Bard Biopsy EnCor Probe, Stereotactic/Ultrasound 7G, REF number ECP017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot numbers VTCT0268, VTCT0269, VTCT0270, VTCT0271, VTCU0340, and VTCU0341. UDI numbers: (01)00801741086281(17)200601(10)VTCT0268, (01)00801741086281(17)200601(10)VTCT0269, (01)00801741086281(17)200601(10)VTCT0270, (01)00801741086281(17)200601(10)VTCT0271, (01)00801741086281(17)200601(10)VTCU0340, and (01)00801741086281(17)200601(10)VTCU0341.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Bard Peripheral Vascular Inc
- Reason for Recall:
- Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Bard Biopsy EnCor Probe, Stereotactic/Ultrasound 7G, REF number ECP017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.
Product Codes/Lot Numbers:
Lot numbers VTCT0268, VTCT0269, VTCT0270, VTCT0271, VTCU0340, and VTCU0341. UDI numbers: (01)00801741086281(17)200601(10)VTCT0268, (01)00801741086281(17)200601(10)VTCT0269, (01)00801741086281(17)200601(10)VTCT0270, (01)00801741086281(17)200601(10)VTCT0271, (01)00801741086281(17)200601(10)VTCU0340, and (01)00801741086281(17)200601(10)VTCU0341.
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1431-2020
Related Recalls
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Bard Peripheral Vascular
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.
Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239
Bard Peripheral Vascular
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.