, RTG601591S, RTG601592S, RTG601598S, RTG601609S, RTG601667S, RTG601686S, RTG601701S, RTG601702S, RTG601703S, RTG601704S, RTG601705S, RTG601707S, RTG601708S, RTG601709S, RTG601710S, RTG601711S, RTG601712S, RTG601713S, RTG601714S, RTG601715S, RTG601716S, RTG601717S, RTG601718S, RTG601719S, RTG601720S, RTG601721S, RTG601722S, RTG601723S, RTG601724S, RTG601725S, RTG601726S, RTG601727S, RTG601728S, RTG601729S, RTG601730S, RTG601731S, RTG601732S, RTG601733S, RTG601734S, RTG601735S, RTG601736S, RTG601737S, RTG601738S, RTG601739S, RTG601740S, RTG601741S, RTG601742S, RTG601743S, RTG601744S, RTG601745S, RTG601746S, RTG601747S, RTG601748S, RTG601749S, RTG601750S, RTG601751S, RTG601752S, RTG601753S, RTG601754S, RTG601755S, RTG601756S, RTG601757S, RTG601758S, RTG601759S, RTG601760S, RTG601766S, RTG601769S, RTG601770S, RTG601772S, RTG601774S, RTG601777S, RTG601779S, RTG601780S, RTG601781S, RTG601785S, RTG601788S, RTG601789S, RTG601792S, RTG601793S, RTG601795S, RTG601799S, RTG601800S, RTG601803S, RTG601804S, RTG601811S, RTG601815S, RTG601819S, RTG601831S, RTG601833S, RTG601836S, RTG601918S, RTG601924S, RTG601925S, RTG601935S, RTG601936S, RTG601939S, RTG601940S, RTG601941S, RTG601952S, RTG601954S, RTG601962S, RTG601985S, RTG601987S, RTG601988S, RTG601990S, RTG602000S, RTG602001S, RTG602002S, RTG602004S, RTG602005S, RTG602007S, RTG602013S, RTG602021S, RTG602022S, RTG602023S, RTG602026S, RTG602029S, RTG602030S, RTG602032S, RTG602038S, RTG602041S, RTG602043S, RTG602045S, RTG602046S, RTG602048S, RTG602052S, RTG602055S, RTG602059S, RTG602065S, RTG602066S, RTG602067S, RTG602071S, RTG602072S, RTG602074S, RTG602075S, RTG602078S, RTG602079S, RTG602080S, RTG602081S, RTG602082S, RTG602085S, RTG602086S, RTG602088S, RTG602089S, RTG602090S, RTG602094S, RTG602095S, RTG602099S, RTG602102S, RTG602103S, RTG602104S, RTG602105S, RTG602106S, RTG602107S, RTG602109S, RTG602111S, RTG602113S, RTG602114S, RTG602115S, RTG602116S, RTG602119S, RTG602120S, RTG602121S, RTG602122S, RTG602124S, RTG602129S, RTG602133S, RTG602169S, RTG602243S, RTG602245S, RTG602246S, RTG602247S, RTG602248S, RTG602250S, RTG602251S, RTG602254S, RTG602257S, RTG602259S, RTG602261S, RTG602263S, RTG602264S, RTG602265S, RTG602267S, RTG602268S, RTG602269S, RTG602270S, RTG602272S, RTG602274S, RTG602278S, RTG602279S, RTG602281S, RTG602283S, RTG602286S, RTG602287S, RTG602288S, RTG602289S, RTG602291S, RTG602292S, RTG602294S, RTG602297S, RTG602299S, RTG602300S, RTG602301S, RTG602307S, RTG602311S, RTG602328S, RTG602356S, RTG602362S, RTG602363S, RTG602366S, RTG602367S, RTG602368S, RTG602370S, RTG602371S, RTG602372S, RTG602374S, RTG602375S, RTG602376S, RTG602377S, RTG602378S, RTG602379S, RTG602380S, RTG602381S, RTG602382S, RTG602383S, RTG602384S, RTG602386S, RTG602387S, RTG602388S, RTG602389S, RTG602390S, RTG602391S, RTG602392S, RTG602393S, RTG602394S, RTG602395S, RTG602396S, RTG602397S, RTG602398S, RTG602399S, RTG602400S, RTG602401S, RTG602403S, RTG602404S, RTG602405S, RTG602406S, RTG602407S, RTG602408S, RTG602409S, RTG602410S, RTG602411S, RTG602412S, RTG602413S, RTG602414S, RTG602415S, RTG602416S, RTG602417S, RTG602418S, RTG602419S, RTG602420S, RTG602424S, RTG602429S, RTG602452S, RTG602518S, RTG602547S, RTG602558S, RTG602560S, RTG602564S, RTG602567S, RTG602568S, RTG602572S, RTG602574S, RTG602575S, RTG602589S, RTG602591S, RTG602592S, RTG602593S, RTG602596S, RTG602598S, RTG602599S, RTG602603S, RTG602604S, RTG602607S, RTG602608S, RTG602609S, RTG602610S, RTG602611S, RTG602613S, RTG602614S, RTG602621S, RTG602624S, RTG602626S, RTG602628S, RTG602632S, RTG602633S, RTG602635S, RTG602636S, RTG602637S, RTG602638S, RTG602639S, RTG602640S, RTG602641S, RTG602642S, RTG602643S, RTG602644S, RTG602645S, RTG602646S, RTG602647S, RTG602648S, RTG602649S, RTG602650S, RTG602651S, RTG602652S, RTG602653S, RTG602655S, RTG602656S, RTG602657S, RTG602658S, RTG602659S, RTG602660S, RTG602661S, RTG602664S, RTG602668S, RTG602670S, RTG602672S, RTG602673S, RTG602674S, RTG602676S, RTG602677S, RTG602678S, RTG602679S, RTG602680S, RTG602681S, RTG602682S, RTG602683S, RTG602684S, RTG602685S, RTG602688S, RTG602689S, RTG602690S, RTG602691S, RTG602694S, RTG602695S, RTG602696S, RTG602699S, RTG602701S, RTG602703S, RTG602705S, RTG602707S, RTG602708S; GTIN 00763000402211, Serial Numbers: RTG600302S, RTG600337S, RTG600344S, RTG600353S, RTG600355S, RTG600357S, RTG600362S, RTG600365S, RTG600371S, RTG600376S, RTG600379S, RTG600382S, RTG600387S, RTG600389S, RTG600391S, RTG600392S, RTG600396S, RTG600397S, RTG600407S, RTG600409S, RTG600410S, RTG600413S, RTG600414S, RTG600416S, RTG600417S, RTG600419S, RTG600420S, RTG600421S, RTG600422S, RTG600425S, RTG600427S, RTG600429S, RTG600431S, RTG600432S, RTG600433S, RTG600439S, RTG600442S, RTG600775S, RTG600776S, RTG600777S, RTG600778S, RTG600785S, RTG600789S, RTG600790S, RTG600797S, RTG600802S, RTG600804S, RTG600807S, RTG600814S, RTG600815S, RTG600818S, RTG600822S, RTG600823S, RTG600824S, RTG600825S, RTG600826S, RTG600827S, RTG600828S, RTG600830S, RTG600832S, RTG600833S, RTG600835S, RTG600836S, RTG600847S, RTG600861S, RTG600868S, RTG600870S, RTG600875S, RTG600992S, RTG600994S, RTG600995S, RTG601007S, RTG601009S, RTG601012S, RTG601025S, RTG601029S, RTG601031S, RTG601032S, RTG601040S, RTG601091S, RTG601093S, RTG601094S, RTG601095S, RTG601097S, RTG601099S, RTG601100S, RTG601101S, RTG601104S, RTG601105S, RTG601107S, RTG601110S, RTG601113S, RTG601114S, RTG601115S, RTG601116S, RTG601117S, RTG601118S, RTG601119S, RTG601120S, RTG601121S, RTG601122S, RTG601123S, RTG601124S, RTG601125S, RTG601126S, RTG601128S, RTG601131S, RTG601132S, RTG601133S, RTG601188S, RTG601208S, RTG601214S, RTG601220S, RTG601233S, RTG601250S, RTG601251S, RTG601254S, RTG601259S, RTG601261S, RTG601266S, RTG601267S, RTG601278S, RTG601279S, RTG601280S, RTG601282S, RTG601290S, RTG601291S, RTG601297S, RTG601305S, RTG601311S, RTG601569S, RTG601823S, RTG601825S, RTG601835S, RTG601838S, RTG601946S, RTG601947S, RTG602003S, RTG602009S, RTG602010S, RTG602044S, RTG602050S, RTG602056S, RTG602062S, RTG602063S, RTG602068S, RTG602076S, RTG602084S, RTG602087S, RTG602091S, RTG602097S, RTG602098S, RTG602100S, RTG602101S, RTG602110S, RTG602252S, RTG602255S, RTG602256S, RTG602262S, RTG602276S, RTG602277S, RTG602280S, RTG602285S, RTG602293S, RTG602298S, RTG602323S, RTG602337S, RTG602343S, RTG602344S, RTG602402S, RTG602541S, RTG602570S, RTG602578S, RTG602597S, RTG602606S, RTG602612S, RTG602618S, RTG602629S, RTG602667S, RTG602715S, RTG602717S, RTG602727S, RTG602729S; GTIN