Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R. Stopcocks and manifolds used to control the direction of IV fluid flow.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a. Model Number MX1431MR, Lot Numbers: 3995544 b. Model Number MX1431MRL, Lot Numbers: 3995545, 4011838, 4027758, 4033501, 4044959 c. Model Number MX4331R, Lot Numbers: 3971467, 3987918
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Smiths Medical ASD Inc.
Reason for Recall:
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R. Stopcocks and manifolds used to control the direction of IV fluid flow.

Product Codes/Lot Numbers:

a. Model Number MX1431MR, Lot Numbers: 3995544 b. Model Number MX1431MRL, Lot Numbers: 3995545, 4011838, 4027758, 4033501, 4044959 c. Model Number MX4331R, Lot Numbers: 3971467, 3987918

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1477-2021

Related Recalls

CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR FUKUOKA 21-2111-0300-09_FUK CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR KYOTO 21-2111-0300-09_KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OKAYAMA 21-2111-0300-09_OKA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OOMIYA 21-2111-0300-09_OMI CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OSAKA 21-2111-0300-09_OSA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR SAPPORO 21-2111-0300-09_SAP CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR TOKYO 21-2111-0300-09_TOK CADD_SOLIS_MODEL 2110_ENGLISH_1/EA 21-2111-0200-50JP CADD-Solis VIP CADD-SOLIS V4.2 LOAN FUK 21-2111-0402-09-FUK CADD-SOLIS V4.2 LOAN NAG 21-2111-0402-09-NAG CADD-SOLIS V4.2 LOAN OKA 21-2111-0402-09-OKA CADD-SOLIS V4.2 LOAN OSA 21-2111-0402-09-OSA CADD-SOLIS V4.2 LOAN SAI 21-2111-0402-09-SAI CADD-SOLIS V4.2 LOAN SAP 21-2111-0402-09-SAP CADD-SOLIS V4.2 LOAN TOKE 21-2111-0402-09-TOKE CADD-SOLIS V4.2 LOAN TOKW 21-2111-0402-09-TOKW CADD-SOLIS V4.2 LOAN YELLOW KYO 21-2112-0402-09-KYO CADD-SOLIS V4.2 LOAN YELLOW NAG 21-2112-0402-09-NAG CADD-SOLIS V4.2 LOAN YELLOW YOK ¿ 21-2112-0402-09-YOK CADD-SOLIS V4.2 LOAN YOK 21-2111-0402-09-YOK Number:21-2111-0403-02Part Type:List Number/Finished GoodLifecycle Phase:03-ReleasedDescription:PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, FRENCH, 1/EA 21-2111-0403-02 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, YELLOW, ENGLISH, 1/EA 21-2112-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA 21-2111-0300-00 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DANISH 1/EA ¿ 21-2111-0300-12 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DUTCH 1/EA ¿ 21-2111-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, FRENCH 1/EA ¿ 21-2111-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, GERMAN 1/EA ¿ 21-2111-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, ITALIAN 1/EA ¿ 21-2111-0300-07 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, JAPANESE 21-2111-0300-09JP PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, FRENCH 1/EA ¿ 21-2111-0300-232 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA ¿ 21-2111-0300-231 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, SWEDISH 1/EA 21-2111-0300-244 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 Not PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, PORTUGUESE 1/EA ¿ 21-2111-0300-17 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SPANISH 1/EA ¿ 21-2111-0300-08 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SWEDISH 1/EA ¿ 21-2111-0300-14 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, DUTCH 1/EA 21-2112-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, GERMAN 1/EA 21-2112-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA 21-2112-0300-00 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, DANISH 1/EA ¿ 21-2101-0200-12 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, FRENCH 1/EA ¿ 21-2101-0200-02 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, GERMAN 1/EA ¿ 21-2101-0200-03 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, ITALIAN 1/EA ¿ 21-2101-02

Smiths Medical ASD

Class I - Dangerous

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Apr 10, 2025 Infusion Pumps Nationwide View Details →

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Apr 10, 2025 Infusion Pumps Nationwide View Details →

CADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE, 1/EA, Model Number: 21-2127-0105-01; 2) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED, 1/EA, Model Number: 21-2120-0105-01; 3) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH 1/EA, Model Number: 21-2120-0103-01; 4) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH 1/EA, Model Number: 21-2120-0102-51; 5) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-01; 6) PUMP KIT, CADD-SOLIS VIP, MDL 2120, CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-51; 7) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE 1/EA, Model Number: 21-2127-0104-01; 8) PUMP KIT, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER, NA ENGLISH, PHARMGUARD, 1/EA, Model Number: 21-2120-0105-00; 9) PUMP, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-78; 10) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, TRIAL/LOANER 1/EA, Model Number: 21-2120-0102-78; 11) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-231; 12) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249; 13) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-249; 14) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, FACTORY LIBRARY 1/EA, Model Number: 21-2125-0104-01; 15) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249.

Smiths Medical ASD

Class I - Dangerous

There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

Apr 10, 2025 Infusion Pumps Nationwide View Details →