Arkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon Anesthesia Delivery System is designed to provide emergency oxygen, vaporized agent delivery and manual ventilation in the event of a power failure scenario. Clinicians are able to use Emergency O2 and manually ventilate the patient, providing gas and agent. However, the anesthesiologist will not have access to mechanical ventilation or ventilator monitoring.
Class I - DangerousWhat Should You Do?
- Check if you have this product: US Serial numbers: ARKN-000003, ARKN-000004, ARKN-000011, ARKN-000012, ARKN-000013, ARKN-000014, ARKN-000016, ARKN-000017, ARKN-000019, ARKN-000020, ARKN-000021, ARKN-000022, ARKN-000023, ARKN-000024, ARKN-000025, ARKN-000026, ARKN-000027, ARKN-000028, ARKN-000029, ARKN-000030, ARKN-000031, ARKN-000037, ARKN-000038, ARKN-000039, ARKN-000040, ARKN-000041, ARKN-000042, ARKN-000044, ARKN-000045, ARKN-000046, ARKN-000047, ARKN-000048, ARKN-000049, ARKN-000051, ARKN-000054, ARKN-000055, ARKN-000056, ARKN-000057, ARKN-000058, ARKN-000059, ARKN-000060, ARKN-000061, ARKN-000062, ARKN-000063, ARKN-000064, ARKN-000065, ARKN-000066, ARKN-000067, ARKN-000068, ARKN-000069, ARKN-000070, ARKN-000071, ARKN-000072, ARKN-000073, ARKN-000074, ARKN-000075, ARKN-000076, ARKN-000077, ARKN-000078, ARKN-000082, ARKN-000091, ARKN-000092, ARKN-000093, ARKN-000094, ARKN-000095, ARKN-000096, ARKN-000097, ARKN-000098, ARKN-000099, ARKN-000100, ARKN-000101, ARKN-000102, ARKN-000103, ARKN-000104, ARKN-000105, ARKN-000106, ARKN-000107, ARKN-000108, ARKN-000109, ARKN-000110, ARKN-000111, ARKN-000112, ARKN-000113, ARKN-000114, ARKN-000115, ARKN-000116, ARKN-000117, ARKN-000119, ARKN-000120, ARKN-000123, ARKN-000129, ARKN-000130, ARKN-000132, ARKN-000133, ARKN-000134, ARKN-000135, ARKN-000136, ARKN-000137, ARKN-000138, ARKN-000139, ARKN-000140, ARKN-000141, ARKN-000142, ARKN-000143, ARKN-000144, ARKN-000145, ARKN-000149, ARKN-000150, ARKN-000151, ARKN-000152, ARKN-000153, ARKN-000154, ARKN-000155, ARKN-000157, ARKN-000167, ARKN-000168, arkn-000169, arkn-000170, ARKN-000171, arkn-000173, arkn-000174, arkn-000175, arkn-000176, arkn-000177, arkn-000178, arkn-000179, arkn-000180, arkn-000181, arkn-000182, arkn-000183, arkn-000184, arkn-000185, ARKN-000186, ARKN-000187, ARKN-000188, ARKN-000189, ARKN-000190, ARKN-000191, ARKN-000192, ARKN-000193, ARKN-000194, ARKN-000196, ARKN-000197, ARKN-000198, ARKN-000199, ARKN-000200, ARKN-000201, ARKN-000202, ARKN-000203, ARKN-000204, ARKN-000205, ARKN-000206, ARKN-000207, ARKN-000208, ARKN-000209, ARKN-000210, ARKN-000211, ARKN-000212, ARKN-000213, ARKN-000214, ARKN-000215, ARKN-000216, ARKN-000217, ARKN-000220, ARKN-000221, ARKN-000222, arkn-000223, arkn-000224, arkn-000226, arkn-000227, arkn-000229, arkn-000230, arkn-000231, arkn-000232, arkn-000233, ARKN-000236, arkn-000237, arkn-000238, arkn-000239, arkn-000240, arkn-000241, arkn-000242, arkn-000243, arkn-000244, ARKN-000245, arkn-000248, arkn-000249, ARKN-000250, ARKN-000251