none, Serial Numbers: RTG601829S, RTG602061S, RTG602258S, RTG602669S, RTG602724S; DTPA2D1: GTIN 00763000178147, Serial Numbers: RTH600598S, RTH601437S, RTH601544S, RTH601557S, RTH601590S, RTH601762S, RTH601782S, RTH601832S, RTH601844S, RTH602085S, RTH602090S, RTH602094S, RTH602104S, RTH602105S, RTH602150S, RTH602168S, RTH602172S, RTH602198S, RTH602220S, RTH602255S, RTH602261S, RTH602265S, RTH602267S, RTH602282S, RTH602312S, RTH602320S; GTIN 00763000178161, Serial Numbers: RTH600013S, RTH600014S, RTH600015S, RTH600016S, RTH600021S, RTH600030S, RTH600031S, RTH600032S, RTH600033S, RTH600034S, RTH600035S, RTH600036S, RTH600037S, RTH600039S, RTH600041S, RTH600042S, RTH600044S, RTH600046S, RTH600048S, RTH600049S, RTH600051S, RTH600052S, RTH600054S, RTH600055S, RTH600056S, RTH600058S, RTH600059S, RTH600061S, RTH600062S, RTH600063S, RTH600064S, RTH600065S, RTH600066S, RTH600071S, RTH600072S, RTH600073S, RTH600075S, RTH600078S, RTH600079S, RTH600080S, RTH600083S, RTH600085S, RTH600086S, RTH600087S, RTH600090S, RTH600276S, RTH600283S, RTH600299S, RTH600328S, RTH600344S, RTH600374S, RTH600442S, RTH600526S, RTH600531S, RTH600539S, RTH600571S, RTH600575S, RTH600691S, RTH600693S, RTH600695S, RTH600720S, RTH600721S, RTH600727S, RTH600730S, RTH600733S, RTH600739S, RTH600740S, RTH600741S, RTH600745S, RTH600750S, RTH600759S, RTH600760S, RTH600761S, RTH600762S, RTH600763S, RTH600764S, RTH600765S, RTH600771S, RTH600772S, RTH600773S, RTH600774S, RTH600775S, RTH600778S, RTH600779S, RTH600780S, RTH600782S, RTH600785S, RTH600788S, RTH600789S, RTH600790S, RTH600791S, RTH600792S, RTH600793S, RTH600795S, RTH600796S, RTH600797S, RTH600813S, RTH600848S, RTH600858S, RTH600861S, RTH600864S, RTH600865S, RTH600866S, RTH600867S, RTH600871S, RTH600874S, RTH600880S, RTH600886S, RTH600890S, RTH600892S, RTH600895S, RTH600896S, RTH600899S, RTH600900S, RTH600901S, RTH600902S, RTH600903S, RTH600904S, RTH600907S, RTH600908S, RTH600909S, RTH600910S, RTH600911S, RTH600912S, RTH600913S, RTH600914S, RTH600915S, RTH600916S, RTH600917S, RTH600919S, RTH600920S, RTH600921S, RTH600924S, RTH600925S, RTH600927S, RTH600928S, RTH600929S, RTH600930S, RTH600931S, RTH600932S, RTH600933S, RTH600934S, RTH600936S, RTH600938S, RTH600939S, RTH600941S, RTH600942S, RTH600943S, RTH600946S, RTH600948S, RTH600950S, RTH600951S, RTH600952S, RTH600955S, RTH600957S, RTH600958S, RTH600959S, RTH600960S, RTH600962S, RTH600963S, RTH600964S, RTH600967S, RTH600968S, RTH600970S, RTH600971S, RTH600972S, RTH600973S, RTH600974S, RTH600975S, RTH600976S, RTH600977S, RTH600978S, RTH600980S, RTH600981S, RTH600984S, RTH600986S, RTH600988S, RTH600989S, RTH600993S, RTH600994S, RTH600996S, RTH600998S, RTH601000S, RTH601002S, RTH601004S, RTH601005S, RTH601006S, RTH601007S, RTH601008S, RTH601011S, RTH601012S, RTH601013S, RTH601014S, RTH601015S, RTH601016S, RTH601017S, RTH601018S, RTH601019S, RTH601020S, RTH601021S, RTH601022S, RTH601023S, RTH601026S, RTH601028S, RTH601031S, RTH601032S, RTH601033S, RTH601034S, RTH601035S, RTH601036S, RTH601038S, RTH601039S, RTH601040S, RTH601041S, RTH601042S, RTH601043S, RTH601044S, RTH601045S, RTH601046S, RTH601047S, RTH601048S, RTH601049S, RTH601050S, RTH601051S, RTH601052S, RTH601054S, RTH601058S, RTH601059S, RTH601060S, RTH601062S, RTH601063S, RTH601065S, RTH601066S, RTH601067S, RTH601068S, RTH601069S, RTH601070S, RTH601071S, RTH601073S, RTH601074S, RTH601075S, RTH601076S, RTH601077S, RTH601080S, RTH601081S, RTH601082S, RTH601083S, RTH601084S, RTH601085S, RTH601086S, RTH601087S, RTH601088S, RTH601089S, RTH601090S, RTH601091S, RTH601092S, RTH601094S, RTH601096S, RTH601097S, RTH601098S, RTH601099S, RTH601100S, RTH601101S, RTH601102S, RTH601103S, RTH601105S, RTH601106S, RTH601107S, RTH601108S, RTH601109S, RTH601110S, RTH601111S, RTH601112S, RTH601113S, RTH601114S, RTH601115S, RTH601116S, RTH601117S, RTH601118S, RTH601119S, RTH601120S, RTH601121S, RTH601122S, RTH601123S, RTH601124S, RTH601126S, RTH601127S, RTH601131S, RTH601133S, RTH601135S, RTH601136S, RTH601137S, RTH601138S, RTH601140S, RTH601141S, RTH601143S, RTH601164S, RTH601165S, RTH601166S, RTH601169S, RTH601171S, RTH601172S, RTH601173S, RTH601174S, RTH601176S, RTH601178S, RTH601179S, RTH601180S, RTH601181S, RTH601183S, RTH601184S, RTH601186S, RTH601187S, RTH601188S, RTH601189S, RTH601190S, RTH601191S, RTH601194S, RTH601195S, RTH601196S, RTH601197S, RTH601198S, RTH601199S, RTH601200S, RTH601201S, RTH601202S, RTH601203S, RTH601204S, RTH601215S, RTH601276S, RTH601279S, RTH601280S, RTH601281S, RTH601283S, RTH601284S, RTH601285S, RTH601286S, RTH601287S, RTH601288S, RTH601289S, RTH601290S, RTH601291S, RTH601292S, RTH601293S, RTH601314S, RTH601315S, RTH601320S, RTH601324S, RTH601328S, RTH601335S, RTH601336S, RTH601338S, RTH601345S, RTH601349S, RTH601351S, RTH601358S, RTH601359S, RTH601360S, RTH601362S, RTH601363S, RTH601384S, RTH601461S, RTH601462S, RTH601464S, RTH601465S, RTH601467S, RTH601468S, RTH601469S, RTH601471S, RTH601472S, RTH601473S, RTH601474S, RTH601476S, RTH601477S, RTH601478S, RTH601481S, RTH601488S, RTH601489S, RTH601490S, RTH601491S, RTH601492S, RTH601493S, RTH601495S, RTH601496S, RTH601497S, RTH601498S, RTH601499S, RTH601500S, RTH601501S, RTH601503S, RTH601505S, RTH601506S, RTH601507S, RTH601512S, RTH601513S, RTH601515S, RTH601519S, RTH601523S, RTH601524S, RTH601527S, RTH601528S, RTH601529S, RTH601530S, RTH601531S, RTH601532S, RTH601535S, RTH601536S, RTH601537S, RTH601539S, RTH601540S, RTH601592S, RTH601593S, RTH601594S, RTH601595S, RTH601596S, RTH601597S, RTH601598S, RTH601599S, RTH601600S, RTH601601S, RTH601602S, RTH601603S, RTH601604S, RTH601605S, RTH601606S, RTH601607S, RTH601608S, RTH601609S, RTH601612S, RTH601628S, RTH601630S, RTH601639S, RTH601650S, RTH601654S, RTH601659S, RTH601660S, RTH601668S, RTH601672S, RTH601674S, RTH601676S, RTH601678S, RTH601680S, RTH601681S, RTH601682S, RTH601683S, RTH601684S, RTH601685S, RTH601689S, RTH601690S, RTH601691S, RTH601692S, RTH601693S, RTH601698S, RTH601700S, RTH601709S, RTH601717S, RTH601720S, RTH601722S, RTH601723S, RTH601775S, RTH601824S, RTH601834S, RTH601841S, RTH601847S, RTH601850S, RTH601855S, RTH601860S, RTH601867S, RTH601869S, RTH601870S, RTH601881S, RTH601882S, RTH601887S, RTH601893S, RTH601898S, RTH601900S, RTH