Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI (GTIN): 07613327294910 Affected Lots: 22011017 21330017 20216017 20140017 20139027
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Stryker Instruments
Reason for Recall:
There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.

Product Codes/Lot Numbers:

UDI-DI (GTIN): 07613327294910 Affected Lots: 22011017 21330017 20216017 20140017 20139027

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1478-2022