, ARKN-000252, ARKN-000253, arkn-000254, arkn-000257, arkn-000258, ARKN-000259, ARKN-000260, arkn-000261, ARKN-000262, arkn-000263, arkn-000264, arkn-000266, arkn-000269, arkn-000270, arkn-000271, arkn-000272, arkn-000273, arkn-000274, arkn-000275, arkn-000276, arkn-000277, ARKN-000278, ARKN-000279, ARKN-000280, ARKN-000281, ARKN-000282, ARKN-000283, ARKN-000284, ARKN-000285, ARKN-000286, ARKN-000287, ARKN-000288, ARKN-000289, ARKN-000290, ARKN-000293, ARKN-000294, ARKN-000295, ARKN-000296, ARKN-000297, ARKN-000300, ARKN-000301, ARKN-000302, ARKN-000303, ARKN-000304, ARKN-000305, ARKN-000306, ARKN-000307, ARKN-000308, ARKN-000309, ARKN-000310, ARKN-000311, ARKN-000312, ARKN-000313, arkn-000314, arkn-000315, arkn-000316, arkn-000317, arkn-000318, ARKN-000319, arkn-000320, arkn-000321, arkn-000322, arkn-000323, arkn-000324, arkn-000325, arkn-000327, arkn-000328, arkn-000329, arkn-000330, ARKN-000332, ARKN-000333, ARKN-000334, ARKN-000335, ARKN-000336, ARKN-000337, ARKN-000338, ARKN-000339, ARKN-000340, ARKN-000341, ARKN-000342, ARKN-000343, ARKN-000344, ARKN-000345, ARKN-000353, ARKN-000360, ARKN-000361, ARKN-000362, ARKN-000363, ARKN-000364, ARKN-000365, ARKN-000366, ARKN-000367, ARKN-000369, ARKN-000370, ARKN-000371, ARKN-000372, ARKN-000373, ARKN-000374, ARKN-000376, ARKN-000377, ARKN-000378, ARKN-000379, ARKN-000380, ARKN-000381, ARKN-000382, ARKN-000383, ARKN-000384, ARKN-000385, ARKN-000386, ARKN-000387, ARKN-000388, ARKN-000389, ARKN-000390, ARKN-000391, ARKN-000392, ARKN-000393, ARKN-000394, ARKN-000395, ARKN-000396, ARKN-000397, ARKN-000398, ARKN-000399, ARKN-000400, ARKN-000401, ARKN-000402, ARKN-000403, ARKN-000404, ARKN-000405, ARKN-000406, ARKN-000407, ARKN-000408, ARKN-000409, ARKN-000410, ARKN-000411, ARKN-000412, and ARKN-000413. ******************************************************************************** INTERNATIONAL SERIAL NUMBERS: ARKN-000001, ARKN-000002, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000018, ARKN-000032, ARKN-000034, ARKN-000035, ARKN-000052, ARKN-000053, ARKN-000079, ARKN-000081, ARKN-000083, ARKN-000084, ARKN-000085, ARKN-000086, ARKN-000087, ARKN-000089, ARKN-000090, ARKN-000121, ARKN-000131, ARKN-000146, ARKN-000147, ARKN-000148, ARKN-000156, ARKN-000158, ARKN-000159, ARKN-000160, ARKN-000161, ARKN-000162, ARKN-000163, ARKN-000164, ARKN-000165, ARKN-000166, ARKN-000172, arkn-000218, arkn-000219, Arkn-000225, arkn-000234, arkn-000235, ARKN-000255, ARKN-000256, ARKN-000268, ARKN-000291, ARKN-000292, ARKN-000298, ARKN-000299, ARKN-000326, ARKN-000331, ARKN-000346, ARKN-000347, ARKN-000348, ARKN-000349, ARKN-000350, ARKN-000351, ARKN-000352, ARKN-000354, ARKN-000355, ARKN-000356, ARKN-000357, ARKN-000358, ARKN-000359, ARKN-000368, ARKN-000375, ARKN-000414, ARKN-000415, ARKN-000416, ARKN-000417, ARKN-000418, ARKN-000419, ARKN-000420, ARKN-000421, ARKN-000422, ARKN-000423, and ARKON-000088.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Del Mar Reynolds Medical, Ltd.