601901S, RTH601903S, RTH601914S, RTH601917S, RTH601918S, RTH601919S, RTH601920S, RTH601921S, RTH601922S, RTH601924S, RTH601925S, RTH601926S, RTH601927S, RTH601928S, RTH601931S, RTH601933S, RTH601935S, RTH601940S, RTH601941S, RTH601943S, RTH601945S, RTH601946S, RTH601947S, RTH601953S, RTH601955S, RTH601957S, RTH601960S, RTH601961S, RTH601968S, RTH601969S, RTH601971S, RTH601972S, RTH601980S, RTH601983S, RTH601984S, RTH601985S, RTH601987S, RTH601989S, RTH601990S, RTH602001S, RTH602004S, RTH602006S, RTH602007S, RTH602010S, RTH602011S, RTH602013S, RTH602120S, RTH602124S, RTH602125S, RTH602132S, RTH602145S, RTH602157S, RTH602158S, RTH602161S, RTH602163S, RTH602166S, RTH602184S, RTH602253S, RTH602273S, RTH602274S, RTH602283S, RTH602285S, RTH602286S, RTH602287S, RTH602290S, RTH602291S, RTH602292S, RTH602293S, RTH602294S, RTH602295S, RTH602296S, RTH602297S, RTH602299S, RTH602300S, RTH602304S, RTH602307S, RTH602308S, RTH602309S, RTH602310S, RTH602311S, RTH602313S, RTH602314S, RTH602315S, RTH602316S, RTH602318S, RTH602322S, RTH602323S, RTH602324S, RTH602325S, RTH602326S, RTH602328S, RTH602329S, RTH602330S, RTH602331S, RTH602332S, RTH602333S, RTH602334S, RTH602335S, RTH602336S, RTH602337S, RTH602338S, RTH602339S, RTH602340S, RTH602341S, RTH602342S, RTH602343S, RTH602344S, RTH602345S, RTH602346S, RTH602347S, RTH602348S, RTH602349S, RTH602350S, RTH602351S, RTH602358S, RTH602359S, RTH602362S, RTH602364S, RTH602367S, RTH602370S, RTH602371S, RTH602373S, RTH602376S, RTH602378S, RTH602379S, RTH602382S, RTH602384S, RTH602386S, RTH602393S, RTH602398S, RTH602400S, RTH602403S, RTH602404S, RTH602405S, RTH602406S, RTH602410S, RTH602411S, RTH602416S, RTH602422S, RTH602423S, RTH602424S, RTH602425S, RTH602426S, RTH602432S, RTH602433S, RTH602435S, RTH602436S, RTH602438S, RTH602440S; GTIN 00763000402204, Serial Numbers: RTH600724S, RTH600725S, RTH600726S, RTH600729S, RTH600731S, RTH600732S, RTH600735S, RTH600736S, RTH600737S, RTH600738S, RTH600742S, RTH600743S, RTH600744S, RTH600746S, RTH600747S, RTH600749S, RTH600751S, RTH600752S, RTH600755S, RTH600756S, RTH600757S, RTH600758S, RTH600766S, RTH600767S, RTH600769S, RTH600777S, RTH600783S, RTH600786S, RTH600794S, RTH600798S, RTH600799S, RTH600800S, RTH600801S, RTH600802S, RTH600803S, RTH600804S, RTH600805S, RTH600807S, RTH600808S, RTH600809S, RTH600810S, RTH600812S, RTH600814S, RTH600815S, RTH600816S, RTH600817S, RTH600818S, RTH600819S, RTH600820S, RTH600821S, RTH600822S, RTH600824S, RTH600825S, RTH600826S, RTH600827S, RTH600828S, RTH600829S, RTH600830S, RTH600831S, RTH600832S, RTH600833S, RTH600834S, RTH600837S, RTH600838S, RTH600839S, RTH600840S, RTH600841S, RTH600842S, RTH600843S, RTH600844S, RTH600846S, RTH600849S, RTH600850S, RTH600851S, RTH600853S, RTH600854S, RTH600855S, RTH600857S, RTH600859S, RTH600860S, RTH600862S, RTH600863S, RTH600868S, RTH600869S, RTH600870S, RTH600872S, RTH600873S, RTH600875S, RTH600876S, RTH600877S, RTH600878S, RTH600879S, RTH600881S, RTH600884S, RTH600887S, RTH600888S, RTH600889S, RTH600893S, RTH600897S, RTH600898S, RTH600905S, RTH600906S, RTH600918S, RTH600922S, RTH600923S, RTH600926S, RTH600937S, RTH600940S, RTH600944S, RTH600945S, RTH600947S, RTH