Integra¿ Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is comprised of the following components which are provided in a sealed and sterilized Tyvek / plastic tray: - An empty plastic syringe with Luer-Lok" tip - A plastic syringe with Luer-Lok" tip containing the Flowable Wound Matrix material - A syringe-to-syringe Luer-Lok" adapter / connector - A flexible plastic tube (injector) with Luer-Lok" connector Instructions for Use (IFU) and the sterilized package are subsequently packaged in a labeled and sealed box that is distributed to customers.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number FWD301 (US) & FDR301 (Non-US) The following Lot Numbers on the packaged kits are affected: - 305000293250, 305000293251, 305000298616 (foreign) The syringes within the packaged kits are also Lot numbered. For the packaged kit Lot Numbers above, the corresponding respective syringe Lot Numbers are: - 10500028786, 105000288118, 105N00288965 (foreign)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Integra LifeSciences Corp.
- Reason for Recall:
- A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified as not meeting stability acceptance criteria for the attributes of visual appearance during dispensing and force vs. displacement (force needed for product to be extruded from the syringe).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Integra¿ Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is comprised of the following components which are provided in a sealed and sterilized Tyvek / plastic tray: - An empty plastic syringe with Luer-Lok" tip - A plastic syringe with Luer-Lok" tip containing the Flowable Wound Matrix material - A syringe-to-syringe Luer-Lok" adapter / connector - A flexible plastic tube (injector) with Luer-Lok" connector Instructions for Use (IFU) and the sterilized package are subsequently packaged in a labeled and sealed box that is distributed to customers.
Product Codes/Lot Numbers:
Model Number FWD301 (US) & FDR301 (Non-US) The following Lot Numbers on the packaged kits are affected: - 305000293250, 305000293251, 305000298616 (foreign) The syringes within the packaged kits are also Lot numbered. For the packaged kit Lot Numbers above, the corresponding respective syringe Lot Numbers are: - 10500028786, 105000288118, 105N00288965 (foreign)
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1452-2015
Related Recalls
AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's Product Numbers (Catalog Number): (1) ASX15/60 and (2) ASX15/80
Integra LifeSciences
Possibility for the obturator to break (separate).
Potential that the induction seal is not completely sealed to the device tube packaging.
Potential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.