FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI: 04055207009635; Lot Numbers: 4500377840, 4500377840, 4500377840, 4500376297, 4500371970, 4500363663, 4500363098, 4500361404, 4500353562, 4500349168, 4500347747, 4500338560, 4500333999
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Richard Wolf GmbH
- Reason for Recall:
- Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051
Product Codes/Lot Numbers:
UDI/DI: 04055207009635; Lot Numbers: 4500377840, 4500377840, 4500377840, 4500376297, 4500371970, 4500363663, 4500363098, 4500361404, 4500353562, 4500349168, 4500347747, 4500338560, 4500333999
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1423-2023
Related Recalls
The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.
FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 828.051
Richard Wolf GmbH
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
FLEX. GRASP. FORCEPS 6.6FR WL 550MM, Product Number 8736.685
Richard Wolf GmbH
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.