600953S, RTH600956S, RTH600961S, RTH600965S, RTH600966S, RTH600969S, RTH600985S, RTH600990S, RTH600992S, RTH600995S, RTH601001S, RTH601027S, RTH601029S, RTH601030S, RTH601037S, RTH601053S, RTH601061S, RTH601095S, RTH601129S, RTH601132S, RTH601139S, RTH601142S, RTH601205S, RTH601206S, RTH601207S, RTH601208S, RTH601209S, RTH601210S, RTH601211S, RTH601212S, RTH601214S, RTH601216S, RTH601217S, RTH601218S, RTH601219S, RTH601220S, RTH601221S, RTH601222S, RTH601223S, RTH601224S, RTH601225S, RTH601226S, RTH601227S, RTH601228S, RTH601229S, RTH601230S, RTH601231S, RTH601232S, RTH601233S, RTH601234S, RTH601235S, RTH601236S, RTH601237S, RTH601238S, RTH601239S, RTH601240S, RTH601241S, RTH601242S, RTH601243S, RTH601244S, RTH601245S, RTH601246S, RTH601247S, RTH601248S, RTH601249S, RTH601250S, RTH601251S, RTH601252S, RTH601305S, RTH601308S, RTH601311S, RTH601312S, RTH601326S, RTH601327S, RTH601329S, RTH601332S, RTH601339S, RTH601346S, RTH601347S, RTH601348S, RTH601352S, RTH601355S, RTH601381S, RTH601385S, RTH601389S, RTH601958S, RTH601962S, RTH601964S, RTH601970S, RTH601974S, RTH601976S, RTH601979S, RTH601982S, RTH601986S, RTH601993S, RTH601996S, RTH602008S, RTH602009S, RTH602114S, RTH602115S, RTH602117S, RTH602119S, RTH602121S, RTH602123S, RTH602126S, RTH602127S, RTH602128S, RTH602130S, RTH602131S, RTH602289S, RTH602352S, RTH602360S, RTH602361S, RTH602363S, RTH602366S, RTH602368S, RTH602369S, RTH602372S, RTH602374S, RTH602375S, RTH602377S, RTH602380S, RTH602381S, RTH602387S, RTH602391S, RTH602392S, RTH602394S, RTH602395S, RTH602399S, RTH602402S, RTH602408S, RTH602415S, RTH602417S, RTH602419S; GTIN none, Serial Numbers: RTH602129S, RTH602327S, RTH602409S; b. Model Numbers: DTPA2QQ: GTIN 00763000178093, Serial Numbers: RTC600339S, RTC600388S, RTC600657S, RTC600779S, RTC602354S, RTC605066S, RTC605335S, RTC605388S, RTC605468S, RTC606108S, RTC606304S, RTC606480S, RTC606523S, RTC606525S, RTC606531S, RTC606532S, RTC607084S, RTC607516S, RTC607842S, RTC607849S, RTC607859S, RTC608327S, RTC608331S, RTC608332S, RTC609193S, RTC609471S, RTC609652S, RTC609694S, RTC609704S, RTC610036S, RTC610579S, RTC610587S, RTC610594S, RTC610597S, RTC610612S, RTC610616S, RTC610620S, RTC610627S, RTC610632S, RTC610634S, RTC610640S, RTC610642S, RTC610644S, RTC610663S, RTC610674S, RTC610678S, RTC610744S, RTC610783S, RTC610910S, RTC610914S, RTC610963S, RTC610992S, RTC611007S, RTC611013S, RTC611037S, RTC611044S, RTC611058S, RTC611214S, RTC611215S, RTC611268S, RTC611287S, RTC611350S, RTC611710S, RTC611717S, RTC611748S; GTIN 00763000178116, Serial Numbers: RTC600048S, RTC600050S, RTC600053S, RTC600055S, RTC600056S, RTC600057S, RTC600058S, RTC600059S, RTC600060S, RTC600061S, RTC600062S, RTC600063S, RTC600064S, RTC600065S, RTC600066S, RTC600067S, RTC600068S, RTC600069S, RTC600070S, RTC600071S, RTC600072S, RTC600073S, RTC600074S, RTC600075S, RTC600076S, RTC600077S, RTC600078S, RTC600079S, RTC600080S, RTC600081S, RTC600082S, RTC600083S, RTC600085S, RTC600087S, RTC600093S, RTC600094S, RTC600095S, RTC600096S, RTC600098S, RTC600099S, RTC600100S, RTC600101S, RTC600102S, RTC600103S, RTC600104S, RTC600105S, RTC600106S, RTC600107S, RTC600108S, RTC600109S, RTC600110