API Listeria (REF 10300). ZYM B reagent is used and included inside of API NH (REF 10400) and API Listeria (10300). All of the APIs are standardized systems combining several biochemical tests that enable group or species identification of most microorganisms.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Code - 1002569040, Exp. Date 8/29/2014, 1002569041, Exp. Date 8/296/2014, 1002587190, Exp. Date 9/3/2014, 1002856790, Exp. Date 7/18/2014, 1002856800, Exp. Date 6/12/2014
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- BioMerieux SA
- Reason for Recall:
- bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
API Listeria (REF 10300). ZYM B reagent is used and included inside of API NH (REF 10400) and API Listeria (10300). All of the APIs are standardized systems combining several biochemical tests that enable group or species identification of most microorganisms.
Product Codes/Lot Numbers:
Code - 1002569040, Exp. Date 8/29/2014, 1002569041, Exp. Date 8/296/2014, 1002587190, Exp. Date 9/3/2014, 1002856790, Exp. Date 7/18/2014, 1002856800, Exp. Date 6/12/2014
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1478-2014
Related Recalls
Under certain conditions, there is a risk for a false negative result.
Software anomaly - Under certain conditions, unwanted alterations to results could be applied when using the BCI Connect scripting feature.
NucliSENS¿ Magnetic Extraction Reagents
BioMerieux SA
Data loggers showed that one shipment to the United States experienced low temperatures that may impact performance of this product.