Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot 416245 ***Updated 10/5/20*** 483123 ***Updated 8/5/22*** Additional released lots are subject to the same workaround: 46449, 50474,525510, and 568976
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Roche Diagnostics Operations, Inc.
- Reason for Recall:
- The firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer
Product Codes/Lot Numbers:
Lot 416245 ***Updated 10/5/20*** 483123 ***Updated 8/5/22*** Additional released lots are subject to the same workaround: 46449, 50474,525510, and 568976
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1444-2020
Related Recalls
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine
Roche Diagnostics Operations
Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination of creatine kinase (CK) in human serum and plasma. Catalog Number: 07190794190
Roche Diagnostics Operations
Creatine Kinase (CK) Assay Abnormal calibrations on the cobas¿ c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.