Medline Convenience Kits: 1) NAIL KIT, Model Number: DYKM1528
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25BBO366; 2) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25ABI804; 3) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25ABF375; 4) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24JBL155; 5) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24IBG900; 6) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24FBT801; 7) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24FBC584; 8) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24DBB692; 9) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24BBL239; 10) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23KBO692; 11) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23IBT788; 12) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23HBJ026; 13) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23EBK780; 14) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23DBL149; 15) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23CBJ576 Additional Lot as of 3/27/26: 1) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25IBP134
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medline Industries, LP
- Reason for Recall:
- Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Medline Convenience Kits: 1) NAIL KIT, Model Number: DYKM1528
Product Codes/Lot Numbers:
1) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25BBO366; 2) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25ABI804; 3) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25ABF375; 4) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24JBL155; 5) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24IBG900; 6) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24FBT801; 7) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24FBC584; 8) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24DBB692; 9) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 24BBL239; 10) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23KBO692; 11) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23IBT788; 12) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23HBJ026; 13) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23EBK780; 14) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23DBL149; 15) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 23CBJ576 Additional Lot as of 3/27/26: 1) DYKM1528, UDI-DI: 10889942689927(each), 40889942689928(case), Lot Number: 25IBP134
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1434-2026
Related Recalls
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.