Enzymatic Creatinine Reagent (CR-E 2 X 200), REF A60298, For in Vitro Diagnostic Use, Rx Only, UDI: 15099590625733

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Lot Numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Beckman Coulter Inc
Reason for Recall:
It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Enzymatic Creatinine Reagent (CR-E 2 X 200), REF A60298, For in Vitro Diagnostic Use, Rx Only, UDI: 15099590625733

Product Codes/Lot Numbers:

All Lot Numbers

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1422-2020

Related Recalls

It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.

Nov 7, 2025 Infusion Pumps Nationwide View Details →

Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.

Sep 22, 2025 Diagnostic Equipment Nationwide View Details →