SlingBar Wide 670. Designed to meet the needs for lifting humans.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product # 3156012, 3156017
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Hill-Rom, Inc.
- Reason for Recall:
- Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SlingBar Wide 670. Designed to meet the needs for lifting humans.
Product Codes/Lot Numbers:
Product # 3156012, 3156017
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1475-2013
Related Recalls
Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients
Missing bushing on Liko M220 and M230 causing wear and potential for patient falling
Missing bushing on Liko M220 and M230 causing wear and potential for patient falling