8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip cover accessory, to be used in conjunction with the Intuitive Surgical Endoscopic Instrument Control System.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Numbers 420179-09 and -10 and 400179-09 and -10.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Intuitive Surgical, Inc.
- Reason for Recall:
- Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip cover accessory, to be used in conjunction with the Intuitive Surgical Endoscopic Instrument Control System.
Product Codes/Lot Numbers:
Model Numbers 420179-09 and -10 and 400179-09 and -10.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1442-2013
Related Recalls
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
8MM,MEGA SUTURECUT ND,IS4000 REF 471309
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.