S, RTC600111S, RTC600112S, RTC600113S, RTC600114S, RTC600115S, RTC600116S, RTC600117S, RTC600118S, RTC600119S, RTC600120S, RTC600121S, RTC600122S, RTC600123S, RTC600124S, RTC600125S, RTC600126S, RTC600127S, RTC600128S, RTC600129S, RTC600130S, RTC600131S, RTC600132S, RTC600134S, RTC600135S, RTC600136S, RTC600137S, RTC600138S, RTC600139S, RTC600140S, RTC600141S, RTC600142S, RTC600143S, RTC600144S, RTC600145S, RTC600146S, RTC600147S, RTC600148S, RTC600149S, RTC600150S, RTC600151S, RTC600152S, RTC600153S, RTC600154S, RTC600155S, RTC600156S, RTC600157S, RTC600158S, RTC600159S, RTC600160S, RTC600161S, RTC600162S, RTC600163S, RTC600164S, RTC600168S, RTC600169S, RTC600170S, RTC600171S, RTC600172S, RTC600173S, RTC600174S, RTC600175S, RTC600176S, RTC600177S, RTC600179S, RTC600180S, RTC600181S, RTC600182S, RTC600183S, RTC600184S, RTC600185S, RTC600186S, RTC600187S, RTC600188S, RTC600189S, RTC600200S, RTC600313S, RTC600403S, RTC600544S, RTC600550S, RTC600557S, RTC600558S, RTC600559S, RTC600560S, RTC600561S, RTC600563S, RTC600564S, RTC600565S, RTC600566S, RTC600568S, RTC600569S, RTC600572S, RTC600575S, RTC600584S, RTC600585S, RTC600589S, RTC600590S, RTC600596S, RTC600599S, RTC600600S, RTC600603S, RTC600606S, RTC600608S, RTC600609S, RTC600611S, RTC600612S, RTC600616S, RTC600618S, RTC600622S, RTC600626S, RTC600627S, RTC600629S, RTC600630S, RTC600631S, RTC600632S, RTC600634S, RTC600635S, RTC600636S, RTC600637S, RTC600640S, RTC600641S, RTC600644S, RTC600645S, RTC600646S, RTC600651S, RTC600652S, RTC600654S, RTC600661S, RTC600665S, RTC600681S, RTC600733S, RTC600734S, RTC600735S, RTC600737S, RTC600739S, RTC600747S, RTC600748S, RTC600749S, RTC600752S, RTC600753S, RTC600754S, RTC600758S, RTC600760S, RTC600764S, RTC600766S, RTC600770S, RTC600771S, RTC600772S, RTC600775S, RTC600777S, RTC600803S, RTC600804S, RTC600805S, RTC600806S, RTC600807S, RTC600808S, RTC600809S, RTC600810S, RTC600811S, RTC600812S, RTC600813S, RTC600814S, RTC600815S, RTC600816S, RTC600817S, RTC600818S, RTC600819S, RTC600820S, RTC600821S, RTC600822S, RTC600823S, RTC600824S, RTC600825S, RTC600826S, RTC600827S, RTC600828S, RTC600913S, RTC600918S, RTC600919S, RTC600922S, RTC600927S, RTC600931S, RTC600932S, RTC600933S, RTC600934S, RTC600935S, RTC600936S, RTC600937S, RTC600938S, RTC600939S, RTC600940S, RTC600941S, RTC600942S, RTC600943S, RTC600944S, RTC600945S, RTC600946S, RTC600947S, RTC600948S, RTC600949S, RTC600950S, RTC600951S, RTC600952S, RTC600953S, RTC600954S, RTC600955S, RTC600956S, RTC600957S, RTC600958S, RTC600959S, RTC600960S, RTC600961S, RTC600962S, RTC600963S, RTC600964S, RTC600965S, RTC600966S, RTC600967S, RTC600968S, RTC600969S, RTC600970S, RTC600971S, RTC600972S, RTC600973S, RTC600974S, RTC600975S, RTC600976S, RTC600977S, RTC600978S, RTC600979S, RTC600981S, RTC600982S, RTC600983S, RTC600984S, RTC600985S, RTC600986S, RTC600987S, RTC600988S, RTC600989S, RTC600990S, RTC600991S, RTC600992S, RTC600993S, RTC600994S, RTC600995S, RTC600996S, RTC601003S, RTC601009S, RTC601010S, RTC601011S, RTC601012S, RTC601013S, RTC601014S, RTC601015S, RTC601016S, RTC601017S, RTC601018S, RTC601019S, RTC601020S, RTC601021S, RTC601022S, RTC601023S, RTC601024S, RTC601025S, RTC601026S, RTC601027S, RTC601028S, RTC601029S, RTC601030S, RTC601031S, RTC601032S, RTC601033S, RTC601034S, RTC601035S, RTC601036S, RTC601037S, RTC601038S, RTC601039S, RTC601040S, RTC601041S, RTC601042S, RTC601043S, RTC601044S, RTC601045S, RTC601046S, RTC601047S, RTC601048S, RTC601049S, RTC601050S, RTC601051S, RTC601052S, RTC601053S, RTC601054S, RTC601055S, RTC601056S, RTC601057S, RTC601058S, RTC601059S, RTC601060S, RTC601061S, RTC601063S, RTC601064S, RTC601065S, RTC601066S, RTC601067S, RTC601068S, RTC601069S, RTC601070S, RTC601071S, RTC601072S, RTC601073S, RTC601074S, RTC601075S, RTC601076S, RTC601077S, RTC601078S, RTC601079S, RTC601080S, RTC601081S, RTC601084S, RTC601091S, RTC601092S, RTC601095S, RTC601096S, RTC601098S, RTC601105S, RTC601120S, RTC601122S, RTC601133S, RTC601136S, RTC601139S, RTC601140S, RTC601169S, RTC601220S, RTC601230S, RTC601238S, RTC601286S, RTC601289S, RTC601294S, RTC601295S, RTC601296S, RTC601300S, RTC601302S, RTC601306S, RTC601307S, RTC601315S, RTC601321S, RTC601325S, RTC601335S, RTC601347S, RTC601364S, RTC601365S, RTC601369S, RTC601374S, RTC601377S, RTC601378S, RTC601380S, RTC601382S, RTC601406S, RTC601411S, RTC601415S, RTC601437S, RTC601441S, RTC601462S, RTC601468S, RTC601480S, RTC601481S, RTC601492S, RTC601499S, RTC601500S, RTC601501S, RTC601502S, RTC601503S, RTC601504S, RTC601505S, RTC601507S, RTC601508S, RTC601509S, RTC601510S, RTC601511S, RTC601512S, RTC601513S, RTC601514S, RTC601515S, RTC601516S, RTC601517S, RTC601518S, RTC601519S, RTC601520S, RTC601521S, RTC601522S, RTC601523S, RTC601524S, RTC601525S, RTC601526S, RTC601527S, RTC601528S, RTC601529S, RTC601530S, RTC601531S, RTC601532S, RTC601533S, RTC601534S, RTC601535S, RTC601536S, RTC601537S, RTC601538S, RTC601539S, RTC601540S, RTC601541S, RTC601542S, RTC601543S, RTC601544S, RTC601545S, RTC601546S, RTC601547S, RTC601548S, RTC601549S, RTC601550S, RTC601551S, RTC601552S, RTC601553S, RTC601554S, RTC601555S, RTC601556S, RTC601557S, RTC601558S, RTC601559S, RTC601560S, RTC601561S, RTC601562S, RTC601563S, RTC601564S, RTC601565S, RTC601566S, RTC601567S, RTC601568S, RTC601569S, RTC601570S, RTC601571S, RTC601572S, RTC601573S, RTC601574S, RTC601575S, RTC601578S, RTC601579S, RTC601580S, RTC601581S, RTC601582S, RTC601583S, RTC601584S, RTC601585S, RTC601586S, RTC601587S, RTC601588S, RTC601589S, RTC601590S, RTC601591S, RTC601592S, RTC601593S, RTC601594S, RTC601595S, RTC601596S, RTC601597S, RTC601598S, RTC601599S, RTC601601S, RTC601602S, RTC601603S, RTC601604S, RTC601605S, RTC601606S, RTC601607S, RTC601608S, RTC601609S, RTC601610S, RTC601611S, RTC601612S, RTC601613S, RTC601614S, RTC601615S, RTC601616S, RTC601617S, RTC601618S, RTC601619S, RTC601620

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1486-2022

Related Recalls

Class I - Dangerous

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

May 10, 2023 Implants & Prosthetics Nationwide View Details →
Class I - Dangerous

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

May 10, 2023 Implants & Prosthetics Nationwide View Details →
Class I - Dangerous

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

May 10, 2023 Implants & Prosthetics Nationwide View Details →