- Reason for Recall:
- Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Arkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon Anesthesia Delivery System is designed to provide emergency oxygen, vaporized agent delivery and manual ventilation in the event of a power failure scenario. Clinicians are able to use Emergency O2 and manually ventilate the patient, providing gas and agent. However, the anesthesiologist will not have access to mechanical ventilation or ventilator monitoring.
Product Codes/Lot Numbers:
US Serial numbers: ARKN-000003, ARKN-000004, ARKN-000011, ARKN-000012, ARKN-000013, ARKN-000014, ARKN-000016, ARKN-000017, ARKN-000019, ARKN-000020, ARKN-000021, ARKN-000022, ARKN-000023, ARKN-000024, ARKN-000025, ARKN-000026, ARKN-000027, ARKN-000028, ARKN-000029, ARKN-000030, ARKN-000031, ARKN-000037, ARKN-000038, ARKN-000039, ARKN-000040, ARKN-000041, ARKN-000042, ARKN-000044, ARKN-000045, ARKN-000046, ARKN-000047, ARKN-000048, ARKN-000049, ARKN-000051, ARKN-000054, ARKN-000055, ARKN-000056, ARKN-000057, ARKN-000058, ARKN-000059, ARKN-000060, ARKN-000061, ARKN-000062, ARKN-000063, ARKN-000064, ARKN-000065, ARKN-000066, ARKN-000067, ARKN-000068, ARKN-000069, ARKN-000070, ARKN-000071, ARKN-000072, ARKN-000073, ARKN-000074, ARKN-000075, ARKN-000076, ARKN-000077, ARKN-000078, ARKN-000082, ARKN-000091, ARKN-000092, ARKN-000093, ARKN-000094, ARKN-000095, ARKN-000096, ARKN-000097, ARKN-000098, ARKN-000099, ARKN-000100, ARKN-000101, ARKN-000102, ARKN-000103, ARKN-000104, ARKN-000105, ARKN-000106, ARKN-000107, ARKN-000108, ARKN-000109, ARKN-000110, ARKN-000111, ARKN-000112, ARKN-000113, ARKN-000114, ARKN-000115, ARKN-000116, ARKN-000117, ARKN-000119, ARKN-000120, ARKN-000123, ARKN-000129, ARKN-000130, ARKN-000132, ARKN-000133, ARKN-000134, ARKN-000135, ARKN-000136, ARKN-000137, ARKN-000138, ARKN-000139, ARKN-000140, ARKN-000141, ARKN-000142, ARKN-000143, ARKN-000144, ARKN-000145, ARKN-000149, ARKN-000150, ARKN-000151, ARKN-000152, ARKN-000153, ARKN-000154, ARKN-000155, ARKN-000157, ARKN-000167, ARKN-000168, arkn-000169, arkn-000170, ARKN-000171, arkn-000173, arkn-000174, arkn-000175, arkn-000176, arkn-000177, arkn-000178, arkn-000179, arkn-000180, arkn-000181, arkn-000182, arkn-000183, arkn-000184, arkn-000185, ARKN-000186, ARKN-000187, ARKN-000188, ARKN-000189, ARKN-000190, ARKN-000191, ARKN-000192, ARKN-000193, ARKN-000194, ARKN-000196, ARKN-000197, ARKN-000198, ARKN-000199, ARKN-000200, ARKN-000201, ARKN-000202, ARKN-000203, ARKN-000204, ARKN-000205, ARKN-000206, ARKN-000207, ARKN-000208, ARKN-000209, ARKN-000210, ARKN-000211, ARKN-000212, ARKN-000213, ARKN-000214, ARKN-000215, ARKN-000216, ARKN-000217, ARKN-000220, ARKN-000221, ARKN-000222, arkn-000223, arkn-000224, arkn-000226, arkn-000227, arkn-000229, arkn-000230, arkn-000231, arkn-000232, arkn-000233, ARKN-000236, arkn-000237, arkn-000238, arkn-000239, arkn-000240, arkn-000241, arkn-000242, arkn-000243, arkn-000244, ARKN-000245, arkn-000248, arkn-000249, ARKN-000250, ARKN-000251, ARKN-000252, ARKN-000253, arkn-000254, arkn-000257, arkn-000258, ARKN-000259, ARKN-000260, arkn-000261, ARKN-000262, arkn-000263, arkn-000264, arkn-000266, arkn-000269, arkn-000270, arkn-000271, arkn-000272, arkn-000273, arkn-000274, arkn-000275, arkn-000276, arkn-000277, ARKN-000278, ARKN-000279, ARKN-000280, ARKN-000281, ARKN-000282, ARKN-000283, ARKN-000284, ARKN-000285, ARKN-000286, ARKN-000287, ARKN-000288, ARKN-000289, ARKN-000290, ARKN-000293, ARKN-000294, ARKN-000295, ARKN-000296, ARKN-000297, ARKN-000300, ARKN-000301, ARKN-000302, ARKN-000303, ARKN-000304, ARKN-000305, ARKN-000306, ARKN-000307, ARKN-000308, ARKN-000309, ARKN-000310, ARKN-000311, ARKN-000312, ARKN-000313, arkn-000314, arkn-000315, arkn-000316, arkn-000317, arkn-000318, ARKN-000319, arkn-000320, arkn-000321, arkn-000322, arkn-000323, arkn-000324, arkn-000325, arkn-000327, arkn-000328, arkn-000329, arkn-000330, ARKN-000332, ARKN-000333, ARKN-000334, ARKN-000335, ARKN-000336, ARKN-000337, ARKN-000338, ARKN-000339, ARKN-000340, ARKN-000341, ARKN-000342, ARKN-000343, ARKN-000344, ARKN-000345, ARKN-000353, ARKN-000360, ARKN-000361, ARKN-000362, ARKN-000363, ARKN-000364, ARKN-000365, ARKN-000366, ARKN-000367, ARKN-000369, ARKN-000370, ARKN-000371, ARKN-000372, ARKN-000373, ARKN-000374, ARKN-000376, ARKN-000377, ARKN-000378, ARKN-000379, ARKN-000380, ARKN-000381, ARKN-000382, ARKN-000383, ARKN-000384, ARKN-000385, ARKN-000386, ARKN-000387, ARKN-000388, ARKN-000389, ARKN-000390, ARKN-000391, ARKN-000392, ARKN-000393, ARKN-000394, ARKN-000395, ARKN-000396, ARKN-000397, ARKN-000398, ARKN-000399, ARKN-000400, ARKN-000401, ARKN-000402, ARKN-000403, ARKN-000404, ARKN-000405, ARKN-000406, ARKN-000407, ARKN-000408, ARKN-000409, ARKN-000410, ARKN-000411, ARKN-000412, and ARKN-000413. ******************************************************************************** INTERNATIONAL SERIAL NUMBERS: ARKN-000001, ARKN-000002, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000018, ARKN-000032, ARKN-000034, ARKN-000035, ARKN-000052, ARKN-000053, ARKN-000079, ARKN-000081, ARKN-000083, ARKN-000084, ARKN-000085, ARKN-000086, ARKN-000087, ARKN-000089, ARKN-000090, ARKN-000121, ARKN-000131, ARKN-000146, ARKN-000147, ARKN-000148, ARKN-000156, ARKN-000158, ARKN-000159, ARKN-000160, ARKN-000161, ARKN-000162, ARKN-000163, ARKN-000164, ARKN-000165, ARKN-000166, ARKN-000172, arkn-000218, arkn-000219, Arkn-000225, arkn-000234, arkn-000235, ARKN-000255, ARKN-000256, ARKN-000268, ARKN-000291, ARKN-000292, ARKN-000298, ARKN-000299, ARKN-000326, ARKN-000331, ARKN-000346, ARKN-000347, ARKN-000348, ARKN-000349, ARKN-000350, ARKN-000351, ARKN-000352, ARKN-000354, ARKN-000355, ARKN-000356, ARKN-000357, ARKN-000358, ARKN-000359, ARKN-000368, ARKN-000375, ARKN-000414, ARKN-000415, ARKN-000416, ARKN-000417, ARKN-000418, ARKN-000419, ARKN-000420, ARKN-000421, ARKN-000422, ARKN-000423, and ARKON-000088.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1460-2017
Related Recalls
Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.
SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes a ventilator.
Del Mar Reynolds Medical
The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.